K213515 · Wesper, Inc. · DQA · Jun 10, 2022 · Cardiovascular
Device Facts
Record ID
K213515
Device Name
WesperO2
Applicant
Wesper, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Jun 10, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Indications for Use
WesperO2 is a finger-worn pulse oximeter in measuring, displaying, storing and transmitting functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for adult patients. It is intended for spot-check and/or continuous data collection, and not continuous monitoring. It can be used in sleep labs and home use.
Device Story
WesperO2 is a finger-worn pulse oximeter for adult patients; used in home or sleep lab settings. Device collects reflected light from finger skin to form PPG waveforms; extracts SpO2 and pulse rate. Hardware consists of battery-powered PCB in silicone thimble-style housing with two LEDs and photodiode. Device transmits data via Bluetooth to companion mobile app for real-time display, recording, and sharing with healthcare providers. Healthcare providers use output to assess patient oxygenation and pulse rate; aids in clinical decision-making for sleep-related conditions. Benefits include convenient, non-invasive, continuous data collection for patient monitoring.
Clinical Evidence
Clinical study (N=223 for SpO2, N=235 for PR) compared WesperO2 to arterial blood CO-Oximetry and reference oximeter. SpO2 accuracy (ARMS) was 2.56% (range 70-100%), meeting the <3.5% specification. Pulse rate accuracy (ARMS) was 2.2 BPM (range 59-122 BPM), meeting the <3 BPM specification. Bench testing validated the full range of 30-230 BPM.
Technological Characteristics
Finger-worn pulse oximeter; silicone housing; optical sensor (2 LEDs, photodiode). Sensing principle: PPG via reflected light. Connectivity: Bluetooth to mobile app. Power: Battery-powered PCB. Standards: IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, ISO 80601-2-61, IEEE/ANSI C63.27. Biocompatibility: ISO 10993-1 compliant for 24h skin contact.
Indications for Use
Indicated for adult patients for measurement, display, storage, and transmission of functional oxygen saturation (SpO2) and pulse rate. Intended for spot-check or continuous data collection in sleep labs or home settings. Not for continuous monitoring.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 10, 2022
Wesper Inc. Amir Reuveny CEO 234 5th Ave New York, New York 10001
Re: K213515
Trade/Device Name: WesperO2 Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: May 11, 2022 Received: May 12, 2022
Dear Amir Reuveny:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of Sleep Disordered Breathing. Respiratory and Anesthsia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K213515
Device Name WesperO2
#### Indications for Use (Describe)
WesperO2 is a finger-worn pulse oximeter in measuring, displaying, storing and transmitting functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for adult patients. It is intended for spot-check and/or continuous data collection, and not continuous monitoring. It can be used in sleep labs and home use.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the word "WESPER" in black font. To the left of the word is a gray circle that is partially filled in, resembling a crescent moon. The font is simple and sans-serif, giving the logo a clean and modern look. The overall design is minimalist and professional.
# Section 5. 510(k) Summary
WesperO2 Pulse Oximeter Date Prepared: October 29, 2021
#### Name of Device and Sponsor Information
Device Name: WesperO2 Sponsor Name: Wesper Inc. (formerly Tatch Inc.) Sponsor Address: 234 5th Ave., New York, USA, 10001 Phone: 917-841-4830 Contact Person: Amir Reuveny
#### Classification Name
Classification Name: Oximeter Product Code: DQA Regulation: 21 CFR 870.2700 Device Class: Class II
#### Predicate Device
Name: Oxiband (Checkme) O2 Pulse Oximeter K-Number: K191088 Classification Name: Oximeter Product Code: DQA Regulation: 21 CFR 870.2700 Device Class: Class II Sponsor Name: Shenzhen Viatom Technology Co., Ltd.
#### Indications for Use
WesperO2 is a finger-worn pulse oximeter indicated for use in measuring, displaying, storing and transmitting functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for adult patients. It is intended for spot-check and/or continuous data collection, and not continuous monitoring. It can be used in sleep labs and home use.
#### Device Description
WesperO2 ("the device") is a pulse oximeter, meant to be worn on the finger of an adult patient over the age of 21. The device is intended for spot checking or continuous collection of functional oxygen saturation of arterial hemoglobin by way of sensing peripheral capillary perfusion ("SpO2") and Pulse Rate. It is intended to be used in the patient's home or in a sleep lab.
The device's core technology is based on collecting reflected light from the intact skin of the patient's finger to form Photoplethysmography ("PPG") waveforms. These waveforms can then be used to extract a measurement of SpO2 and Pulse Rate.
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the word "WESPER" in a sans-serif font. To the left of the word is a gray circle that is partially filled in, resembling a crescent moon. The text is black and the background is white. The image is simple and clean, with a focus on the text and the circular graphic.
The sensing component is carried by an electronics assembly, which is comprised of a batterypowered Printed Circuit Board ("PCB"). The electronics assembly is housed inside of a soft, flexible casing, which encapsulates the entirety of the assembly, except for the actual optical front-end of the sensing component, which exposes an array, consisting of two LEDs and a photodiode. The entire assembly is referred to as "the WesperO2 probe."
The silicone housing has a "thimble-style" form factor, such that the patient can wear it on their fingertip, with the optical sensor pointed at the pad of their finger. The probe is then meant to create a snuq, yet comfortable fit to ensure consistent positioning of the optical sensor throughout the usage period.
The probe has a Bluetooth interface through which it sends SpO2 and Pulse Rate measurements to a companion mobile application ("WesperO2 app"), which resides on the patient's mobile device. The application can both record the data and display it in real-time to the patient. Recordings are saved on the mobile device and can be shared later through the application with a healthcare provider over the internet.
#### Performance Data
#### Software
Wesper's software development life cycle is in accordance with the following standards:
- FDA Guidance: General Principles of Software Validation, January 2002 ●
- FDA Guidance: Guidance for the Content of Premarket Submissions for Software ● Contained in Medical Devices, May 2005
- FDA Guidance: Off-The-Shelf Software Use in Medical Devices, September 2019 ●
- IEC 62304:2006+AMD1:2015 CSV, Medical Device Software Software Life-Cycle ● Processes
- ISO 14971:2019, Medical Devices Application of Risk Management to Medical Devices ●
Based on this process, Wesper has established it has a Moderate Level of Concern for its software used in the WesperO2 pulse oximeter. Therefore, Wesper has performed a detailed Risk Analysis and developed a Software Requirement Specification and Software Detailed Design which were verified and validated and shown to have full traceability back to all risk mitigations, which reduce the level of residual risk in the software to an acceptable level. Wesper has also performed Off-The-Shelf (OTS) Software validation.
In addition, Wesper has performed a full Cybersecurity Threat Analysis based on recommendations provided in FDA Guidance: "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices," October 2014.
#### Electromagnetic Compatibility (EMC) and Electrical Safety
The following standards were met for EMC and Electrical Safety:
- IEC 60601-1:2005+AMD 2012, 3rd Ed. Part 1: General Requirements for Basic Safety . and Essential Performance
- IEC 60601-1, Part 11:2015 General requirements for basic safety and essential ● performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
{5}------------------------------------------------
# OWESPER
- IEC 60601-1-2:2014 General requirements for basic safety and essential performance -● Collateral Standard: Electromagnetic disturbances - Requirements and tests
- ISO 80601-2-61:2017 - Particular requirements for basic safety and essential performance of pulse oximeter equipment [selected sections pertaining to safety]
- IEEE/ANSI C63.27-2017 American National Standard for Evaluation of Wireless . Coexistence
#### Biocompatibility
All skin-contacting materials in the WesperO2 probe have been tested to comply with ISO 10993-1 with conditions which meet Wesper's intended use of up to 24 hours at 37°C on intact skin.
### Performance Testing - Bench
Validation of the Range of Heart Rate Measurement: Testing was performed to ensure that the WesperO2 pulse oximeter measures heart rate accurately over its targeted range. One heart rate measurement accuracy test was applied to three independently generated datasets. The test passed for every dataset. Based on these three successful tests. WesperO2 is validated for measuring heart rate across its target range of 30 to 230 BPM. This full range was not observed in clinical testing of the device. Therefore, bench testing was performed to supplement clinical testing in demonstrating the accuracy across the full tarqet range.
### Performance Testing - Animal
No animal testing was conducted for this 510(k) submission.
### Performance Testing - Clinical
An SpO2 accuracy comparison was conducted as part of the final validation of the WesperO2 pulse oximeter. The study was conducted in accordance with the following standards:
- ISO 80601-2-61:2019, Medical electrical equipment Particular requirements for basic . safety and essential performance of pulse oximeter equipment
- FDA-2007-D-0205, Pulse Oximeters Premarket Notification Submissions [510(k)s] ● Guidance for Industry and Food and Drug Administration Staff
The objective of the study was to validate the SpO2 accuracy of the WesperO2 pulse oximeter over the range of 70-100% SaQ2 as compared to arterial blood sample CO-Oximetry. It was expected that the Accuracy Root Mean Square (ARMS) performance of the oximeter would meet the required specification of ARMS of 3.5% or less.
A secondary goal was to evaluate the Pulse Rate (PR) performance simultaneously collected over the SpO2 range covered. The PR ARMs performance of the WesperO2 pulse oximeter was expected to meet a specification of 3 Beats per Minute (BPM).
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the word "WESPER" in black, bold, sans-serif font. To the left of the word is a gray circle that is partially filled in, resembling a crescent moon. The word is horizontally aligned and centered.
#### Table 1: Summary of SpO2 Accuracy Results
| Comparison to Reference CO-Oximetry<br>(functional SaO2) | SpO2<br>Range | SpO2 ARMS | ARMS Spec <<br>3.5% |
|----------------------------------------------------------|---------------|--------------|---------------------|
| WesperO2 | 70-100% | 2.56 (N=223) | Pass |
#### Table 2: Summary of Pulse Rate Accuracy Data
| Comparison to Reference Oximeter<br>Pulse Rate | Pulse Rate<br>Range | Pulse Rate<br>ARMS | ARMS Spec <<br>3BPM |
|------------------------------------------------|---------------------|--------------------|---------------------|
| WesperO2 | 59-122 BPM | 2.2 (N=235) | Pass |
The results of the study show that the WesperO2 pulse oximeter outperforms the ARMS specification of 3.5% for the range 70-100%. WesperO2 provides accurate PR values over the range 59-122 BPM, meeting a specification of 3 BPM.
The validated pulse rate range was extended beyond the limits of clinical observation to 30-230 BPM using bench testing.
#### Substantial Equivalence
Both devices use light absorption in the finger to measure SpO2 and Pulse Rate, and both use Red and Infra-Red as their working wavelength bands. The subject device uses reflected energy to measure absorption, while the predicate device uses transmitted energy to accomplish that. Both methods are well understood and accepted methodologies to accurately measure Sp02 and Pulse Rate.
Both devices have identical SpO2 and Pulse Rate ranges (70% - 100%, and 30BPM – 250BPM, respectively), with the subject device having a maximum resolution of 2BPM against the predicate's 1BPM resolution. This difference in resolutions is not clinically significant. The subject device has an overall Arms of 2.56%, while the predicate has an Arms of 1.88%. Both results are within the necessary acceptance criteria, making them clinically equivalent.
Both devices use silicone-based housings and are applied to the finger. The subject device is entirely contained in the fingertip unit and the predicate device attaches to a wrist-worn unit. Since both devices have biocompatible patient-contacting materials and similar form factors, they are functionally equivalent.
Both devices have capabilities to record, and display in real time, SpO2 and Pulse Rate. The subject device uses the companion mobile application via Bluetooth to achieve this, while the predicate has an on-board LCD display. The displayed measurements are the same both units perform the same function, and the subject device's software has been fully validated to meet the necessary functional requirements.
A substantial equivalence chart comparing the similarities and differences between WesperO2, and its predicate device is provided in Table 1. Differences in technology as described above do not raise new type questions of safety and effectiveness. Furthermore, clinical performance testing per ISO 80601-2-61:2019 and Pulse Oximeters - Premarket Notification Submissions [510(k)s] - Guidance for Industry and Food and Drug Administration
{7}------------------------------------------------
## O WESPER
Staff, March 4, 2013 has shown that WesperO2 successfully meets the accuracy criteria for Pulse Oximeters intended for marketing in the US.
| Characteristic | Subject Device<br>WesperO2 | Predicate Device<br>Oxiband (Checkme)<br>O2 Pulse Oximeter<br>(K191088) | Comparison |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | DQA | DQA | Same. |
| Class | II | II | Same. |
| Indications For Use | WesperO2 is a finger-<br>worn pulse oximeter<br>indicated for use in<br>measuring, displaying,<br>storing and transmitting<br>functional oxygen<br>saturation of arterial<br>hemoglobin (SpO2) and<br>pulse rate for adult<br>patients. It is intended<br>for spot-check and/or<br>continuous data<br>collection, and not<br>continuous monitoring.<br>It can be used in sleep<br>labs and home use. | The Oxiband Pulse<br>Oximeter is a wrist<br>pulse oximeter<br>indicated for use in<br>measuring, displaying,<br>storing, and<br>transmitting functional<br>oxygen saturation of<br>arterial hemoglobin<br>(SpO2) and pulse rate<br>for adult patients. It is<br>intended for spot-check<br>and/or continuous data<br>collection, and not<br>continuous monitoring.<br>It can be used in sleep<br>labs, long-term care,<br>hospitals and home<br>use. | Indications for use are<br>similar between the<br>two devices. The<br>subject device is only<br>indicated for sleep<br>labs and home use<br>which is a subset of<br>the predicate's<br>indications. |
| Intended application site | Finger | Finger | Same. |
| Continuous collection | Yes | Yes | Same. |
| Sensor form<br>factor | Self-contained, wireless<br>finger unit | Detachable, wired<br>finger unit | Both devices target<br>the finger, where the<br>detachable design of<br>the predicate does not<br>affect the accuracy of<br>the measurement. |
| Table 3: Substantial Equivalence Table | | | |
|----------------------------------------|---------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sp02<br>measurement<br>accuracy | 70 - 100%: Arms = 2.56<br>70 - 80%: Arms = 3.6<br>80 - 90%: Arms = 2.05<br>90 - 100%: Arms = 1.84 | 70 - 100%: Arms = 1.88<br>70- 80%: ±3%<br>80- 90%: ±2%<br>90- 100%: ±2% | Both devices meet the<br>necessary accuracy<br>requirements for 70-<br>100% and have similar<br>error profiles decade-<br>by-decade, therefore<br>they have equivalent<br>SpO2 measurement<br>accuracies. |
| SpO2 display<br>range | 70 - 100% | 0 - 100% | The predicate's<br>extended display<br>range has no practical<br>significance, as both<br>devices claim the<br>same SpO2 range,<br>which is the same as<br>the subject's display<br>range. The subject<br>device's display range<br>is within the range of<br>the predicate device. |
| SpO2 resolution | 1% | 1% | Same. |
| Pulse Rate<br>measurement<br>range | 30 - 230 BPM | 30 - 250 BPM | Similar. |
| Pulse Rate<br>accuracy | ±3 BPM RMSE | ±2 BPM or ± 2%<br>(whichever is greater) | In the worst-case<br>scenario, the two<br>devices differ in 1<br>BPM on average,<br>which is not clinically<br>significant. |
| Pulse Rate<br>resolution | 2 BPM | 1 BPM | 1 BPM does not<br>constitute a clinically<br>significant difference. |
| Display method | In-app via Bluetooth | On-unit OLED screen | Both devices will<br>display the same type<br>of live data. Both<br>displays rely on<br>software components<br>to render the<br>measurements, and<br>Wesper performed<br>software validation of<br>the integrity and real-<br>time response of the<br>app display to<br>demonstrate<br>equivalence. |
| Wireless<br>connectivity | Bluetooth | Bluetooth | Same. |
| Storage | Yes, on-probe and in-<br>app | Yes, on-probe | Similar. |
| Contact duration | Up to 24h | Up to 24h | Same. |
{8}------------------------------------------------
Image /page/8/Picture/2 description: The image shows the word "WESPER" in black font. To the left of the word is a gray circle that is partially shaded, resembling a crescent moon. The font is simple and sans-serif, giving the logo a clean and modern look. The overall design is minimalist and professional.
{9}------------------------------------------------
Image /page/9/Picture/2 description: The image shows the word "WESPER" in black font. To the left of the word is a gray circle that is partially filled in, resembling a crescent moon. The font is simple and sans-serif. The image is clean and minimalist.
#### Conclusion
The information provided in this submission, in accordance with the requirements of 21 CFR 807 Subpart E, supports the conclusion that WesperO2 is substantially equivalent to the predicate device, the Oxiband (Checkme) O2 Pulse Oximeter.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.