LIFESENSE, MODEL LS1-9R

K063752 · Nonin Medical, Inc. · DQA · May 4, 2007 · Cardiovascular

Device Facts

Record IDK063752
Device NameLIFESENSE, MODEL LS1-9R
ApplicantNonin Medical, Inc.
Product CodeDQA · Cardiovascular
Decision DateMay 4, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2700
Device ClassClass 2
AttributesPediatric

Indications for Use

The LifeSense® Model LS1-9R Capnography/Pulse Oximeter monitor indicated for use in simultaneously measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2), end tidal carbon dioxide (ETC02), respiration and pulse rate of well or poorly perfused adult, pediatric and infant patients. It is intended for use in environments where patients require continuous, non-invasive monitoring of these parameters by a healthcare professional, e.g. hospitals, medical facilities, post-operative care, patient transport, home-use, or any emergency medical services and environments.

Device Story

Portable monitor for continuous, non-invasive measurement of SpO2, ETCO2, respiration, and pulse rate. Inputs include SpO2 sensors and ETCO2 sampling lines (airway adapter, nasal cannula, or nasal cannula with oxygen input). Device processes physiological signals to display real-time numerical values and waveforms on an LCD touch panel. Used by healthcare professionals in hospitals, post-operative care, transport, and home environments. Provides clinicians with continuous monitoring data to assist in patient assessment and clinical decision-making. Benefits include early detection of respiratory or oxygenation changes in intubated or spontaneously breathing patients.

Clinical Evidence

Bench and clinical testing performed to demonstrate functional equivalence to predicate devices. No specific performance metrics (e.g., sensitivity, specificity) provided in the summary document.

Technological Characteristics

Portable LCD monitor with touch panel interface. Measures SpO2 via pulse oximetry and ETCO2 via sampling lines (airway adapter/cannula). Connectivity includes display of graphs and numerical parameters. Intended for continuous monitoring of adult, pediatric, and infant patients.

Indications for Use

Indicated for continuous, non-invasive monitoring of SpO2, ETCO2, respiration, and pulse rate in well or poorly perfused adult, pediatric, and infant patients in clinical, transport, or home-use settings.

Regulatory Classification

Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K063752 # SECTION 2. SUMMARY AND CERTIFICATION #### 510(k) Summary A. MAY - 4 2007 | Submitter: | Nonin Medical, Inc. | |-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Lori M. Roth<br>Clinical/Regulatory Specialist<br>Nonin Medical, Inc.<br>13700 1st Ave. North<br>Plymouth, MN 55441-5443 | | Date Prepared: | December 15, 2006 | | Trade Name: | LifeSense® Model LS1-9R Capnography/Pulse Oximeter<br>Monitor | | Classification Name:<br>and Number: | Class II, 21 CFR 868.1400 | | Product Code: | DQA, CCK | | Predicate Device(s): | Nonin's LifeSense® Model LS1-9R is substantially equivalent to the MicroCap Plus/NPB-75 Capnograph/Pulse Oximeter manufactured by Oridion Medical, LTD cleared by the FDA under K024300 on 4/03/03 and the Capnocheck® Plus manufactured by BCI International, Inc. cleared under K970209 on 4/18/97. | | Device Description: | The LifeSense® Model LS1-9R is a lightweight, portable Capnography/Pulse Oximeter monitor indicated for use in simultaneously measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2), end tidal carbon dioxide (ETC02), respiration and pulse rate of adult, pediatric, and infant patients. It is intended for use in any environment where patients require continuous, non-invasive monitoring of intubated or spontaneous breathing patients. | | | When measuring ETCO2 the patient is attached to the monitor by a simple sample line that can be either an airway adapter for an endotracheal tube, a nasal cannula, or a nasal cannula with oxygen input. | | | The liquid crystal display (LCD) monitor of the Model LS1-9R displays factory default or operator defined | {2}------------------------------------------------ parameters and alarm settings, ETCO2 and respiration graphs, battery status and fault messages. Operator defined settings can be made on the touch panel display. LifeSense Model LS1-9R monitor is intended for prescription use with adult, pediatric, and infant patients. Intended Use: The LifeSense® Model LS1-9R Capnography/Pulse Oximeter monitor indicated for use in simultaneously measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2), end tidal carbon dioxide (ETC02), respiration and pulse rate of well or poorly perfused adult, pediatric and infant patients. It is intended for use in environments where patients require continuous, non-invasive monitoring of these parameters by a healthcare professional, e.g. hospitals, medical facilities, post-operative care, patient transport, home-use, or any emergency medical services and environments. Safety Testing: Nonin's Model LS1-9R Capnography/Pulse Oximeter monitor has undergone both bench and clinical testing in order to demonstrate that it has appropriate functional features and is substantially equivalent to the predicate devices. Conclusion: Nonin's LifeSense Model LS1-9R is substantially equivalent to the MicroCap Plus/NPB-75 Capnograph/Pulse Oximeter manufactured by Oridion Medical, LTD cleared by the FDA under K024300 and the Capnocheck® Plus manufactured by BCI International, Inc. under K970209. Functional and {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its wings. The eagle is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA". Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Lori M. Roth Clinical/Regulatory Specialist Nonin Medical, Incorporated 13700 1st Avenue North Plymouth, Minnesota 55441-5443 MAY - 4 2007 Re: K063752 Trade/Device Name: Nonin LifeSense® Model LS1-9R Capnography/ Pulse Oximeter Monitor Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DOA, CCK Dated: April 25, 2007 Received: April 27, 2007 Dear Ms. Roth: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ ### Page 2 – Ms. Roth Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Susan Runne Chiu Lin, Ph.D. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health #### Enclosure {5}------------------------------------------------ ## Indications for Use Statement ### 510(K) Number: K063752 ### Device Name: ### Nonin LifeSense® Model LS1-9R Capnography/Pulse Oximeter Monitor _____________________________________________________________________________________________________________ ### Indications for Use: The LifeSense® Model LS1-9R Capnography/Pulse Oximeter monitor indicated for use in simultaneously measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2), end tidal carbon dioxide (ETCO2), respiration and pulse rate of well or poorly perfused adult, pediatric and infant patients. It is intended for use in environments where patients require continuous, non-invasive monitoring of these parameters by a healthcare professional, e.g. hospitals, medical facilities, post-operative care, patient transport, home-use, or any emergency medical services and environments. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) | Concurrence of CDRH, Office of Device Evaluation (ODE) | | |--------------------------------------------------------|---------------------------------------------| | | | | | <div style="text-align:left;">Cheffto</div> | | (Division Sign-Off) | | | Division of Anesthesiology, General Hospital, | | | Infection Control, Dental Devices | | | | | | 510(k) Number: | K063752 | Nonin Medical, Inc. LifeSense® Model LS1-9R Traditional 510(K): Premarket Notification Confidential
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...