This SpO2 Sensor is intended for the continuous noninvasive monitoring and spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR, measured by an SpO2 sensor) for use with the finger of adult and pediatric patients. It acts as re-usable accessory of a Patient Monitor.
Device Story
SpO2 sensor (models U401-A/B/C/D) functions as reusable accessory for patient monitors; utilizes red (660 nm) and infrared (905 nm) LEDs to transmit light through finger tissue; photodetector measures light absorption variations caused by blood pulse and hemoglobin oxygenation; microprocessor calculates SpO2 and pulse rate values. Used in clinical settings by healthcare providers; output displayed on host patient monitor to assist in patient status assessment and clinical decision-making regarding oxygenation levels.
Clinical Evidence
Clinical validation conducted per ISO 80601-2-61:2011 Annex EE. Study involved healthy volunteers (ages 18-55) with varied skin pigmentations, including three darkly pigmented subjects. Accuracy validated against arterial blood samples measured via CO-oximeter. Results demonstrate SpO2 accuracy of ±2% and pulse rate accuracy of ±2 bpm.
Technological Characteristics
Reusable finger-clip SpO2 sensor; 2-wavelength optical absorption (660 nm/905 nm); photodetector sensing. Complies with ISO 10993-1 (biocompatibility), IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and ISO 80601-2-61 (performance). No embedded software.
Indications for Use
Indicated for continuous or spot-check monitoring of functional arterial hemoglobin oxygen saturation (SpO2) and pulse rate in adult and pediatric patients via finger sensor. Contraindications: none stated.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Shenzhen Med-Link Pulse Oximeter Probe, model S0136J-L (K113727)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Shenzhen Upnmed Equipment Co., Ltd. Fu Jian Management Representative 4th Floor, Building #1 East, Huihuang Industrial Area Xitian Community, Gongming Town. Guangming District Shenzhen, 518107 Cn
Re: K183277
Trade/Device Name: SpO2 Sensor U401-A, U401-B, U401-C, U401-D Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA Dated: April 8, 2019 Received: April 16, 2019
Dear Fu Jian:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Michael Ryan Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183277
Device Name
SpO2 Sensor, models U401-A, U401-B, U401-C, U401-D
#### Indications for Use (Describe)
This SpO2 Sensor is intended for the continuous nonitoring and spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR, measured by an SpO2 sensor) for use with the finger of adult and pediatric patients. It acts as re-usable accessory of a Patient Monitor.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <span style="font-family: Wingdings;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size: 10pt;"><span style="font-family: Wingdings;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# K183277
# Prepared in accordance with the requirements of 21 CFR Part 807.92
| 1. Submitter: | Shenzhen Upnmed Equipment Co., Ltd.<br>4th Floor, Building #1 East, Huihuang Industrial Area, Xitian Community,<br>Gongming Town, Guangming District, 518107, Shenzhen, P.R. China<br>Tel.: +86 -755-29728789<br>Fax: +86 -755-29423789 |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Fu Jian |
| Prepare date: | 2019-04-08 |
| 2. Device name<br>and classification: | Device Name: SpO2 Sensor<br>Models: U401-A, U401-B, U401-C, U401-D<br>Classification Name: 21 CFR 870.2700 Oximeter<br>Product code: DQA<br>Regulatory Class: Class II |
| 3. Reason for<br>Submission | New Application. |
| 4. Predicate<br>Device(s): | Shenzhen Med-link Electronics Tech Co., Ltd., Shenzhen Med-Link Pulse Oximeter<br>Probe, model S0136J-L/ K113727 |
| 5. Device<br>Description: | The SpO2 Sensor consists of compatible connectors, cable, patient side sensor. And<br>one side of sensor is designed to locate light emitting diodes (LEDs) and a light<br>detector (called a photo-detector). Red and Infrared lights are shone through the<br>tissues from one side of the probe to the other. Then parts of the light emitted<br>absorbed by blood and tissues. The light absorbed by the blood varies with the<br>oxygen saturation of haemoglobin. After that, the photo-detector detects the light<br>volume transmitted through the tissues which depends on blood pulse, Hereafter, the<br>microprocessor calculates a value for the oxygen saturation (SpO2).<br><br>The U401 series SpO2 Sensor acts as re-usable accessory of Mindray PM-8000<br>Express Patient Monitor, which has been cleared by K070791. And the function<br>alarms and interface accessories are controlled by the monitor, and such information<br>is not available in this submission. |
{4}------------------------------------------------
This SpO2 Sensor is intended for the continuous noninvasive monitoring and 6. Indications for spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and Use: pulse rate (PR, measured by an SpO2 sensor) for use with the finger of adult and pediatric patients. It acts as re-usable accessory of a Patient Monitor.
# 7. Predicate Device Comparison
Comparison to the predicate devices, the subject device has same intended use, similar product design, same performance effectiveness as the predicate device.
Please refer to following table to find differences between the subject device and predicate device. All the differences do not affect the basic design principle, usage, effectiveness of the subject device. And the differences do not raise different questions of safety of effectiveness.
| ITEM | Proposed Device<br>U401 Series SpO2 Sensor | Predicate Device<br>S0136J-L/K113727 | Comparison<br>Result |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Manufacture | Shenzhen Upnmed Equipment<br>Co., Ltd. | Shenzhen Med-link Electronics<br>Tech Co., Ltd. | --- |
| Indications for Use | This SpO2 Sensor is intended<br>for the continuous noninvasive<br>monitoring and spot-checking<br>of functional oxygen saturation<br>of arterial hemoglobin (SpO2)<br>and pulse rate (PR, measured<br>by an SpO2 sensor) for use<br>with adult and pediatric<br>patients. | Shenzhen Med-link disposable<br>probe, model S0136J-L, is<br>indicated for single patient use<br>when continuous, non-invasive<br>arterial oxygen saturation and<br>pulse rate monitoring are<br>required for adult patients or<br>pediatric patients weighing more<br>than 30 kg. The S0136J-L is<br>contraindicated for use with<br>patients during motion<br>condition. | Different 1 |
| Intended patient<br>population | Pediatric, Adult | Pediatric, Adult | Same |
| Intended<br>application site | Finger | Finger | Same |
| use under motion<br>and low perfusion<br>conditions | No | No | Same |
| Measurement<br>Principles | 2-wavelength Relative<br>Optical Absorption | 2-wavelength Relative<br>Optical Absorption | Same |
| Light Emitting<br>Diodes (LEDs)<br>wavelengths | RED: 660 nm, nominal<br>IRED: 905nm, nominal | RED: 660 nm, nominal<br>IRED: 905 nm, nominal | Same |
| Signal Detection<br>Method | Photodetector | Photodetector | Same |
| SpO2 Range | 70% -100% | 70%-100% | Same |
| SpO2 Accuracy | ±2% | ±3% | Different 2 |
| Pulse Rate Range | 20 bpm - 250 bpm | 20 bpm – 250 bpm | Same |
| Pulse Rate<br>Accuracy | ±2 bpm | ±3 bpm | Different |
| Measuring Mode | spot-checking or continuous<br>monitoring | spot-checking or continuous<br>monitoring | Same |
| Shipped Sterile | No | No | Same |
| Storage and<br>Transport | Temperature: -10℃to 50°C<br>Atmospheric Pressure: 50 kPa<br>to 106 kPa<br>Relative Humidity: ≤93% (no<br>condensation) | Temperature: -10℃to 40°C<br>Atmospheric Pressure: 86 kPa to<br>106 kPa<br>Relative Humidity: 0%-80% (no<br>condensation) | Different 3 |
| Operating<br>Temperature | Temperature: 5°C to 40°C<br>Atmospheric Pressure: 50 kPa<br>to 106 kPa<br>Relative Humidity: ≤85% (no<br>condensation) | Temperature: -10℃to 40°C<br>Atmospheric Pressure: 86 kPa to<br>106 kPa<br>Relative Humidity: 0%-80% (no<br>condensation) | |
| Physical Specifications | | | |
| Cable Length | 1.0 m | 0.9 m | Different 4 |
Table 1 Comparison between main predicate S0136J-L and the subject device
{5}------------------------------------------------
### Justification for the differences:
- 1) Different Indications for Use
As indicated in the comparison table, the difference of indications for Use is caused by different language, so the difference do not raise different questions of safety and effectiveness.
- 2) Different SpO2 Accuracy
The accuracy of the subject device is ±2%, which of the predicate is ±3%, which means the performance of the subject is better, and such specification is verified per the international standard ISO 80601-2-61, so the different accuracy will be acceptable for the subject probe.
- 3) Different Storage and Transport
Minor difference to Storage & Transport and operation environments (including Temperature, Atmospheric Pressure and Relative Humidity) for the subject device, but the system has been proved to be safe and effective since the testing was conducted under the suggested environment; Moreover, environment testing data shows the device can work as declared under the suggested conditions. So those changes do not raise different questions of safety and effectiveness.
- 4) Different Physical Specifications
The subject and the predicate are of different size but proximity. Moreover, such engineering design has been verified during the international standards, so such minor differences do not raise different questions of safety and effectiveness.
As seen in the comparison tables, the subject and predicate devices have same design principle, similar design features and performance specifications. The main differences between the subject and predicate devices are minor, including only the operating/storage environment, cable length which do not raise different questions of safety or effectiveness. Moreover, as demonstrated in the non-clinical, the different technological characteristics do not raise different questions of safety and effectiveness.
# 8. Performance Testing:
Performance data includes "Non-Clinical Data" and "Clinical Data", brief description of which are shown as below.
# Non-Clinical Data:
{6}------------------------------------------------
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility testing
The biocompatibility evaluation for the SpO2 Sensor was conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The worst case of the whole system is considered tissue contacting for duration of less than 24 hours. And the battery of testing included the following tests:
- Cytotoxicity
- Skin Sensitization
- Skin Irritation
### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted, and the results show that the subject device complies with the IEC 60601-1: 2005+CORR. 1 (2006)+CORR. 2 (2007)+AM1 (2012) Medical electrical equipment Part 1: General requirements for basic safety and essential performance for safety and the IEC 60601-1-2: 2007 Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests standard for EMC.
# Bench Testing
Bench testing was conducted and the results show that the subject device complies with the ISO 80601-2-61: 2011 Medical electrical equipment — Part 2-61: Particular requirements for basic safety and essential performance of Pulse Oximeter Equipment standard for performance effectiveness.
### Software Verification and Validation Testing
Not applicable, no software is embedded.
# Clinical data:
Clinical testing is conducted per Annex EE Guideline for evaluating and documenting SpO2 ACCURACY in human subjects of ISO 80601-2-6:2011 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment. The SpQ2 accuracy has been validated in human studies against arterial blood sample reference measured with a CO-oximeter. Pulse oximeter measurements are statistically distributed, only about two-thirds of the measurements can be expected to fall within the specified accuracy compared to CO-oximeter measurements. The volunteer population in the studies composed of local healthy men and women from age 18 to 55, with variations of skin pigmentations and per FDA's guidance for Pulse Oximeters, darkly pigmented subjects are included in the desaturation study is three.
# Summary
Based on the non-clinical performance and clinical data as documented in the device development, the subject devices were found to be as safe and as effective as the predicate device.
# 9. Conclusion:
Verification and validation testing was conducted on the subject device and all testing passed pre-specified criteria. This premarket notification submission demonstrates that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.