K083557 · Smiths Medical Pm, Inc. · DQA · May 1, 2009 · Cardiovascular
Device Facts
Record ID
K083557
Device Name
BCI WW1030 PULSE OXIMETER
Applicant
Smiths Medical Pm, Inc.
Product Code
DQA · Cardiovascular
Decision Date
May 1, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The BC1® Model WW1030 pulse oximeter is intended to be used for continuous monitoring of a patient's functional oxygen saturation (%SpO2), pulse rate, pulse signal strength, and pulse amplitude index readings. It is equipped with audible and visual alarms. It may be used by physicians, respiratory therapists, nurses, certified nurse assistants, emergency medical technicians, sleep technicians, clinicians and home use. The intended patient population ranges from neonatal to adult. It can be used on patients with low perfusion or during patient motion. The WW 1030 may be used in the hospital or clinical environment, during emergency land transport, and the home.
Device Story
Handheld pulse oximeter for continuous monitoring of SpO2, pulse rate, perfusion index, and signal strength; utilizes Digital Micro Power OEM oximeter board (31402B2) to process sensor inputs; displays data via blue LED; features audible/visual alarms; operated by clinicians or home users; compatible with various BCI and Nellcor sensors; supports accessories including docking station, thermal printer, and USB connectivity; provides real-time physiological data to assist clinical decision-making; benefits patients through continuous monitoring in diverse environments including transport and home.
Clinical Evidence
Clinical evidence consists of three desaturation studies (Desat 37, 38, 39) using the Digital Micro Power OEM oximeter board (31402B2). Studies 37 and 38 evaluated SpO2 accuracy across the 70-100% SaO2 range compared to a reference CO-oximeter. Study 39 evaluated SpO2 accuracy under controlled motion conditions across the 70-100% SaO2 range. Results confirmed performance within acceptable limits.
Technological Characteristics
Handheld pulse oximeter; utilizes Digital Micro Power OEM oximeter board (31402B2); LED display; 5-button keypad; battery-powered (disposable or rechargeable pack); compatible with BCI and Nellcor sensors; includes audible/visual alarms; connectivity via USB and nurse call cables; tested for EMC, electrical, and mechanical durability.
Indications for Use
Indicated for continuous monitoring of functional oxygen saturation (%SpO2), pulse rate, pulse signal strength, and perfusion index in patients ranging from neonates to adults. Suitable for use in hospital, clinical, home, and emergency land transport settings. Includes patients with low perfusion or during motion.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a handwritten string of alphanumeric characters. The string appears to be "K08 3357". The characters are written in a simple, slightly slanted style, with some variation in line thickness.
## 510(k) Summary of Safety and Effectiveness
| Manufacturer: | Smiths Medical |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | N7 W22025 Johnson Drive<br>Waukesha, WI 53186 |
| | MAY - 1 2009 |
| Telephone Number: | (262) 542-3100 |
| Fax Number: | (262) 542-3325 |
| Contact Person: | Donald Alexander<br>Director Regulatory Affairs |
| Date Prepared: | December 1, 2008 |
| Proprietary Name: | BCI® model WW1030 Handheld Pulse Oximeter |
| Common/Classification Name: | Pulse Oximeter |
| Product Code: | DQA |
| Regulation Number: | 870.2700 (Class II) |
| Predicate Devices: | Masimo SET Radical pulse oximeter (K040214)<br>BCI® Autocorr model 3304 pulse oximeter (K070732)<br>BCI® MiniCorr model 3402 pulse oximeter (K991410) |
#### Device Description:
The aim of this submission is to obtain market clearance for the BCI model WW1030 pulse oximeter.
The BCI model WW1030 is a handheld pulse oximeter intended for continuous monitoring. The WW1030 monitors and displays patient functional oxygen saturation %SpO2, pulse rate, perfusion index, and pulse signal strength information. It is equipped with audible and visual patient alarms. The user interface includes a blue LED display, speaker, a 5-button keypad control, and an ON/OFF button.
The WW1030 comes with disposable batteries, a reusable pulse oximetry sensor with extension cable and relevant manuals. Optional accessories include other oximetry sensors, docking station, thermal printer (attaches to dock), rechargeable battery pack, universal mains AC charger, patient isolated USB cable, nurse call cables, universal mounting bracket and protective glove. The WW1030 is compatible with BCI oximetry sensors, Nellcor DS100A oximetry sensor and extension cables.
#### Indications for Use
The BC10 model WW1030 handheld pulse oximeter is intended to be used for continuous monitoring of a patient's functional oxygen saturation (%SpO2), pulse rate, pulse signal strength, and perfusion index readings. It is equipped with audible and visual alarms. It may be used by physicians, respiratory therapists, nurses, certified nurse assistants, emergency medical technicians, sleep technicians, clinicians and home users. The intended patient population ranges from neonatal to adult. It can be used on patients with
## FDA CDRH DMC
DEC 2 2008
Received
{1}------------------------------------------------
low perfusion or during patient motion. The WW1030 may be used in the hospital or clinical environment, during emergency land transport and in the home.
#### Risk Mitigation Table
Below is a summary of risks common to pulse oximeters and how this submission addresses those risks.
| Identified Risk | Mitigation Measures |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Inadequate Device Performance | Accuracy Testing (Section 9)<br>Benchtop Performance Testing (Section 9)<br>Software Documentation and Testing<br>(Section 8) |
| Electrical or Mechanical Failure | Electrical, Mechanical, and Environmental<br>Testing (Section 9) |
| Electromagnetic Interference | Electromagnetic Compatibility Testing<br>(Section 9) |
| Adverse Tissue Reactivity | Biocompatibility Information (Section 10) |
| Cross-contamination or Infection | Cleaning and Disinfection Testing (Section<br>9) |
| Improper Use | Labeling (Section 7) |
An in-depth risk management analysis, including mitigation measures, was performed on the BCI WW1030 system. The results are provided under Section 5.
#### Performance Testing
The BCI WW1030 pulse oximeter passed all performance bench top testing including EMC, electrical, mechanical durability, safety (operator and patient), and temperature/humidity. Test reports, including test protocols, pass/fail criteria, results and conclusions, are provided under Section 9.
#### Clinical Testing
The Digital Micro Power OEM oximeter board (31402B2) is a satellite board capable of monitoring a patient's SpO2 level, pulse rate and plethysmogram waveform. The Digital Micro Power OEM oximeter board is the pulse oximetry board in the BCI WW1030 pulse oximeter. This board has undergone three separate desaturation clinical studies: Desat 37. Desat 38, and Desat 39. The intent of the Desat Studies 37 and 38 was to determine the SpO2 accuracy of the pulse oximeter module using various sensors over the range 70-100% SaO2 as determined by the reference CO-oximeter. The intent of Desat Study 39 was to obtain clinical data to determine the SpO2 accuracy over the range 70-100% SaO2 as determined by a reference CO-oximeter under controlled motion conditions.
{2}------------------------------------------------
### Conclusion
Supporting information per this premarket submission confirms that the BCI® WW1030 Pulse Oximeter is substantially equivalent to its predicate devices.
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
{3}------------------------------------------------
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## MAY - 1 2009
Mr. Donald Alexander Director Regulatory Affairs Smiths Medical PM, Incorporated N7 W22025 Johnson Drive Waukesha, Wisconsin 53186-1856
Re: K083557
Trade/Device Name: BCI® Model WW1030 Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: December 1, 2008 Received: December 2, 2008
Dear Mr. Alexander:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration
{4}------------------------------------------------
Page 2- Mr. Alexander
and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Parl 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Susan Rinnes
Susan Runner, D.D.S., MA Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number: K083557
Device Name: BCI Model® WW1030 Handheld Pulse Oximeter
Indications for Use: The BC1® Model WW1030 pulse oximeter is intended to be used for continuous monitoring of a patient's functional oxygen saturation (%SpO2), pulse rate, pulse signal strength, and pulse amplitude index readings. It is equipped with audible and visual alarms. It may be used by physicians, respiratory therapists, nurses, certified nurse assistants, emergency medical technicians, sleep technicians, clinicians and home use. The intended patient population ranges from neonatal to adult. It can be used on patients with low perfusion or during patient motion. The WW 1030 may be used in the hospital or clinical environment, during emergency land transport, and the home.
L. Schutte
(Division Sign-Off) Division of Anesthestology, General Hospital Infection Control, Dentai Devices
K08355 510(k) Number:
Prescription Use XXX (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.