The AS-3XX Series Fingertip Pulse Oximeter (with models AS-301, AS-302, AS-301-L, AS-302-L, AS-303, AS-304, AS-304-L and AS-311) is a portable non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult in hospital-type facilities, and home environments.
Device Story
Portable fingertip pulse oximeter; uses dual-wavelength spectrophotometry (660nm red, 895nm near-infrared) to measure light transmission through finger tissue. Photosensitive element captures signal; internal microprocessor processes data to calculate SpO2 and pulse rate. Results displayed on OLED screen. Operated by patient or clinician for spot-checking in home or hospital settings. Battery-powered (2x AAA). Provides non-invasive, real-time physiological monitoring to assist in assessing patient oxygenation status.
Clinical Evidence
Clinical hypoxia accuracy testing (controlled desaturation study) performed on 10 healthy, nonsmoking, light-to-dark-skinned subjects. Subject device SpO2 compared against CO-oximeter SaO2 reference across 70-100% range. Accuracy calculated using root-mean-squared (Arms) per ISO 80601-2-61. Bland-Altman analysis showed majority of data within 95% limits of agreement; outliers deemed occasional and non-concerning for safety/performance.
Technological Characteristics
Materials: ABS and TPR. Sensing: Dual-wavelength spectrophotometry (660nm/895nm). Power: 3.0V DC (2x AAA batteries). Form factor: Fingertip clamp. Connectivity: Standalone. Sterilization: Non-sterile. Standards: ISO 80601-2-61, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 10993-5/10.
Indications for Use
Indicated for spot-checking arterial hemoglobin oxygen saturation (SpO2) and pulse rate in adult patients within hospital-type facilities and home environments.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right is written in blue and white. The logo is commonly used to identify the FDA and its role in regulating food and drugs in the United States.
Xiamen Acurio Instruments Co., Ltd Angel Liu General Manager Assistant 3rd-4th Floors, BI Park, No.2028 Wengjiao West Road, Haicang District Xiamen, 361026 Cn
Re: K180685
Trade/Device Name: AS-3XX Series Fingertip Pulse Oximeter (with models AS-301, AS-302, AS-301-L, AS-302-L, AS-303, AS-304, AS-304-L, AS-311) Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA Dated: October 19, 2018 Received: October 25, 2018
Dear Angel Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Todd D. Courtney -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K180685
#### Device Name
The AS-3XX Series Fingertip Pulse Oximeter (with models AS-301, AS-302-L, AS-302-L, AS-303, AS-304, AS-304-L and AS-311)
#### Indications for Use (Describe)
The AS-3XX Series Fingertip Pulse Oximeter (with models AS-301-L, AS-301-L, AS-302-L, AS-303, AS-304, AS-304-L and AS-311) is a portable non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult in hospital-type facilities, and home environments.
Type of Use (Select one or both, as applicable)
| <div> <span></span>Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|---------------------------------------------------------------------------|
| <div> <span></span>Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) SUMMARY AS-3XX Series Fingertip Pulse Oximeter 510(k) Number: K180685
This summary of 510(k) information is being submitted in accordance with the requirements of 21 CFR 807.87(h) and 21 CFR 807.92.
#### Submitter:
Xiamen Acurio Instruments Co., Ltd 3rd-4th Floors, Biomedical Industry Park, No.2028, Wengjiao West Road, Haicang District, Xiamen, Fujian, 361026, CHINA
#### ● Contact Person:
Angel Liu General Manager Assistant Tel: 86-592-6312776 Fax: 86-592-6312776 Email: liushuzhen@acurio.cn
- Date Prepared: ● November 28, 2018
## Name of the Devices:
- Device Common Name: Oximeter
- Device Proprietary Name: AS-3XX Series Fingertip Pulse Oximeter (with models AS-301, AS-302, AS-301-L, AS-302-L, AS-303, AS-304, AS-304-L, AS-311)
- Classification Name: Oximeter
- Classification: Class II
- Regulation Number: 21CFR 870.2700
- Product Code: DQA
{5}------------------------------------------------
- Review Panel: Anesthesiology
#### Legally Marketed Predicate Device(s):
K140582 Fingertip Pulse Oximeter, Model JPD-500
#### Indications for Use:
The AS-3XX Series Fingertip Pulse Oximeter (with models AS-301, AS-302, AS-301-L, AS-302-L, AS-303, AS-304, AS-304-L and AS-311) is a portable non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult in hospital-type facilities, and home environments.
## Device Description:
The AS-301/AS-302/AS-301-L/AS-302-L/AS-303/AS-304/AS-304-L/AS-311 Fingertip Pulse Oximeter manufactured by ACURIO provides noninvasive blood oxygen measurement by the dual-wavelength spectrophotometric technique, and shows the results by the OLED. The oximeter is easy to operate, small in volume, light in weight, convenient in carrying, low consumption in design and with strong resistance to ambient light interference. 2pcs of AAA batteries can be continuously used for 20 hours, and the battery voltage can be indicated. The fingertip pulse oximeter is a portable non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adults in hospitals, hospital-type facilities, and home environments.
The components of the oximeter include the body part and a hanging rope.
Arterial oxygen saturation is measured by a method called pulse oximetry. It is a continuous, non-invasive method based on the different spectra absorption of hemoglobin and oxyhemoglobin. It measures how much light, sent from light sources on one side of the sensor, is transmitted through patient tissue (such as a finger), to a receiver on the other side. Two beams of different wavelength of lights (660nm red and 895nm near infrared light) can be focused onto a human nail tip through c clamping finger-type sensor. A measured signal obtained by a photosensitive element, will be shown on the oximeter's display through process in electronic circuits and microprocessor.
#### Comparison to Predicate Devices:
The Substantial Equivalence Comparison Chat is provided as follows:
Table 1: Substantial Equivalence Comparison Chart
{6}------------------------------------------------
| Comparison items | Manufacturer (aCurio) | Manufacturer (Jumper) | Comparison results | |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|-----------|
| Name | Fingertip pulse oximeter | Fingertip pulse oximeter | Identical | |
| Model | AS-301/AS-302/AS-301-L/AS-<br>302-L/AS-303/AS-304/AS-304-<br>L/AS-311 | JPD-500A | | |
| 510(K) number | K180685 | K140582 | | |
| Regulatory Class | Class II | Class II | Identical | |
| Intended use | This product is used for spot-<br>checking of oxygen saturation of<br>arterial hemoglobin (SpO2) and<br>pulse rate for adults. | This product is used for spot-<br>checking of oxygen saturation<br>of arterial hemoglobin (SpO2)<br>and pulse rate for adult and<br>pediatric users. | Identical | |
| Indications for use | The AS-3XX Series Fingertip<br>Pulse Oximeter (with models AS-<br>301, AS-302, AS-301-L, AS-302-<br>L, AS-303, AS-304, AS-304-L and<br>AS-311) is a portable non-<br>invasive device intended for spot-<br>checking of oxygen saturation of<br>arterial hemoglobin (SpO2) and<br>pulse rate of adult in hospitals,<br>hospital-type facilities, and home<br>environments. | The JPD-500A Fingertip Pulse<br>Oximeter is non-invasive<br>device intended for spot-<br>checking of functional oxygen<br>saturation of arterial<br>hemoglobin (SpO2) and pulse<br>rate. The portable fingertip<br>device is indicated for adult<br>and pediatric patients in home<br>and hospital environments<br>(including clinical use in<br>internist/surgery, anesthesia,<br>intensive care, etc) | Similar | |
| Target population | Adults only | Adult and pediatric users | Similar | |
| Anatomical sites | Fingertip | Fingertip | Identical | |
| Where used | hospitals, hospital-type facilities,<br>and home environments. | Home and hospital (including<br>clinical use in<br>internist/surgery, Anesthesia,<br>intensive care etc. | similar | |
| Energy used and/or<br>delivered | D.C.3.0V (2xAAA alkaline<br>batteries) | D.C. 3.0V (2×AAA alkaline<br>batteries) | Identical | |
| Human factors | | | | |
| Design | Body part and a lanyard | Body part and a lanyard | Identical | |
| Performance | Measuring<br>principle | Dual-wavelength<br>spectrophotometr<br>y | Dual-wavelength<br>spectrophotometry | Identical |
| | | | | |
| | AS-<br>301/AS302/AS30<br>3 AS-304/AS-<br>311:<br>SpO2: 36% ~<br>99%<br>Pulse rate:<br>30bpm ~ 250bpm<br>AS-301-<br>L/AS302-L:<br>SpO2: 36% ~<br>100%<br>Pulse rate:<br>30bpm ~ 250bpm | SpO2: 35%~100%<br>Pulse Rate: 25bpm ~ 250bpm | | |
| Measurement<br>range | | | | |
| | | | | |
| Measurement<br>accuracy | AS-301/AS-<br>302/AS-303/AS-<br>304/AS-311:<br>SpO2: ±3% (70%<br>- 99%), ±3%<br>(70% - 80%),<br>±2% (80% -<br>90%), ±2%<br>(90% - 99%)<br>less than 70% no<br>definition<br>Pulse rate<br>accuracy: ±2bpm<br>AS-301-L/AS-<br>302-L/AS-304-L:<br>SpO2: ±3% (70%<br>- 100%) ±3%<br>(70% - 80%),<br>±2% (80% -<br>90%), ±2%<br>(90%-99%)<br>less than 70% no<br>definition<br>Pulse rate<br>accuracy: ±2bpm | SpO2: ±2% (70% ~100%), no<br>definition (35%~69%)<br>Pulse rate: ±2bpm | Similar | |
| | | | | |
| Low<br>perfusion | SpO2 and pulse<br>rate can be | Measurement performance in<br>low perfusion condition is | Similar | |
| | | | | |
| | error | shown correctly when pulse-filling ratio is 0.2% | 0.3%. | |
| | Light sensor | RED wavelength 660nm<br>IR wavelength 940nm | RED wavelength 660nm<br>IR wavelength 940nm | Identical |
| | | | | |
| | Protection against electric shock | Internal power supply;<br>Type BF | Internal power supply;<br>Type BF | Identical |
| | Operating environment | Operation temperature: 5-40°C<br>Relative humidity: 15% - 95% (Non condensing) | Operation temperature: 5-40°C<br>Relative humidity: 15% - 80 %( Non condensing) | Similar |
| | | | | |
| | Storage environment | Storage temperature: - 20~55°C<br>Relative humidity: 15% - 95% (Non condensing) | Storage temperature: - 10°C~50°C<br>Relative humidity: 10% - 95 %( Non condensing) | Similar |
| | | | | |
| | Weight | 50g (including batteries) | 57g (including batteries) | Similar |
| | Size | AS-301, AS-301-L:<br>56(L)×34(W)×32(H)mm<br>AS-302, AS-302-L:<br>58(L)×35(W)×32(H)mm<br>AS-303:<br>57(L)×35(W)×37(H)mm | 64(L)×35(W)×34(H) mm | Similar |
| | | | | |
| | | | | |
| | | | | |
| Standards met | ISO 80601-2-61: 2011<br>IEC 60601-1: 2005<br>IEC 60601-1-2: 2007/AC | | ISO 80601-2-61: 2011<br>IEC 60601-1: 2005<br>IEC 60601-1-2: 2007/AC | Identical |
| | | | | |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
| | 510(K) Summary |
|--|----------------|
|--|----------------|
| Materials | ISO 10993-5: 2009<br>ISO 10993-10: 2010<br>ISO 13485: 2003<br>ISO 15223-1: 2012<br>IEC 62304: 2006<br>IEC 62366: 2007<br>ISO 10993-1: 2009/AC: 2010<br>IEC 60601-1-11: 2010<br>IEC 60601-1-6: 2010<br>ABS+TPR | ISO 10993-5: 2009<br>ISO 10993-10: 2010<br>ISO 13485: 2003<br>ISO 15223-1: 2012<br>IEC 62304: 2006<br>IEC 62366: 2007<br>ISO 10993-1: 2009/AC: 2010<br>IEC 60601-1-11: 2010<br>IEC 60601-1-6: 2010<br>ABS+TPR | Conform to the<br>bio-compatibility<br>requirements. |
|------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------|
| Biocompatibility | Passed the detection of ISO10993-<br>5 and ISO10993-10. | Passed the detection of<br>ISO10993-5 and ISO10993-<br>10. | Conform to the<br>bio-compatibility<br>requirements. |
| Compatibility with<br>the environment and<br>other devices | The pulse oximeter can't be used<br>in an MRI or CT environment, in<br>situations where alarms are<br>required, or in an explosive<br>atmosphere. | The pulse oximeter can't be<br>used in an MRI or CT<br>environment, in situations<br>where alarms are required, or<br>in an explosive atmosphere. | Identical |
| Sterility | Non-sterile | Non-sterile | Identical |
| Electrical safety | Same | Same | Identical |
## Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence:
The Sponsor performed a risk assessment based on the changes made to their predicate device and determined that the following non-clinical testing was necessary to demonstrate that the changes did not introduce any new risks to the subject device.
Laboratory testing was conducted to validate and verify that the Pulse Oximeters continued to meet all requirements of related international standards, including electrical safety, EMC, software, bio-compatibility. Results of these tests demonstrated compliance to the requirements of the below consensus standards and FDA Guidance documents.
#### Electrical Safety and Performance:
1. IEC 60601-1:2005+A1:2012;
2. ISO 80601-2-61:2011.
Medical Electrical Equipment and medical Electrical Systems Used in the Home
{10}------------------------------------------------
#### Healthcare Environment:
1. IEC 60601-1-11:2010+CORR.1:2011.
Electromagnetic Compatibility:
1. IEC 60601-1-2:2007.
#### Biocompatibility:
- 1. ISO 10993-5:2009
- 2. ISO 10993-10:2010
#### Software:
- 1. IEC 62304: 2006
- 2. General Principles of Software Validation Final Guidance for Industry and FDA Staff.
## Pulse Oimeters Guidance:
Pulse Oximeters - Premarket Notification Submissions [510(k)s]: Guidance for Industry and Food and Drug Administration Staff
Therefore, we conclude that the AS-3XX Series Fingertip Pulse Oximeter (with models AS-301, AS-302, AS-301-L, AS-302-L, AS-303, AS-304, AS-304-L, AS-311) is both safe and effective for its intended use.
## Discussion of Clinical Tests Performed:
Clinical hypoxia accuracy testing (controlled desaturation study) was conducted during induced hypoxia studies on 10 healthy, nonsmoking, light-to-dark-skinned subjects in an independent research laboratory. The measured arterial hemoglobin saturation value (SpO2) of the proposed device was compared with arterial hemoglobin oxygen (SaO2) value, determined from blood samples with a CO-oximeter. The accuracy of the device is in comparison with the CO-oximeter samples measured over the SpO2 range of 70-100%.
Accuracy data was calculated using the root-mean-squared (Arms value) for all subjects, per ISO 80601-2-61, the result showed that the error is far less than the scope specified in the STANDARD ; and the Agreement between Methods of Measurement with Multiple Observations per each subject was analyzed with the Bland and Altman statistics, the analysis demonstrated that the vast majority of data is within ±95% limit of agreement, the data points beyond or below this scope were regarded as outliers. By analyzing, these few outliers are occasional, which does not raise safety and performance concerns regarding the accuracy of the device.
#### Conclusions:
The AS-3XX Series Fingertip Pulse Oximeter(with models AS-301. AS-302. AS-301-L, AS-302-L, AS-303, AS-304, AS-304-L, AS-311) has the same intended use and
{11}------------------------------------------------
similar characteristics as the predicate device. Moreover, bench testing contained in this submission demonstrates that any difference in their technological characteristics do not raise any different questions of safety or effectiveness. Thus, the AS-3XX Series Fingertip Pulse Oximeter (with models AS-301, AS-302, AS-301-L, AS-302-L, AS-303, AS-304, AS-304-L, AS-311) is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.