The Asthma Monitor AM3/AM3 BT/AM3 GSM is an electronic measurement device to monitor the lung function (determination of the respiratory flows and volume) with high reproducibility wherever and whenever there is a need of. The AM3 measures the flow during expiration serving for the calculation of further parameters as FEV1. The AM3 is used to monitor the respiratory status of adult human beings in the areas of asthma, chronic obstructive pulmonary disorder and in areas like occupational medicine, clinical trials and disease management. The patient is informed of the results by numeric values for the selected parameters (e.g. PEF, FEV1). Furthermore, a visual control unit, displayed in the form of traffic lights, allows an immediate indication of the measurement based on criteria defined by the patient's physician. The device saves the results of a measurement (always with date and time) automatically in an internal database. In addition, questionnaire functionality can be called up by the use of a software package (AMOS) to record e.g. the "Quality of Life" status. When enabled, the AM3 can be programmed with a couple of questions, where the patient can then select from a couple of different answers. This information is also stored in the internal database and can be transmitted for evaluation to a standard PC using the software package AMOS. The AM3 is designed to replace an ordinary peak flow meter, diary and pencil by a single system. Easy handling, sturdy and handy design allow the Asthma Monitor AM3 to be used in healthcare, clinical and home use environments/settings.
Device Story
Handheld electronic spirometer; measures expiratory flow to calculate PEF and FEV1; includes patient-reported outcome questionnaire functionality. Used in clinical, healthcare, and home settings by adult patients. Device stores up to 1200 measurements and 400 questionnaire sets in internal non-volatile memory. Data transmitted to PC/database via USB, Bluetooth, serial (RS232), or integrated GSM module. Provides immediate visual feedback via traffic-light display based on physician-defined criteria. Replaces manual peak flow meters and paper diaries; facilitates remote monitoring and clinical trial data collection. Data transfer is offline, ensuring no interference with active measurements.
Clinical Evidence
Bench testing only. No clinical data presented. Verification included FMEA risk analysis, EMC testing per EN 60601-1-2, and GSM module testing per R&TTE and FCC guidelines. Software verified per IEC 62304.
Technological Characteristics
Handheld diagnostic spirometer; flow sensor and mouthpiece (single patient use); Polysterol 454C and Bormed RG835 MO materials. Energy source: Li-Ion Polymer battery (3.7V, 1700 mAh). Connectivity: USB, RS232, Bluetooth (WML-C46), and GSM (Sierra Wireless WISMO288). Dimensions: 112x82x37 mm; Weight: 120g. Software developed per IEC 62304.
Indications for Use
Indicated for adult patients with asthma, COPD, or those in occupational medicine, clinical trials, and disease management to monitor lung function (PEF, FEV1) and record patient-reported outcomes via questionnaires.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
{0}------------------------------------------------
# MAY 0 8 2014
# 510(k) Summary
# GENERAL INFORMATION
#### 5.1 Type of Submission
Special 510(k) Submission
Submission date: 12/02/2013
#### 5.2 Submitter
Name: eResearchTechnology GmbH
Address:
Sieboldstrasse3 D-97230 Estenfeld Germany
# Contact person in Germany:
Address:
Phone:
E-mail
FAX:
Johannes Thalhammer eResearchTechnology GmbH Sieboldstrasse 3, 97230 Estenfeld Germany +49 9305 720 6376 +49 9305 720 60 johannes.thalhammer@ert.com
Contact person in the U.S .: (Official Correspondent) Address
# Richard Miller
eResearchTechnology 1818 Market Street, Suite 1000 Philadelphia, PA 19103
Phone/Fax: E-mail
215-2825588 RMiller@ert.com
,
{1}------------------------------------------------
### 5 510(k) Summary
- 5.3 Establishment Registration Number 3008505660
- 5.4 Common Name or Classification Name Diagnostic Spirometer (CFR 868.1840, Product Code BZG)
#### 5.5 Trade Name
Asthma Monitor AM3 / AM3 BT / AM3 GSM
### 5.6 Device Classification
This is a Class II device
#### 5.7 Classification Panel
73 Anesthesiology Part 868 Code BZG
#### 5.8 Reason for Premarket Notification
Device modification to an existing device regarding "The New 510(k) Paradigm"
- Additional data transfer to Database by GSM;
- Li-lon battery - is used as energy type
#### 5.9 Legally predicate marketed device
Asthma Monitor AM3/AM3 BT K092890 Code BZG
#### Predicate Device Company 5.10
eResearchTechnology GmbH (device was formally listed by Carefusion Germany 234 GmbH)
#### Device Description 5.11
The Asthma Monitor AM3 GSM is a medical device (peak flow meter with symptom diary) providing following characteristics:
- Handheld device ●
- Battery operation .
- Storing capacity of 1200 measurements .
- Storing capacity of 400 sets of questionnaires (max. 20 questions each ) .
- . Measurement Parameters: PEF and FEV1
- Accuracy Flow: ± 5% or ± 20 1/min .
- . Accuracy Volume: ± 3% or ± 0.05 liter
- Data transmission to computer/database via USB, Bluetooth, GSM and . Serial (RS232)
- Flow sensor (single patient use) .
- . Mouthpiece (single patient use)
{2}------------------------------------------------
#### Intended Use Statement 5.12
The Asthma Monitor AM3/AM3 BT/AM3 GSM is an electronic measurement device to monitor the lung function (determination of the respiratory flows and volume) with high reproducibility wherever and whenever there is a need of. The AM3 measures the flow during expiration serving for the calculation of further parameters as FEV1.
The AM3 is used to monitor the respiratory status of adult human beings in the areas of asthma, chronic obstructive pulmonary disorder and in areas like occupational medicine, clinical trials and disease management.
The patient is informed of the results by numeric values for the selected parameters (e.g. PEF, FEV1). Furthermore, a visual control unit, displayed in the form of traffic lights, allows an immediate indication of the measurement based on criteria defined by the patient's physician.
The device saves the results of a measurement (always with date and time) automatically in an internal database. In addition, questionnaire functionality can be called up by the use of a software package (AMOS) to record e.g. the "Quality of Life" status. When enabled, the AM3 can be programmed with a couple of questions, where the patient can then select from a couple of different answers. This information is also stored in the internal database and can be transmitted for evaluation to a standard PC using the software package AMOS.
The AM3 is designed to replace an ordinary peak flow meter, diary and pencil by a single system. Easy handling, sturdy and handy design allow the Asthma Monitor AM3 to be used in healthcare, clinical and home use environments/settings.
#### 5.13 Required Components
AM3 / AM3 BT /AM3 GSM measurement device Accessories User Manual
# 5.14 Summary Table of Comparison
- a) Comparison with Asthma Monitor AM3/AM3 BT with 510(k) K092890
| | Asthma Monitor<br>AM3/AM3 BT<br>(K092890) | Asthma Monitor<br>AM3 GSM | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | The Asthma Monitor AM3 / AM3 BT from Cardinal Health is an<br>electronic measurement device<br>to monitor the lung function | The Asthma Monitor AM3/AM3/AM3 GSM is an electronic<br>measurement device to monitor<br>the lung function (determination | |
| (determination of the respiratory<br>flows and volume) with high<br>reproducibility wherever and<br>whenever is a need of. The<br>AM3 / AM3 BT measures the<br>flow during expiration serving<br>for the calculation of further<br>parameters as FEV1. | of the respiratory flows and<br>volume) with high reproducibility<br>wherever and whenever is a<br>need of. The AM3 measures the<br>flow during expiration serving far<br>the calculation of further<br>parameters as FEV1. | | |
| The AM3 / AM3 BT is used to<br>monitor the respiratory status of<br>human beings in the areas<br>asthma, chronic obstructive<br>pulmonary disorder and in<br>areas like occupational<br>medicine, clinical trials and<br>disease management. | The AM3 is used to monitor the<br>respiratory status of human adult<br>beings in the areas asthma,<br>chronic obstructive pulmonary<br>disorder and in areas like<br>occupational medicine, clinical<br>trials and disease management. | | |
| The patient is informed of the<br>results by numeric values for<br>selected parameters (e.g. PEF,<br>FEV1). Furthermore a visual<br>control unit, displayed as a kind<br>of traffic lights, allows an<br>immediate indication of the<br>measurement based on criteria<br>defined by the patient's<br>physician. | The patient is informed of the<br>results by numeric values for<br>selected parameters (e.g. PEF,<br>FEV1). Furthermore a visual<br>control unit, displayed as a kind<br>of traffic lights, allows an<br>immediate indication of the<br>measurement based on criteria<br>defined by the patient's<br>physician. | | |
| The device saves the results of<br>a measurement (always with<br>date and time) automatically in<br>an internal database.<br>The memory capacity is<br>designed to store up to 400<br>measurements.<br>In addition, a questionnaire<br>functionality can be called up by<br>the use of a software package<br>(AMOS) to record e.g. the<br>"Quality of Life" status. When<br>enabled, the AM3 / AM3 BT can<br>be programmed with a couple of<br>questions, where the patient<br>can select then from a couple of<br>different answers | designed to store up to 400<br>measurements.<br>In addition, a questionnaire<br>functionality can be called up by<br>the use of a software package<br>(AMOS) to record e.g. the<br>"Quality of Life" status. When<br>enabled, the AM3 / AM3 BT can<br>be programmed with a couple of<br>questions, where the patient<br>can select then from a couple of<br>different answers.<br>This information is also stored<br>in the internal database and can<br>be transmitted for evaluation to<br>a standard PC using the<br>software package AMOS.<br>The AM3 / AM3 BT is designed<br>to replace ordinary peak flow<br>meter, diary and pencil by a<br>single system. Easy handling,<br>sturdy and handy design allow<br>the AM3 / AM3 BT being used<br>almost everywhere: at work, at<br>home, in school, for experts<br>opinion, research or clinical trial<br>purposes and in occupational<br>medicine. | The device saves the results of a<br>measurement (always with date<br>and time) automatically in an<br>internal database.<br>In addition, a questionnaire<br>functionality can be called up by<br>the use of a software package<br>(AMOS) to record e.g. the<br>"Quality of Life" status. When<br>enabled, the AM3 can be<br>programmed with a couple of<br>questions, where the patient can<br>select then from a couple of | internal database.<br><br>In addition, a questionnaire<br>functionality can be called up by<br>the use of a software package<br>(AMOS) to record e.g. the<br>"Quality of Life" status. When<br>enabled, the AM3 can be<br>programmed with a couple of<br>questions, where the patient can<br>select then from a couple of<br>different answers.<br>This information is also stored in<br>the internal database and can be<br>transmitted for evaluation to a<br>standard PC using the software<br>package AMOS.<br>The AM3 is designed to replace<br>ordinary peak flow meter, diary<br>and pencil by a single system.<br>Easy handling, sturdy and handy<br>design allow the Asthma Monitor<br>AM3 being used in healthcare,<br>clinical and home use<br>environments/settings, for expert<br>opinion, research or clinical trial<br>purposes and in occupational<br>medicine. |
| | | > Besides minor wording<br>changes the Intended Use is<br>identical to K092890 (see italic<br>text in Intended Use of to<br>K092890) | |
| Patient<br>population | The Asthma Monitor can be<br>used for patients from 4 years<br>on and older. | identical | |
| | Length x Width x Height:<br>112*82*37 mm | Length x Width x Height:<br>112*82*37 mm | |
| Dimensions<br>(housing) | Weight:<br>167 g (batteries included) | Weight:<br>120 g (batteries included) | |
| Display | LCD module<br>Size: 54,0 x 33,5 mm<br>255 x 160 dots | Identical | |
| | Foil Key-panel (4 keys): | | |
| Key-panel | - ESC (on/off)<br>- UP-ARROW<br>- DOWN-ARROW<br>- OK | identical | |
| Integrated<br>mouthpiece<br>(material) | Polysterol 454C | identical | |
| Single Use<br>mouthpiece<br>(material) | Bormed RG835 MO | identical | |
| | Parameters: | | |
| Performance<br>(measurements) | PEF<br>FEV1 | identical | |
| Interface | Serial RS 232 & USB &<br>Blue-tooth | Serial RS 232 & USB &<br>Blue-tooth & GSM | |
| Energy type | 3 x 1,5 (Micro AAA) | LI-ION Polymer battery 3.7 V,<br>1700 mAh | |
| Operating<br>Require-<br>ments | PC software AMOS | identical | |
| Bluetooth | WML-C46 (Mitsumi) | identical | |
| | | | |
| interface | | | |
| GSM<br>Interface | none | Sierra Wireless<br>WISMO288 | |
{3}------------------------------------------------
# 5 510(k) Summary
.
.
{4}------------------------------------------------
.
Page 5 of 7
ﻧﮯ
.
{5}------------------------------------------------
### Discussion of the table above:
The insignificant difference to the AM3/AM3 BT (K092890) is:
- GSM is used for wireless data transfer to a Database as an additional . possibility besides the serial, USB and BlueTooth interface communication. GSM (Global System for Mobile communications) is an open, digital cellular technology used for transmitting mobile voice and data services. The in-built GSM module is used instead of an external cell phone paired to the AM3 BT to transmit the data to a database.
- The device performance is not affected by adding the GSM function, as the ● data transfer is offline and will not take place when a measurement is active or a questionnaire is performed.
- . As the data transfer is offline to the measurement or questionnaire functionality and all data is stored in a nonvolatile memorv. The availability of reception to the GSM is not critical, as the data transfer could be done at any time without any influence to the device functionality.
- . The wireless transfer is based on the use of the GSM module Sierra Wireless WISMO288. The GSM module uses techniques to guarantee the integrity of the transferred data. Latency and throughput is part of the GSM module implementation. The use of the qualified GSM module ensures that the latency and throughput requirements were met.
- Minor modifications of the software concerning the GSM module are . verified. The software requirements and design specifications submitted, include requirements and specifications of the entire device. Based on the recommendations of the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" only verification reports related to the modifications which caused the need for a Special 510(k) are submitted.
- . The Applicable EMC and telecommunications standards and regulations, including device emissions that may cause EMI with other equipment are fulfilled.
- · Li-lon battery is used as energy type. The battery has to be charged by a power supply. The Li-Ion battery and the power supply fulfil the common standards and were implemented and tested for safe and effective use.
#### 5.15 Summary of Device Testing
The following practices were followed and monitored for development of the Asthma Monitor AM3 with the GSM data transmission:
- . The GSM data transmission for the above device was developed in accordance with the eResearchTechnology development standard operating procedures.
{6}------------------------------------------------
- The risk analysis method used to assess the impact of Asthma Monitor . AM3 / AM3 BT with the additional GSM data transmission was a Failure Modes and Effects Analysis (FMEA).
- Safety test procedures demonstrate satisfaction of all safety . requirements and mitigation of all identified hazards.
- The EMC testing was performed according EN 60601-1-2. .
- The GSM module was tested according to R&TTE and FCC guidelines. .
- The software was developed according to the IEC 62304 Standard. .
#### 5.16 Conclusions
Based on the above, eResearchTechnology concludes that the Asthma Monitor AM3 with the "GSM" data transmission module does not raise new questions of safety and effectiveness and is as safe and as effective for its intended use, and performs at least as well as the predicate devices.
{7}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
#### Public I lealth Service
Food and Drug Administration 0903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD. 20993-0002
May 8, 2014
eResearchTechnology GmbH Richard Miller Vice President, Quality Management 1818 Market Street, Suite 1000 Philadelphia, PA 19103
Re: K133722
Trade/Device Name: Asthma Monitor AM3 GSM Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: Class II Product Code: BZG Dated: April 07, 2014 Received: April 08, 2014
Dear Mr. Miller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
Page 2 - Mr. Miller
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/8/Picture/7 description: The image contains a logo with text. The text includes the words "Clinical Deputy Director" and "DAGRID/ODE/CDRH". The logo appears to be for an organization or department, possibly related to clinical or regulatory affairs. The image quality is low, making it difficult to discern finer details.
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
# Indications for Use
510(k) Number (if known): KI33722
Device Name:
АМЗ AM3 BT AM3 GSM
Indications for Use:
The Asthma Monitor AM3/AM3 GSM is an electronic measurement device to monitor the lung function (determination of the respiratory flows and volume) with high reproducibility wherever and whenever there is a need of. The AM3 measures the flow during expiration serving for the calculation of further parameters as FEV1.
The AM3 is used to monitor the respiratory status of adult human beings in the areas of asthma. chronic obstructive pulmonary disorder and in areas like occupational medicine, clinical trials and disease management.
The patient is informed of the results by numeric values for the selected parameters (e.g. PEF, FEV1). Furthermore, a visual control unit, displayed in the form of traffic lights, allows an immediate indication of the measurement based on criteria defined by the patient's physician.
The device saves the results of a measurement (always with date and time) automatically in an internal database. In addition, questionnaire functionality can be called up by the use of a software package (AMOS) to record e.g. the "Quality of Life" status. When enabled, the AM3 can be programmed with a couple of questions, where the patient can then select from a couple of different answers. This information is also stored in the internal database and can be transmitted for evaluation to a s…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.