K072469 · Digiflo, Inc. · CCL · Jan 3, 2008 · Anesthesiology
Device Facts
Record ID
K072469
Device Name
DIGIFLO CONCENTRATOR ANALYZER
Applicant
Digiflo, Inc.
Product Code
CCL · Anesthesiology
Decision Date
Jan 3, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1720
Device Class
Class 2
Indications for Use
The DigiFLO Concentrator ANALYZER is a tool used by service personnel to measure Oxygen purity and Flow at the outlet of an oxygen concentrator. Also, it measures gas pressure inside and at the outlet of an oxygen concentrator. It is not intended to be used by patients who are prescribed oxygen, nor is it intended to continuously monitor or confirm oxygen delivery to a patient. Intended Environment: The DigiFLO Concentrator ANALYZER is intended to be used in an environment where oxygen concentrators are being serviced or repaired. This includes Hospitals, Nursing Homes, Extended Care Facilities, Patient Homes, and Respiratory Device Service and Repair Centers. Caution: The Federal Law restricts this device to sale by or on order of a physician, or any other practitioner licensed by the law of the state in which he practices to use or order the use of this device.
Device Story
DigiFLO Concentrator ANALYZER is a portable, battery-operated diagnostic tool for oxygen concentrator service personnel; used in hospitals, nursing homes, and repair centers. Input: oxygen concentrator output gas. Mechanism: ultrasonic wave transit-time measurement between piezoelectric transducers for flow/purity; piezoresistive pressure transducer for internal/outlet pressure. Output: digital display of oxygen concentration, flow rate, and pressure. Operation: connects to concentrator via standard therapy tubing; requires clean, dry, unhumidified gas. Clinical utility: enables technicians to verify concentrator performance and internal pressures during maintenance. Benefit: ensures proper functioning of oxygen concentrators before clinical use; prevents use of malfunctioning equipment.
Clinical Evidence
No clinical data. Bench testing only. Device performance verified against specified ranges: flow (0-20 L/M ±0.2 L/M), oxygen content (20.8%-95.7% ±1.8%), and pressure (0-35 PSI ±0.5%).
Technological Characteristics
Battery-operated (9V N-type) handheld analyzer. Sensing: ultrasonic wave transit-time (piezoelectric transducers) for flow/purity; piezoresistive transducer for pressure. Operating range: +10°C to +40°C. Standalone device; no accessories or external power. Designed for clean, dry, unhumidified gas.
Indications for Use
Indicated for use by professional service personnel to measure oxygen purity, flow rate, and gas pressure in oxygen concentrators during maintenance or repair. Not for patient use; not for continuous monitoring or confirmation of oxygen delivery to patients.
Regulatory Classification
Identification
An oxygen gas analyzer is a device intended to measure the concentration of oxygen in respiratory gases by techniques such as mass spectrometry, polarography, thermal conductivity, or gas chromatography. This generic type of device also includes paramagnetic analyzers.
{0}------------------------------------------------
# 510(k) Summary
### JAN - 3 2008
K072469
- Trade name DigiFLO Concentrator ANALYZER .
- Common name DigiHH ●
- Classification name Oxygen gas analyzer (Product Code: CCL)
### Equivalence claimed to:
- Invacare Check O2 Plus K984295 .
- PRO2 CHECK OXYGEN INDICATOR K983500. .
### description of the device
The DigiFLO Concentrator ANALYZER is a lightweight, battery operated device, which measures oxygen concentrator outlet oxygen gas concentration and flow rate by means of ultrasonic wave, which travels from one piezoelectric transducer to another, and Oxygen pressure by means of a piezoresistive pressure transducer. The device is a standalone product and does not have any accessories, power supply adapters and fittings. The device is intended as a tool for use by oxygen concentrator service personnel, to evaluate the output and internal pressures in an oxygen concentrator. It is not intended to be used by patients who are prescribed oxygen, nor is it intended to continuously monitor or confirm oxygen delivery to a patient.
The device connects to an oxygen concentrator through existing therapy tubing presently in use throughout the industry. The device is designed and calibrated for use only on the species gas mixture inherent with the output flow of a concentrator. Information provided by DigiFLO Concentrator ANALYZER is the concentrator's oxygen concentration, flow rate, and pressure. Continuous monitoring of output gas flow between an operating oxygen concentrator and the patient is not possible. Use with any downstream devices or patient accessories is also not possible.
It is further intended that this device be used on clean, dry and unhumidified concentrator output gas. In conjunction with the operating temperature of oxygen concentrators, the ambient temperature range of DigiFLO Concentrator ANALYZER is within +10 to +40 degrees Celsius and should remain in dry indoor environment.
{1}------------------------------------------------
The concentrator gas is expected to fall within the following ranges and will be measured with accuracy's listed below.
| Flow Range | Concentrator outlet: 0 to 20 Liters per Minute ±0.2 L/MAir: 0 to 10 Liters per Minute ±0.2 L/M |
|-------------------|------------------------------------------------------------------------------------------------|
| Oxygen Content | 20.8% to 95.7% oxygen ±1.8% |
| Pressure Range | 0 to 35 PSI, ±0.5% |
| Battery | 9V "N" type battery |
| Temperature range | +10°C to +40°C |
## Intended Use:
The DigiFLO Concentrator ANALYZER is a tool used by service personnel to measure Oxygen purity and Flow at the outlet of an oxygen concentrator. Also, it measures gas pressure inside and at the outlet of an oxygen concentrator. It is not intended to be used by patients who are prescribed oxygen, nor is it intended to continuously monitor or confirm oxygen delivery to a patient.
Intended Environment: The DigiFLO Concentrator ANALYZER is intended to be used in an environment where oxygen concentrators are being serviced or repaired. This includes Hospitals, Nursing Homes, Extended Care Facilities, Patient Homes, and Respiratory Device Service and Repair Centers.
### "Caution:
The Federal Law restricts this device to sale by or on order of a physician, or any other practitioner licensed by the law of the state in which he practices to use or order the use of this device."
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and head. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN - 3 2008
Mr. Michael Stern DigiFLO, Incorporated 6942 96th Avenue, SE Mercer Island, Washington 98040
Re: K072469
Trade/Device Name: DigiFLO Concentrator ANALYZER Regulation Number: 868.1720 Regulation Name: Oxygen Gas Analyzer Regulatory Class: II Product Code: CCL Dated: December 21, 2007 Received: December 28, 2007
Dear Mr. Stern:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Stern
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Suita y. M. Aain Das
Shih-Hsin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indication for Use
510(k) Number K072469
Device Name: DigiFLO Concentrator ANALYZER
Indication For Use:
The DigiFLO Concentrator ANALYZER is a tool used by service personnel to measure Oxygen purity, Flow at the outlet of an oxygen concentrator. Also, it measures gas pressure inside and at the outlet of an oxygen concentrator. It is not intended to be used by patients who are prescribed oxygen, nor is it intended to continuously monitor or confirm oxygen delivery to a patient. The DigiFLO Concentrator ANALYZER is intended to be used in an environment where oxygen concentrators are being serviced or repaired. This includes Hospitals, Nursing Homes, Extended Care Facilities, Patient Homes, and Respiratory Device Service and Repair Centers.
Prescription Use X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Uing Thaler
Lester W. Clark
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.