EasyMed Instruments Co., LTD. Peripheral Nerve Stimulator is a battery-powered device intended for monitoring the magnitude of neuromuscular blocks in general anesthesia, by delivering an electrical stimuls near a peripheral motor nerve.
Device Story
Battery-powered peripheral nerve stimulator; delivers low electrical direct current (DC) stimulation near peripheral motor nerves to assess anesthetic nerve block levels. Features variable current control (0-70 mA into 2K ohm load); output stimulus pulse indicator; modes include Train-of-Four, Twitch, and 100Hz Tetanus. Used in clinical settings (e.g., OR) by anesthesia providers to ensure supramaximal stimulation. Output allows clinicians to monitor pharmacological effects of anesthetic drugs/gases, aiding clinical decision-making regarding neuromuscular blockade management.
Clinical Evidence
Bench testing only. Testing included electromagnetic compatibility (EN 60601-1-2: 2007), radiated emissions, ESD, radiated immunity, and power-frequency magnetic field immunity. Physical performance testing verified pulse amplitude, pulse width, pulse frequency, and low battery indicator functionality.
Technological Characteristics
Battery-powered electrical stimulator; provides 0-70 mA DC output into 2K ohm load. Modes: Train-of-Four, Twitch, 100Hz Tetanus. Complies with EN 60601-1-2: 2007 for EMC and ISO 14971:2007 for risk management. Manual control via dial.
Indications for Use
Indicated for monitoring neuromuscular block magnitude in patients undergoing general anesthesia by delivering electrical stimulation near peripheral motor nerves.
Regulatory Classification
Identification
An electrical peripheral nerve stimulator (neuromuscular blockade monitor) is a device used to apply an electrical current to a patient to test the level of pharmacological effect of anesthetic drugs and gases.
{0}------------------------------------------------
i21743
# MAR 28 2013
#### EasyMed Instruments Co., LTD.
5/F - 6/F, block A, Gupo Gongmao Building, Fengxin Road, Fengxiang Industrial District, Daliang, Shunde, Foshan, CHINA 528300
Phone: +86 757 2809 1300
## 510(k) SUMMARY 21 CFR 807.92(c)
#### SUBMITTER INFORMATION
510k Owner:
Company Address:
EasyMed Instruments Co., LTD.
+86 757 2809 1300
Wu Tingjie/ Owner
April 17, 2012
March 27, 2013
5/F ~ 6/F, block A, Gupo Gongmao Building, Fengxin Road, Fengxiang Industrial District, Daliang, Shunde, Foshan, CHINA 528300
Company Phone:
Contact Person:
Date Summary Prepared:
Date of Updated Summary:
#### DEVICE INFORMATION
Trade/Proprietary Name:
Common/Usual Names:
Classification Name:
Product Code:
EasyMed Instruments Co., LTD. Peripheral Nerve Stimulator
Peripheral Nerve Stimulator Battery-Powered Nerve Stimulator
Electrical Peripheral Nerve Stimulator 21 CFR 868.2775
BXN
EasyMed Instruments Co., LTD. Peripheral Nerve Stimulator 510(k) Premarket Notification
{1}------------------------------------------------
#### IDENTIFICATION OF PREDICATE DEVICE
| Company | Device | 510k No. |
|-----------------------|----------|----------|
| Life-Tech Intl., Inc. | MiniStim | K913184 |
#### DEVICE DESCRIPTION
The EasyMed Peripheral Nerve Stimulator devices are battery powered peripheral nerve stimulators which provide low electrical direct current (DC) stimulation in order to determine the level of anesthetic nerve block. The Stimulus Amplitude control dial provides variable current control (0 to 70 mA into a 2K ohm load).
The Output Stimulus Pulse Indicator flashes each time current passes through the patient. Functions include: Train-of-Four, Twitch, and 100Hz Tetanus. The device offers sufficient output to ensure supramaximal stimulation: from 0 to 70 mA.
All finished products are tested and must meet all required release specifications before distribution. Testing includes physical testing (e.g. Pulse Amplitude, Pulse Width, Pulse Frequency and Low Battery Voltage Indicator) and visual examination.
#### INTENDED USE
EasyMed Instruments Co., LTD. Peripheral Nerve Stimulator is a battery-powered device intended for monitoring the magnitude of neuromuscular blocks in general anesthesia, by delivering an electrical stimuls near a peripheral motor nerve.
#### TECHNOLOGICAL CHARACTERISTICS
The technical characteristics of the EasyMed Peripheral Nerve Stimulators are similar to those of the predicate device in design, energy source, intended use and function. Like the predicate device, the EasyMed Peripheral Nerve Stimulators are devices used to apply an electric current to a patient to test the level of pharmacological effect of anesthetic drugs and gases. A comparative summary of the EasyMed device to the predicate device is provided in Section 12.0, Table 1.
EasyMed Instruments Co., LTD. Peripheral Nerve Stimulator 510(k) Premarket Notification
{2}------------------------------------------------
#### PERFORMANCE DATA (e.g. non-clinical testing)
The EasyMed Peripheral Nerve Stimulator underwent the tests noted in the below table and results are indicated as well. The full test report can be found in Section 21.0 Appendix -Appendix VIII.
| Test | Test Requirement | Test Method | Result |
|-----------------------------------------------------------------------------------|--------------------|------------------------------------------------|--------|
| Electromagnetic<br>Interference -<br>Radiated Emission<br>(30 MHz to 1 GHz) | EN 60601-1-2: 2007 | EN 55011: 2009<br>+ A1: 2010 | Pass |
| Electromagnetic<br>Susceptibility - ESD | EN 60601-1-2: 2007 | EN 61000-4-2: 2009 | Pass |
| Electromagnetic<br>Susceptibility -<br>Radiated Immunity (90<br>MHz to 2.5 GHz) | EN 60601-1-2: 2007 | EN 61000-4-3: 2006<br>+ A1: 2008<br>+ A2: 2010 | Pass |
| Electromagnetic<br>Susceptibility -<br>Power-frequency magnetic<br>field immunity | EN 60601-1-2: 2007 | EN 61000-4-8:1993<br>+ A1: 2001 | Pass |
### BASIS OF SUBSTANTIAL EQUIVALENCE/CONCLUSION
EasyMed Peripheral Nerve Stimulators have been tested and found to perform as intended. EasyMed Peripheral Nerve Stimulators have been compared to legally cleared predicate device and found to be substantially equivalent. Substantial equivalence testing results can be found in section 12.0.
The EasyMed Peripheral Nerve Stimulator and Stimulator Plus models have the same functions with the following exceptions: the Peripheral Nerve Stimulator functions do not include double burst and 2 tetanus options (50 Hz and 100 Hz).
#### STANDARDS
ISO 14971:2007 Medical Devices - Application of Risk Management to Medical . Devices
. EN 60601-1-2: 2007 (Second edition, 2001), Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance- Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
EasyMed Instruments Co., LTD. Peripheral Nerve Stimulator 510(k) Premarket Notification
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
#### March 28, 2013
SunMed Limited Liability Company C/O Mr. Wu Tingjie EasyMed Instruments Company, Limited 5/F - 6/F, Block A, Gupo Gongmao Building Fengxin Road, Fengxiang Industrial District Daliang, Shunde, Foshan China 528300
Re: K121743
Trade/Device Name: Peripheral Nerve Stimulator Regulation Number: 21 CFR 868.2775 Regulation Name: Electrical Peripheral Nerve Stimulator Regulatory Class: II Product Code: BXN Dated: February 22, 2013 Received: February 25, 2013
Dear Mr. Tingjie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (2) CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Image /page/4/Picture/7 description: The image shows the words Kwame Ulmer-S in a stylized font. The letters are black and bolded, and the words are stacked on top of each other. The letters are slightly distorted, and the image has a grainy texture. There is also a logo to the right of the words.
for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use
510(k) Number (if known):
Device Name:
Peripheral Nerve Stimulator
K121743
Indications for Use:
The EasyMed InstrumentsCo., LTD. Peripheral Nerve Stimulator is a battery -powered device intended for monitoring the magnitude of neuromuscular blocks in general anesthesia; by delivering an electrical stimulus near a peripheral motor nerve.
Prescription Use ਮ (Part 21 CFR 801 Subpart D)
+
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Lester W. Schultheis Jr 2013.03.27 12:42:42:44:204'00'
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hosphal
Division of Anesthesiology, General Devices Division of Allestheolorology.
Infection Control, Dental Devices
Kreizer
510(k) Number:_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.