The pulse oximeter is indicated for the noninvasive spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult and pediatric (≥10Kg) patients during both no motion and motion conditions, and for patients who are well or poorly perfused. Pulse Oximeter is intended for hospital-type facilities, home environments.
Device Story
Portable, battery-powered digit pulse oximeter; measures light absorption (red/infrared) through perfused tissue; calculates SpO2, pulse rate, and perfusion index. Used in hospitals and home environments for spot checks. Output displayed on OLED/LED screen as numerical values, plethysmogram, and bargraph. Provides real-time physiological data to clinicians or patients to assess oxygenation status; supports clinical decision-making regarding patient respiratory health.
Clinical Evidence
No new clinical data provided. Substantial equivalence is based on the device being identical to the predicate. Bench testing confirmed compliance with ISO 80601-2-61 for SpO2 and pulse rate accuracy under motion, no-motion, and low-perfusion conditions.
Technological Characteristics
Plastic enclosure; 62mm x 35mm x 31mm; <75g; powered by two AAA alkaline batteries. Sensing via red/infrared light absorption. Standards: ISO 10993-1/5/10 (biocompatibility), IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60601-1-11 (home use), ISO 80601-2-61 (pulse oximeter performance).
Indications for Use
Indicated for noninvasive spot checking of SpO2 and pulse rate in adult and pediatric (≥10Kg) patients, including those with or without motion and those who are well or poorly perfused.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 19, 2022
Unimed Medical Supplies, Inc. Hui Zhang General Manager Bld#8, Nangang 3rd Industrial Park Tangtou, Shiyan Shenzhen, Guangdong 518108 China
### Re: K222213
Trade/Device Name: Pulse Oximeter (UC-200, UC-201, UC-202, UC-203, UC-204, UC-100, UC-101, UC-102, UC-103, UC-104) Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: November 10, 2022 Received: November 10, 2022
Dear Hui Zhang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James J. Lee -S
James J. Lee, Ph.D. Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K222213
Device Name Pulse Oximeter
Indications for Use (Describe)
The pulse oximeter is indicated for the noninvasive spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult and pediatic (≥10Kg) patients during both no motion and for patients who are well or poorly perfused.
Pulse Oximeter is intended for hospitals, hospital-type facilities, home environments.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Unimed medical supplies. The logo includes the text "A brand of Xinwell" above the word "Unimed" in gray letters. To the right of the word "Unimed" is a red oval with a white design inside.
# 510(k) Summary
In accordance with 21 CFR 807.92 the 510(k) Summary for the Unimed pulse oximeters are provided below.
#### 1. SUBMITTER
| Applicant: | Unimed Medical Supplies, Inc.<br>Bld #8,Nangang 3rd Industrial Park, Tangtou, Shiyan, Baoan<br>Shenzhen Guangdong, CHINA 518108<br>Tel: +86 755 2669 5165 |
|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Contact Person: HuiZhang<br>Title: General Manager<br>Phone: +86 15013808983<br>Tel: +86 755 2669 5165<br>E-mail: FrankZhang@unimed.cn |
Date Prepared: June 24, 2022
#### 2. DEVICE
510(k) Number: K222213
- Device Trade Name: Oximeter (including UC-200, UC-201, Pulse UC-202, UC-203, UC-204, UC-100, UC-101, UC-102, UC-103, UC-104)
- Classification Name: 21 CFR 870.2700, Oximeter
- Regulatory Class: Class II
- Product Code, Panel: DQA, Anesthesiology
- Reason for Submission: New Application. No prior submission associated with the current submission.
- Predicate Device: Nariq Bio-Medical Pulse Oximeter, K211632
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Unimed medical supplies. The logo includes the text "A brand of Xinwell" in smaller font above the word "UNIMED" in a larger, gray font. Below "UNIMED" is the text "medical supplies" also in gray. To the right of the text is a red circle containing a white stylized "UM".
### DEVICE DESCRIPTION 3.
The Pulse Oximeter is a small, lightweight, portable, digit pulse oximeter that displays numerical values for functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR) and perfusion index (PI) by measuring the absorption of red and infrared light passing through perfused tissue. It indicated for adult and pediatric (≥10Kg) patients during both no motion and motion conditions, and for patients who are well or poorly perfused.
The device consists of probe, electronic circuits, OLED/LED display (differentiated by models) and plastic housing which powered by two alkaline AAA batteries.
The function difference of models is listed as follows:
### Function Difference of Models
Abbreviation description:
- A sup: support function
- A opt: optional function
- > n/a: disable/not support function
| Function | UC-200 | UC-201 | UC-202 | UC-203 | UC-204 | UC-100 | UC-101 | UC-102 | UC-103 | UC-104 |
|-------------------------|--------|--------|--------|--------|--------|--------|--------|--------|--------|--------|
| Display Type | OLED | OLED | OLED | OLED | OLED | LED | LED | LED | LED | LED |
| Key Function | sup | sup | sup | sup | sup | sup | sup | sup | sup | sup |
| Automatic Shut-Down | sup | sup | sup | sup | sup | sup | sup | sup | sup | sup |
| Low Power Tip | sup | sup | sup | sup | sup | sup | sup | sup | sup | sup |
| Low Power Shut-Down | sup | sup | sup | sup | sup | sup | sup | sup | sup | sup |
| Display Light Intensity | opt | opt | opt | opt | opt | opt | opt | opt | opt | opt |
| Direction Rotation | opt | opt | opt | opt | opt | n/a | n/a | n/a | n/a | n/a |
| SpO2 Parameter | sup | sup | sup | sup | sup | sup | sup | sup | sup | sup |
| Pulse Parameter | sup | sup | sup | sup | sup | sup | sup | sup | sup | sup |
| Perfusion Parameter | opt | opt | opt | opt | opt | opt | opt | opt | opt | opt |
| Bargraph | sup | sup | sup | sup | sup | sup | sup | sup | sup | sup |
| Plethysmogram | sup | sup | sup | sup | sup | n/a | n/a | n/a | n/a | n/a |
| Battery Display | sup | sup | sup | sup | sup | n/a | n/a | n/a | n/a | n/a |
| User Menu Setup | opt | opt | opt | opt | opt | n/a | n/a | n/a | n/a | n/a |
| Trend Graph | opt | opt | opt | opt | opt | n/a | n/a | n/a | n/a | n/a |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Unimed medical supplies. The logo includes the text "A brand of Xinwell" in smaller font above the word "UNIMED" in a larger, gray font. Below "UNIMED" is the text "medical supplies" in a smaller, gray font. To the right of the text is a red circle with a white, stylized "UM" inside.
#### INTENDED USE/INDICATIONS FOR USE 4.
The pulse oximeter is indicated for the noninvasive spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult and pediatric (≥10Kg) patients during both no motion and motion conditions, and for patients who are well or poorly perfused.
Pulse Oximeter is intended for hospital-type facilities, home environments.
### SUBSTANTIAL EQUIVALENCE 5.
### Device Specification
| Feature | Specification | Models |
|-------------------------------------------------------|------------------------------------------------|------------------------------------------|
| Display | | |
| Display Range | Oxygen Saturation (SpO2): 0-100% | All models |
| | Pulse Rate (PR): 25-250 beats per minute (BPM) | All models |
| | Perfusion Index (PI): 0.02-20% | All models |
| Display Waveform | Plethysmogram | UC-200/ UC-201/ UC-202/<br>UC-203/UC-204 |
| | Bargraph | All models |
| Display Resolution | SpO2: 1% | All models |
| | PR: 1 BPM | |
| Measurement Accuracy in Accordance with ISO80601-2-61 | | |
| SpO2, No Motion | 70 – 100%, 2%, ARMS, adults/pediatrics | All models |
| SpO2, Motion | 70 – 100%, 3% ARMS, adults/pediatrics | All models |
| SpO2, Low Perfusion | 70 – 100%, 2%, ARMS, adults/pediatrics | UC-200/ UC-100 |
| Pulse Rate, No Motion | 25 – 250 BPM, 2 BPM ARMS, adults/pediatrics | All models |
| Pulse Rate, Motion | 25 – 250 BPM, 4 BPM ARMS, adults/pediatrics | All models |
| Pulse Rate, Low Perfusion | 25 – 250 BPM, 2 BPM ARMS, adults/pediatrics | UC-200/ UC-100 |
| Power | | |
| Internal battery | Alkaline “AAA” batteries | All models |
The pulse oximeter specifications are as follows:
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for Unimed medical supplies, a brand of Xinwell. The word "Unimed" is in large, gray, sans-serif font. Below it, the words "medical supplies" are in a smaller, lighter font. To the right of the word "Unimed" is a red oval with a white, stylized "um" inside.
| Mechanical | | |
|----------------------------|----------------------------------|------------|
| Enclosure Material | Plastic | All models |
| Dimensions/Weight | 62mm × 35mm × 31mm | All models |
| Weight | Less than 75g (contain battery) | All models |
| Environmental | | |
| Operating Temperature | 0°C to +40°C, ambient humidity | All models |
| Storage Temperature | -20°C to +60°C, ambient humidity | All models |
| Operating/Storage Humidity | 15% to 95%, non-condensing | All models |
| Atmospheric Pressure | 79.5kPa~107.4kPa | All models |
| Mode of Operation | | |
| Mode of Operation | Spot Check | All models |
In conclusion, the minor differences of the models do not change the fundamental intended use of the pulse oximeter.
### Technological Comparison
The table below compares the key technological feature of the subject devices to the predicate device (the Narig Bio-Medical's Pulse Oximeter, K211632).
| Item | Predicate Devices(K211632) | Subject Devices | Comparison |
|-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|------------|
| Display | | | |
| Display Type | OLED display (FRO-200, FRO-201, FRO-202, FRO-203, FRO-204)<br>LED display (FRO-100, FRO-101, FRO-102, FRO-103, FRO-104) | OLED display (UC-200, UC-201, UC-202, UC-203, UC-204)<br>LED display (UC-100, UC-101, UC-102, UC-103, UC-104) | Different |
| Display Range | Oxygen Saturation (SpO2): 0-100%<br>Pulse Rate (PR): 25-250 beats per minute (BPM)<br>Perfusion Index (PI): 0.02-20% | Oxygen Saturation (SpO2): 0-100%<br>Pulse Rate (PR): 25-250 beats per minute (BPM)<br>Perfusion Index (PI): 0.02-20% | Same |
| Display Waveform | Plethysmogram (FRO-200, FRO-201, FRO-202, FRO-203, FRO-204)<br>Bargraph | Plethysmogram (UC-200/ UC-201/ UC-202/UC-203/UC-204)<br>Bargraph | Same |
| Display Resolution | SpO2: 1%<br>PR: 1 BPM | SpO2: 1%<br>PR: 1 BPM | Same |
| Measurement Accuracy in Accordance with ISO80601-2-61 | | | |
| | | | |
| SpO2, No Motion | 70 - 100%, 2%, ARMS, adults/pediatrics | 70 - 100%, 2%, ARMS, adults/pediatrics | Same |
| SpO2, Motion | 70 - 100%, 3% ARMS, adults/pediatrics | 70 - 100%, 3% ARMS, adults/pediatrics | Same |
| SpO2, Low Perfusion | Only for <i>FRO-200 / FRO-100</i> | Only for <i>UC-200 / UC-100</i> | Different |
| | 70 - 100%, 2%, ARMS, adults/pediatrics | 70 - 100%, 2%, ARMS, adults/pediatrics | |
| Pulse Rate, No Motion | 25 - 250 BPM, 2 BPM ARMS, adults / | 25 - 250 BPM, 2 BPM ARMS, adults / | Same |
| | pediatrics | pediatrics | |
| Pulse Rate, Motion | 25 - 250 BPM, 4 BPM ARMS, adults / | 25 - 250 BPM, 4 BPM ARMS, adults / | Same |
| | pediatrics | pediatrics | |
| Pulse Rate, Low Perfusion | Only for <i>FRO-200/ FRO-100</i> | Only for <i>UC-200/ UC-100</i> | Different |
| | 25 - 250 BPM, 2 BPM ARMS, adults / | 25 - 250 BPM, 2 BPM ARMS, adults / | |
| | pediatrics | pediatrics | |
| Power | | | |
| Internal battery | Alkaline "AAA" batteries | Alkaline "AAA" batteries | Same |
| Mechanical | | | |
| Enclosure Material | Plastic | Plastic | Same |
| Dimensions/Weight | 62mm × 35mm × 31mm | 62mm × 35mm × 31mm | Same |
| Weight | Less than 75g (contain battery) | Less than 75g (contain battery) | Same |
| Environmental | | | |
| Operating Temperature | 0°C to +40°C, ambient humidity | 0°C to +40°C, ambient humidity | Same |
| Storage Temperature | -20°C to +60°C, ambient humidity | -20°C to +60°C, ambient humidity | Same |
| Operating/Storage Humidity | 15% to 95%, non-condensing | 15% to 95%, non-condensing | Same |
| Atmospheric Pressure | 79.5kPa-107.4kPa | 79.5kPa-107.4kPa | Same |
| Mode of Operation | | | |
| Mode of Operation | Spot Check | Spot Check | Same |
| Compliance Standards | | | |
| Bio-compatibility | ISO 10993-1 | ISO 10993-1 | Same |
| | ISO 10993-5 | ISO 10993-5 | |
| | ISO 10993-10 | ISO 10993-10 | |
| Electrical Safety | IEC 60601-1 | IEC 60601-1 | Same |
| | IEC 60601-1-11 | IEC 60601-1-11 | |
| EMC | IEC 60601-1-2 | IEC 60601-1-2 | Same |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Unimed medical supplies. The logo includes the text "A brand of Xinwell" in smaller font above the word "Unimed" in a larger, bold font. To the right of "Unimed" is a red oval shape with a white "U" inside. Below "Unimed" is the text "medical supplies" in a smaller font.
Image /page/7/Picture/1 description: The image shows a red oval with white letters inside. The letters appear to be 'u' and 'm' intertwined. The logo is simple and modern, with a focus on the color red.
As seen in the comparison tables, the subject and predicate devices have same design principle, same design features and performance specifications. In conclusion, the different technological characteristics between the subject and predicate devices will not impact the substantial equivalence.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Unimed medical supplies, a brand of Xinwell. The word "Unimed" is written in a gray sans-serif font, with the words "medical supplies" written in a smaller font below. To the right of the word "Unimed" is a red oval with a white design inside.
#### PERFORMANCE DATA 6.
The devices and the predicate are identical devices. The predicate devices have been conducted functional and system level test. The functional and system level testing showed that the devices continue to meet specifications and meets relevant consensus standards.
# Biocompatibility Testing
The devices and the predicate are identical devices. Testing of the predicate devices have been completed per FDA's "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" guidance document and the requirements of ISO 10993-1:2018.
# Software Verification and Validation Testing
The predicate devices were tested for Software verification and validation, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Verification of the pulse oximeter was conducted to ensure that the product works as designed. Validation was conducted to check the design and performance of the product.
# Electromagnetic Compatibility and Electrical Safety
The pulse oximeter was assessed for conformity with the relevant requirements of the following standards and found to comply:
- ANSI/AAMI ES 60601-1:2005/(R) 2012 and A1:2012, C1:2009/(R) 2012 and . A2:2010/(R) 2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
- . IEC 60601-1-2:2014 (Fourth Edition) Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: electromagnetic disturbances - Requirements and tests.
- . IEC 60601-1-11:2015 Medical electrical equipment – Part 1-11: General requirements for basic safety and essential performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- . IEC 62471: 2006 Photobiological Safety of Lamps and Lamp Systems.
# Bench Testing
The device and the predicate are identical devices. The predicate devices have been conducted functional and system level testing. The results of the bench testing show that the subject device meets its accuracy specification and meet relevant consensus standards.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Unimed medical supplies. The logo includes the text "A brand of Xinwell" in smaller font above the word "Unimed" in a larger, bolder font. To the right of the text is a red oval containing a white stylized design.
- IEC 60601-1-6:2010, AMD1:2013 Medical electrical equipment-part 1-6: general . requirements for basic safety and essential performance- collateral standard: usability.
- IEC 62366-1:2015 Medical devices Part 1: Application of usability engineering to ● medical devices.
- . ISO 80601-2-61:2011 Medical electrical equipment - part 2-61: particular requirements for basic safety and essential performance of pulse oximeter equipment.
- EN ISO 15223-1:2016 Medical devices. Symbols to be used with medical device labels, . labelling and information to be supplied. General requirements.
# Cleaning Validation
Cleaning and disinfection validation testing of the predicate devices were conducted as recommended by FDA's Guidance for Industry and FDA Staff, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling". The test results meet relevant standards. The device and the predicate are identical devices.
# Clinical Data
Clinical testing of the predicate devices have shown that pulse oximeters meet relevant consensus standards.
- . ISO 80601-2-61:2017 Medical electrical equipment – Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment.
### SUBSTANTIAL EQUIVALENCE CONCLUSION 7.
The subject and predicate devices are exactly the same. The pulse oximeter is identical to the cleared version and is not modified. Therefore, the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.