AIRWATCH II ZONE

K982676 · Enact Health Management Systems · BZH · Oct 29, 1998 · Anesthesiology

Device Facts

Record IDK982676
Device NameAIRWATCH II ZONE
ApplicantEnact Health Management Systems
Product CodeBZH · Anesthesiology
Decision DateOct 29, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.1860
Device ClassClass 2
AttributesPediatric

Intended Use

AirWatch II Zone is a medical instrument which is intended to be used for the monitoring of asthma and other chronic respiratory diseases. The instrument measures both Peak Expiratory Flow (PEF) and Forced Expiratory Volume-One Second (FEV1). The instrument is designed for use by consumers from 5 years of age to adults.

Device Story

Pocket-sized personal electronic peak flow meter; measures Peak Expiratory Flow (PEF) and Forced Expiratory Volume in one second (FEV1). Device includes LCD display, control buttons, removable mouthpiece, and RJ-11C telephone connector for remote data transfer. Used by consumers (patients) for home monitoring of respiratory status. Features an 'effort check' algorithm that performs three tests per measurement to detect poor technique; displays warning symbols if measurement quality is suspect. Healthcare providers use the stored data to assess patient respiratory status and guide clinical decision-making. Benefits include improved patient self-management of chronic respiratory conditions through accurate, quality-controlled monitoring.

Clinical Evidence

Four studies conducted with 20 subjects each: Validation/Measurement Quality Study, Home Use Study, User Guide Study, and Quick Reference Guide Study. Validation study confirmed the device accurately provides PEF/FEV1 measurements and identifies suspect quality measurements. Other studies demonstrated consumer ability to safely/effectively use the device, interpret display readings, and follow instructions for use.

Technological Characteristics

Electronic peak flow meter; dimensions 4.00 x 2.56 x 0.93 in; weight 3.3 oz. Powered by two CR2032 3V lithium batteries. Features LCD display, RJ-11C telephone connector for data transfer, and 480-session memory. Conforms to American Thoracic Society 1994 standards for accuracy, precision, linearity, and back pressure. Includes effort check algorithm for measurement quality validation.

Indications for Use

Indicated for monitoring asthma and chronic respiratory diseases in patients aged 5 years to adults.

Regulatory Classification

Identification

A peak-flow meter for spirometry is a device used to measure a patient's maximum ventilatory flow rate.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K982676 OCT 2 9 1998 510(k) Summary (as per 21CFR807.92) July 31, 1998 Submitter and Contact Person: James Lyon Director, Engineering and Manufacturing Operations ENACT Health Management Systems, Inc. 1975 El Camino Real, Suite 306 Mountain View, CA 94040-2218 (650) 967-0379 (650) 967-9223 Fax Trade Name: AirWatch II Zone Common Name: Peak Flow Meter Classification Name: Peak-flow meter for spirometry (as per 21 CFR868.1860) Predicate Device: ENACT Products, Inc. AirWatch Monitor (K941226) Description of Device: The AirWatch II Zone is a pocket-sized, personal electronic device which monitors respiratory status by measuring peak expiratory flow (PEF) rate in liters per minute and forced expiratory volume in one second (FEV1) in liters. The AirWatch II Zone includes an LCD for displaying PEF and FEV1 information to the consumer, three control buttons, a removable mouthpiece, and an RJ-11C telephone connector which can be used to transfer data stored in the AirWatch to a remote computer. Indications for Use: AirWatch II Zone is a medical instrument which is intended to be used for the monitoring of asthma and other chronic respiratory diseases. The instrument measures both Peak Expiratory Flow (PEF) and Forced Expiratory Volume-One Second (FEVI). The instrument is designed for use by consumers from 5 years of age to adults. Summary of Technological Characteristics: There is one technological difference between the AirWatch II Zone and its predicate device, the addition of an effort check warning feature. {1}------------------------------------------------ measurements which do not accurately represent airway status. The AirWatch II Zone checks every measurement with three effort check tests. Warning symbols appear on the display when a measurement fails any one or more of these tests. The appearance of the warning symbols will alert the consumer that the monitor has detected a measurement of suspect quality as a result of poor technique. The AirWatch II Zone also differs from its predicate device by conforming to the new American Thoracic Society 1994 Standards versus the 1987 Standards. Summary of Non-Clinical Performance Data: The AirWatch II Zone and its predicate device meet all applicable electrical, mechanical and environmental performance requirements given in the Reviewers Guidance for Premarket Notification Submissions, Appendix A. The AirWatch II Zone and its predicate device also conform to the National Asthma Education Program's Statement on Technical Standards for peak flow meters. The AirWatch II Zone meets the American Thoracic Society 1994 Standards for accuracy, precision, linearity, and back pressure. The predicate device meets the American Thoracic Society 1987 Standards. Summary of Clinical Performance Data: The AirWatch II Zone was tested on 20 study subjects in each of the following studies: Validation/Measurement Quality Study; Home Use Study; User Guide Study; and Quick Reference Guide Study. In summary, the Validation/Measurement Quality Study demonstrated that the AirWatch II Zone can appropriately provide information regarding a subject's PEF and FEV1 measurements and can appropriately identify PEF measurement of suspect quality. Other studies demonstrated that consumers can safely and effectively use the AirWatch II Zone under conditions of actual use; that consumers can interpret the AirWatch II Zone's display readings and take appropriate actions; and that the Quick Reference Guide, User Guide, AirWatch II Zone's physical design, and other human factors characteristics are appropriate for consumers. Technical Specifications: - Dimensions: L x 4.00 in, W x 2.56 in, D x 0.93 in. - Weight: . . . - 3.3 oz. - Temperature Range: 41°F to 104°F (5°C to 40°C) - Storage Temperature: -4'F to 140'F (-20°C to 60°C) - Operating Humidity: 0% to 95%, non-condensing - . Display: LCD Display {2}------------------------------------------------ - Battery Symbol: ● - Battery Life: . - Automatic Turn-Off . Period: - Test Session Memory ● Capacity: - . Memory Life With Batteries Removed or While Battery Drawer Is Open: - Telephone Connector: ◆ Two Type CR2032 3-Volt lithium Batteries Low Battery Approximately 1 year with an average daily usage of 3 test sessions per day 30 to 90 seconds, depending on display mode 480 test sessions (maximum) 2 minutes (minimum) Type RJ-11C (standard household jack) {3}------------------------------------------------ Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & H. MAN. TPITS, USA" around the perimeter. Inside the seal is an abstract symbol of three human profiles facing right, with flowing lines representing hair or clothing. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 29 1998 Mr. James Lyon ENACT Health Management Systems, Inc. 1975 El Camino Real, Suite 306 Mountain View, CA 94040-2218 Re: K982676 AirWatch II Zone Regulatory Class: II (two) Product Code: 73 BZH Dated: July 31, 1998 Received: July 31, 1998 Dear Mr. Lyon: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ Page 2 - Mr. James Lyon This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Thomas J. Callahan Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Page *of* 510(k) Number (if known):_K982676 Device Name: ANR WATOH II ZONE Indications For Use: AirWatch II Zone is a medical instrument which is intended to be used for the monitoring of asthma and other chronic respiratory diseases. The instrument measures both Peak Expiratory Flow (PEF) and Forced Expiratory Volume-One Second (FEV1). The instrument is designed for use by consumers from 5 years of age to adults. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) : . · OR тиате Over-The-Counter Use 1 (Optional Format 1-2-96) (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number _
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...