K032755 · Respironics Novametrix, Inc. · DQA · May 21, 2004 · Cardiovascular
Device Facts
Record ID
K032755
Device Name
MODEL 509M PULSE OXIMETER, MODEL 509M
Applicant
Respironics Novametrix, Inc.
Product Code
DQA · Cardiovascular
Decision Date
May 21, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Model 509M Pulse Oximeter is intended to provide continuous, non-invasive monitoring of functional arterial oxygen saturation and pulse rate in neonatal, pediatric and adult patients during both no motion and motion conditions and for patients who are well or poorly perfused in hospital, hospital-type facilities and intra-hospital transport environments such as the operating room, emergency department and intensive care units. The Model 509M Pulse Oximeter and its sensors are intended to be used by trained operators when pulse oximetry monitoring is required in the judgement of a physician.
Device Story
Model 509M Pulse Oximeter provides continuous, non-invasive monitoring of functional arterial oxygen saturation (SpO2) and pulse rate. Device uses ratiometric technique; red and infrared LEDs beam light through pulsating vascular bed (finger/toe); photodiode on opposing side captures non-absorbed light. Dual microprocessor system processes signals; digital signal processor (DSP) performs filtering, pulse rate/saturation calculations, and noise reduction. Device displays digital SpO2 and pulse rate values; transmits plethysmogram to Philips Monitoring System via VueLink interface. Used in hospitals, emergency departments, ICUs, and intra-hospital transport by trained operators. Healthcare providers use displayed numerical values and waveforms to assess patient oxygenation status, aiding clinical decision-making regarding respiratory support or intervention. Benefits include real-time monitoring of patient perfusion and oxygen saturation levels.
Clinical Evidence
No clinical data provided. Substantial equivalence established via inter-device comparison studies to ensure accuracy and sensor performance against the predicate Model 2001.
Technological Characteristics
Dual microprocessor-based data acquisition system. Sensing principle: ratiometric red and infrared light absorbance. Components: red/infrared LEDs, photodiode. Output: digital SpO2, pulse rate, plethysmogram waveform. Connectivity: VueLink interface for Philips Monitoring System. Power: external DC supply. Software: dual-microprocessor architecture with DSP for signal filtering and proprietary algorithm processing.
Indications for Use
Indicated for continuous, non-invasive monitoring of functional arterial oxygen saturation and pulse rate in neonatal, pediatric, and adult patients, including those well or poorly perfused, in hospital and transport settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
K032755
# MAY 2 1 2004
## 510(K) SUMMARY
December 15, 2003
#### Applicant's Name and Address a.
Respironics Novametrix, Inc. 5 Technology Drive Wallingford, CT 06492
## b. Contact Person
Michael J. Malis Q.A. and Regulatory Manager (203) 697-6442 (203) 284-0753 (facsimile)
- Name of Device ပ
| Device Names (Proprietary/Trade Names): | Model 509M Pulse Oximeter |
|-----------------------------------------|----------------------------------------------|
| Device Name (Common Name): | Pulse Oximeter |
| Classification: | Class II, 21 C.F.R. 870.2700<br>74DQA/74 DPZ |
#### Equivalent Devices d.
Substantial equivalence to the following legally marketed predicate devices with the same or similar indications for use has been demonstrated by a comparison of product features as described in the labeling and promotional literature for the Model 509M, as well as testing to accepted industry standards. In addition, inter-device comparison studies were conducted to establish the Model 509Ms accuracy and to ensure that the sensors meet their currently published accuracy specifications with the Model 2001. The predicate device is as follows:
- Model 2001 Pulse Oximeter, Novametrix Medical Systems, Inc., K993979,K000794) 1.
#### Device Description e.
The Model 509M Pulse Oximeter Module is designed for continuous, non-invasive monitoring of the functional oxygen saturation of arterial hemoglobin (SpO₂) and pulse rate. Oxygen saturation is measured with ratiometric technique using red and infrared absorbance of oxy- and deoxyhemoglobin and pulse rate is measured using the time between successive pulses. The O2 saturation sensors are already legally marketed as accessories to the Model 2001 monitor. The Model 509M displays digital values of SpO₂ and pulse rate and transmits these values as well as the plethysmogram to a Philips Monitoring System via the VueLink Interface. The Model 509M module consists of a dual microprocessor based data acquisition system that measures oxygen saturation data. The firmware for the primary microprocessor is responsible for handling the user interface and communications with external devices via VueLink interface. The firmware for the second microprocessor, a digital signal processor, performs the filtering, pulse rate and saturation calculations of the algorithms which analyze the incoming signals and perform noise reduction on that signal when the presence of noise is detected.
{1}------------------------------------------------
The Model 509M is powered by an external DC power supply. Audible and visual alarms for high/low saturation and pulse rate are available. Visual alerts are provided as a status for highlow outdren and Philips monitor's display (providing the alarms on the Philips dipslay are not disabled) The Model 509M also includes a temporary (2 minute) and diseay are not alsublour, he other configurable settings. The Model 2001 Pulse Oximeter has visual indicators for pulse error conditions, and alarm silence.
#### Intended Use f.
The Model 509M Pulse Oximeter is intended to provide continuous, non-invasive monitoring of functional arterial oxygen saturation and pulse rate in neonatal, pediatric and adult patients during both no motion and motion conditions and for patients who are well or poorly perfused in hospital, hospital-type facilities and intra-hospital transport environments such as the operating room, emergency department and intensive care units. The Model 509M Pulse Oximeter and its sensors are intended to be used by trained operators when pulse oximetry monitoring is required in the judgement of a physician. The intended use, patient population and environments of use are the same or similar to the predicate device, the Novametrix Model 2001.
## g. Technological Characteristics
The Model 509M Pulse Oximeter measures functional oxygen saturation and pulse rate with sensors that contain red and infrared light sources. Since oxygen saturated blood absorbs different amounts of light at each wavelength (red and infrared) as compared with unsaturated blood, the amount of light absorbed at each wavelength by the blood in each pulse can be used to calculate oxygen saturation. The light energy from red and infrared LEDs is beamed through a sample cell- a pulsating vascular bed, the patient's finger or toe for example. The remaining light energy not absorbed by the sample cell reaches a photodiode, on the opposing side of the sensor. The signal received by the photodiode is split into its red and infrared components, sampled, software filtered, processed using proprietary algorithms and displayed as a numerical value for functional oxygen saturation and as a waveform, the plethysmogram.
The Model 509M uses the identical SpO2 and pulse rate software algorithm to process the information from the sensor as the predicate device, Model 2001 Pulse Oximeter, cleared under K993979 and K000794.
### h. Certification Statement
In accordance with the requirements of 21 CFR 807.87(j), the following certification is provided:
Respironics Novametrix, Inc. believes that all data and information submitted in this notice are truthful and accurate and no material fact has been omitted.
Michael Means
Michael J. Malis Q.A. and Regulatory Manager
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized design featuring three abstract human figures, possibly representing individuals or families, with flowing lines suggesting movement or connection.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# MAY 21 2004
Mr. Kevin Mader Manager of Quality Assurance and Regulatory Affairs Respironics Novametrix, Inc. 5 Technology Drive Wallingford, Connecticut 06492
Re: K032755
Trade/Device Name: Modcl 509M Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Codc: DQA, DPZ Dated: April 15, 2004 Received: April 19, 2004
Dear Mr. Mader:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amerouse over and have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 NET), to may of each be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration
{3}------------------------------------------------
## Page 2 -- Mr. Kevin Mader
and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice and hoting (21 CF R Part 807), as ality systems (QS) regulation (21 CFR Part 820); and if requirements us bet form in all ation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
21 CFX 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter wifi are n you ve organ substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), If you desire specific at not to research (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Inter State State Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Ches Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ K032755
Model 509M Pulse Oximeter Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications For Usc:
The Model 509M Pulse Oximeter is intended to provide continuous, non-invasive monitoring The Model 505M F alse Oxinced is intended to protein neonatal, pediatric and adult patients of functional artenal oxygen saturation and for patients who are well or poorly perfused dultrig both no motion and motion ochalities and intra-hospital transport environments such as the in nospital, noopkar type for department and intensive care units.
Prescription Use_X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Söng Hams 5/20/2004
esthesiology, General Hospital, Infection Control. Dental Devices
510(k) Number: K032755
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.