Linshom Respiratory Monitoring Device (LRMD) is indicated for use by healthcare professionals in healthcare facilities, such as procedural areas and recovery rooms, to monitor breathing in adult (at least 22 years of age) patients. LRMD is a non-invasive system that graphically displays temperature changes against time and reports values of respiratory rate and seconds since last breath, along with a trend of tidal volume. LRMD measurements are used as an adjunct to other clinical information sources.
Device Story
LRMD is a thermistor-based respiratory monitor used in healthcare facilities by professionals. It utilizes a Thermistor Sensor Assembly (TSA) to detect temperature changes associated with the breathing cycle; inputs are processed by the Integrated Linshom Module (ILM) Core. The device outputs respiratory rate, seconds since last breath, and a trend of tidal volume via a GUI. It serves as an adjunct to other clinical information sources, assisting clinicians in monitoring patient respiratory status. By providing real-time respiratory data, the device aids in clinical decision-making regarding patient stability, potentially benefiting patients by enabling earlier detection of respiratory distress.
Clinical Evidence
Evidence includes bench testing (lifetime 24hr, mechanical lung, movement simulation) and clinical studies. Clinical data involved 38 volunteers and 12 endoscopy patients for respiratory rate, and 40 volunteers for tidal volume. Performance was compared against capnography and ventilators. Results showed respiratory rate accuracy of ± 1 BPM and tidal volume trend correlation (r²) of 0.97 to a ventilator, supporting substantial equivalence.
Technological Characteristics
Thermistor-based sensing; includes Thermistor Sensor Assembly (TSA) and Integrated Linshom Module (ILM) Core. Complies with IEC 60601-1 (safety) and IEC 60601-1-2 (EMC). Biocompatibility per ISO 10993 and ISO 18562. GUI interface. Operating range: 18.3°C–29.4°C. Respiration range: 5-60 BPM; Tidal volume range: 120-1000mL.
Indications for Use
Indicated for adult patients (≥22 years) in healthcare facilities (procedural areas, recovery rooms) for non-invasive breathing monitoring by healthcare professionals.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
January 30, 2020
Linshom Management LLC Ronen Feldman President 2922 Excelsior Springs Court Ellicott City, Maryland 21042
Re: K190734
Trade/Device Name: Linshom Respiratory Monitoring Device (LRMD) Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: Class II Product Code: BZQ Dated: January 1, 2020 Received: January 2, 2020
#### Dear Ronen Feldman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Michael Ryan Division Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190734
Device Name
Linshom Respiratory Monitoring Device (LRMD)
Indications for Use (Describe)
Linshom Respiratory Monitoring Device (LRMD) is indicated for use by healthcare professionals in healthcare facilities, such as procedural areas and recovery rooms, to monitor breathing in adult (at least 22 years of age) patients.
LRMD is a non-invasive system that graphically displays temperature changes against time and reports values of respiratory rate and seconds since last breath, along with a trend of tidal volume.
LRMD measurements are used as an adjunct to other clinical information sources.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summarv – K190734
(per 21 CFR 807.92)
#### Submitter Information
| Name | Linshom Management LLC |
|----------------|---------------------------------------------------------|
| Address | 2922 Excelsior Springs Court<br>Ellicott City, MD 21042 |
| Phone Number | (410) 480-2700 |
| Contact Person | Ronen Feldman<br>President |
| Date Prepared | January 30, 2020 |
## Device Information
| Trade Name | Linshom Respiratory Monitoring Device (LRMD) |
|----------------|-------------------------------------------------------------------|
| Common Name | Respiratory Monitor |
| Classification | Breathing frequency monitor<br>21 CFR 878.2375 (Product Code BZQ) |
#### Predicate Device Information
| Primary Predicate | ExSpiron, K120087 |
|-------------------|--------------------|
| Reference Device | ReDe Mask, K161953 |
#### Device Description
The Linshom Respiratory Monitoring Device (LRMD) is a medical device designed to monitor a patient's respiratory rate, tidal volume trend and seconds since last breath within a healthcare setting. The LRMD is a thermistor based respiratory monitor which includes the Thermistor Sensor Assembly (TSA), Integrated Linshom Module (ILM) Core, the power supply and software.
## Indications for Use
Linshom Respiratory Monitoring Device (LRMD) is indicated for use by healthcare professionals in healthcare facilities, such as procedural areas and recovery rooms, to monitor breathing in adult (at least 22 years of age) patients.
LRMD is a non-invasive system that graphically displays temperature changes against time and reports values of respiratory rate and seconds since last breath, along with a trend of tidal volume.
LRMD measurements are used as an adjunct to other clinical information sources.
# Technological Characteristics
As shown in Table 1, many technological similarities exist between the subject and predicate devices. The operating mechanism utilized in the LRMD is thermistor-based, similar to the cited reference predicate. Other technological characteristics of LRMD, including software, EMC and ES, and usability aspects, are similar to predicate devices.
{4}------------------------------------------------
| Characteristic | Subject Device<br>LRMD | Predicate Device<br>(K120087)<br>ExSpiron | Reference Device<br>(K161953)<br>ReDe Mask | Comparison |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Linshom Respiratory<br>Monitoring Device<br>(LRMD) | ExSpiron | ReDe Mask | |
| Common Name | Respiratory Monitor | Respiratory Monitor | Breathing Monitor | |
| 510(K) Number | K190734 | K120087 | K161953 | |
| Regulation Classification<br>(Product Code) | 21 CFR 868.2375 (BZQ) | 21 CFR 868.2375 (BZQ)<br>AND<br>21 CFR 868.1850 (BZK) | 21 CFR 879.2375 (PRK) | Same |
| Intended Use | Non-invasive monitoring<br>of respiration and tidal<br>volume trends for adults<br>in healthcare settings | Non-invasive monitoring<br>of respiration and tidal<br>volume for adults in<br>healthcare settings | Non-invasive monitoring<br>of respiration for adults<br>in healthcare settings | Same (ExSpiron) |
| Indications for Use | Linshom Respiratory<br>Monitoring Device (LRMD)<br>is indicated for use by<br>healthcare professionals in<br>healthcare facilities, such<br>as procedural areas and<br>recovery rooms, to<br>monitor breathing in adult<br>(at least 22 years of age)<br>patients.<br>LRMD is a non-invasive<br>system that graphically<br>displays temperature<br>changes against time and<br>reports values of<br>respiratory rate and<br>seconds since last breath,<br>along with a trend of tidal<br>volume.<br>LRMD measurements are<br>used as an adjunct to other<br>clinical information<br>sources. | ExSpiron is indicated for<br>use by healthcare<br>professionals in healthcare<br>facilities, such as post-<br>operative care and critical<br>care units, to monitor<br>breathing in adult (at least<br>21 years of age) patients.<br>ExSpiron is a non-invasive<br>system that graphically<br>displays lung volume<br>against time and reports<br>an approximate value of:<br>· Tidal volume,<br>· Respiratory rate, and<br>· Minute ventilation.<br>ExSpiron measurements<br>are used as an adjunct to<br>other clinical information<br>sources. | The ReDe Mark is indicated<br>for use by healthcare<br>professionals in healthcare<br>facility procedural areas<br>and recovery rooms as an<br>adjunct to monitor<br>breathing in adult patients<br>who are sedated for a<br>diagnostic or therapeutic<br>procedure. The ReDe Mask<br>measures the time period<br>between the current and<br>previous exhalation and<br>illuminates a colored light<br>during the exhalation that<br>reflects the interval of time<br>between breaths. If the<br>interval is less than 7.5<br>seconds, the green light<br>illuminates during<br>exhalation; if the interval is<br>greater than 7.5 seconds<br>but up to 20 seconds, the<br>yellow light illuminates<br>during exhalation; and if<br>the interval between<br>breaths is 20 seconds or<br>longer, the red light flashes<br>continuously. The ReDe<br>Mask is only to be used<br>when supplemental<br>oxygen is provided by the<br>facemask. The ReDe Mask<br>is not a standalone device<br>and is only to be used as<br>an adjunct to pulse<br>oximetry. | Equivalent (ExSpiron)<br>Minor text differences to not suggest<br>a new or difference intended use.<br>Justification: minor differences reflect minor<br>differences in devices but does not change<br>the use, users, or conditions of use |
| Mechanism (General) | Thermistor | Thoracic Bioimpedence | Thermistor | Same (ReDe Mask) |
| Measurements | Respiratory Rate<br>Seconds Since Last<br>Breath<br>Tidal Volume Trend | Respiratory Rate<br>Tidal Volume<br>Minute Volume | Seconds Since Last<br>Breath | Equivalent (All)<br>Justification: these parameters are all related<br>to the breathing cycle |
| Characteristic | Subject Device<br>LRMD | Predicate Device<br>(K120087)<br>ExSpiron | Reference Device<br>(K161953)<br>ReDe Mask | Comparison |
| Communication Method | GUI Interface | GUI Interface | Light Display (Red /<br>Yellow / Green) | Same (ExSpiron) |
| ES / EMC Testing | IEC 60601-1 & 60601-1-2 | IEC 60601-1 & 60601-1-2 | IEC 60601-1 & 60601-1-2 | Same |
| Material Biocompatibility | ISO 10993<br>ISO 18562 | unknown | Tested (mask<br>components) | Equivalent<br>Justification: all devices address<br>biocompatibility risks (albeit different<br>mechanisms) |
| Mask Type | Face mask | ExSpiron 1Xi PadSets | Face mask | Equivalent (ReDe Mask)<br>Justification: LRMD utilizes a legally marketed<br>facemask while the ReDe Mask incorporates<br>its own mask; regardless, both devices attach<br>to the patient face via an oxygen mask |
| Mounting Design | Sensor attached to<br>separate electronics box | Electrodes attached to<br>separate electronics box | Face mask and<br>electronics housing<br>placed on patient's face | Equivalent (ExSpiron)<br>Justification: these are similar concepts just<br>have different electronics |
| Ambient Operating<br>Temperature | 65°F -- 85°F<br>(18.3°C - 29.4°C) | Unknown | 16°C to 31°C | Equivalent (ReDe Mask)<br>Justification: both devices demonstrate safe<br>and effective performance at the extremes of<br>its stated operating range |
| Working Range | Respiration: 5-60 BPM<br>Tidal Volume: 120-<br>1000mL | Respiration: 5-30 BPM<br>Tidal Volume: unknown | Unknown | Equivalent (ExSpiron)<br>Justification: both devices have performance<br>testing that demonstrates safety and<br>effectiveness against a gold standard (e.g.,<br>capnography) for the entire working range |
| Accuracy | Respiration: ± 1 BPM<br>Tidal Volume Trend: 0.97<br>(r² correlation to<br>ventilator) | Respiration: 2%<br>Tidal Volume: 3.9% | Unknown | Equivalent (ExSpiron)<br>Justification: although LRMD reports its<br>accuracy across the working range (via<br>correlation), performance data showed that<br>accuracy is equivalent across working range<br>for both respiration and tidal volume |
| Weight (at point of<br>measurement) | ~15g (Thermistor Sensor<br>Assembly only) | unknown | 40g | Within Range of Existing Products |
| Dimensions (of unit as<br>point of measurement) | 60mm H<br>254mm L<br>158mm W<br>(ILM Core) | unknown | 43.43mm H<br>30.48mm W<br>14.99mm D | Within Range of Existing Products |
| Performance Testing | Lifetime Test (24hr)<br>Breathing Rate Test<br>(Mechanical Lung)<br>Movement Test<br>38 volunteer subjects<br>(for respiratory rate)<br>12 endoscopy subjects<br>(for respiratory rate)<br>40 volunteer subjects<br>(for tidal volume)<br>Human Factors Test<br>Tidal Volume Trend<br>Bench Test | 20 subjects (for all 3<br>measures, compared<br>against spirometer) | Lifetime Test (8hr)<br>Human Factors (for<br>overall use and for light<br>interpretation)<br>38 volunteer subjects<br>(for detection of<br>exhalation and low<br>respiratory rate<br>compared to<br>capnography)<br>50 volunteer subjects<br>(for detection of<br>exhalation and low<br>respiratory rate<br>compared to<br>capnography) | Demonstrates subject device is as<br>safe and as effective as predicate<br>device. |
{5}------------------------------------------------
{6}------------------------------------------------
## Performance Data
Bench and clinical performance data were collected to support a substantial equivalence determination. This testing included mechanical lung data, patient movement simulation, 24-hour testing, clinical comparison of respiratory rate to capnography, and clinical comparison of tidal volume trend to a ventilator. These data suggest the LRMD is as safe and as effective as the identified predicate devices.
## Conclusions
The LRMD and predicate devices have intended use and similar indications, both serving as medical devices to monitor respiration and tidal volume trend. As described above, the minor technological differences between the LRMD and its predicates do not present any new or different issues of safety or effectiveness. Based upon analysis and valid scientific evidence the LRMD is substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.