K982956 · Spirometrics Medical Equipment Co. · BZG · Nov 24, 1998 · Anesthesiology
Device Facts
Record ID
K982956
Device Name
FLOWMATE II PLUS, MODEL #4000
Applicant
Spirometrics Medical Equipment Co.
Product Code
BZG · Anesthesiology
Decision Date
Nov 24, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1840
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Intended use for the Flowmate II PLUS is for Pulmonary Function Testing (measuring a person's ability to move air into and out of their lungs). The volume and flow rates measured for the effort and then reported as numerical results with a printout of the results if selected.
Device Story
Flowmate II PLUS is a desktop diagnostic spirometer used in hospitals, clinics, and industrial screening. Device utilizes a flow-sensing pneumotach base to measure differential pressure as air passes through; calculates pulmonary function parameters including FVC, FEV1, FEV1/FVC%, FEF25-75%, PEF, and MVV. System performs automatic test calculations and selects best test results based on ATS 1994 criteria. Clinicians operate device to assess lung function; results are displayed and printed on thermal or standard paper. Cross-contamination is managed via external filter system. Device provides objective data to assist physicians in diagnosing respiratory conditions, monitoring therapeutic efficacy, and evaluating disability or occupational health status.
Clinical Evidence
Bench testing only. Performance validated using a Pulmonary Waveform Generator (PWG) to simulate 24 BTPS waveforms and 26 PEF waveforms. Accuracy and precision for FVC, FEV1, FEF 25-75, PEF, and MVV compared against known 3-liter syringe inputs and ATS 1994 criteria. Results showed device met or exceeded ATS recommendations for accuracy and precision.
Technological Characteristics
Desktop design; molded high-impact UL-rated ABS plastic. Flow-sensing pneumotach; measures differential pressure. Capacity 10L; volume accuracy +/- 3%; max flow rate +/- 16 L/s; back pressure < 1.5 cm H2O. Connectivity: internal/external printer support. Calibration: field-calibrated via 1-9L syringe. Safety: IEC 601-1, UL 2601, EN60601-1-2. Software: automatic calculation and best-test selection based on ATS 1994 criteria.
Indications for Use
Indicated for pulmonary function testing in children and adults. Used for diagnostic evaluation of respiratory symptoms (dyspnea, wheezing, cough, chest pain), signs (cyanosis, diminished breath sounds), or abnormal lab results. Used to measure disease effects, screen high-risk individuals (smokers, occupational exposure), assess preoperative risk, prognosis, and fitness for physical activity. Used for monitoring therapeutic interventions (bronchodilators, steroids), disease progression (obstructive/interstitial lung, cardiac, neuromuscular), and drug toxicity. Used for disability/impairment evaluations (rehabilitation, insurance, legal) and public health/epidemiologic surveys.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date November 24, 1998. The month is abbreviated to NOV. The numbers are clearly printed in a bold, sans-serif font. The date is likely extracted from a document or photograph.
K982956
Image /page/0/Picture/2 description: The image shows the logo for Spirometrics Medical Equipment Company. The logo consists of a stylized graphic on the left, followed by the company name in a stylized font. Below the company name is the address "415 Rodman Road Auburn, Maine 04210".
## 510(k) SUMMARY
Submitted by: New VED Inc. d.b.a. Spirometrics Medical Equipment Co. 415 Rodman Road Auburn, ME 04210 207-784-0906, Fax 207-784-1481
Donald Henton, Official Correspondent Contact Person: August 21, 1998 Date Prepared:
- Classification Name: Diagnostic Spirometer �
- Common or Usual Name: Diagnostic or Screening Spirometer �
- Proprietary Name: Flowmate II PLUS �
- Establishment Registration Number: 1720605 �
- Classification: Diagnostic Spirometers have been classified as Class II devices. �
- � Performance Standard: None applicable from FDA, used conformance with
American Thoracic Society (ATS) 1994 Spirometry criteria
Spirometrics has been marketing flow sensing pneumotachs with models Flowmate, Flowmate LTE, and PC Flow + devices. This listed product is substantially equivalent to the Flowmate 2500LTE premarket notification number K954759, Flowmate 2500 premarket notification number K863953/A, and PC Flow + 3350 premarket notification number K900673 manufactured by Spirometrics, the Renaissance from Puritan Bennett premarket notification number K911143, and the Presto Flash Portable Spirometry System from Burdick Inc. premarket notification number K894997.
For a description of the device see Summary Exhibit 1.
Description of the device:
- � Flow sensing pneumotach and base that measures differential pressure as air passes through it (the same principle as the Flowmate LTE, Flowmate, and PC Flow + devices)
510(k) SUMMARY
"Dedicated to Improving the Quality of Life"
{1}------------------------------------------------
- Desktop design molded from high impact UL rated ABS plastic �
- Automatic calculations of test results, and best test selection based on � American Thoracic Society (ATS) criteria
- Testing capabilities include; FVC, FEV1, FEV3, FEV1/FVC%, FEF.2-1.2, � FEF25-75%, FEF75-85, PEF, FEF50, FIVC, FEF50/FIF50, PIF, MVV, Methacholine Challenge, and Lung Age
- Testing Capabilities: Capacity 10 Liters �
Volume Accuracy +/- 3% Max. Flow Rate +/- 16 L/s
Back Pressure Less than 1.5 cm H2O
- � Population Predicted Normals included: Composite
Knudson - 1976 Knudson - 1983 Crapo (ITS) ECCS Hsu Lam
- � Customization allows user to change report format, input temperature and barometric readings, and select internal / external printer and associated driver for printer selected
- � Reports printed on thermal paper 112mm wide or standard 8 1/2 " x 11" paper
- � Field calibration with a 1 to 9 Liter calibration syringe
- Cross-contamination control via external filter system �
Intended use for the Flowmate II PLUS is for Pulmonary Function Testing (measuring a person's ability to move air into and out of their lungs). The volume and flow rates measured for the effort and then reported as numerical results with a printout of the results if selected.
Waveform performance testing was conducted using a Pulmonary Waveform Generator (PWG), and showed that the Flowmate II PLUS exceeded the ATS recommendations for FVC, FEV1, FEF 25-75 and PEF accuracy and precision at all of the 24, testing under Body Temperature Pressure Saturated with water (BTPS) (see Summary Exhibit 2), and the 26 PEF waveforms (see Summary Exhibit 5).
The exhale / inhale performance of the device was tested using exponential waveforms created by the PWG using the same exponential curve in both exhale / inhale maneuvers. FVC, FIVC, PEF and PIF were recorded from four progressively faster exponential waveforms, with the accuracy and precision compared against the known three (3) liter syringe inputs. See Summary Exhibit 3 & 4.
MVV testing with the PWG also showed acceptable results compared with the ATS criteria for accuracy and precision to the delivered values (See Exhibit 3 & 4).
## 510(k) SUMMARY
{2}------------------------------------------------
EMC and safety testing has shown the device to meet the EN60601-1-2 for emmissions and immunity, and the IEC 601-1 and UL 2601 safety standard for medical equipment. See Summary Exhibit 6.
Our conclusion is that the Flowmate II PLUS is substantially equivalent to the predicate device the FlowmateLTE. It is also substantially equivalent to the Flowmate, PC Flow+, Renaissance, and Presto Flash Portable Spirometry System.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the seal for the Department of Health & Human Services USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an abstract image of three faces in profile, stacked on top of each other.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 2 4 1998
Mr. Donald Henton Spirometrics Medical Equipment Company 415 Rodman Road Auburn, ME 04210
K982956 Re: Flowmate II PLUS Regulatory Class: II (two) Product Code: 73 BZG Dated: October 30, 1998 Received: November 5, 1998
Dear Mr. Henton:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Mr. Donald Henton
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callehan Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number: K982956
Device Name: Flowmate II PLUS
The intended environments for use are hospitals, physician / clinician offices, and industrial health screening locations.
The intended patient population is from children to adults.
Indications For Use: Pulmonary Function Testing
Diagnostic
To evaluate symptoms, signs, or abnormal laboratory tests
- Symptoms: dyspnea, wheezing, orthopnea, cough, phlegm production, chest pain
- Signs: diminished breath sounds, overinflation, expiratory slowing, cyanosis, chest deformity, unexplained crackles
- Abnormal laboratory tests: hypoxemia, hypercapnia, polycythemia, abnormal chest radiographs
To measure the effects of disease on pulmonary function
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Optional Format 3-10-98)
Prescription Use
illone
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number
{6}------------------------------------------------
510(k) Number: K982956
Device Name: Flowmate II PLUS
Indications For Use: Pulmonary Function Testing - continued
Diagnostic - continued
To screen individuals at risk of having pulmonary diseases
- Smokers
- Individuals in occupations with exposures to injurious substances
- Some routine physical examinations
- To assess preoperative risk
- To assess prognosis (lung transplant, etc.)
- To assess health status before enrollment in strenuous phsical activity programs
Monitoring
To assess therapeutic interventions
- Bronchodilator therapy
- Steroid treatment for asthma, interstitial lung disease, etc.
- Management of congestive heart failure
- Other (antibiotics in cystic fibrosis, etc.)
- To describe the course of diseases affecting lung function
- Pulmonary diseases
- Obstructive airways diseases
- Interstitial lung diseases
- Cardiac diseases
Congestive heart failure
- Neuromuscular diseases
Guillain-Barré
To monitor persons in occupations with exposure to injurious agents
To monitor for adverse reactions to drugs with known pulmonary toxicity
## Disability / Impairment Evaluations
To assess patients as part of a rehabilitation program
- Medical
- Industrial
- Vocational
To assess risks as part of an insurance evaluation
To assess individuals for legal reasons
- Social Security or other government compensation programs
- Personal injury lawsuits
- Others
krame
{7}------------------------------------------------
510(k) Number: K982956
Indications For Use: Pulmonary Function Testing - continued
Public Health
Epidemiologic surveys
- Comparison of health status of populations living in different environments
- Validation of subjective complaints in occupational / environmental settings Derivation of reference equations
1. American Thoracic Society. 1995. Standardization of Spirometry 1994 Update. Am. J. Respir. Crit. Care. Med. 152:1108.
ert thanle
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.