OXIPROBE, MODELS BM-100, BM-200, BM-400, BM-300, BM-300S, BM-600P

K092549 · Bio Medical Technologies Co., Ltd. · DQA · Nov 16, 2009 · Cardiovascular

Device Facts

Record IDK092549
Device NameOXIPROBE, MODELS BM-100, BM-200, BM-400, BM-300, BM-300S, BM-600P
ApplicantBio Medical Technologies Co., Ltd.
Product CodeDQA · Cardiovascular
Decision DateNov 16, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2700
Device ClassClass 2
AttributesPediatric

Indications for Use

The OxiprobE is indicated for non-invasive, continuus, beat-by-beat monitoring of oxygen saturation of functional arterial hemoglobin, pulse rate and pulse amplitude. Individual models are labeled specifically for either adult or pediatric use. It is for prescription use only.

Device Story

OxiprobE consists of pulse oximetry sensors and cables; inputs include optical signals from patient tissue; device measures blood oxygen saturation, pulse rate, and pulse amplitude; disposable and reusable sensor configurations available; disposable sensors use adhesive for secure positioning; reusable sensors intended for short-term/intermittent monitoring of relatively immobile patients; device connects to compatible pulse oximetry monitors; healthcare providers use output for continuous patient monitoring; facilitates clinical decision-making regarding patient oxygenation status; benefits include non-invasive, real-time monitoring and reduced cross-contamination risk with single-use options.

Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and design similarities to predicate devices.

Technological Characteristics

Pulse oximetry sensor system consisting of connector, cable, and sensor. Available in disposable (adhesive) and reusable configurations. Sensing principle based on optical light source and photo detector for non-invasive measurement of arterial hemoglobin saturation. Designed for specific patient populations (adult/pediatric) based on model. Non-sterile.

Indications for Use

Indicated for non-invasive, continuous, beat-by-beat monitoring of arterial hemoglobin oxygen saturation, pulse rate, and pulse amplitude in adult and pediatric patients. Specific models are designated for adult use (BM-100, BM-200, BM-300, BM-300s) or pediatric use (BM-400, BM-600p). Pediatric models are weight-restricted (BM-600p for 10-50 kg). Prescription use only.

Regulatory Classification

Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K092549 Image /page/0/Picture/1 description: The image shows a diagram of a molecule. The molecule is made up of three atoms, labeled as M, B, and M. The atoms are connected by lines, indicating that they are bonded together. The text below the diagram says 'Bimetallic homogeneous molecular catalyst'. Medical Technologies Co., Ltd 1 Bldg. #431-17. Shindang-dong, Jung-gu,Seoul, 100-452. Korea cl.822.2234.0780~2 Fax: 822.2234.0748 ## NOV 1 6 2009 ## 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92(c). Date: Aunistrator 6, 2009 1. Company and Correspondent making the submission: | | Company | |--------------|-----------------------------------------------------------------------------------| | Name | Bio Medical Technologies Co., Ltd. | | Address | Megu Bldg. #431-17. Shindang-dong, Jung-<br>gu, Seoul, 100-452, Republic of Korea | | Phone<br>Fax | +82 2 2234-0780-2<br>+82 2 2234-0748 | | Contact | K.N. Kim | #### 2. Device: Proprietary Name - OxiprobE Common Name - Oximetry Sensor Classification Name - Oximeter #### 3. Predicate Device: Model D-25, N-25, D-20, Nellcor Puritan Bennett Inc., K993637 Philips Reusable SpO2 Sensors-M1191T, M1192T, Philips Medical Systems, K063783 ## 4. Classifications Names & Citations: Regulation Number: 21 CFR 870.2700 Regulatory Class: Class II Regulation Name: Oximeter Product Code: DQA 5. Description: The Oxiproble is Pulse Oximetry sensors and cables used for non-invasive measurement of blood oxygen saturation and pulse rute measurement. It consists of Connector, Cable and sensor. Bio Medical Technologies Co., Ltd. supplies both disposable and reusable sensors. Disposable sensors are designed to enable the sensor's light source and photo detector to be securely and properly positioned on the patient. The adhesive stabilizes {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows a logo with three circles connected by lines. The top circle contains the letters 'm' and 'm', the left circle contains the letters 'b' and 'b', and the right circle contains the letters 't' and 't'. Below the circles, the text 'An Medical Technologies' is written, followed by 'www.am-tech.co.kr'. ### Bio Medical Technologies Co.,Ltd legu Bldg. #431-17. Shindang-dong, Jung-gu,Seoul, 100-452, Korea Tel.822.2234.0780~2 Fax.822.2234.0748 these important optical components and provides a comfortable fit. Adhesive sensors are patient-dedicated and can travel with your patients. Single-use sensors do not present the risk of cross-contamination caused by products that are reused from patient up patient. Reusable sensors are designed to monitor various sizes of relatively immobile patients. When short-term or intermittent monitoring is necessary, these reusable, nonsterile sensors are an effective monitoring alternative. The Oxiproble models BM-100, BM-200 and BM-300 are indicated for adult patients. The Oxiproble model BM-400 is indicated for pediatric patients. The Oxiproble models BM-300s is indicated for adult patients over 30 kg. The OxiproblE models BM -600p is indicated for pediatric patients over 10 kg and less than 50 kg. - ర్. Indication for use: The OxiprobE is indicated for non-invasive, continuus, beat-by-beat monitoring of oxygen saturation of functional arterial hemoglobin, pulse rate and pulse amplitude. Individual models are labeled specifically for either adult or pediatric use. It is for prescription use only. #### 7. Review: The OXIPROBE has the same device characteristics as the predicate device, the intended use, materials, design and use concept are similar. Based on the comparison of intended use and technical features, the OXIPROBE is substantially equivalent to the predicate device. ## 8. Conclusions: In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification Bio Medical Technologics Co., Ltd. concludes that the OXIPROBE are safe and effective and substantially equivalent to predicate devices as described herein. - 9. Bio Medical Technologies Co., Ltd. will update and include in this summary any other information deemed reasonably necessary by the FDA. END {2}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird, with its wings forming a flowing, abstract design. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002 Bio Medical Technologies Company, Limited C/O Ms. Cathryn Cambria Arkin Consulting Group, L.L.C. 5536 Trowbridge Drive Dunwoody, Georgia 30338 NOV 1 6 2009 Re: K092549 Trade/Device Name: OxiprobE Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II . Product Code: DQA Dated: August 6. 2009 Received: August 19, 2009 Dear Ms. Cambria: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ #### Page 2- Ms. Cambria Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, for Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ # Indications for Use 510(k) Number K Device Name: OxiprobE Indication for use: The OxiprobE is indicated for non-invasive, continuus, beat-by-beat monitoring of oxygen saturation of functional arterial hemoglobin, pulse rate and pulse amplitude. Individual models are labeled specifically for either adult or pediatric use. It is for prescription use only. Prescription Use N ાર (Per 21CFR801 Subpart D) Over-The-Counter Use (Per 21CFR807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) L. Schultheis (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: k 092549
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...