OXIMAX NEOMAX ADHESIVE FOREHEAD REFLECTANCE OXIMETRY SENSOR

K051271 · Nellcor Puritan Bennett, Inc. · DQA · Jan 10, 2006 · Cardiovascular

Device Facts

Record IDK051271
Device NameOXIMAX NEOMAX ADHESIVE FOREHEAD REFLECTANCE OXIMETRY SENSOR
ApplicantNellcor Puritan Bennett, Inc.
Product CodeDQA · Cardiovascular
Decision DateJan 10, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2700
Device ClassClass 2
AttributesPediatric

Indications for Use

The Nellcor® OxiMax® adhesive forehead reflectance sensor, model NeoMAX, is indicated for single patient use when continuous noninvasive arterial oxygen saturation and pulse rate monitoring are required for neonatal and pediatric patients (≤ 40 kg).

Device Story

OxiMax NeoMAX is a sterile, latex-free, single-patient-use adhesive forehead reflectance sensor. Designed for neonatal and pediatric patients (≤ 40 kg). Device functions as a pulse oximetry sensor; inputs optical signals from forehead tissue; contains an internal memory chip storing sensor-specific data (Advanced Signal Evaluation, lot code, revision, model). Operates by connecting to compatible OxiMax pulse oximetry monitors. Monitors process signals to provide continuous noninvasive arterial oxygen saturation and pulse rate data. Used in clinical settings; operated by healthcare professionals. Output allows clinicians to monitor patient oxygenation status, facilitating clinical decision-making regarding respiratory support or intervention. Benefits include noninvasive, continuous monitoring of vital signs in vulnerable patient populations.

Clinical Evidence

Human oxygenation evaluations were conducted to confirm conformance to accuracy and precision specifications. Bench testing was also performed. No specific numerical results (e.g., sensitivity, specificity) provided in the summary.

Technological Characteristics

Adhesive forehead reflectance sensor; sterile; latex-free. Contains integrated memory chip for sensor data storage. Compatible with OxiMax pulse oximetry monitors. Designed for neonatal/pediatric use (≤ 40 kg).

Indications for Use

Indicated for neonatal and pediatric patients (≤ 40 kg) requiring continuous noninvasive arterial oxygen saturation and pulse rate monitoring.

Regulatory Classification

Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K05/27/ 4280 Haclenda Drive Tele: 925 463-4000 Fax: 925 463-4020 Pleasanton, CA 94588 tyco Healthcare Nellcor 510(k) Summary JAN 1 0 2006 | Submitted by: | Nellcor Puritan Bennett, Inc.<br>4280 Hacienda Drive<br>Pleasanton. CA 94588 | |--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Company Contact: | James Patrick Garvey II, RRT<br>Regulatory Affairs Manager<br>(925) 463-4479<br>(925) 463-4020 - FAX | | Date Summary Prepared: | May 4, 2005 | | Trade Name: | OxiMax NeoMAX Adhesive Forehead Reflectance<br>Oximetry Sensor | | Common/Usual Name: | Oxygen Sensor | | Classification Name: | Oximeter (DQA) per 21 CFR §870.2700 | | Substantially Equivalent<br>Devices: | • Nellcor Puritan Bennett, Inc.. OxiMax Pulse Oximet<br>System with N-595 Pulse Oximeter and OxiMax<br>Sensors, K012891<br>• Nellcor Puritan Bennett, Inc., OxiMax MAX-FAST<br>Adhesive Forehead Sensor, K021089 | ### DEVICE DESCRIPTION The OxiMax NeoMAX is a sterile, latex-free, single patient use forehead sensor, sized appropriately to fit the forehead of a neonatal or pediatric patient. The OxiMax NeoMAX sensor contains a memory chip carrying information about the sensor which the oximeter needs for correct operation, including in-sensor data. Advanced Signal Evaluation, lot code and data set revision, and sensor model. The OxiMax NeoMAX sensor is compatible with OxiMAX monitors. ### INTENDED USE The Nellcor® OxiMax® adhesive forehead reflectance sensor, model NeoMAX, is indicated for single patient use when continuous noninvasive arterial oxygen saturation and pulse rate monitoring are required for neonatal and pediatric patients (≤ 40 kg). {1}------------------------------------------------ # SUMMARY OF TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE The OxiMax NeoMAX sensor has the same technological characteristics as the OxiMax MAX-FAST. The differences relate to dimensions. lighter adhesive on the patient contact surface, and labeling. # TESTS PERFORMED TO SUPPORT DETERMINATION OF SUBSTANTIAL EQUIVALENCE Human and bench tests were performed to support the determination of substantial equivalence. Human oxygenation evaluations were conducted to confirm conformance to accuracy and precision specifications. ## CONCLUSIONS The technological characteristics of the OxiMax NeoMAX sensor and the results of testing do not raise new questions of safety or effectiveness when compared to the legally marketed predicate device. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized image of an eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is surrounded by the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" in a circular arrangement. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN I 0 2006 Mr. James P. Garvey Regulatory Affairs Manager Nellcor Puritan Bennett, Incorporated 4280 Hacienda Drive Pleasanton, California 94588 Re: K051271 Trade/Device Name: OxiMax NeoMax Adhesive Forehead Reflectance Regulation Number: 21 CFR 870.2700 Regulation Name: Oximcter Regulatory Class: II Product Code: DQA Dated: December 22, 2005 Received: December 23, 2005 Dear Mr. Garvey: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ ### Page 2 - Mr. Garvey Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-frec number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Sutte if. Michin Omd. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ ### Indications for Use 510(k) Number (if known): Device Name: Indications for Use: The Nellcor® OxiMax® NeoMAX adhesive forchead reflectance sensor is indicated for single patient use when continuous noninvasive arterial oxygen saturation and pulse rate monitoring are required for neonatal and pediatric patients (≤ 40 kg). ﻟﺴ Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Suette y. Michael Ond Henry General Hospital, Rehabilitation Services
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...