MD2000A VITAL SIGN MONITOR, MODEL MD2000A

K083466 · Beijing Choice Electronic Technololgy Co., Ltd. · DQA · Apr 3, 2009 · Cardiovascular

Device Facts

Record IDK083466
Device NameMD2000A VITAL SIGN MONITOR, MODEL MD2000A
ApplicantBeijing Choice Electronic Technololgy Co., Ltd.
Product CodeDQA · Cardiovascular
Decision DateApr 3, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2700
Device ClassClass 2
AttributesPediatric

Indications for Use

The MD2000A vital sign monitor is portable device indicated for use in non-invasively measuring and displaying functional oxygen saturation of arterial haemoglobin (SpO2), and pulse rate of adult, pediatric and neonate patients via finger in hospitals, medical facilities, and subacute environments. The vital sign monitor is intended for spot-checking and/or continuous monitoring of patients, the sensor of device is reusable

Device Story

MD2000A Vital Sign Monitor is a portable, prescription-use pulse oximeter for clinical and subacute settings. It utilizes a finger-applied sensor with dual light sources (660 nm red; 940 nm ultra-red) and a photodetector to measure light absorption in pulsating arteriolar vascular beds. The device calculates SpO2 based on the ratio of light absorption during systole and diastole. It provides real-time SpO2, pulse rate, and plethysmogram data via LCD and LED displays. Features include three-level visual/audible alarms, 72-hour data storage (recorded every 4 seconds), and trend review capabilities. Powered by AC or rechargeable Ni-MH battery. Clinicians use the output for patient monitoring to assess oxygenation status and heart rate, facilitating clinical decision-making regarding patient respiratory and cardiovascular health.

Clinical Evidence

Clinical testing conducted in a laboratory setting following ISO 9919:2005 standards to validate SpO2 and pulse rate measurement accuracy. No further clinical study details provided.

Technological Characteristics

Portable pulse oximeter; dual-wavelength (660nm/940nm) optical sensing; Ni-MH battery (7.2V) or AC power (100-230V, 50/60Hz); LCD and LED display interfaces; reusable finger sensors (adult/pediatric/neonate); non-sterile; data storage capacity of 72 hours. Complies with IEC 60601-1 and IEC 60601-1-2.

Indications for Use

Indicated for non-invasive measurement and display of SpO2 and pulse rate in adult, pediatric, and neonate patients. Used for spot-checking or continuous monitoring in hospitals, medical facilities, and subacute environments.

Regulatory Classification

Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

Submission Summary (Full Text)

{1}------------------------------------------------ # Chapter III 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 9807.92 The Assigned 510(k) Number is: K083466 #### 1. Device Information Device Common Name: Oximeter Device Trade/Proprietary Name: MD 2000A Vital Sign Monitor #### Classification Information: (1) Classification Name: Oximeter - (2)Regulation Number:870.2700 - (3)Product Code:DQA (4) Class: II - (5)Review Panel: Anesthesiology #### 2. Submitter Information #### Manufacturer: Beijing Choice Electronic Technology Co., Ltd. Room 1127-1128 Building B, Bailangyuan Fuxing Road , No. A36,Beijing, China 100039 #### Contact Person of the Submission Ms. Diana Hong Mr. Tarzan Wang Suite 8D, Zhongxin Zhongshan Mansion, No.19, Lane 999, Zhong Shan Nan Er Road Shanghai, China 20020 Phone: +86-21-64264467 Fax: (240) 238-7587 Email: Diana.hong@mid-link.net #### 3. Device Description The device works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of ### APR - 3 2009 {2}------------------------------------------------ ・・ light source is 660 nm, which is red light; the other is 940 nm, which is ultra red light. Skin, bone, tissue, and venous vessels normally absorb a constant amount of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2. The applicant device of MD2000A Vital Sign Monitor is the Desktop Pulse Oximeter Monitor, which mainly function are measurement, display, alarm, data storage. #### Measurement function That's the mainly function of the device, which use the method describe above to measure the SpO2 value and Pulse Rate value of lay user. #### Display function The display function of the device have two display area dividually, the one display area use the LCD display mode, which displays the date & time, battery capacity indicator, SpO2 and PR alarm limit value, ID number, auditory alarm state indication, pulse beep state indication and pulse plethysmogram. The another display area use the LED display mode, which displays the SpO2 value, PR value and blip bar. #### Alarm function The applicant device has there-level priorities alarm, they are High priority, Medium priority and Low priority. Each alarm of device contains VISUAL and AUDIBLE alarm. #### Data storage function The measured record is stored automatically every four seconds. The monitor can store 72 hours records. The new record will be stored with the initial records being erased when the stored records are full. The device also has the function as "record review", "SpO2 Trend Review" and "PR Trend review", the user can use those function to review stored data easily. #### Power The applicant device has two kinds of power supply as AC power supply and Battery power supply. The AC power supply complies with following specification: 100-230 (VAC), {3}------------------------------------------------ #### 50/60 (Hz). The battery used in the applicant device is a Ni-MH battery with 7.2V d.c. output, the battery be able to recharged by AC power supply. The battery be able to provide the power to make the device run in normal, when the AC power supply connected, the battery will turn into charge state from power supply state automatically. The applicant device have 2 models detachable sensor as the accessory, they are listed below: M-50A (for Adult) M-50C (for Pediatric and Neonate) - . The applicant device is not for life-supporting or life-sustaining, not for implant. The device is not sterile and does not need sterilization or re-sterilization. The device is for prescription. The device does contain drug or biological product. #### 4. Intended Use The MD2000A vital sign monitor is portable device indicated for use in non-invasively measuring and displaying functional oxygen saturation of arterial haemoglobin (SpO2), and pulse rate of adult, pediatric and neonate patients via finger in hospitals, medical facilities, and subacute environments. The vital sign monitor is intended for spot-checking and/or continuous monitoring of patients, the sensor of device is reusable #### 5. Substantially Equivalence Determination The applicant devices Substantially Equivalent (SE) to the predicate device which is US legally market device. Therefore, the applicant devices are determined as safe and effectiveness. #### 6. Test Summary The device is electrically operated and the electrical safety and electromagnetic compatibility following IEC 60601-1 and IEC60601-1-2 were conducted. The Clinical Test Report following ISO 9919:2005, Medical electrical equipment-Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use are conducted in the Clinical Laboratory. {4}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Beijing Choice Electronic Technology Company, Limited C/o Ms. Diana Hong General Manager Shanghai Midlink Business Consulting Company, Limited Suite 8D, Zhongxin Zhongshan Mansion Number 19, Lane 999, Zhong Shan Nan Er Road Shanghai, CHINA 20020 APR - 3 2009 Re: K083466 > Trade/Device Name: MD 2000A Vital Sign Monitor Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: February 27, 2009 Received: March 3, 2009 Dear Ms. Kong: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Page 2 - Ms. Kong Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Susan Loar Ginette Y. Michaud, M.D. Ginette Y. Michaud, M.D. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ Premarket Notification 510(k) Chapter II Indication for Use Statement Report No .: TE20080902 ## Indication For Use 510(k) Number (if known): _ Pending __________________________________________________________________________________________________________________________________________ Device Name: MD 2000A Vital Sign Monitor Indications for Use: The MD2000A vital sign monitor is portable device indicated for use in non-invasively measuring and displaying functional oxygen saturation of arterial haemoglobin (SpO2), and pulse rate of adult, pediatric and neonate patients via finger in hospitals, medical facilities, and subacute environments. The vital sign monitor is intended for spot-checking and/or continuous monitoring of patients, the sensor of device is reusable .. Prescription Use (Part 21 CFR 801 Subpart D) 510(k) Number: AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Susankunnes (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...