ENLIGHT 2100 is a non-invasive, non-radiation medical device that provides information of local impedance variation within a cross-section of a patient's thorax. This information is presented to the clinician user as an adjunctive tool to other clinical information in order to support the user's assessment of variations in regional air content within a cross section of a patient's lungs. It is intended for mechanically ventilated adult and pediatric patients in a hospital setting, whose thorax perimeter is within the range of 37.5 - 134cm. ENLIGHT 2100 does not measure regional ventilation of the lungs.
Device Story
Ventilatory electrical impedance tomograph (EIT) using 16-32 surface electrodes placed around patient thorax; inputs are surface electrode voltage measurements; device estimates local impedance variation (LIV) within lung cross-section; LIV is linearly related to regional air content variations; high temporal resolution (50 Hz); used in hospital settings by clinicians; output is visual representation of LIV; serves as adjunctive tool for assessing regional air content; aids clinical decision-making for mechanically ventilated patients; benefits include non-invasive, non-radiation monitoring of lung air content variations.
Clinical Evidence
Bench testing only. Performance metrics compared between ENLIGHT 2100 and ENLIGHT 1810 included Signal to Noise Ratio (50-95dB vs 50-85dB), Voltage Accuracy (80-100% vs 85-100%), Drift (Allan Variance <100pV2 for both), Reciprocity Accuracy (95-100% vs 96-100%), Amplitude response (90-104% vs 94-106%), Position error (<4% radius), Ringing (<0.6), Resolution (<0.42), and Plethysmogram error (<5%). Results demonstrate substantial equivalence in imaging quality.
Technological Characteristics
Ventilatory electrical impedance tomograph; 16-32 surface electrodes; non-invasive; non-radiation; transportable monitor-like form factor; digital/analog hardware with driven shields and common mode circuits; includes pneumotachograph OEM module; DC-DC power supplies; no battery; software uses reconstruction meshes for different patient sizes; non-sterile; single-patient use accessories.
Indications for Use
Indicated for mechanically ventilated adult and pediatric patients in a hospital setting with a thorax perimeter of 37.5 - 134 cm.
Regulatory Classification
Identification
A ventilatory electrical impedance tomograph is a prescription non-invasive, non-radiological ventilatory device that provides an assessment of local impedance variation within a cross-section of a patient's thorax.
Special Controls
In combination with the general controls of the FD&C Act, the ventilatory electrical impedance tomograph is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including the following:
(i) Characterization of device parameters, including signal-to-noise ratio, voltage accuracy, drift, reciprocity accuracy, amplitude response, position error, and ringing;
(ii) Real time evaluation of local impedance variation;
(iii) Plethysmogram accuracy testing; and
(iv) Use life testing of reusable components.
(3) Performance data must validate reprocessing instructions for any reusable components of the device.
(4) Performance data must demonstrate the electrical, thermal, and mechanical safety and the electromagnetic compatibility of the device.
(5) Software verification, validation, and hazard analysis must be performed.
(6) Labeling must include the following:
(i) Guidance for interpretation of the images generated;
(ii) A warning that the device should be removed before use of a defibrillator, or defibrillator interaction information based on defibrillator performance testing with the device;
(iii) A use life for any reusable components; and
(iv) Instructions for reprocessing any reusable components.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" being larger and bolder than the rest of the text.
January 6, 2022
Timpel S.A. % Paul Dryden Consultant ProMedic Consulting LLC 131 Bay Point Dr NE Saint Petersburg, Florida 33704
Re: K211135
Trade/Device Name: Enlight 2100 Regulation Number: 21 CFR 868.1505 Regulation Name: Ventilatory Electrical Impedance Tomograph Regulatory Class: Class II Product Code: OEB Dated: December 6, 2021 Received: December 8, 2021
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brandon Blakely, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
## K21135
Device Name
Enlight 2100
Indications for Use (Describe)
ENLIGHT 2100 is a non-invasive, non-radiation medical device that provides information of local impedance variation within a cross-section of a patient's thorax. This information is presented to the clinician user as an adjunctive tool to other clinical information in order to support the user's assessment of variations in regional air content within a cross section of a patient's lungs.
It is intended for mechanically ventilated adult and pediatric patients in a hospital setting, whose thorax perimeter is within the range of 37.5 - 134cm.
ENLIGHT 2100 does not measure regional ventilation of the lungs.
Type of Use (Select one or both, as applicable)
XX Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number. "
FORM FDA 3881 (6/20)
Page 1 of 1
{3}------------------------------------------------
| Date Prepared: | 6-Jan-22 |
|---------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | TIMPEL S.A.<br>Rua Simão Álvares, 356 Cj. 41/42<br>Pinheiros, São Paulo/SP - Brazil<br>T - +55-113088-0305<br>E - Rafael.holzhacker@timpelmedical.com |
| Submission Correspondent: | Paul Dryden<br>ProMedic, LLC<br>131 Bay Point Dr NE<br>St. Petersburg, FL 33704 |
| Proprietary or Trade Name: | ENLIGHT 2100 US |
| Regulation Number:<br>Regulation Name:<br>Product code: | 21 CFR 868.1505<br>Ventilatory electrical impedance tomograph<br>QEB |
| Predicate Device: | Timpel – ENLIGHT 1810 – DEN170072 |
### Device Description:
ENLIGHT 2100 is a Ventilatory electrical impedance tomograph that uses several electrodes (usually between 16 and 32) placed around the patient's thorax to assess regional impedance variation in a lung slice (tomography). It provides a relative measurement, so it only provides information on variations in local impedance.
ENLIGHT 2100 estimates Local Impedance Variation, occurring in a cross section of the thorax during a respiratory cycles, and which are linearly related to Variations in Regional Air Content within the lung.
## Principle of Operation:
Electrical impedance tomography (EIT) is a technique in which the electrical properties of tissues are estimated from surface electrode voltage measurements and used to provide information on Local Impedance Variation (LIV) within a cross section of a patient's thorax. Considerable electrical impedivity variations are imposed on lung tissue by variations in the lung's air content.
There is a linear relationship between the variations in air content and the percentage change in lung tissue impedance. This is a characteristic of lung tissue due to the properties of alveoli walls. This linear relationship is explored in Electrical Impedance Tomography, supporting its use as a tool to support the user's assessment of the variations in regional air content within a cross section of a patient's lungs.
The electrode voltage measurement can be made at high rates as 50 times / second, making it possible to display Local Impedance Variations with a high temporal resolution, almost continuously.
#### Indications for Use:
ENLIGHT 2100 is a non-invasive, non-radiation medical device that provides information of local impedance variation within a cross-section of a patient's thorax. This information is presented to the clinician user as an adjunctive tool to other clinical information in order to
{4}------------------------------------------------
support the user's assessment of variations in regional air content within a cross section of a patient's lungs.
It is intended for mechanically ventilated adult and pediatric patients in a hospital setting, whose thorax perimeter is within the range of 37.5 - 134 cm.
ENLIGHT 2100 does not measure regional ventilation of the lungs.
#### Patient Population:
Mechanically ventilated adult and pediatric patients whose thorax perimeter is within the range of 37.5 -134 cm.
Environments of use:
Hospital setting.
#### Substantial Equivalence Discussion
Indications - There are no differences in the indication for use, besides the patient population.
Patient Population - The intended Patient Population for ENLIGHT 2100 are adult and pediatric patients, whose thorax perimeter is within the range of 37.5 - 134 cm. ENLIGHT 2100 is indicated for mechanically ventilated patients, so the patient is usually sedated, and the application of the electrode belt is performed in uncooperative patients, regardless being adult or pediatric patients. There are no differences in the algorithm or software between different sizes of electrode belt, only the reconstruction meshes are different, but with the same performance.
Environment of Use - There is no difference in the environment of use.
Technological Characteristics - The Operating Principle is the same for ENLIGHT 2100 and ENLIGHT 1810. The hardware system was redesigned, reducing the module volume with improved hardware lavout. The digital part of the hardware was redesigned and the analog hardware (driven shields and common mode circuits of acquisition channels) have the same wellproven topology and only layout was improved.
The enclosure also includes the pneumotachograph OEM Module (the same part number used in ENLIGHT 1810) and DC-DC power supplies, that were previously presented within other mechanical enclosures - Hub and power supply enclosure, respectively.
Improvements on power dissipation were implemented, reducing static consumption of operational amplifiers by reducing the voltage of power supplies without compromising performance. ENLIGHT 1810 OEM AC-DC power supply was replaced by a lower power version OEM module of same manufacturer, as the power consumption of ENLIGHT 2100 US is lower than its predecessor. There is no battery in ENLIGHT 2100 US.
{5}------------------------------------------------
## Non-clinical testing
Biocompatibility – There is no difference in the patient contact between the proposed device and the predicate.
Bench testing - We have performed the following performance tests, with similar results between the 2 devices.
#### Substantial Equivalence Conclusion
Data presented indicated that there is substantial equivalence between ENLIGHT 2100 and ENLIGHT 1810.
{6}------------------------------------------------
| | | | Explanation of Differences |
|---------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Attributes | Subject<br>ENLIGHT 2100 | Predicate<br>ENLIGHT 1810 | |
| 510(k) | DEN170072 | | |
| Product Classification<br>CFR | 868.1505 - QEB | 868.1505 - QEB | Similar |
| Indications for Use | ENLIGHT 1810 is a non-invasive, non-<br>radiation medical device that provides<br>information of local impedance variation<br>within a cross-section of a patient's thorax.<br>This information is presented to the clinician<br>user as an adjunctive tool to other clinical<br>information in order to support the user's<br>assessment of variations in regional air content<br>within a cross section of a patient's lungs.<br>It is intended for mechanically ventilated adult<br>and pediatric patients in a hospital setting,<br>whose thorax perimeter is within the range of<br>37.5 -134 cm.<br>ENLIGHT 2100 does not measure regional<br>ventilation of the lungs. | ENLIGHT 2100 is a non-invasive, non-<br>radiation medical device that provides<br>information of local impedance variation<br>within a cross-section of a patient's thorax.<br>This information is presented to the clinician<br>user as an adjunctive tool to other clinical<br>information in order to support the user's<br>assessment of variations in regional air content<br>within a cross section of a patient's lungs.<br>It is intended for mechanically ventilated adult<br>patients in a hospital setting, whose thorax<br>perimeter is within the range of 78-122 cm. | Inclusion of new sizes of accessories and<br>software modifications to expand the patient<br>range. |
| Patient Population | Pediatric patients, whose thorax perimeter is<br>within the range of 37.5 -134 cm. | Adult patients, whose thorax perimeter is<br>within the range of 78-122 cm. | Inclusion of pediatric patients. |
| Patient type | Mechanically ventilated patients | Mechanically ventilated patients | No differences. |
| Prescriptive | Yes | Yes | No differences. |
| Principle of Operation | Electrical Impedance Tomography based on<br>voltage measures to estimate local impedance<br>variation within a cross-section of a patient's<br>thorax. | Electrical Impedance Tomography based on<br>voltage measures to estimate local impedance<br>variation within a cross-section of a patient's<br>thorax. | No differences. |
| Contraindications | Declared on Instructions for use | Declared on Instructions for use | No differences. |
| Environment of Use | Hospital Setting | Hospital Setting | No differences. |
| Duration of Use | Up to 30 days, with Addere Change each 48<br>hours. | Up to 30 days, with Addere Change each 48<br>hours. | No differences. |
| Attributes | Subject<br>ENLIGHT 2100 | Predicate<br>ENLIGHT 1810 | Explanation of Differences |
| Useful life | ENLIGHT 2100 has 7 years of useful life. | ENLIGHT 1810 has 7 years of useful life. | No differences. |
| Shelf life | Addere has 2 year of shelf life.<br>Electrode Belt has 1 year of shelf life. | Addere has 1 year of shelf life.<br>Electrode Belt has 1 year of shelf life. | There is no change of shelf life for the<br>Electrode Belts, but new validation test data<br>with real-time aged samples increased Addere<br>shelf life from 1 to 2 years. |
| Non-sterile | There are no sterile components or accessories. | There are no sterile components or accessories. | No differences. |
| Cleaning methods | Cleaning and disinfection prescribed only for<br>the device.<br>Accessories are single patient use. | Cleaning and disinfection prescribed only for<br>the device.<br>Accessories are single patient use. | No differences. |
| Features | | | |
| Available sizes | Electrode Belt sizes P0, P1, P2, 4S, 5S, XXS,<br>XS, S, M, L, XL<br>Addere sizes P0, P1, P2, 4S, 5S, XXS, XS, S,<br>M, L, XL | Electrode Belt sizes XS, S, M, L<br>Addere sizes XS, S, M, L | New Electrode Belts and Addere sizes<br>included to expand the patient range. |
| | Shaper No 2<br>Reference Cable (Single size) | Shaper No 2<br>Reference Cable (Single size) | |
| Shape | Monitor-like device, transportable | Monitor-like device, mobile, with incorporated<br>cart | Device size is smaller to facilitate the shipping<br>logistics and device handling inside the<br>hospital. |
| Patient Contact per<br>ISO 10993-1 | Surface contact<br>Intact skin<br>Prolonged duration (Up to 30 days) | Surface contact<br>Intact skin<br>Prolonged duration (Up to 30 days) | No differences. |
| Attributes | Subject<br>ENLIGHT 2100 | Predicate<br>ENLIGHT 1810 | Explanation of Differences |
| Performance Characteristics – Bench Test with Electrode Belt size S | | | |
| Signal to Noise Ratio<br>(SNR) | (50dB - 95dB) | (50dB - 85dB) | SNR and Drift are both parameters that reflect<br>the precision of data acquisition. SNR<br>indicates the system noise in data acquisition. |
| Voltage Accuracy | (80% - 100%) | (85% - 100%) | indicates the system noise in data acquisition. |
| Drift | Allan Variance converges to zero (below<br>100pV2) | Allan Variance converges to zero (below<br>100pV2) | Drift indicates a long-term stability in data<br>acquisition.<br>ENLIGHT 2100 has slight better performance<br>considering the Signal to Noise Ratio when<br>compared to ENLIGHT 1810. The drift of<br>both devices is similar, as Allan Variance of<br>both devices converge to less than 100pV2,<br>which is a negligible value when considering<br>the background noise measured by SNR.<br>ENLIGHT computes normalized difference<br>voltages from the acquired voltages to<br>generate local impedance variation images and<br>not on the absolute voltages itself. The<br>subtraction of a reference voltage reduces the<br>effect of hardware imperfections related to<br>bias and the normalization by a reference<br>voltage reduces the effect of imperfections<br>related to gain. Voltage Accuracy and<br>Reciprocity Accuracy reflect those<br>imperfections even though attenuated by the<br>imaging algorithm.<br>Voltage Accuracy and Reciprocity Accuracy<br>of ENLIGHT 2100 are similar to ENLIGHT<br>1810, but as indicated above, the imaging |
| Reciprocity Accuracy | (95% - 100%) | (96% - 100%) | quality is not reflected by these parameters…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.