K202851 · Jkh USA, LLC · DQA · Nov 27, 2020 · Cardiovascular
Device Facts
Record ID
K202851
Device Name
Spo2 Sensor
Applicant
Jkh USA, LLC
Product Code
DQA · Cardiovascular
Decision Date
Nov 27, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
As an accessory of the legally marketed oximeters or patient monitors in the United States, the proposed device SpO2 Sensor is intended for hospital use and continuous monitoring of functional arterial oxygen saturation and pulse rate.
Device Story
SpO2 sensor probe attaches to patient finger; connects to compatible pulse oximeter monitor. Device uses 2-wavelength optical absorption (red 660-666nm, infrared 880-950nm) to detect oxygenated vs. deoxygenated hemoglobin. Monitor calculates SpO2 percentage and pulse rate; displays values for clinician. Used in hospital settings for continuous monitoring. Benefits include non-invasive, real-time assessment of patient oxygenation status to support clinical decision-making.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing and comparison to identical predicate devices.
Technological Characteristics
Materials: ABS, PVC, TPU, silicone, 3M medical tape, sponge. Sensing: 2-wavelength optical absorption. Power: Provided by host monitor. Form factors: Reusable soft tip, reusable finger clip, disposable non-adhesive, disposable adhesive. Standards: IEC 60601-1, IEC 60601-1-2, ISO 80601-2-61, ISO 10993-5, ISO 10993-10. Non-sterile.
Indications for Use
Indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult patients >40 kg, pediatric patients 10-50 kg, and neonatal patients ≥3 kg.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
JKH USA, LLC Bill Dai Manager 14271 Jeffrey Rd. #246 Irvine, California 92620
Re: K202851
Trade/Device Name: Spo2 Sensor Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: September 18, 2020 Received: September 28, 2020
Dear Bill Dai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K202851
Device Name SpO2 Sensor
Indications for Use (Describe)
SpO2 Sensors are indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult patients weighing greater than 40 kg, pediatric patients weighing 10-50 kg, and neonatal patients weighing no less than 3 kg.
Type of Use (Select one or both, as applicable)
| <div> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
|------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## K202851
| Submitter: | Name: JKH USA, LLC<br>Mailing Address: 14271 Jeffrey Rd. #246, Irvine, CA 92620 |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Name: Bill Quanqin Dai<br>Phone Number: 909-929-9896<br>Email Address: Bill@jkhUSA.com |
| Date Prepared: | 09/18/2020 |
| Device Trade Name: | Spo2 Sensor |
| Device Common Name: | Spo2 Sensor |
| Model: | N543-01, U103-01, U103S-01, U403-01, U403S-01, U503-01,<br>U543-01, U410-03, U403-07, U410-02, U403-08, U403-125, U403-06,<br>U410-09, U403S-91 |
| Classification Names: | Oximeter |
| Regulation Number: | 21 CFR 870.2700 |
| Product Code: | DQA |
| Predicate Device 1: | |
| 510(k) Number: | K082546 |
| Device Name: | Unimed compatible oximeter sensors |
| Manufacturer: | UNIMED MEDICAL SUPPLIES INC |
| Predicate Device 2: | |
| 510(k) Number: | K142832 |
| Device Name: | Unimed Reusable & Disposable Spo2 Sensor |
| Manufacturer: | UNIMED MEDICAL SUPPLIES INC |
#### Description of Devices:
As an accessory of the legally marketed oximeters or patient monitors in the United States, the proposed device SpO2 Sensor is intended for hospital use and continuous monitoring of functional arterial oxygen saturation and pulse rate.
The SpO2 sensor consist of a probe attached to the patient's finger. The sensor shall be connected to its corresponding pulse oximeter monitor with a data acquisition system which is used to calculate and display oxygen saturation levels and heart rate conditions. Oxygenation of blood is measured by detecting the infrared and red light absorption characteristics of deoxygenated hemoglobin and oxygenated hemoglobin.
. The saturation value is determined by the percentage ratio of the oxygenated hemoglobin (HbO2) to the total amount of hemoglobin (Hb).
Four types of sensor housings are described in this submission:
{4}------------------------------------------------
- Reusable soft tip sensor comprised of an integrated silicone rubber tip.
- Reusable finger clip sensor with rigid halves and silicone pads
- Disposable non- adhesive sensor with sponge and velcro backing.
- Disposable adhesive sensors constructed of a medical tape laminate.
Each SpO2 sensor has unique labeling and specifications designed for compatibility with the specific oximeters/monitors that have been legally marketed in the United States. The compatible oximeters/monitors are listed in the labeling of each SpO2 sensor.
#### Indications for Use:
SpO2 Sensors are indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult patients weighing greater than 40kg, pediatric patients weighing 10-50 kg, and neonatal patients weighing no less than 3 kg.
#### Comparison to predicate device:
The subject and predicate devices are exactly the same, and there is no difference between them.
| Description | Subject Device (K202851) | Predicate Device (K082546 and<br>K142832) |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | SpO2 Sensors are indicated for<br>continuous non-invasive monitoring of<br>functional oxygen saturation of arterial<br>hemoglobin (SpO2) and pulse rate (PR)<br>for adult patients weighing greater<br>than 40 kg, pediatric patients weighing<br>10-50 kg, and neonatal patients<br>weighing no less than 3 kg. | Unimed Disposable and Reusable<br>SPO2 Sensors are indicated for<br>continuous non-invasive monitoring of<br>functional oxygen saturation of<br>arterial hemoglobin (SpO2) and pulse<br>rate(PR) for adult patients weighing<br>greater than 40 kg, pediatric patients<br>weighing 10 -50 kg, and neonatal<br>patients weighing no less than 3 Kg. |
| Prescription /<br>over-the-counter<br>use | Prescription | Prescription |
| Energy source | Powered by compatible devices | Powered by compatible devices |
| Measurement<br>Method | 2-wavelength Relative Optical<br>Absorption | 2-wavelength Relative Optical<br>Absorption |
| Light Emitting | Red: 660-666nm, infrared: 880-950nm | Red: 660-666nm, infrared: 880-950nm |
| Sensor Material | ABS, PVC, TPU, Silicone, 3M medical<br>tape and sponge | ABS, PVC, TPU, Silicone, 3M medical<br>tape and sponge |
| Distal connector | finger clip , soft tip, textile adhesive<br>and sponge non- adhesive | finger clip , soft tip, textile adhesive<br>and sponge non- adhesive |
| Design | and sponge non- adhesive | and sponge non- adhesive |
| SpO2<br>Measurement<br>Range | 70-100% | 70-100% |
| Table 1 Substantial Equivalence Table | |
|---------------------------------------|--|
| | |
{5}------------------------------------------------
| SpO2 Accuracy | $\pm$ 3% | $\pm$ 3% |
|-----------------------------------------------|---------------------------------------|---------------------------------------|
| PR Measurement<br>Range | 30-250bpm | 30-250bpm |
| PR Accuracy | $\pm$ 3 | $\pm$ 3 |
| Usage | Reusable and disposable | Reusable and disposable |
| Sterile | Non-sterile | Non-sterile |
| Biocompatibility<br>contact<br>classification | Surface Skin contact, less than 24hrs | Surface Skin contact, less than 24hrs |
#### Non-clinical test data:
The subject device meets the following the recognized standards:
- IEC 60601-1 Medical Electrical Equipment-Part 1: General Requirements for Safety
- IEC 60601-1-2 Medical Electrical Equipment-Part 1-2: General Requirements for Safety -Collateral Standard: Electromagnetic Compatibility-Requirements and Tests
- ISO 80601-2-61 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment
- ISO 10993-5, Biological evaluation of medical devices-Part 5: Tests for In Vitro cytotoxicity
- ISO 10993-10, Biological evaluation of medical devices-Part 10: Tests for irritation and delayed-type hypersensitivity
The proposed device belongs to skin contact, and the contact duration is less than 30 days. Biocompatibility tests have been conducted on the proposed device, including cytotoxicity, sensitization, and skin irritation. The test results show that the proposed device has no issue of cytotoxicity, sensitization, or skin irritation.
#### Clinical test data:
The subject and predicate devices are exactly the same. Since the sensors are identical, no further clinical testing is necessary.
#### Substantial Equivalence:
The subject and predicate devices are exactly the same. The sensors are identical to the cleared version and are not modified. Therefore, the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.