The Grab 'n Go system is an integrated delivery system intended to provide supplemental oxygen, by Rx only, to neonates, pediatrics and adults. The device is MR-Conditional, and suitable for use during MR imaging for MRI systems up to 3.0 Tesla. The device is intended for limited duration use, such as would be necessary during patient transports.
Device Story
Integrated medical gas delivery system comprising oxygen cylinder, valve-regulator, pressure gauge, flow meter, and protective shroud. Input: compressed oxygen gas. Operation: single-knob flow selection; provides user-selectable flow settings and 50-PSI DISS connection. Digital version includes pressure transducer for audible/visual low-contents alerts and estimated time-of-use calculation. Used by trained medical personnel in clinical settings, including during patient transport and MRI environments (up to 3.0T). Output: regulated oxygen flow to patient; visual/audible status indicators. Healthcare providers use output to manage patient oxygenation during transport or imaging. Benefits: portable, integrated oxygen delivery with safety monitoring.
Clinical Evidence
No clinical data. Evidence consists of bench testing and an end-user evaluation. End-user study validated that users could operate the device, understand the manual, and interpret alarm functionality per established acceptance criteria.
Technological Characteristics
Integrated valve-regulator with aluminum oxygen cylinder. Single-stage piston regulator. Digital version uses pressure transducer; standard version uses Bourdon-tube gauge. 4 internal filters. MR-conditional (ASTM F2052, F2213, F2503, F2119). Complies with ISO 10524-3, ISO 9170-1, ISO 11117, and CGA E-18. Electrical safety per IEC 60601-1, EMC per IEC 60601-1-2, alarm systems per IEC 60601-1-8.
Indications for Use
Indicated for neonates, pediatrics, and adults requiring supplemental oxygen. Intended for limited duration use, such as patient transport. MR-Conditional up to 3.0 Tesla.
Regulatory Classification
Identification
A pressure regulator is a device, often called a pressure-reducing valve, that is intended for medical purposes and that is used to convert a medical gas pressure from a high variable pressure to a lower, more constant working pressure. This device includes mechanical oxygen regulators.
Predicate Devices
LIV (Linde Integrated Valve) Portable Oxygen System (K101792)
Submission Summary (Full Text)
{0}------------------------------------------------
K132778 510(k) Summary
for the Praxair Grab 'n Go Plus / Grab 'n Go Digital
(per 21CFR 807.92 and http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/default.htm)
# 1. SUBMITTER/510(K) HOLDER
Praxair Healthcare Services 175 East Park Drive Tonawanda, NY 14150 800-772-9247 Phone: Internet: www.praxair.com
Establishment Registration No: 1000448358
Contact: Mike Skrjanc, Director FDA Compliance and Project Safety Contact's Phone: +1 (330) 949-3324 Contact's Fax: +1 (203) 837-2503 Contact's Email: Mike Skrjanc@praxair.com Date Prepared: September 3, 2013
## 2. DEVICE NAME
Trade Name:
Common Name:
1. Grab 'n Go Plus - Portable Medical Gas Delivery System (D, E cylinders)
2. Grab 'n Go Digital - Portable Medical Gas Delivery System (D, E cylinders)
Medical Gas Delivery System
Device Name: Integrated Valve-Regulator and Compressed Gas Cylinder Classification Panel: Anesthesiology Classification Number: Unclassified
Product Code: ECX
## 3. PREDICATE DEVICES
The following device is a legally marketed device to which equivalence is being claimed:
- · LIV (Linde Integrated Valve) Portable Oxygen System (Linde North America, Inc.), K101792
{1}------------------------------------------------
# 4. DEVICE DESCRIPTION
The Grab 'n Go Plus / Grab 'n Go Digital (collectively referred to as the Grab 'n Go system) integrated delivery system supplies oxygen using a device comprising an integrated valve-regulator with pressure gauge, flow meter and oxygen gas cylinder with protective shroud all integrated into a single unit. The device is able to provide a range of user selectable flow settings, including low and high flows that may be clinically appropriate for certain classes of patients. An additional DISS connection provides standard 50-PSI oxygen gas delivery.
The optional digital pressure gauge (Grab 'n Go Digital) provides both audible and visual low contents alerts and alarms as well as estimated time of use.
The system must only be used by trained medical personnel.
# 5. INDICATION FOR USE/INTENDED USE
The Grab 'n Go system is an integrated delivery system intended to provide supplemental oxygen, by Rx only, to neonates, pediatrics and adults. The device is MR-Conditional, and suitable for use during MR imaging for MRI systems up to 3.0 Tesla. The device is intended for limited duration use, such as would be necessary during patient transports.
### SUMMARY OF TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE 6. DEVICE
A summary comparison of technological characteristics, including design and materials is provided in the table below:
| Parameter | Grab 'n Go Plus / Grab 'n Go Digital<br>(Praxair Healthcare Services) | LIV (Linde Integrated Valve)<br>(Linde North America, Inc.) |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory Status | Proposed device | K101792 |
| Indications for Use | The Grab 'n Go system is an integrated delivery<br>system intended to provide supplemental<br>oxygen, by Rx only, to neonates, pediatrics and<br>adults. The device is MR-Conditional, and<br>suitable for use during MR imaging for MRI<br>systems up to 3.0 Tesla. The device is intended<br>for limited duration use, such as would be<br>necessary during patient transports. | The LIV is an integrated portable oxygen<br>delivery system intended to provide<br>supplemental oxygen to pediatric and adults.<br>The device is MR-conditional (per ASTM<br>standard 2503-05), and intended for use during<br>MR imaging for MRI systems up to 3.0T. Rx<br>only. Compressed gas cylinders in service or in<br>storage shall be stabilized or otherwise secured<br>to prevent falling and rolling. |
| Low Flow Setting | Yes | Yes |
| Flow Between Settings | No | No |
| One or Two Knob Operation | One | Two |
Praxair Healthcare Services, Traditional 510(k) September 3, 2013 Grab 'n Go Plus / Grab 'n Go Digital
{2}------------------------------------------------
| Parameter | Grab 'n Go Plus / Grab 'n Go Digital<br>(Praxair Healthcare Services) | LIV (Linde Integrated Valve)<br>(Linde North America, Inc.) |
|----------------------------------------|-----------------------------------------------------------------------|-------------------------------------------------------------|
| DISS Connection | Yes | Yes |
| Filling Port | Yes | Yes |
| Contents Gauge | Active | Active |
| Contents Gauge Type | Bourdon-tube or Transducer, Digital | Bourdon-tube |
| Filters | 4 | 3 |
| Service Pressure Maximum | 3335 PSI (max) | 3000 PSI (max) |
| Regulator Style | Single stage piston | Single stage piston |
| MR Compatibility | Yes, up to 3.0 T. | Yes, up to 3.0 T |
| Hand Grip | 1 grip | 2 grips |
| Access Ports | Yes | Yes |
| Flow Selector | Yes | Yes |
| Color | Green | Green |
| Height<br>(valve-regulator and shroud) | 7.5" | 6.5" |
| Width | 4.25" | 4.75" |
| Cylinder Sizes | D, E | D, E |
| Medical Gases | Oxygen | Oxygen |
| Cylinder Material | Aluminum | Aluminum |
The manufacturer believes that the technological characteristics of the Grab 'n Go system is substantially equivalent to those of the predicate device. Both the proposed and predicate devices are oxygen gas delivery systems comprised of an integrated valveregulator with pressure gauge, flow meter and oxygen gas cylinder with protective shroud all integrated into a single unit. Differences of the Grab 'n Go system include the design and dimensions of the shroud, single knob operation, optional digital gauge and number of internal filters.
#### 7. SUMMARY OF NON-CLINICAL TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE
The Grab 'n Go system complies with the following industry and international standards that are applicable to medical gas regulators:
- · ISO 13485:2003, "Medical devices -- Quality management systems -- Requirements for regulatory purposes"
- · ISO 14971: 2007, "Medical devices -- Application of risk management to medical devices"
- · IEC 60601-1:2012, "Medical electrical equipment Part 1: General requirements 'for basic safety and essential performance"
- · IEC 60601-1-2:2007, "Medical electrical equipment Part. 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests"
{3}------------------------------------------------
- · IEC 60601-1-8:2006, "Medical electrical equipment -- Part 1-8: General requirements for basic safety and essential performance -- Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems"
- · IEC 60529:2001, "Degrees of protection provided by enclosures (IP Code)"
- · IEC 60086-4:2007, "Primary batteries Part 4: Safety of lithium batteries"
- · CGA E-18:2008, "Medical Gas Valve Integrated Pressure Regulators"
- · ISO 10524-3:2005/Amd 1:2013, "Pressure regulators for use with medical gases --Part 3: Pressure regulators integrated with cylinder valves"
- · ISO 9170-1:2008, "Terminal units for medical gases pipeline systems Part 1: terminal units for use with compressed medical gases and vacuum"
- · ISO 11117:2008 / Corr:2009, "Gas cylinders -- Valve protection caps and valve guards -- Design, construction and tests"
- · ASTM G175 03(2011), "Standard Test methods for Evaluating the Ignition Sensitivity and Fault Tolerance of Oxygen Regulators Used for Medical and Emergency Applications"
The Grab 'n Go system is MR-conditional and complies with:
- . ASTM F2052 - 06e1, "Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment"
- ASTM F2213 06(2011), "Standard Test Method for Measurement of . Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment"
- ASTM F2503 13, "Standard Practice for Marking Medical Devices and Other . Items for Safety in the Magnetic Resonance Environment"
- . ASTM F2119-01, "Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants"
#### SUMMARY OF CLINICAL TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE 8.
An end-user evaluation of the Grab 'n Go system was performed to validate the intended use of the device in the intended environment. Data shows that responses from endusers met the established acceptance criteria. End users were able to understand the user manual, properly use the device and understand the alarm functionality. Therefore, the Grab 'n Go system met the validation requirements.
{4}------------------------------------------------
# 9. STATEMENT OF SUBSTANTIAL EQUIVALENCE
The indications for use and overall design of the Grab 'n Go systems are identical to the predicate device. The proposed Grab 'n Go'system complies with internationally recognized standards and meets all established safety and performance criteria. Therefore, the manufacturer believes that the Grab 'n Go system is substantially equivalent to the predicate device and the differences between the two products are minor, and raise no new issues of safety and effectiveness.
1
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes extending from its wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 7, 2014
Praxair Healthcare Services c/o Tina Wu, Ph.D., RAC AptivSolutions 62 Forest Street, Suite 300 Marlborough, MA 01752
Re: K132778
Trade/Device Name: Grab 'n Go Plus I Grab 'n Go Digital Regulation Number: Unclassified Regulation Name: Medical Gas Delivery System Regulatory Class: Unclassified Product Code: ECX Dated: June 2, 2014 Received: June 3, 2014
Dear Dr. Tina Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Dr. Wu
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
MaryFBAner-S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K132778
Device Name
Grab 'n Go Plus / Grab 'n Go Digital
## Indications for Use (Describe)
The Grab 'n Go system is an integrated delivery system intended to provide supplemental oxygen, by Rx only, to neonates, pediatrics, and adults. The device is MR Conditional, and suitable for use during MR imaging for MRI systems up to 3.0 Tesla. The device is intended for limited duration use, such as would be necessary during patient transports.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
HART . THE THE FORFORIEDA USE ONLY . THE CHANTE
Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.