K202837 · Monitored Therapeutics, Inc. · BZG · Dec 4, 2020 · Anesthesiology
Device Facts
Record ID
K202837
Device Name
GoSpiro
Applicant
Monitored Therapeutics, Inc.
Product Code
BZG · Anesthesiology
Decision Date
Dec 4, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.1840
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The GoSpiro® is intended to be used by adults and children over 5 years old in physician's offices, clinics and home settings to conduct basic lung function and spirometry testing.
Device Story
GoSpiro is a diagnostic spirometer measuring lung function via a vertical turbine volume sensor. It captures real-time inspiratory and expiratory flow-volume loop data. The device transmits data via Bluetooth to a smartphone or tablet running the GoSpiro App. The internal program performs calculations meeting ATS/ERS requirements, including quality control indices like time to peak flow, back-extrapolated volume, and total expiratory time. Used in clinics or homes by patients or clinicians, the device provides flow-volume and volume-time curves to assist healthcare providers in assessing lung function. The turbine is designed for multi-patient, multi-use via cleaning and disinfection. It is powered by an internal rechargeable Lithium battery.
Clinical Evidence
Bench testing only. Validation included high-level disinfection validation, post-disinfecting performance testing, post-disinfecting biocompatibility, and material evaluation per ISO 10993-1.
Technological Characteristics
Vertical turbine volume sensor; electronic temperature sensor for BTPS correction; Bluetooth connectivity; rechargeable 3.7V 500 mAh Lithium battery; IP22 ingress protection. Complies with AAMI ANSI ES 60601-1, IEC 60601-1-2, and IEC 60601-1-11. Materials identical to predicate.
Indications for Use
Indicated for adults and children over 5 years old for basic lung function and spirometry testing in physician offices, clinics, and home settings.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
December 4, 2020
Monitored Therapeutics, Inc. % Paul Dryden President - consultant Monitored Therapeutics, Inc. c/o ProMedic, LLC 131 Bay Point Dr. NE St. Petersburg, Florida 33704
Re: K202837
Trade/Device Name: GoSpiro® Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: Class II Product Code: BZG Dated: November 3, 2020 Received: November 4, 2020
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Malvina B. Eydelman, M.D. Director OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K202837
Device Name
# GoSpiro®
Indications for Use (Describe)
The GoSpiro® is intended to be used by adults and children over 5 years old in physician's offices, clinics and home settings to conduct basic lung function and spirometry testing.
Type of Use (Select one or both, as applicable)
XX Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
PSC Publishing Services (301) 443-6740
вн
{3}------------------------------------------------
| Sponsor: | Michael Taylor - CEO<br>Monitored Therapeutics, Inc.<br>5995 Shier Rings Road, Ste A<br>Dublin, OH 43016<br>Tel: +614-761-3555 |
|------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Submission Correspondent: | Paul Dryden<br>ProMedic, LLC |
| Proprietary or Trade Name: | GoSpiro® |
| Common/Usual Name: | Diagnostic Spirometer |
| Regulation Number:<br>Regulation Code:<br>Product Code:<br>Regulatory Class: | 21CFR 868.1840<br>Diagnostic Spirometer<br>BZG<br>II |
| Predicate Device: | Monitored Therapeutics, Inc. – K163249 – GoSpiro® |
#### Modification:
Changing the turbine from single patient, multi-patient, multi-patient, multi-use.
#### Device Description:
The GoSpiro® is intended to be used by adults and children over 5 years old in physician's offices, clinics and home settings to conduct basic lung function and spirometry testing.
The GoSpiro spirometer transmits real-time lung function data to computers, tablets or smartphones over a Bluetooth connection for tele-healthcare applications. The GoSpiro performs full flow-volume loops including inspiratory and expiratory data. The internal program performs all of the calculations for measurements to meet American Thoracic Society and European Respiratory Society requirements. It has built-in quality control measurements and transmits indices of measurement quality including time to peak flow, back-extrapolated volume, total expiratory time, and end-expiratory flow detection.
It is used with the GoSpiro App display and communications software on a smartphone or tablet.
The GoSpiro is powered by an internal rechargeable Lithium battery and is charged via its USB charging station connected to a USB power source. The device complies with ES 60601-1, IEC 60601-1-2, and IEC 60601-1-11.
The fundamental technology to measure flow is a vertical turbine volume sensor. The turbine transducer measures expired air directly at B.T.P.S. (body temperature and pressure with saturated water vapor) thus avoiding the requirement for temperature correction on exhalation. An electronic temperature sensor on the device PCB measures atmospheric temperature, thus enabling correction of inspired volumes and flows. This transducer is insensitive to the effects of condensation and temperature and avoids the need for individual calibration prior to performing a test.
{4}------------------------------------------------
# Indications for Use:
The GoSpiro® is intended to be used by adults and children over 5 years old in physician's offices, clinics and home settings to conduct basic lung function and spirometry testing.
# Device Comparison:
Table 1 compares the subject device to the predicate Monitored Therapeutics – GoSpiro®, K163249.
| Table 1 - Comparison Predicate to Proposed Device | | | | | |
|---------------------------------------------------|--|--|--|--|--|
|---------------------------------------------------|--|--|--|--|--|
| Technical feature/specification | Predicate<br>K163249 - GoSpiro® | Proposed<br>GoSpiro® |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The GoSpiro® is intended to be used by<br>adults and children over 5 years old in<br>physician's offices, clinics and home<br>settings to conduct basic lung function<br>and spirometry testing. It is single-<br>patient use device. | The GoSpiro® is intended to be used<br>by adults and children over 5 years old<br>in physician's offices, clinics and<br>home settings to conduct basic lung<br>function and spirometry testing. |
| Environment of use | Physician's offices, clinics and home<br>settings | Physician's offices, clinics and home<br>settings |
| Patient Population | Adults and pediatric patients over 5<br>years old | Adults and pediatric patients over 5<br>years old |
| Use with PFT filter after each use | Yes | Yes |
| Turbine may be cleaned / disinfected | No, had to be replaced if no filter was<br>used | Cleaned and Disinfected |
| Technology for measure flow and<br>volume | Bidirectional Turbine | Bidirectional Turbine |
| Energy Type | Rechargeable 3.7V, 500 mAh Lithium<br>battery | Rechargeable 3.7V, 500 mAh Lithium<br>battery |
| Physical configuration | Touchscreen on smartphone or tablet and<br>hard buttons | Touchscreen on smartphone or tablet<br>and hard buttons |
| Operating Conditions | Temperature - 17 to 35°C<br>Humidity - 30%RH to 75%RH | Temperature - 17 to 35°C<br>Humidity - 30%RH to 75%RH |
| Weight | 10.5 oz. | 10.5 oz. |
| Size | 3.5 x 4.5" | 3.5 x 4.5" |
| Flow Range | ±14 l/s | ±14 l/s |
| Flow Accuracy | ±5% or 200 mL/s | ±5% or 200 mL/s |
| Flow Resistance | 137 Pascals (Pa) per Liter per second,<br>measured at 14 Liters per second (Lps) | 137 Pascals (Pa) per Liter per second,<br>measured at 14 Liters per second (Lps) |
| Volume Range | 0-8 liters | 0-8 liters |
| Technical feature/specification | Predicate | Proposed |
| | K163249 - GoSpiro® | GoSpiro® |
| Volume accuracy | ±3% of reading, or 0.05 liters, whichever<br>is greater. | ±3% of reading, or 0.05 liters,<br>whichever is greater. |
| Used with PFT filter and / or<br>mouthpiece | Yes | Yes |
| Communication | Bluetooth to Wi-Fi or Cellular | Bluetooth to Wi-Fi or Cellular |
| Coaching | Display of flow-time curves, 6 second<br>count-down timer, tone at 6 seconds. | Display of flow-time curves, 6 second<br>count-down timer, tone at 6 seconds. |
| Feedback of test quality | Yes- Trend data | Yes- Trend data |
| Test stored | Yes | Yes |
| Typical battery life per charge | Approximately 140 measurements | Approximately 140 measurements |
| Auditory / visual alarms | Yes | Yes |
| Allows inspiration before forced<br>exhalation | Yes | Yes |
| Tests Performed | *FVC | *FVC |
| | FEV0.75, *FEV1, FEV3, FEV6 | FEV0.75, *FEV1, FEV3, FEV6 |
| | FEV/FVC (FER) for 0.75 /1 /3 / 6 | FEV/FVC (FER) for 0.75 /1 /3 / 6 |
| | *PEF | *PEF |
| | MMEF | MMEF |
| | FEF 25/50/75 | FEF 25/50/75 |
| | FEF25-75 (MEF) | FEF25-75 (MEF) |
| | FIV1 | FIV1 |
| | FIVC | FIVC |
| | PIF | PIF |
| | FIF25-75 (MIF25-75) | FIF25-75 (MIF25-75) |
| | FIF 25/50/75 | FIF 25/50/75 |
| | MET25-75 | MET25-75 |
| | FEV0.75/FEV6 | FEV0.75/FEV6 |
| | FEV1/FEV6 | FEV1/FEV6 |
| | FEF50/FVC | FEF50/FVC |
| | MMEF/FVC (FEF25-75/FVC) | MMEF/FVC (FEF25-75/FVC) |
| | FIV1/FIVC (FIR) | FIV1/FIVC (FIR) |
| | R50 (FEF50/FIF50) | R50 (FEF50/FIF50) |
| | FET | FET |
| | MVV (ind) | MVV (ind) |
| | Vext | Vext |
| | Time to PEF | Time to PEF |
| | Possible Cough | Possible Cough |
| | LAST500V | LAST500V |
| | Flow / Volume Curve (on server) | Flow / Volume Curve (on server) |
| | Volume / Time Curve (on server) | Volume / Time Curve (on server) |
| | Flow / Time Curve | Volume / Time Curve (on server) |
| Page 4 of 4 | | |
| Technical feature/specification | Predicate | Proposed |
| | K163249 - GoSpiro® | GoSpiro® |
| ATS Spirometry guidelines | Yes | Yes |
| Standards Compliance | AAMI ANSI ES 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11 | AAMI ANSI ES 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11 |
| Water Ingress Protection | IP22 | IP22 |
| Biocompatibility and Patient Contact | Externally communicating (Indirect),<br>Tissue and Surface Contact, Mucosa,<br>limited exposure | Similar patient contact and identical<br>materials |
| Cleaning and Disinfection validation | Cleaning of mouthport | Same mouthport<br>Cleaning and disinfection of Turbine |
{5}------------------------------------------------
# 510(k) Summary
December 4, 2020
{6}------------------------------------------------
# 510(k) Summary December 4, 2020
* Indicates parameters displayed for patients only.
#### Differences vs. Predicate:
The only difference between the predicate and the subject device is the labeling to allow for cleaning, disinfection and reuse of the turbine. The validation testing has been performed demonstrating that the turbine may be cleaned, disinfected and reused.
All other features and specifications are identical to the cleared K163249 device.
## Substantial Equivalence Discussion:
As the devices are identical the only discussion relates to the change in the ability to clean, disinfect and reuse the turbine for multi-patient, multi-use. Disinfection is only required when a filter is not used with each patient.
All other comparative features are identical.
- Indications for Use / Patient Population / Environment of Use ●
- Prescriptive
- Design and Technology ●
- Performance and Specifications ●
- Compliance with Standards
- Performance Testing ●
- High level disinfection validation
- Post-disinfecting performance o
- Post-disinfecting biocompatibility о
- . Biocompatibility
- The materials have been evaluated per ISO 10993-1 o
## Substantial Equivalence Conclusion:
The GoSpiro® is substantially equivalent to the cleared GoSpiro® predicate, K163249.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.