REPROCESSED PULSE OXIMETER SENSORS

K012677 · Sterilmed, Inc. · NLF · Mar 28, 2003 · Cardiovascular

Device Facts

Record IDK012677
Device NameREPROCESSED PULSE OXIMETER SENSORS
ApplicantSterilmed, Inc.
Product CodeNLF · Cardiovascular
Decision DateMar 28, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2700
Device ClassClass 2

Indications for Use

Reprocessed pulse oximeter sensors are used when continuous external monitoring of arterial oxygen saturation and pulse rate is required.

Device Story

Reprocessed pulse oximeter sensors; electro-optical sensors measuring light absorption of functional arterial hemoglobin. Components: two LEDs (light sources), one photodiode (receiver). Sensor positioned with LEDs and photodiode opposing across tissue. Connected via cable to pulse oximeter monitor. Provides continuous, non-invasive, self-calibrated measurements of oxygen saturation and pulse rate. Used in clinical settings; operated by healthcare professionals. Reprocessing involves cleaning and sterilization of single-use devices to restore functionality equivalent to original predicate sensors.

Clinical Evidence

Clinical study and bench testing performed on representative samples. Process validation conducted for cleaning, sterilization, and packaging. Manufacturing process includes 100% visual and functional testing of all units to ensure performance equivalence to predicate devices.

Technological Characteristics

Electro-optical sensor; two LEDs and one photodiode. Non-invasive, external application. Connects to external pulse oximeter monitor via cable. Reprocessed single-use device.

Indications for Use

Indicated for patients requiring continuous external monitoring of arterial oxygen saturation and pulse rate.

Regulatory Classification

Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 1 2004 NOV Dr. Bruce Lester Vice President of Research and Development SterilMed, Incorporated 11400 73Rd. Avenue North Minneapolis, Minnesota 55369 Re: K012677 - Supplemental Validation Submission Trade/Device Name: See Enclosed List Regulation Number: 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: NLF Dated: August 13, 2001 Received: August 14, 2001 Dear Dr. Lester: The above-referenced premarket notification (510(k)) was cleared by the Office of Device Evaluation (ODE) on March 28, 2003. We have received your supplemental validation data as required for reprocessed single-use devices by the Medical Device User Fee and Modernization Act of 2002. After reviewing your supplemental validation data, we have determined the devices listed in the enclosure accompanying this letter are substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market these devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your devices are classified (see above) into either class II (Special Controls) or class III (PMA) they may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your devices in the Federal Register. {2}------------------------------------------------ Page 2 - Dr. Bruce Lester Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your devices comply with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's applicable requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. The FDA finding of substantial equivalence of your devices to legally marketed predicate devices results in classification for your devices and thus, permits you to legally market the devices. This letter will allow you to continue marketing the devices listed in the enclosure accompanying this letter. If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Page 3 - Dr. Bruce Lester ## Trade/Device Names SterilMed Reprocessed Nellcor Pulse Oximeter Oxisensor II Sensor N-25 SterilMed Reprocessed Nellcor Pulse Oximeter Oxisensor II Sensor N-25LF {4}------------------------------------------------ # Indications for use Page 510(K) Number Kolze 77 Device Name: Reprocessed Pulse Oximeter Sensors ### Indications for Use: Reprocessed Pulse Oximeter Sensors are intended for use when continuous external monitoring of arterial oxygen saturation and pulse rate are required. Concurrence of CDRH, Office of Device Evaluation (ODE) \$\underset{-}{\checkmark}\$ PRESCRIPTION SAWesterman on of Anesthesiology, General Hospital Infection Control, Dental Devi 510(k) Number: K012677 SterilMed, Inc. Pulse Oximeter Probes Confidential Premarket Notification {5}------------------------------------------------ # SECTION 2. SUMMARY AND CERTIFICATION #### 510(k) Summary A. | Submitter: | SterilMed, Inc. | |-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Patrick Fleischhacker<br>Submitted by SterilMed, Inc.<br>11400 73rd Avenue North<br>Minneapolis, MN 55369<br>Ph: 888-856-4870<br>Fax: 763-488-3350 | | Date Prepared: | August 13, 2001 | | Trade Name: | SterilMed Reprocessed Pulse Oximeter Sensors | | Classification Name:<br>and Number: | Class II, 21 CFR 870.2700 | | Product Code: | DQA | | Predicate Device(s): | The reprocessed pulse oximeter sensors are substantially<br>equivalent to the Nellcor Oxisensors™, which are<br>compatible with the Nellcor N-200 pulse oximeter<br>(K863784). | | Device Description: | The reprocessed pulse oximeter sensor is an electro-optical<br>sensor that uses an optical means to determine the light<br>absorption of functional arterial hemoglobin. The sensor<br>contains three optical components: two light emitting<br>diodes (LED's) that serve as light sources, and one<br>photodiode, that acts as a light receiver. The oximeter<br>sensor is positioned so that the LED's and photodiode<br>oppose one another across the tissue. The sensor is<br>connected via cable to a pulse oximeter, which provides<br>continuous non-invasive, self-calibrated measurements of<br>both oxygen saturation of functional hemoglobin and pulse<br>rate. Please note that this submission only pertains to the<br>sensor. It does not pertain to the pulse oximeter or<br>connecting cable. | | Intended Use: | Reprocessed pulse oximeter sensors are used when<br>continuous external monitoring of arterial oxygen<br>saturation and pulse rate is required. | | Functional and<br>Safety Testing: | Representative samples of reprocessed sensors underwent<br>bench testing and a clinical study to verify functional<br>characteristics which are substantially equivalent to the<br>predicate devices'. Process validation testing was done to<br>validate the cleaning and sterilization procedures as well as<br>the device's packaging. In addition, the manufacturing<br>process includes visual and functional testing of all<br>products produced. | | Conclusion: | The pulse oximeter sensors reprocessed by SterilMed are<br>substantially equivalent the Nellcor Oxisensors™ which<br>are compatible with the Nellcor N-200 pulse oximeter.<br>This conclusion is based upon the fact that the reprocessed<br>sensors are substantially equivalent to their predicate<br>devices in terms of functional design, materials, indications<br>for use, and methods of construction. | --------- {6}------------------------------------------------ ・
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...