MASIMO SET RAD-5 PULSE OXIMETER, MODEL RAD-5; MASIMO SET RAD-5V PULSE OXIMETER, MODEL RAD-5V
Applicant
Masimo Corp.
Product Code
DQA · Cardiovascular
Decision Date
Feb 19, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Masimo SET® RAD-5 Pulse Oximeter is intended for continuous nonitoring of functional oxygen saturation of arterial hemoglobin (SpOz) and pulse rate (measured by an SpO2 sensor) for adult, pediatric, and neonatal patients in hospitals, hospital-type facilities, and mobile environments.
Device Story
Handheld pulse oximeter; inputs red/infrared light signals via finger-attached sensor; utilizes Masimo SET® technology to perform spectrophotometry and plethysmography; decomposes pulsatile absorbance into arterial signal and noise components; calculates SpO2 and pulse rate; displays data on LED screen; includes Signal IQ™ for signal quality, FastSat™ for rapid tracking, and adjustable sensitivity (Max, Normal, APOD); operated by clinicians in hospitals, mobile, or home environments; provides continuous monitoring to assist in patient assessment and intervention decisions; benefits patients by maintaining accurate monitoring during motion and low perfusion states.
Clinical Evidence
Clinical studies conducted on healthy adult volunteers and neonates under induced hypoxia. Evaluated performance during no-motion, motion, and low-perfusion conditions. Primary endpoints: SpO2 accuracy (vs. CO-Oximeter) and pulse rate accuracy (vs. ECG). Results: Adult/pediatric SpO2 accuracy ±2 digits (no motion), ±3 digits (motion); neonatal SpO2 accuracy ±3 digits. Pulse rate accuracy ±3 digits (no motion), ±5 digits (motion). Studies followed 21 CFR Parts 50, 56, and 812.
Technological Characteristics
Handheld pulse oximeter; LED 7-segment display; powered by 4 AA alkaline batteries (36-hour life). Sensing principle: spectrophotometry and plethysmography. Features Signal IQ™ and FastSat™ algorithms. Connectivity: sensor-to-device cable. Environmental: 5°C to 40°C operating range. Microprocessor-controlled with automatic self-test and alarm management.
Indications for Use
Indicated for continuous monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate in adult, pediatric, and neonatal patients. Suitable for use during motion and non-motion conditions, and for patients who are well or poorly perfused in hospital-type facilities, mobile, and home environments.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Masimo SET Radical Pulse Oximeter with SatShare™ and LNOP series of Sensors and Cables (K031330)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the text string "K033296" in a handwritten style. The text is oriented diagonally, running from the lower-left to the upper-right corner of the image. The characters appear to be written with a thick marker or pen, giving them a bold appearance.
FEB 1 9 2004
| Submitted by: | Masimo Corporation |
|----------------------------------|---------------------------------------------------------------------------------|
| | 2852 Kelvin Ave |
| | Irvine, CA 92614-5826 |
| | (714) 250-9688 |
| | FAX (714) 250-9686 |
| Company Contact: | James J. Cronin, Vice President, Regulatory Affairs/Quality Assurance |
| Date Summary Prepared: | December 5, 2003 |
| Trade Name | Masimo SET® Rad-5 Pulse Oximeter |
| Common Name | Pulse Oximeter |
| Classification Name | Oximeter (74DQA) (870.2700) |
| Substantially Equivalent Devices | Masimo SET Radical Pulse Oximeter with SatShare™ and LNOP series of Sensors and |
| | Cables<br>510(k) Number - K031330 |
| Features and Benefits | |
| | Clinically proven Masimo SET <sup>TM</sup> technology performance |
| | Applicable for use on neonate, pediatric and adult patients |
| | Proven for accurate monitoring in motion and low perfusion environments |
| | SpO <sub>2</sub> , pulse rate, alarm, and perfusion index displays |
| | Signal IQ <sup>TM</sup> for signal identification and quality indication |
| | Lightweight, convenient handheld design |
| | Long battery life: over 36 hours on 4 "AA" alkaline batteries |
| | Audible Alarm for sensor-off and low battery |
| | Alarms for Hi/Low saturation and Hi/Low pulse rate |
| | FastSat <sup>TM</sup> |
| | Three sensitivity levels - Max, Normal and APOD <sup>TM</sup> |
| | Adjustable alarm volume |
Adjustable averaging 2 to 16 seconds
## Intended use
The Masimo SET® RAD-5 Pulse Oximeter is intended for continuous nonitoring of functional oxygen saturation of arterial hemoglobin (SpOz) and pulse rate (measured by an SpO2 sensor) for adult, pediatric, and neonatal patients in hospitals, hospital-type facilities, and mobile environments.
{2}------------------------------------------------
## Indications for use
The Rad-5 Handheld Pulse Oximeter is indicated for the continuous nonitoring of functional oxygen saturation of arterial hemoglobin (Sp() ) and pulse rate (measured by an SpO2 sensor). The Rad-5 Handheld Pulse Oximeter is indicated for use with adult, pediatric and neonatal patients during both motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, and mobile environments.
## Principles of Operation
The principles of operation of the Masimo SET® Rad-Spulse oximeter are that oxyhemoglobin differ in their absorption of red and infrared light (spectrophotomery), the volume of arterial blood in tissue (and hence, light absorption by that blood) changes during the pulse (plethysmography), and that arterio-venous shunting is highly variable and that fluctuating absorbance by venous blood is a major component of noise. Because oxyhemoglobin and deoxyhemoglobin differ in light absorption, the amount of red and infrared light absorbed to hemoglobin oxygen saturation. The Masimo SET® Rad-Spulse oximeter decomposes the red and infrared pulsatile absorbance signal into an arterial signal plus a noise component and calculates the ratio of the arterial signals without noise. The ratio of the two arterial pulse-added absorbance signals and its value is used to find the SpO2 saturation into the Masimo SET Rad-5 software. The values in the look-up table are based upon human blood studies against a laboratory co-oximeters in induced hypoxia states during motion and non-motion conditions.
## Method of Operation
The Masimo SET® Rad-5 pulse oximeter is turned on. An oximetery sensor is attached to a patient's finger and one end of a patient cable is connected to the sensor and the other end connected to the Rad-5 pulse oximeter.
The monitor will begin continuously displaying the patient's pulse . The practitioner can adjust the high and low alarm limits to their desired value, if required. The practitioner can then use the information that is continuously displayed on the monitor, and hear if an alarm limit is reached, to help assess the condition of the patient and as an aide in determining if any intervention is required by the practitioner.
Once the practitioner determines the patient no longer requires monitoring, the cable is discomected from the oximetry sensor is removed (and disposed of if it is a single use device), and the power to the monitor is turned off.
#### Power Source
The Masimo SET® Rad 5 pulse oximeter is powered by 4 AA batteries with an operating time of 36 hours'.
## Specifications and Operating Ranges
| Range | |
|----------------------------------------------------|----------------------|
| Saturation (% SpO2) | 1% - 100% |
| Pulse Rate (bpm) | 25 - 240 |
| Perfusion | 0.02% - 20% |
| Accuracy | |
| Saturation (% SpO2) - During No Motion Conditions1 | |
| Adults, Pediatrics | 70% - 100% ± 2 digit |
| | 0% - 69% unspecified |
| Neonates | 70% - 100% ± 3 digit |
| | 0% - 69% unspecified |
| Saturation (% SpO2) - During Motion Conditions2,3 | |
| Adults, Pediatrics2 | 70% - 100% ± 3 digit |
| | 0% - 69% unspecified |
| Neonates3 | 70% - 100% ± 3 digit |
| | |
{3}------------------------------------------------
#### 0% - 69% unspecified
| Pulse Rate (bpm) - During No Motion Conditions1 | |
|-------------------------------------------------|----------------------|
| Adults, Pediatric, Neonates | 25 to 240 ± 3 digits |
Pulse Rate (bpm) - During Motion Conditions233 25 to 240 ± 5 digits Adults, Pediatric, Neonates
#### Resolution
| Saturation (% SpO2) | 1% |
|---------------------|----|
| Pulse Rate (bpm) | 1 |
## Low Perfusion Performance4
> 0.02% Pulse Amplitude and % Transmission > 5%
Saturation (% SpO2) ± 2 digits Pulse Rate ± 3 digits
#### Interfering Substances
Carboxyhemoglobin may erroneously increase readings. The level of increase is approximately equal to the amount of carboxyhemoglobin present. Dyes, or any substance containing dyes, that change usual arterial pigmentation may cause erroneous readings.
#### Power
Internally powered by 4 "AA" Alkaline batteries
#### Isolation
No external power or ground connection, internally powered only
#### Environmental
| Operating Temperature | 41°F to + 104°F (5°C to +40°C) |
|-----------------------|------------------------------------------------------------------------------|
| Storage Temperature | -40°F to + 158°F (-40°C to +70°C) |
| Relative Humidity | 5% to 95% noncondensing |
| Operating Altitude | 500 mbar to 1060 mbar pressure<br>-1,000 ft to 18,000 ft (-304 m to 5,486 m) |
## Circuitry
Microprocessor controlled Automatic self-test of oximeter when powered on Automatic setting of parameters Automatic alarm messages
### Display
| Type | LED, 7-segment |
|----------------|--------------------------------------------------------------------------------------------------------------------------------|
| Data Displayed | Pulse Rate, SpO2 %, Alarm status, alarm silenced status, Perfusion<br>Index Bar, Signal IO Bar, Battery Status, APOD. FastSat. |
## Audio indicators
Adjustable volume audible pulse: OFF and 33% to 100% in 3 steps Adjustable volume audible alarm tone: levels and 33% to 100% in 3 steps Alarm silence (120 seconds); all mute (continuous silence) Pulse rate out-of-limits alarm SpO2 level out-of limits alarm Sensor condition alarms System failure and battery low alarms
{4}------------------------------------------------
| Physical characteristics | |
|--------------------------|-----------------------------------------------|
| Dimensions: | 6.2" x 3.0" x 1.4" (15.8 cm x 7.6cm x 3.6 cm) |
| Weight: | 13oz. (0.32 kg) |
Modes
Averaging mode: Sensitivity
2, 4, 8, 10, 12, and 16 seconds Normal, APOD, and MAX
- 1 The Masimo SET Technology with LNOP Adt sensors has been validated for no motion accuracy in human blood studies on healthy adult volunteers in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory cooximeter and ECG monitor. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
- 2 The Masimo SET Technology with LNOP Adt sensors has been validated for motion accuracy in human blood studies on healthy adult volunteers in induced hypoxia studies while performing rubbing and tapping motions, at 2 to 4 Hz at an amplitude of 1 to 2 cm and a non-repetitive motion between 1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory co-oximeter and ECG monitor. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
- 3 The Masimo SET Technology with LNOP Neo and Neo Pt sensors has been validated for neonatal motion accuracy in human blood studies on neonates while moving the neonate's foot at 2 to 4 Hz at an amplitude of 1 to 2 cm against a laboratory co-oximeter and ECG monitor. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
- 4 The Masimo SET Technology has been validated for low perfusion accuracy in bench top testing against a Biotek Index 2 simulator and Masimo's simulator with signal strengths of greater than 0.02% and a % transmission of greater than 5% for saturations ranging from 70 to 100%. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
- રે This represents approximate run time at lowest indicator brightness, using a new, fully charged battery.
## Environmental Testing
Applicable environmetal testing per the Reviewers Guidance for Premarket Submissions - November 1993, i.e. electrical, mechanical and environmental were performed and all tests passed.
## Nonclinical tests performed that support a determination of substantial equivalence.
The Masimo SET® Rad-5 Pulse Oximeters was subjected to bench testing using a simulator that determined the performance accuracy of the instruments against the simulator under the range of saturation and pulse rates that both devices specify.
The results of the bench testing showed that the Masimo SET® Rad-5 Pulse Oximeters returned the same saturation accuracy values within ± 2 digits and pulse rate values within ± 3 digits when compared to the simulators used.
## Clinical tests performed that support a determination of substantial equivalence.
Clinical studies were performed using the Masimo SET® technology on healthy adult volunteer subjects during no motion and motion conditions who were subjected to a progressive induced hypoxia and measuring the arterial hemoglobin saturation value with the instruments against the arterial hemoglobin oxygen determined from arterial blood samples with a CO-Oximeter.
Clinical studies were performed using the technology on neonates during no motion and motion conditions and measuring the arterial hemoglobin saturation value with the instruments against the arterial hemoglobin oxygen determined from arterial blood samples with a CO-Oximeter.
Clinical studies were performed using the technology on healthy adult volunteer subjects who were subjected to low perfusion conditions and to a progressive induced hypoxia and measuring the arterial hemoglobin saturation value with the instruments against the arterial hemoglobin oxygen determined from arterial blood samples with a CO-Oximeter.
{5}------------------------------------------------
Clinical studies were performed using the Masimo SET technology on neonates with low perfusion and measuring the arterial hemoglobin saturation value with the instruments against the arterial hemoglobin oxygen determined from arterial blood samples with a CO-Oximeter.
Clinical studies were conducted following regulations under Title 21 of the Code of Federal Regulations (21CFR), Part 812 -Investigational Device Exemptions, Part 50 - Protection of Human Subjects and Part 56 - Institutional Review Boards.
The results from the clinical studies show that the Masimo SET® technology saturation accuracy values for adults and pediatrics within ± 2 digils during no motions and ± 3 digits during motion conditions when compared to the CO-Oximeter and the pulse rate accuracy values within ± 3 digits during no motion conditions and ± 5 digits during motion compared to the ECG.
The specified saturation accuracy from 70% - 100% for neonates is based on the results from clinical studies with saturations down to 83% combined with clinical studies on adults to show that the Masimo SET® technology to be within ± 3 digits during both motion and no motion conditions when compared to the CO-Oximeter, and the pulse rate accuracy values for neonates to be within ± 3 digits during no motion and ± 5 digits during motion conditions when compared to the ECG.
## Conclusions
The results of the environmetal testing demonstrated that the Masmo SET® Rad-5Pulse Oximeter of Reviewers Guidance for Premarket Submissions - November 1993.
The results of the bench testing demonstrates that the Masimo SET® Rad-5 Pulse Oximeters met its performance requirements.
The results of the clinical testing demonstrates that the Masimo SET® technology meets its performance requirements during no motion and motion conditions and low perfusion conditions.
The non-clinical and clinical testing performed demonstrates that the Masimo SET® Rad-5 Pulse Oximeters is safe, effective.
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure with three stylized lines representing its wings or body.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 9 2004
Mr. James J. Cronin Vice President, Regulatory Affairs/Quality Assurance Masimo Corporation 2852 Kelvin Avenue Irvine, California 92614-5826
Re: K033296
Trade/Device Name: Masimo SET RAD-5 Pulse Oximeter Regulation Number: 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: January 23, 2004 Received: January 26, 2004
Dear Mr. Cronin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 – Mr. James J. Cronin
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Chris
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## Indications for Use
510(k) Number (if known): K033296
Masimo SET Rad 5 Pulse Oximeter Device Name:
Indications For Use:
The Masimo SET® Rads Pulse Oximeter is indicated for the continuous nonitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor). The Masimo SET® Rad 5 Pulse Oximeter is indicated for use with adult, pediatric, and neonatal patients during both no motion and ition conditions, and for patients who are well or poorly perfused in hospital-type facilities, mobile, and home environments.
Prescription Use _ X (Per 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (Per 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
h
Anesthesiology, General Hospital, Intection Control. Der
510(k) Number:
1080
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.