IRMA C02 MODEL 200101; IRMA AX 200601

K123043 · Phasein AB · CCK · Oct 25, 2012 · Anesthesiology

Device Facts

Record IDK123043
Device NameIRMA C02 MODEL 200101; IRMA AX 200601
ApplicantPhasein AB
Product CodeCCK · Anesthesiology
Decision DateOct 25, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 868.1400
Device ClassClass 2
AttributesPediatric

Indications for Use

The IRMA mainstream gas analyzer is intended to be connected to other medical devices for monitoring of breath rate and the breathing gases CO2, N2O and the anesthetic agents Halothane, Enflurane, Isoflurane, Sevoflurane and Desflurane. It is intended to be connected to a patient breathing circuit for monitoring of inspired/expired gases during anesthesia, recovery and respiratory care. It may be used in the operating suite, intensive care unit and patient room for adult, pediatric and infant patients. IRMA CO2 may also be used in the emergency medical services environment and road ambulances. Note: An IRMA mainstream gas analyzer shall only be connected to medical backboard devices approved by PHASEIN.

Device Story

IRMA is a mainstream respiratory gas analyzer utilizing infrared gas spectrometry; connects to patient breathing circuits via airway adapters. Measures CO2, N2O, and anesthetic agents; provides data to host medical backboard devices for display. Used in OR, ICU, patient rooms, and EMS (IRMA CO2 only) by clinicians. Operates by analyzing gas samples directly in the breathing circuit; no changes to sensor technology in this submission; adds M80 device as host display. Output enables real-time monitoring of respiratory parameters to assist clinical decision-making during anesthesia and respiratory care.

Clinical Evidence

Bench testing only. Performance evaluated by comparing IRMA models against predicate devices across the full operating range using calibrated gas samples and legally marketed anesthesia and ventilation devices. No clinical data presented.

Technological Characteristics

Mainstream infrared gas spectrometry; connects to breathing circuit via airway adapter. Models: 200101 (CO2) and 200601 (CO2, N2O, anesthetic agents). Features automatic agent identification. Connectivity: interfaces with approved medical backboard devices. Single-patient-use airway adapters.

Indications for Use

Indicated for adult, pediatric, and infant patients requiring monitoring of breath rate and inspired/expired gases (CO2, N2O, and anesthetic agents: Halothane, Enflurane, Isoflurane, Sevoflurane, Desflurane) during anesthesia, recovery, and respiratory care in operating suites, ICUs, patient rooms, and (for IRMA CO2) EMS/ambulances.

Regulatory Classification

Identification

A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) SUMMARY K 123043 - 1. 510(k) Owner Name and Address: PHASEIN AB Svärdvägen 15 182 33 Danderyd Sweden Telephone: 46-8-544-98-150 Fax: 46-8-544-98-169 PHASEIN is a subsidiary of Masimo Corporation - Contact Person: 2. David Weissburg Weissburg Associates 808 Williamson St., Suite 402 Madison, Wisconsin, 53703 USA - 3. Date prepared: 28 September, 2012 - Trade Name: Infrared Mainstream Gas Analyzer (IRMA), model/catalogue numbers 4. 200101 and 200601 - Common Name: Multigas Monitor 5. - Classification Names: Carbon-dioxide gas analyzer (21 CFR 868.1400, Product Code 6. CCK) - 7. Substantially equivalent to: ISA (Infrared Sidestream Gas Analyzer, K103604) VEO Multigas Monitor for Pocket PC (K081601) - 8. Device Description: IRMA is a mainstream respiratory gas analyzer based on infrared gas spectrometry. It is intended to be connected to another medical backboard device for display of respiratory parameters. IRMA is connected to the patient breathing circuit via the Airway Adapter. This premarket submission adds the M80 device as a host backboard display to IRMA. There are no changes to the previously cleared sensor technology. The following IRMA models are available: - a) IRMA CO2 (model/catalogue number 200101), measurement of CO2. - IRMA AX+ (model/catalogue number 200601), measurement of CO2, N2O and 5 b) anesthetic agents with automatic agent identification The IRMA product family also includes the adult and pediatric/infant Airway Adapter. A modura holder and velcro holder are offered as optional accessories. - 9. Indications for Use: The IRMA mainstream gas analyzer is intended to be connected to other medical devices for monitoring of breath rate and the breathing gases CO2, N20 and the anesthetic agents Halothane, Enflurane, Isoflurane, Sevoflurane and Desflurane. It is intended to be connected to a patient breathing circuit for monitoring of inspired/expired gases during anesthesia, recovery and respiratory care. It may be used in the operating suite, intensive care unit and patient room for adult, pediatric and infant patients. IRMA CO2 may also be used in the emergency medical services environment and road ambulances. ﮯ ﺗ {1}------------------------------------------------ Note: An IRMA mainstream gas analyzer shall only be connected to medical backboard devices approved by PHASEIN. - 10. Comparison to predicates: The IRMA model/catalogue numbers 200101 and 200601 have the same gas measurement capabilities as its predicate devices. The IRMA uses the same basic technology concepts used in the predicate devices. The intended use of the IRMA and its predicates are the same. All the devices consume equivalent amounts of electric power and utilize single-patient-use connections to interface with gases in the breathing circuit. Labeling and materials used are equivalent. - 11. Testing vs. predicates: Testing in direct comparison to predicates throughout the operating range was conducted using calibrated gas samples and legally marketed anesthesia and ventilation devices. - 12. Conclusions from testing: The IRMA, models 200101 and 200601 demonstrated performance, safety and effectiveness equivalent or superior to its predicates in all characteristics. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo features a stylized eagle with its wings spread, facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Phasein AB C/O Weissburg Associates Mr. David Weissburg Principal 808 Williamson Street, Suite 402 Madison, Wisconsin 53703 OCT 25 2012 Re: K123043 Trade/Device Name: IRMA - Infrared Mainstream Gas Analyzer Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: II Product Code: CCK, CBR, CBS, CBQ, NHQ, NHP, NHO Dated: September 19, 2012 Received: September 28, 2012 Dear Mr. Weissburg: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ ## Page 2- Mr. Weissburg Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, ph for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use Statement 510(k) Number: _______________________________________________________________________________________________________________________________________________________________ Device Name: IRMA - Infrared Mainstream Gas Analyzer. Indications for Use: The IRMA mainstream gas analyzer is intended to be connected to other medical devices for monitoring of breath rate and the breathing gases CO2, N2O and the anesthetic agents Halothane, Enflurane, Isoflurane, Sevoflurane and Desflurane. It is intended to be connected to a patient breathing circuit for monitoring of inspired/expired gases during anesthesia, recovery and respiratory care. It may be used in the operating suite, intensive care unit and patient room for adult, pediatric and infant patients. IRMA CO2 may also be used in the emergency medical services environment and road ambulances. Note: An IRMA mainstream gas analyzer shall only be connected to medical backboard devices approved by PHASEIN. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) signature Page 1 of 1 (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number:
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...