Surgical Instrument Service and Savings Inc (Dba Medline
Product Code
DQA · Cardiovascular
Decision Date
Mar 22, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Medline ReNewal Reprocessed Nellcor OxiMax SpO2 Sensors are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO₂) and pulse rate. They are intended for use with infant, pediatric and adult patients in hospitals, hospital-type facilities, and intra-hospital transport. These devices are for prescription use only.
Device Story
Reprocessed Nellcor OxiMax SpO2 sensors; transmissive sensors using red and infrared LEDs and photodiode detectors. Device attaches to patient via adhesive bandage; connects to Nellcor pulse oximeter via cable/pin connector. Operates by sensing light signal interaction with tissue; pulse oximeter processes electrical signals to calculate SpO2 and pulse rate. Used in hospitals and intra-hospital transport; operated by clinicians. Output provides real-time SpO2, pulse rate, and pulse amplitude values to assist clinical decision-making regarding patient oxygenation status. Benefits include non-invasive, continuous monitoring of patient vital signs.
Clinical Evidence
Clinical trial conducted on 10 healthy adult volunteers (ages 25-36) to evaluate SpO2 accuracy. Study followed ISO 80601-2-61:2011 and ISO 14155:2011. Results demonstrated 2% accuracy for 70%-100% SpO2 range, meeting the Arms specification of 3% under steady-state, non-motion conditions.
Technological Characteristics
Transmissive pulse oximeter sensor; dual-wavelength (red/infrared) LEDs and photodiode detector. Form factor: adhesive bandage with cable and pin connector. Non-sterile. Connectivity: wired to Nellcor-compatible pulse oximeter. Principle: photoplethysmography (time-varying tissue optical properties).
Indications for Use
Indicated for single patient use for continuous noninvasive arterial oxygen saturation and pulse rate monitoring in infant, pediatric, and adult patients in hospitals, hospital-type facilities, and intra-hospital transport. Prescription use only.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Hygia Health Services Reprocessed OxiMax Sensors (K041867)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the FDA acronym followed by the full name of the agency, "U.S. Food & Drug Administration," on the right. The FDA part of the logo is in blue, with the acronym in a solid blue square and the agency's name in a lighter blue font.
Surgical Instrument Service and Savings Inc (dba Medline) % Stephanie Boyle Mays Regulatory Specialist, Quality Assurance and Regulatory Affairs Surgical Instrument Service and Savings Inc (dba Medline ReNewal) 1500 NE Hemlock Ave. Redmond, Oregon 97756
Re: K181738
Trade/Device Name: Medline ReNewal Reprocessed Nellcor OxiMax Spoz Sensors, models MAXA, MAXAL, MAXP, and MAXI Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: NLF Dated: February 18, 2019 Received: February 19, 2019
Dear Stephanie Boyle Mays:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd D. Courtney -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K181738
Device Name
Medline ReNewal Reprocessed Nellcor OxiMax SpO2 Sensors Models MAXAL, MAXP and MAXI
Indications for Use (Describe)
The Medline ReNewal Reprocessed Nellcor OxiMax SpO2 Sensors models MAXAL, MAXP, and MAXI are indicated for single patient use when continuous noninvasive arterial oxygen saturation and pulse rate monitoring are required for patients in the sizes indicated in the respective sensor directions for use. These devices are for prescription use only.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows two logos side by side. The first logo is for Medline, and it features a blue square with a white star-like symbol and the word "MEDLINE" in white letters. The second logo is for ReNewal, and it features the word "ReNewal" in green and blue letters, with the words "Full Circle Reprocessing" in smaller letters below.
# K181738 510(k) Summary
This 510(k) summary is prepared in accordance with 21 CFR § 807.92.
| Submitter/Owner | Surgical Instrument Service and Savings (dba Medline ReNewal)<br>1500 NE Hemlock Ave., Redmond, OR 97756 | |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| Prepared by/Contact Name | Stephanie Boyle Mays<br>Regulatory Affairs Specialist, Quality Assurance/Regulatory Affairs<br>P: 541-516-4205 • F: 541-923-3375 • E:smays@medline.com | |
| Date Prepared | February 18, 2019 | |
| Device Name<br>and<br>Classification | Proprietary/Trade Name: | Medline ReNewal Reprocessed Nellcor OxiMax SpO2<br>Sensors, models MAXA, MAXAL, MAXP, and MAXI |
| | Common or usual name | Oximeter, reprocessed |
| | Regulatory Name/Reference: | Oximeter; 21 CFR § 870.2700 |
| | Regulatory Class: | Class II |
| | Product Code: | NLF |
| | Panel: | Cardiovascular/anesthesiology |
| Predicate<br>Device | 510(k) number: | K052186 |
| | Proprietary/Trade Name: | Nellcor OxiMax Pulse Oximeter Sensors, models<br>MAX-A, MAX-AL, MAX-N, MAX-P, MAX-I and MAX-FAST |
| | Common or usual name | Oxygen sensor |
| | Regulatory Name/Reference: | Oximeter; 21 CFR § 870.2700 |
| | Regulatory Class: | Class II |
| | Product Code: | DQA |
| | Panel: | Cardiovascular/anesthesiology |
| | Manufacturer: | Nellcor Puritan Bennett, Inc.<br>4280 Hacienda Dr., Pleasanton, CA 94588 |
| Reference<br>Device | 510(k) number: | K041867 |
| | Proprietary/Trade Name: | Hygia Health Services Reprocessed OxiMax Sensors<br>Model # HHS-MAX-A, HHS-MAX-AL, HHS-MAX-N |
| | Regulatory Name/Reference: | Oximeter; 21 CFR § 870.2700 |
| | Common or usual name | Oximeter, reprocessed |
| | Regulatory Class: | II |
| | Product Code: | NLF |
| | Panel: | Cardiovascular/anesthesiology |
| | Manufacturer: | Hygia Health Services, Inc. |
| Statement of<br>Indications for<br>Use | The Medline ReNewal Reprocessed Nellcor OxiMax SpO2 Sensors models<br>MAXA, MAXAL, MAXP, and MAXI are indicated for single patient use when<br>continuous noninvasive arterial oxygen saturation and pulse rate monitoring<br>are required for patients in the sizes indicated in the respective sensor<br>directions for use. These devices are for prescription use only. | |
| Device<br>Description | The Medline ReNewal Reprocessed Nellcor OxiMax SpO2 Sensors,<br>models MAXA, MAXAL, MAXP, and MAXI are designed for the continuous<br>non-invasive monitoring of functional oxygen saturation of arterial<br>hemoglobin (SpO2) and pulse rate in conjunction with a Nellcor Pulse<br>Oximeter. The Reprocessed Nellcor OxiMax SpO2 Sensors, models MAXA,<br>MAXAL, MAXP, and MAXI are intended for prescription use with adult,<br>pediatric and infant patients in hospitals, hospital-type facilities, and intra-<br>hospital transport. The proposed device is not provided sterile. | |
| Statement of<br>Intended Use | Medline ReNewal Reprocessed Nellcor OxiMax SpO2 Sensors are<br>indicated for the continuous non-invasive monitoring of functional oxygen<br>saturation of arterial hemoglobin (SpO₂) and pulse rate. They are intended<br>for use with infant, pediatric and adult patients in hospitals, hospital-type<br>facilities, and intra-hospital transport. These devices are for prescription use<br>only. | |
| Technology<br>(including<br>features,<br>materials and<br>principles of<br>operation) | The technological characteristics and the fundamental scientific technology<br>of the subject devices are identical to the predicate and reference devices.<br>The proposed devices are a reprocessed version of the predicate K052186<br>device. These devices use an adhesive bandage, light source (LEDs),<br>photodetector (Faraday cage/photodiode), cable, and connector in the<br>same manner as the predicate devices. The predicate and reference<br>devices were used to support intended use, technological characteristics,<br>and performance specifications. (Also see comparison of technological<br>features in the Summary Table) | |
| Performance<br>Testing -<br>Nonclinical<br>Tests | The functional characteristics of the subject device have been evaluated in<br>accordance with Pulse Oximeters - Premarket Notifications Submissions<br>[510(k)] Guidance for Industry and Food and Drug Administration Staff<br>(March 4, 2013 ) and have been determined to be substantially equivalent<br>to the predicate device based on the following tests:<br>· Biocompatibility: cytotoxicity, sensitization, irritation; acute systemic<br>toxicity<br>· Disinfection<br>• Shelf Life<br>· Electrical<br>· Performance testing:<br>· tissue heating<br>· pulse rate accuracy<br>· active element assessment<br>· adhesive peel and<br>· environment (extreme heat and operating conditions)<br>· Cleaning:<br>· visual inspection;<br>· cleaning efficacy (residual protein and residual hemoglobin). | |
| Performance<br>Testing -<br>Clinical Tests | The purpose of the clinical trial was to perform an oxygen saturation (SpO2)<br>accuracy comparison. The study was conducted in accordance with CFR<br>for Non-significant Risk Investigational Studies, following ISO 14155:2011<br>Clinical Investigation of medical devices for human subjects - Good clinical<br>practice as appropriate and the pulse oximeter guidelines of ISO 80601-2-<br>61:2011 Procedure for invasive laboratory testing on healthy volunteers<br>applicable sections and Pulse Oximeters – Premarket Notifications<br>Submissions [510(k)] Guidance for Industry and Food and Drug<br>Administration Staff (March 4, 2013 ). After Institutional Review Board<br>Approval, 10 healthy adults volunteer subjects (ages 25 to 36 yr.; weight<br>105 – 220 lb.; height 60 – 72 in.; BMI of 20.0 – 33.4) were included in the<br>study which was conducted from May 9 to May 10, 2018 to evaluate the<br>SpO2 accuracy of the proposed devices. The proposed devices achieved<br>an accuracy of 2% for 70% - 100% SpO2. The study concluded that the<br>SpO2 accuracy performance of the proposed devices passed the Arms<br>specification of 3% under steady state and non-motion conditions for the<br>range of 70% to 100%. | |
| Device Models | MAXA, MAXAL, MAXP, and MAXI | |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows two logos side by side. The logo on the left is for Medline, and it features the company name in bold, sans-serif font, accompanied by a stylized star-like graphic. The logo on the right is for ReNewal, with the words "Full Circle Reprocessing" underneath. The ReNewal logo has the "Re" in green and the "Newal" in blue.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains two logos. The first logo is for Medline, and it features the word "MEDLINE" in a sans-serif font, with a blue star-like symbol above it. The second logo is for Renewal, and it features the word "Renewal" in a combination of green and blue, with the words "Full Circle Reprocessing" underneath it in a smaller font.
Summary continued on next page.
[This area of page was left blank intentionally]
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows two logos side by side. The logo on the left is for Medline, and it features a blue square with a white star-like symbol and the word "MEDLINE" in white letters. The logo on the right is for ReNewal, and it features the word "ReNewal" in green and blue letters, with the words "Full Circle Reprocessing" underneath in smaller blue letters.
| | | | Summary Table: Predicate, Reference and Revocessed Covidien Nellor OxiMax pulse oximeter sensor comparison. | | | |
|--|--|--|--------------------------------------------------------------------------------------------------------------|--|--|--|
|--|--|--|--------------------------------------------------------------------------------------------------------------|--|--|--|
| | Predicate | Reference | Proposed | Comparison |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Characteristics | Covidien Nellcor<br>OxiMax Pulse Oximeter<br>Sensors | Hygia Health Services<br>Reprocessed OxiMax Sensors | Medline ReNewal<br>Reprocessed Nellcor OxiMax<br>Sp02 Sensors | Same devices;<br>original and<br>reprocessed |
| 510(k) Number | K052186 | K041867 | K181738 | N/A |
| Common Name | Oximeter | Oximeter | Oximeter | Same |
| Regulation No. | 870.2700 | 870.2700 | 870.2700 | Same |
| Product Code | DQA | NLF | NLF | As stated |
| Models | MAXA, MAXAL, MAXN, MAXP,<br>MAXI and MAXFAST | MAXA, MAXAL, MAXN | MAXA, MAXAL, MAXP, and<br>MAXI | As stated |
| Indications for use | The Nellcor OxiMax Pulse<br>Oximetry Sensors, models<br>MAX-A, MAX-AL, MAX-N, MAX-<br>P, MAX-I and MAX-FAST are<br>indicated for single patient use<br>when continuous noninvasive<br>arterial oxygen saturation and<br>pulse rate monitoring are<br>required for patients in the sizes<br>indicated in the respective<br>sensor directions for use. | The sensor is indicated for use<br>as a non-invasive method to<br>provide continuous SpO2<br>monitoring and pulse rate. | The Medline ReNewal<br>Reprocessed Nellcor OxiMax<br>SpO2 Sensors models MAXA,<br>MAXAL, MAXP, and MAXI are<br>indicated for single patient use<br>when continuous noninvasive<br>arterial oxygen saturation and<br>pulse rate monitoring are<br>required for patients in the sizes<br>indicated in the respective<br>sensor directions for use. These<br>devices are for prescription use<br>only. | The predicate and<br>proposed devices<br>have the same<br>indications. The<br>predicate and<br>reference devices<br>share the same<br>predicate,<br>K012891. |
| Intended use | The OxiMax Pulse Oximeter<br>System is indicated for the<br>continuous non-invasive<br>monitoring of functional oxygen<br>saturation of arterial hemoglobin<br>(SpO2) and pulse rate. It is<br>intended for use with neonatal,<br>pediatric, and adult patients | None listed | The reprocessed Nellcor<br>OxiMax SpO2 Sensors are<br>indicated for the continuous<br>non-invasive monitoring of<br>functional oxygen saturation of<br>arterial hemoglobin (SpO2) and<br>pulse rate. They are intended for<br>use with pediatric and adult | Medline Renewal<br>will only reprocess<br>pulse oximeter<br>sensors. Medline<br>ReNewal will not<br>make claims for<br>neonatal use or for<br>motion or low |
| Device<br>Characteristics | Predicate | Reference | Proposed | Comparison |
| | Covidien Nellcor<br>OxiMax Pulse Oximeter<br>Sensors | Hygia Health Services<br>Reprocessed OxiMax Sensors | Medline ReNewal<br>Reprocessed Nellcor OxiMax<br>SpO2 Sensors | |
| Intended use<br>concluded | during both no motion and<br>motion conditions and for<br>patients who are either well or<br>poorly perfused, in hospitals,<br>hospital-type facilities, intra-<br>hospital transport, and home<br>environments. These devices<br>are for prescription only.…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.