SunStim Peripheral Nerve Stimulator/Sunstim Plus Peripheral Nerve Stimulator
K153045 · Easy Med Instrument Co., Ltd. · BXN · May 13, 2016 · Anesthesiology
Device Facts
Record ID
K153045
Device Name
SunStim Peripheral Nerve Stimulator/Sunstim Plus Peripheral Nerve Stimulator
Applicant
Easy Med Instrument Co., Ltd.
Product Code
BXN · Anesthesiology
Decision Date
May 13, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.2775
Device Class
Class 2
Indications for Use
SunStim™ Peripheral Nerve Stimulator and SunStim™ Plus Peripheral Nerve Stimulator are a battery-powered device intended for monitoring the magnitude of neuromuscular block in general anesthesia, by delivering an electrical stimulus near a peripheral motor nerve.
Device Story
Battery-powered peripheral nerve stimulator; delivers low electrical direct current (DC) stimulation near peripheral motor nerves to assess anesthetic nerve block levels. Clinician-operated in clinical settings. Features variable current control (0-70 mA into 2K ohm load). Output modes include Twitch, Train-of-Four, 100Hz Tetanus; SunStim Plus adds Double Burst and 50Hz Tetanus. Device includes an output pulse indicator. Uses an 8-bit MCU for timing control to replace logic circuits found in predicate. Output stimulus pulse indicator provides visual feedback to clinician. Enables assessment of pharmacological effect of anesthetic drugs/gases; assists in clinical decision-making regarding neuromuscular blockade management.
Clinical Evidence
Bench testing only. Testing included compliance with AAMI/ANSI ES60601-1 (general safety), AAMI/ANSI/IEC 60601-1-2 (EMC), ISO 14971 (risk management), and IEC 62304 (software life cycle).
Technological Characteristics
Battery-powered; 8-bit MCU for timing control; voltage generator for stimulation output. Output: 0-70 mA into 2K ohm load. Standards: AAMI/ANSI ES60601-1, IEC 60601-1-2, ISO 14971, IEC 62304. Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for monitoring neuromuscular block magnitude in patients undergoing general anesthesia by delivering electrical stimulation near peripheral motor nerves.
Regulatory Classification
Identification
An electrical peripheral nerve stimulator (neuromuscular blockade monitor) is a device used to apply an electrical current to a patient to test the level of pharmacological effect of anesthetic drugs and gases.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a human figure with three faces in profile, one behind the other, with flowing lines representing the body.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 13, 2016
EasyMed Instrument Co., Ltd. Jeffery Wu (Tingjie Wu) General Manager 5/F-6/F, Block A, Gupo Gongmao Building Fengxin Road, Fengxiang Industrial District Daliang, 528300 Shunde, Foshan, Guangdong, China
Re: K153045
Trade/Device Name: SunStim™ Peripheral Nerve Stimulator/SunStim™ Plus Peripheral Nerve Stimulator Regulation Number: 21 CFR 868.2775 Regulation Name: Electrical Peripheral Nerve Stimulator Regulatory Class: Class II Product Code: BXN Dated: February 18, 2016 Received: April 08, 2016
Dear Jeffery Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153045
#### Device Name
SunStim Peripheral Nerve Stimulator / SunStim Plus Peripheral Nerve Stimulator
#### Indications for Use (Describe)
SunStim Peripheral Nerve Stimulator and SunStim Plus Peripheral Nerve Stimulator are a battery-powered device intended for monitoring the magnitude of neuromuscular block in general anesthesia, by delivering an electrical stimulus near a peripheral motor nerve.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"> <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
| <div style="display:flex; align-items:center;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart G) </div> |
| X Prescription Use (Part 21 CFR 801 Subpart D)
|_ | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Section 6.0 510(k) Summary
{4}------------------------------------------------
# 510(k) Summary
Date of submission prepared: 12 May 2016
| Submitter: | EasyMed Instruments Co., Ltd. |
|-------------------|----------------------------------------------------------------|
| Address: | 5/F – 6/F, Block A, Gupo Gongmao Building, Fengxin Road, |
| | Fengxiang Industrial District, Daliang, 528300 Shunde, Foshan, |
| | Guangdong, CHINA |
| Tel: | +86 757 2809 1300 |
| Fax : | +86 757 2772 7868 |
| Official Contact: | Jeffery Wu (Wu Tingjie) |
| E-Mail: | jeffery@easymed.com.cn |
Address of the manufacturing facility: The same as above
# SUBMITTED DEVICE:
| Proprietary or Trade Name: | SunStim™ Peripheral Nerve Stimulator / |
|----------------------------|-------------------------------------------|
| | SunStim™ Plus Peripheral Nerve Stimulator |
| Common/Usual Name: | Peripheral Nerve Stimulator |
| | Battery-Powered Nerve Stimulator |
| Classification Name: | Electrical Peripheral Nerve Stimulator |
| | 21 CFR 868.2775 |
| Product Code: | BXN |
| Device Panel: | Anesthesiology |
| Device Classification: | Class II |
| Prior submission: | No |
# PREDICATE DEVICE:
| Device Name: | EasyMed Instruments Co., LTD. Peripheral Nerve Stimulator |
|----------------|-----------------------------------------------------------|
| Manufacturer: | EasyMed Instruments Co., Ltd. |
| 510(k) Number: | K121743 |
| Product Code: | BXN |
| Trade name: | Microstim Peripheral Nerve Stimulator |
{5}------------------------------------------------
### INTENDED USE:
SunStim™ Peripheral Nerve Stimulator and SunStim™ Plus Peripheral Nerve Stimulator are a battery-powered device intended for monitoring the magnitude of neuromuscular block in general anesthesia, by delivering an electrical stimulus near a peripheral motor nerve.
## DEVICE DESCRIPTION:
SunStim™ Peripheral Nerve Stimulator and SunStim™ Plus Peripheral Nerve Stimulator are battery powered peripheral nerve stimulators which provide low electrical direct current (DC) stimulation in order to determine the level of anesthetic nerve block. The Stimulus Amplitude control dial provides variable current control (0 to 70 mA into a 2K ohm load).
The Output Stimulus Pulse Indicator flashes each time current passes through the patient. Functions include: Double Burst (DBS), Train-of-Four, Twitch, and 100Hz Tetanus. The device offers sufficient output to ensure supramaximal stimulation: from 0 to 70 mA. SunStim™ Peripheral Nerve Stimulator and SunStim™ Plus Peripheral Nerve Stimulator models have the same functions with the following exceptions: the SunStim™ Peripheral Nerve Stimulator functions do not include Double Burst and the 50 Hz Tetanus option.
## TECHNOLOGICAL CHARACTERISTICS
The technical characteristics of the modified Peripheral Nerve Stimulators (i.e. SunStim™ Peripheral Nerve Stimulator and SunStim™ Plus Peripheral Nerve Stimulator) are the same as those of the original legally marketed Peripheral Nerve Stimulators (K121743) in energy source, intended use and functions. Like the predicate device, SunStim™ Peripheral Nerve Stimulator and SunStim™ Plus Peripheral Nerve Stimulator are devices used to apply an electric current to a patient to test the level of pharmacological effect of anesthetic drugs and gases.
The devices, both the predicated ones and modified ones, can be divided into two parts:
- 1. The timing control
This part can be considered to be simplified as a switch controlling the device high voltage output on or off.
In our original design this part uses a simple logical circuit; in our modified design this part uses a simple 8 bits MCU (Micro Control Unit) so that the circuitry can be simplified.
### 2. The voltage generator
This part generates voltage to the setting level, and output the stimulating signal at the time when the output of timing control part is set to on.
This part is not changed in this modification.
The design principle and operational principle of the modified Peripheral Nerve Stimulators (i.e.
{6}------------------------------------------------
SunStim™ Peripheral Nerve Stimulator and SunStim™ Plus Peripheral Nerve Stimulator) are the same as those of original legally marketed Peripheral Nerve Stimulators (K121743). The difference is an MCU is used inside the modified Peripheral Nerve Stimulators to take the place of timing control part of the predicate device.
This modification does not affect the device's intended use or alter the device's fundamental scientific technology.
# PERFORMANCE DATA (e.g. non-clinical tests)
The SunStim™ Peripheral Nerve Stimulator and SunStim™ Plus Peripheral Nerve Stimulator underwent the tests and recognized consensus standards noted in the below table.
| FDA recognition No. | Standard Title |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 19-5 | AAMI/ANSI ES60601-1:2005/(R)2012 And A1:2012, C1:2009/(R)2012<br>And A2:2010/(R)2012 (Consolidated Text) Medical Electrical<br>Equipment - Part 1: General Requirements For Basic Safety And<br>Essential Performance (IEC 60601-1:2005, Mod). (General I (QS/RM)) |
| 19-2 | AAMI / ANSI / IEC 60601-1-2:2007/(R) 2012, Medical Electrical<br>Equipment - Part 1-2: General Requirements For Basic Safety And<br>Essential Performance - Collateral Standard: Electromagnetic<br>Compatibility - Requirements And Tests (Edition 3). (General II<br>(ES/EMC)) |
| 5-40 | ISO 14971: Second Edition 2007-03-01 Medical devices- Application of<br>Risk Management To Medical Devices. (General I (QS/RM)) |
| 13-8 | IEC 62304 First Edition 2006-05, Medical devices software- Software<br>life cycle processes (Software/ Informatics) |
Test results show that the modification for the new device do not affect the safety and effectiveness.
# SUBSTANTIAL EQUIVALENCE CONCLUSION
The modified Peripheral Nerve Stimulators (i.e. SunStim™ Peripheral Nerve Stimulator and SunStim" Plus Peripheral Nerve Stimulator) are of the following similarities to the original legally marketed Peripheral Nerve Stimulators (K121743)
- . The same intended use;
- . Use the same operating principle;
{7}------------------------------------------------
- Incorporate the same basic design principle; .
- The same operating modes; ●
- . The same stimulation parameters
The modified Peripheral Nerve Stimulators (i.e. SunStim™ Peripheral Nerve Stimulator and SunStim™ Plus Peripheral Nerve Stimulator) have the same intended use and similar technological characteristics to the original legally marketed Peripheral Nerve Stimulators (K121743). Thus, the modified Peripheral Nerve Stimulators (i.e. SunStim™ Peripheral Nerve Stimulator and SunStim™ Plus Peripheral Nerve Stimulator) are substantially equivalent to the predicate device.
The modifications do not affect the device's intended use or alter the device's fundamental scientific technology. The modified devices are as safe, as effective and are substantially equivalent to the predicated device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.