The Masimo SET RAD-5v Pulse Oximeter is intended for non-continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for adult, pediatric, and neonatal patients in hospitals, hospital-type facilities, and mobile environments.
Device Story
Handheld pulse oximeter; monitors SpO2 and pulse rate noninvasively. Input: optical signals from finger-mounted SpO2 sensor. Operation: uses Masimo SET technology to differentiate arterial blood from venous blood and noise; calculates ratio of pulsatile absorbance signals; compares against empirically derived look-up table. Output: continuous numeric SpO2 and pulse rate values; LED indicator bars for Pulse Amplitude Index (PAI) and Signal Quality (SQ); audible alarms for sensor-off, low battery, and physiological limits. Used by clinicians in hospitals, mobile, or home environments. Provides real-time data to assist in patient assessment and intervention decisions. Benefits: accurate monitoring during motion and low perfusion states.
Clinical Evidence
Clinical studies conducted on healthy adult volunteers and neonates. Evaluated performance during induced hypoxia, motion (rubbing/tapping), and low perfusion conditions. Compared device readings against arterial blood CO-Oximeter measurements and ECG monitors. Results: SpO2 accuracy ±2 digits (no motion), ±3 digits (motion); pulse rate accuracy ±3 digits (no motion), ±5 digits (motion). Neonatal accuracy adjusted for fetal hemoglobin. Studies followed 21 CFR Part 812 and Part 56.
Technological Characteristics
Handheld pulse oximeter; 4 AA battery power source; 36-hour runtime. Microprocessor-controlled; LED 7-segment display. Sensing principle: optical absorption of red and infrared light. Connectivity: sensor cable interface. Environmental: 5°C to 40°C operating range. Software: proprietary Masimo SET algorithm for signal processing and noise reduction.
Indications for Use
Indicated for non-continuous, noninvasive monitoring of functional arterial hemoglobin oxygen saturation (SpO2) and pulse rate in adult, pediatric, and neonatal patients. Applicable during motion and no-motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Masimo SET Radical Pulse Oximeter with SatShare and LNOP series of Sensors and Cables (K031330)
{0}------------------------------------------------
MAR 1 9 2004
# 510(k) SUMMARY
Image /page/0/Picture/2 description: The image shows the logo and contact information for Masimo. The logo is a stylized wordmark with a small circle above the "i". Below the logo is the address "2852 Kelvin Avenue, Irvine, CA 92614". The phone number is "949-250-9688" and the fax number is "949-250-9686".
| Submitted by: | Masimo Corporation<br>2852 Kelvin Ave<br>Irvine, CA 92614-5826<br>(714) 250-9688<br>FAX (714) 250-9686 |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact: | James J. Cronin, Vice President, Regulatory Affairs/Quality Assurance |
| Date Summary Prepared: | February 18, 2004 |
| Trade Name | Masimo SET <sup>*</sup> Rad-5v Pulse Oximeter |
| Common Name | Pulse Oximeter |
| Classification Name | Oximeter (74DQA) (870.2700) |
| Substantially Equivalent Devices | Masimo SET Radical Pulse Oximeter with SatShare™ and LNOP series of Sensors and<br>Cables<br>510(k) Number - K031330<br>Nonin Model 2500 Pulse Oximeter - 510(k) Number - K001930 |
The Masimo SET® Rad-Sv Handheld Pulse Oxineter is noninvasive, arterial oxygen saturation and pulse rate monitor. The Rad-ITIC Mannin SEE - Not of Principely that continuously displays numeric values for SpO2 and pulse rate, as well as LED indicator bars for Pulse Amplitude Index (PAI) and Signal Quality (SQ).
Features and Benefits
ano Delleria
Clinically proven Masimo SETTM technology performance Applicable for use on neonate, pediatric and adult patients Proven for accurate monitoring in motion and low perfusion environments SpO2, pulse rate, alarm, and pulse amplitude index displays SQ for signal identification and quality indication Lightweight, convenient handheld design Long battery life: over 36 hours on 4 "AA" alkaline batteries Audible Alarm for sensor-off and low battery
### Intended use
The Masimo SET " RAD-5v Pulse Oximeter is intended for non-continuous noninvasive monitoring of finctional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO3 sensor) for adult, pediatric, and neonatal patients in hospitals, hospital-type facilities, and mobile environments.
.
{1}------------------------------------------------
### Indications for use
The Rad-5v Handheld Pulse Oximeter is indicated for the non-continuous nonitoring of functions of functions or indicated I he Rad-Sv Handleto Pulse Oxmleer is interest for the Rad-Sv Handheld Pulse Oxinere is indicated of arteral hellogiount (Spo?) and patients during both motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, and mobile environments.
### Principles of Operation
The principles of operation of the Masimo SET® Rad-5v pulse oximeter are that oxyhemoglobin and deoxyhemonion I he price of operation of the Mistino DE - - Las - Parterial blood in tissue (and hence, light absorption ther adsorpound of countine the pulse (specifical starterior renous shunting is highly variable and that fluctuating by mal blood is a major component of noise during the pulse. Because oxylemoglobin and deoxyhemoglobin and deoxyhemoglobin differ absorbance of yenous not is a najor computent of the blood is related to hemoglobin oxygen saturation. The in light at sur hour of readed minered tight aborder of and infraced pulsatile absorbance signal into an arterial included a noise component and calculates the ratio of the artist without noise. The ratio of the two arterial pulse-added a bsorbance signals and its value is used to find the SpO saturation in an empirically derived equation into the Masimo SET® Rad-Sv software. The values in the look-up table are based upon human blood studies against a laboratory co-oximeters in induced hypoxia states during motion and non-motion conditions.
### Method of Operation
The Masino SET * Rad-Sv pulse oximeter is turned on. An oximetery sensor is attached to a patient's finger and of a patient cable is connected to the sensor and the other end connected to the Rad-5v pulse oximeter.
The montor will begin continuously displaying the patient's pulse rate. and SpO2 value. The practitioner can then use the I Intention that is continuously displayed on the monitor, and hear if an alarm limit is reached, to help assess the condition of the patient and as an aide in determining if any intervention is required by the practitioner.
Once the practitioner deternines the patient no longer requires monitoring, the cable is disconnected from the oxinetty sensor is removed (and disposed of if it is a single use device), and the power to the monitor is turned off
#### Power Source
The Masimo SET* Rad-5v pulse oximeter is powered by 4 AA batteries with an operating time of 36 hours .
### Specifications and Operating Ranges
| Range | | |
|----------|----------------------------------------------------|----------------------|
| | Saturation (% SpO2) | 1% - 100% |
| | Pulse Rate (bpm) | 25 - 240 |
| | Perfusion | 0.02% - 20% |
| Accuracy | | |
| | Saturation (% SpO2) - During No Motion Conditions¹ | |
| | Adults, Pediatrics | 70% - 100% ± 2 digit |
| | | 0% - 69% unspecified |
| | Neonates | 70% - 100% ± 3 digit |
| | | 0% - 69% unspecified |
| | Saturation (% SpO2) - During Motion Conditions2,3 | |
| | Adults, Pediatrics | 70% - 100% ± 3 digit |
| | | 0% - 69% unspecified |
| | Neonates | 70% - 100% ± 3 digit |
| | | 0% - 69% unspecified |
| | | |
{2}------------------------------------------------
| Pulse Rate (bpm) - During No Motion Conditions1 | 25 to 240 ± 3 digits |
|-------------------------------------------------|----------------------|
| Adults, Pediatric, Neonates | |
| Pulse Rate (bpm) - During Motion Conditions2,3 | 25 to 240 ± 5 digits |
| Adults, Pediatric, Neonates | |
| Resolution | |
| Saturation (% SpO₂) | 1% |
| Pulse Rate (bpm) | |
|-------------------------------------------------------|--------------------------------|
| Low Perfusion Performance4<br>> 0.02% Pulse Amplitude | Saturation (% SpO2) ± 2 digits |
| and % Transmission > 5% | Pulse Rate ± 3 digits |
## Interfering Substances
Carboxyhemoglobin may erroneously increase readings. The level of increase is approximately equal to the Carousynemoglobin may erroreasily in Dyes, or any substance containing dyes, that change usual arterial pigmentation may cause erroncous readings.
#### Power
Internally powered by 4 "AA" Alkaline batteries
#### Isolation
No external power or ground connection, internally powered only
### Environmental
| Operating Temperature | 41°F to + 104°F (5°C to +40°C) |
|-----------------------|-------------------------------------------------------------------------------|
| Storage Temperature | -40°F to + 158°F (-40°C to +70°C) |
| Relative Humidity | 5% to 95% noncondensing |
| Operating Altitude | 500 mbar to 1060 mbar pressure<br>-1.000 ft to 18.000 ft (-304 m to +5.486 m) |
#### Circuitry
Microprocessor controlled Automatic self-test of oximeter when powered on Automatic setting of parameters Automatic alarm messages
#### Display
: 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、 、
11 - 1 - 1 - - -
Type Data Displayed LED, 7-segment Pulse Rate, SpO2 %, Aların status, alarnı silenced status, Pulse Amplitude Index Bar. Signal Quality Bar. and Battery Status.
#### Audio indicators
Adjustable volume audible pulse: OFF and 33% to 100% in 3 steps Adjustable volume audible alarm tone: levels and 33% to 100% in 3 steps Alam silence (120 seconds): all mute (continuous silence) Sensor condition alarms System failure and battery low alarms
{3}------------------------------------------------
Physical characteristics Dimensions: Weight:
6.2" x 3.0" x 1.4" (15.8 cm x 7.6cm x 3.6 cm) 13oz. (0.32 kg)
Modes
Averaging mode: Sensitivity
8 seconds Normal
- 1 The Masimo SET Technology with LNOP Adt sensors has been validated for no motion accuracy in human blood studies on healthy adult volunters in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory coo studies on nearly admit roluneers in internet equals plus or minus one standard deviation which encompasses 68% of the population.
- ule population.
2 The Masimo SET Technology with LNOP Adt sensors has been validated for motion accuracy in human blood studies on healthy adult volunters in induced hypoxia studies while performing rubbing and tapping motions, at 2 lo 4 Hz at an amplitude of 1 to 2 cm and a non-repetitive motion between 1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory co-oximeter and ECG monitor. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
- The specifical saturation accuracy from 70% 100% for neonates is based on the results from clinical studies on adult volunteers and 1% was added to account for the properties of fetal hemoglobin to be within ± 3 digits during both voluneers and 170 was added to when compared to the CO-Oximeter, and the pulse rate accuracy values for neonates more and no no include on and ± 5 digits during motion conditions when compared to the ECG. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
- The Masico SET Technology has been validated for low perfusion accuracy in bench top testing against a Brotek Index 2 simulator and Masino's simulator with signal strengths of greater than 0.02% and a % transmission of greater than 5% for saturations ranging from 70 to 100%. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
- This represents approximate run time at lowest indicator brightness, using a new, fully charged battery. ה
## Environmental Testing
Applicable environmental testing per the Reviewers Guidance for Premarket Submissions - November 1993, i.e. electrical, mechanical and environmental were performed and all tests passed.
# Nonclinical tests performed that support a determination of substantial equivalence.
The Masimo SET " Rad-5v Pulse Oximeters was subjected to bench testing using a simulator that determined the performance accuracy of the instruments against the simulator under the range of saturation and pulse rates that both devices specify.
The results of the bench testing showed that the Masimo SET* Rad-5v Pulse Oximeters returned the same sauracy values within = 2 digits and pulse rate values within ± 3 digits when compared to the simulators used.
# Clinical tests performed that support a determination of substantial equivalence.
Clinical sudies were performed using SET technology on healthy adult volunteer subjects during no motion and motions who were subjected to a progressive induced hypoxia and measuring the arterial hemoglobin saturation value with the instruments against the arterial hemoglobin oxygen determined from arterial blood samples with a CO-Oximeter.
Clinical studies were performed using the Masimo SET* technology on neonates during no motion conditions and measuring the arterial hemoglobin saturation value with the instruments against the arterial hemoglobin oxygen determined from arterial blood samples with a CO-Oximeter.
Clinical studies were performed using SET* technology on healthy adult volunteer subjects who were subjected to low perfusion conditions and to a progressive induced hypoxia and measuring the arterial hemoglobin saturation value with the mstruments against the arterial hemoglobin oxygen determined from arterial blood samples with a CO-Oximeter.
{4}------------------------------------------------
Clinical studies were performed using SET® technology on neonates with low perfusion conditions and neasuring the arterial hemoglobin saturation value with the instruments against the arterial hemoglobin oxygen determined from arterial blood samples with a CO-Oximeter.
Clinical studies were conducted following regulations under Title 21 of the Code of Federal Regulations (21CFR), Part 812 -Cuncal Studies Neve Charletee roll of Human Subjects and Part 56 - Institutional Review Boards.
The results from the clinical studies show that the Masino SET* technology saturation accuracy values for adults and pediatrics r it is a 2 digits during no motions and ± 3 digits during motion conditions when compared to the CO-Oxineer and the whan = 2 digits dating no motion conditions and = 5 digits during motion conditions when conditions when compared to the ECG.
The specified saturation accuracy from 70% - 100% for neonates is based on the results from clinical studies on adult volunteers and 1% was added to account for the properties of fetal hemoglobin to be within ± 3 digits during both notion and no notion conditions when compared to the CO-Oximeter, and the pulse rate accuracy values for neonates to be within = 3 digits during no motion and = 5 digits during motion conditions when compared to the ECG.
#### Conclusions
:24:00 来
The results of the environmetal testing demonstrated that the Masimo SET* Rad-5vPulse Oximeter met the requirements of Reviewers Guidance for Premarket Submissions - November 1993.
The results of the bench testing demonstrates that the Masimo SET * Rad-5v Pulse Oximeters met its performance requirements.
The results of the clinical testing demonstrates that the Masino SET * technology meets its performance requirements during no motion and motion conditions and low perfusion conditions.
The non-clinical and clinical testing performed demonstrates that the Masino SET* Rad-5v Pulse Oximeters is safe, effective.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of an eagle with three wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 9 2004
Mr. James Cronin Vice President, Regulatory Affairs/Quality Assurance Masimo Corporation 2852 Kelvin Ave. Irvine, CA 92614-5826
Re: K033998
Trade/Device Name: Masimo SET Rad 5v Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA, DPZ Dated: December 22, 2003 Received: December 24, 2003
Dear Mr. Cronin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can he found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2 -- Mr. James Cronin
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic rordlict radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control pro revi marketing your device as described in your Section 510(k) I ms letter witi anow you to ought inding of substantial equivalence of your device to a legally premarket notineation: "The stars in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you attire of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general micronational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincercly yours,
Chris Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## 510(k) Number (if known):
Masimo SET Rad 5v Pulse Oximeter Device Name:
Indications For Use:
The Masimo SET® Rad5v Pulse Oximeter is indicated for the non-continuous noninvasive monitoring of inconal I II Masino SET " Rad Pri unse Oxiniver to Cased pulse rate (measured by an SpO2 sensor). The Masimo SET® Rad Sv oxygell saluration of arterial acult, pediatric, and neonatal patients during both no motion and motion r use Oxfilions, and for patients who are well or porty perfused in hospital-type facilities, mobile, and home environments..
Prescription Use_ X (Per 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (Per 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
thi
Anesthesiology, G
510(k) Number
0020
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.