← Product Code [DQA](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DQA) · K212300

# Pulse Oximeter (K212300)

_Shenzhen Mericonn Technology Co., Ltd. · DQA · Feb 25, 2022 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-c%E2%80%94cardiovascular-monitoring-devices/DQA/K212300

## Device Facts

- **Applicant:** Shenzhen Mericonn Technology Co., Ltd.
- **Product Code:** [DQA](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DQA.md)
- **Decision Date:** Feb 25, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2700
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

This Pulse Oximeter is intended for measuring and recording the functional oxygen saturation (SpO2) and pulse rate (PR). It is intended for spot check of SpO2, PR of adult or pediatric patients (weight >=40kg) in hospitals, clinics, or home. This device is not intended for continuous monitoring. It is intended to be used by both lay person or healthcare professional in the home environment.

## Device Story

Pulse oximeter (models PO101, PO102, PO103) measures SpO2 and pulse rate via fingertip transmittance optical sensor. Device uses dual light sources (660 nm red, 905 nm infrared) and photodetector to measure light absorption in pulsating arteriolar vascular bed; calculates SpO2 based on ratio of light absorption during systole and diastole. Powered by 2 AAA batteries; features OLED/LED screen, PCB, and plastic shell. Operated by lay persons or healthcare professionals for spot checks; no alarms; not for life-support. Output displayed on screen; assists clinicians/users in assessing oxygenation status. Benefits include non-invasive, portable monitoring of oxygen saturation.

## Clinical Evidence

Clinical hypoxia testing conducted on human adult volunteers per ISO 80601-2-61:2017. Accuracy validated against co-oximetry (SaO2). Results support accuracy claims for specified saturation range (70-100% with ARMS ±3%).

## Technological Characteristics

Transmittance optical sensor; dual light source (660 nm red, 905 nm infrared); photodetector. Powered by 2 AAA alkaline batteries. Form factor: fingertip device with OLED/LED display. Standards: ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-61. Software level of concern: moderate.

## Regulatory Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

## Predicate Devices

- Pulse Oximeter, AP-10 ([K201468](/device/K201468.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 25, 2022

Shenzhen Mericonn Technology Co., Ltd. Kevin Wang Consultant Chonconn Medical Device Consulting Co., Ltd. Room 508, Block C, No. 1029 Nanhai Avenue, Nanshan District Shenzhen, Guangdong 518067 China

Re: K212300

Trade/Device Name: Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: February 25, 2022 Received: February 23, 2022

Dear Kevin Wang:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Todd Courtney Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known) K212300

Device Name

Pulse Oximeter (Model: PO101, PO102, PO103)

Indications for Use (Describe)

This Pulse Oximeter is intended for measuring the functional oxygen saturation (SpO2) and pulse rate (PR). It is intended for spot check of SpO2, PR of adult or pediatric patients (weight >=40kg) in home. This device is not intended for continuous monitoring. It is intended to be used by both lay person or healthcare professional in the home environment.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

] Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 5 510(k) Summary

{4}------------------------------------------------

# 510(K) Summary

## Prepared in accordance with the requirements of 21 CFR Part 807.92

## Prepared Date: 2022/2/7

#### 1. Submission sponsor

Name: Shenzhen Mericonn Technology Co., Ltd. Address: Room 20203, Building 6, Donglongxing Kejiyuan, Huaning Road, Xinshishequ, Dalang Street, Longhua district, Shenzhen City, Guangdong, 518109 P.R. China Contact person: Jiang Chuanyuan Title: General manager E-mail: info@digitalems.net Tel: 0086 13928416535

#### 2. Submission correspondent

Name: Chonconn Medical Device Consulting Co., Ltd.

Address: Room 508, Block C, No. 1029 Nanhai Avenue, Nanshan District, Shenzhen, Guangdong, P. R. China 518067

Contact person: Kevin Wang

E-mail: kevin(@chonconn.com

Tel: +86-755 33941160

| Trade/Device Name | Pulse Oximeter                  |
|-------------------|---------------------------------|
| Model             | PO101, PO102, PO103             |
| Common Name       | Fingertip Pulse Oximeter        |
| Regulatory Class  | Class II                        |
| Classification    | 21CFR 870.2700 / Oximeter / DQA |
| Submission type   | Traditional 510(K)              |

#### 3. Subject Device Information

#### 4. Predicate Device

Manufacturer: Shenzhen Creative Industry Co., Ltd. Device name: Pulse Oximeter, AP-10 510(K) Number: K201468

#### ട്. Device Description

The Pulse Oximeter is intended for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR).

The Pulse Oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660 nm, which is red light; the 

{5}------------------------------------------------

other is 905 nm, which is Infrared light. Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photodetector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2. The Pulse Oximeter is powered by 2 AAA alkaline batteries.

The device mainly composed of PCB board, On/Off button, mode button, OLED&LED screen, battery compartment, and plastic shell. The device is a spot-check pulse oximeter and does not include alarms. The device is not intended for life-supporting or life-sustaining.

#### 6. Intended use & Indication for use

This Pulse Oximeter is intended for measuring and recording the functional oxygen saturation (SpO2) and pulse rate (PR).

It is intended for spot check of SpO2, PR of adult or pediatric patients (weight >=40kg) in hospitals, clinics, or home. This device is not intended for continuous monitoring. It is intended to be used by both lay person or healthcare professional in the home environment.

| Features                     | Subject Device                                                                                                                                                                                                                                                      | Predicate Device                                                                                                                                                                                                                                         | Comparison                                                                                                                                                                                                      |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                              | Pulse Oximeter,<br>PO101/PO102/PO103                                                                                                                                                                                                                                | K201468<br>Pulse Oximeter, AP-10                                                                                                                                                                                                                         |                                                                                                                                                                                                                 |
| Applicant                    | Shenzhen Mericonn<br>Technology Co., Ltd.                                                                                                                                                                                                                           | Shenzhen Creative<br>Industry Co., Ltd.                                                                                                                                                                                                                  | /                                                                                                                                                                                                               |
| Classification<br>Regulation | 21CRF 870.2700                                                                                                                                                                                                                                                      | 21CRF 870.2700                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                            |
| Classification<br>and Code   | Class II, DQA                                                                                                                                                                                                                                                       | Class II, DQA                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                            |
| Common<br>name               | Pulse Oximeter                                                                                                                                                                                                                                                      | Pulse Oximeter                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                            |
| Intended use                 | This Pulse Oximeter is<br>intended for<br>measuring and<br>recording the<br>functional oxygen<br>saturation (SpO2) and<br>pulse rate (PR).<br>It is intended for spot<br>check of SpO2, PR of<br>adult or pediatric                                                 | This Pulse Oximeter is<br>intended for measuring<br>and recording the<br>functional oxygen<br>saturation (SpO2) and<br>pulse rate (PR). It is<br>intended for spot check<br>and continuous recording<br>of SpO2, PR of adult or<br>pediatric patients in | The subject devices<br>do not support<br>continuous recording<br>compared with the<br>predicate, but the rest<br>is the same. The<br>difference does not<br>raise any safety and<br>effectiveness<br>questions. |
| Features                     | Subject Device                                                                                                                                                                                                                                                      | Predicate Device<br>K201468                                                                                                                                                                                                                              | Comparison                                                                                                                                                                                                      |
|                              | Pulse Oximeter,<br>PO101/PO102/PO103<br>patients in hospitals,<br>clinics, or home. This<br>device is not intended<br>for continuous<br>monitoring. It is<br>intended to be used by<br>both lay person or<br>healthcare<br>professional in the<br>home environment. | Pulse Oximeter, AP-10<br>hospitals, clinics, or<br>home.<br>This device is not<br>intended for continuous<br>monitoring.                                                                                                                                 |                                                                                                                                                                                                                 |
| Patient<br>populations       | adult or pediatric                                                                                                                                                                                                                                                  | adult or pediatric                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                            |
| Type of<br>SpO2 Sensor       | Transmittance Optical<br>Sensor                                                                                                                                                                                                                                     | Transmittance Optical<br>Sensor                                                                                                                                                                                                                          | Same                                                                                                                                                                                                            |
| Application<br>Site          | Finger                                                                                                                                                                                                                                                              | Finger                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                            |
| Light<br>Emitting            | Red: 660 nm<br>Infrared: 905nm                                                                                                                                                                                                                                      | Red: 660 nm<br>Infrared: 905nm                                                                                                                                                                                                                           | Same                                                                                                                                                                                                            |
| Measuring<br>Mode            | Spot-check                                                                                                                                                                                                                                                          | Spot-check and<br>Continuous recording                                                                                                                                                                                                                   | Different                                                                                                                                                                                                       |
| SpO2<br>Measuring<br>Range   | 35%-100%                                                                                                                                                                                                                                                            | 0%-100%                                                                                                                                                                                                                                                  | Different                                                                                                                                                                                                       |
| SpO2<br>Resolution           | 1%                                                                                                                                                                                                                                                                  | 1%                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                            |
| SpO2<br>Accuracy             | 70~100%,<br>ARMS±3%.<br><70%, unspecified.                                                                                                                                                                                                                          | 70~100%, ±3%.<br><70%, unspecified;                                                                                                                                                                                                                      | Same                                                                                                                                                                                                            |
| PR Range                     | 30 bmp – 250 bmp                                                                                                                                                                                                                                                    | 30 bmp – 250 bmp                                                                                                                                                                                                                                         | Same                                                                                                                                                                                                            |
| PR<br>Resolution             | 1 bpm                                                                                                                                                                                                                                                               | 1 bpm                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                            |
| PR Accuracy                  | ±2bpm or±2% select<br>larger                                                                                                                                                                                                                                        | ±2bpm or ±2%<br>(whichever is greater)                                                                                                                                                                                                                   | Same                                                                                                                                                                                                            |
| Power source                 | 2 AAA alkaline<br>batteries                                                                                                                                                                                                                                         | Rechargeable Lithium-<br>Ion Polymer Battery<br>(3.7V,500mAh)                                                                                                                                                                                            | Different                                                                                                                                                                                                       |
| Data update                  | /                                                                                                                                                                                                                                                                   | USB, Bluetooth                                                                                                                                                                                                                                           | Different                                                                                                                                                                                                       |
| Features                     | Subject Device                                                                                                                                                                                                                                                      | Predicate Device                                                                                                                                                                                                                                         | Comparison                                                                                                                                                                                                      |
|                              | Pulse Oximeter,<br>PO101/PO102/PO103                                                                                                                                                                                                                                | K201468<br>Pulse Oximeter, AP-10                                                                                                                                                                                                                         |                                                                                                                                                                                                                 |
| Type of<br>Protection        | Internal Powered                                                                                                                                                                                                                                                    | Internal Powered                                                                                                                                                                                                                                         | Same                                                                                                                                                                                                            |

#### 7. Comparison to the Predicate Device

{6}------------------------------------------------

{7}------------------------------------------------

In conclusion, the differences in technological characteristics do not raise new questions of safety and effectiveness.

#### 8. Performance Data

The following performance data were provided in support of the substantial equivalence determination. Software

Software verification and validation were provided in compliance with FDA Guidance for the Content of the Premarket Submission for Software Contained in Medical Devices. These verifications and validations demonstrate that the subject device work functionally and the software for the device is considered as a "moderate" level of concern, as defined by the FDA guidance.

## Biocompatibility testing

The biocompatibility evaluation for the Pulse Oximeter was conducted in accordance with the FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The biocompatibility testing included the following tests:

- . Cytotoxicity
- Sensitization .
- Irritation .

The subject devices are considered surface contacting for a duration of not exceed 24 hours.

## Non-clinical data

The Pulse Oximeter has been tested according to the following standards:

- ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 . (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
- IEC 60601-1-2:2014, Medical electrical equipment- Part 1-2: General requirements for basic . safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests
- . IEC 60601-1-11 Edition 2.0 2015-01Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- ISO 80601-2-61: 2017 Medical Electrical Equipment Part 2-61: Particular Requirements for .

{8}------------------------------------------------

Basic Safety and Essential Performance of Pulse Oximeter Equipment.

The test was selected to show substantial equivalence between the subject device and the predicate.

## Clinical data

Clinical studies were conducted to verify the accuracy of proposed device. The clinical studies were conducted per following standards:

- ISO 80601-2-61: 2017 Medical Electrical Equipment Part 2-61: Particular Requirements for . Basic Safety and Essential Performance of Pulse Oximeter Equipment.
- Pulse Oximeters-Premarket Notification Submissions: Guidance for Industry and Food and . Drug Administration Staff

Clinical hypoxia test results were obtained in human adult volunteers to validate the accuracy of Pulse Oximeter versus arterial oxygen saturation (SaO2) as determined by co-oximetry. Clinical test results support device accuracy claims for the specified saturation range.

#### 9. Conclusion

Performance testing and compliance with voluntary standards demonstrate that the proposed subject device is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-c%E2%80%94cardiovascular-monitoring-devices/DQA/K212300](https://fda.innolitics.com/submissions/AN/subpart-c%E2%80%94cardiovascular-monitoring-devices/DQA/K212300)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
