K260910 · Glooko, Inc. · NDC · May 1, 2026 · Anesthesiology
Device Facts
Record ID
K260910
Device Name
EndoTool IV Cloud 1.0
Applicant
Glooko, Inc.
Product Code
NDC · Anesthesiology
Decision Date
May 1, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1890
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
EndoTool Cloud is intended for use by trained healthcare professionals in inpatient care settings to assist in the management of blood glucose levels in adult and pediatric patients aged two years and older who weigh at least 12 kg. The software is intended to recommend patient-specific intravenous insulin dosing, carbohydrate dosing for recovery of or supplementation, and subcutaneous insulin dosing for the transition from intravenous therapy and for ongoing management of hyperglycemia. EndoTool Cloud generates dosing recommendations based on the prescribing healthcare provider's orders for target glucose range, dosing mode and patient-specific clinical characteristics. The software is intended to assist healthcare professionals in adjusting and maintaining a patient's glucose level within a configurable, clinician-determined target range. The EndoTool Cloud logic is not a substitute for clinical judgment but is intended as a decision support tool for trained healthcare professionals. Final dose decisions must be made by the healthcare professional after consideration of the patient's overall clinical status. No medical decision should be made solely based on the recommendations provided by this software.
Device Story
Software system for inpatient glucose management; inputs include current/cumulative glucose values, clinical lab values, and provider-ordered target ranges; processes data via dosing logic to calculate/recommend IV insulin, carbohydrate doses, and subcutaneous basal transition doses; used in hospital settings by trained healthcare professionals (physicians, nurses); output displayed to clinicians to assist in maintaining glucose within target ranges; decision support tool only; final dosing decisions remain with clinician; cloud-native architecture hosted on AWS.
Clinical Evidence
No clinical data. Substantial equivalence supported by software verification and validation testing per IEC 62304, cybersecurity documentation, and comparative use-related risk analysis for UI changes.
Technological Characteristics
Software-based glucose management system; cloud-native architecture (AWS SaaS); interoperable with EMR; utilizes algorithmic dosing logic for insulin/carbohydrate recommendations. Complies with ANSI AAMI IEC 62304 (software lifecycle) and ISO 14971 (risk management).
Indications for Use
Indicated for inpatient management of blood glucose in adult and pediatric patients aged 2+ years weighing ≥12 kg. Contraindicated for use without clinical judgment; requires caution in patients with low/low-normal potassium levels (risk of life-threatening drop).
Regulatory Classification
Identification
A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
## I Background Information:
A 510(k) Number
K260910
B Applicant
Glooko, Inc.
C Proprietary and Established Names
EndoTool IV Cloud 1.0
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| NDC | II | 21 CFR 868.1890 - Predictive Pulmonary-Function Value Calculator | Clinical Chemistry |
## E Purpose for Submission:
Modification to the device to change dosing logic to include nutrition, algorithm changes, updates to the user interface, and integration with the electronic medical record
## II Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
EndoTool Cloud is intended for use by trained healthcare professionals in inpatient care settings to assist in the management of blood glucose levels in adult and pediatric patients aged two years and older who weigh at least 12 kg. The software is intended to recommend patient-specific intravenous insulin dosing, carbohydrate dosing for recovery of or supplementation, and subcutaneous insulin dosing for the transition from intravenous therapy and for ongoing management of hyperglycemia.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
EndoTool Cloud generates dosing recommendations based on the prescribing healthcare provider's orders for target glucose range, dosing mode and patient-specific clinical characteristics. The software is intended to assist healthcare professionals in adjusting and maintaining a patient's glucose level within a configurable, clinician-determined target range.
The EndoTool Cloud logic is not a substitute for clinical judgment but is intended as a decision support tool for trained healthcare professionals. Final dose decisions must be made by the healthcare professional after consideration of the patient's overall clinical status. No medical decision should be made solely based on the recommendations provided by this software.
## C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
When insulin therapy is initiated in patients with a low-to-low normal potassium level, they are at risk for a life-threatening drop in their circulating potassium level. Consult with the patient's Provider to determine if the potassium level is adequate or if potassium treatment should be initiated prior to insulin administration.
As a rule, potassium levels of less than 4.0 mEq/L should be evaluated for treatment, particularly if fluid replacement is occurring in conjunction with or prior to the initiation of insulin therapy.
## III Device Description
EndoTool Cloud 1.0 is a software system for glucose management which uses the current and cumulative glucose values provided by the user to calculate and recommend intravenous insulin or carbohydrate doses, to adjust and maintain the patient's glucose level within a provider-ordered target range. In addition, the application can recommend a subcutaneous Basal transition insulin dose when IV insulin therapy is no longer required.
The EndoTool Cloud 1.0 is a modification of the EndoTool IV 3.1 System cleared under K241088 with the following changes:
- Algorithmic changes (dosing calculations and targeting)
- Additional clinical lab values used to support treatment decisions
- UI enhancements
- Interoperability with Electronic Medical Health Records
- Platform architecture changes to support cloud-native architecture
## IV Substantial Equivalence Information:
A Predicate Device Name(s):
EndoTool IV 3.1
B Predicate 510(k) Number(s):
K241088
K260910 - Page 2 of 5
{2}
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K260910 | K241088 |
| --- | --- | --- |
| Device Trade Name | EndoTool Cloud 1.0 | EndoTool IV 3.1 |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Calculate recommended doses of insulin and carbohydrate, for intravenous and/or subcutaneous delivery, to adjust and maintain the blood glucose level within a configurable clinician-determined target range. | Same |
| Intended user group | 2 yr and 12 kg or greater | Same |
| Functionality | Creates patient-specific IV drug infusion profiles; evaluates current and cumulative glucose values; calculates next dose of insulin, glucose, or saline; calculates carbohydrate intake, nutritional bolus, and meal intake. | Same |
| General Device Characteristic Differences | | |
| Environment of Use | Amazon Web Services (AWS) hosting environment includes cloud-based application services and managed database services. These are hosted in a SaaS virtual network. | Windows hosting environment includes physical or virtual web application server and database server. These may be hosted on client premises or in a private virtual network. |
V Standards/Guidance Documents Referenced:
- ANSI AAMI IEC 62304:2006/A1:2016, Medical device software - Software life cycle processes [Including Amendment 1 (2016)]
K260910 - Page 3 of 5
{3}
- ISO 14971 Third Edition 2019-12, Medical devices - Application of risk management to medical devices
## VI Performance Characteristics:
### A. Non-Clinical Performance
#### Software Documentation
The level of software documentation for this software device is enhanced. Software documentation consistent with an enhanced level was provided as described in the FDA guidance for the content of premarket submissions for device software functions. Documentation included complete software requirements and design specification documentation, depiction of functional units and software modules, system diagrams and flowcharts, revision history of design changes for each release after the predicate device with date and version number, configuration management plan, traceability of verification and validation to test cases for requirements and improvements, and verification and validation of test cases with protocols, pass/fail criteria, and results.
Software testing was conducted per IEC 62304 and included the following:
- Verification and validation of test cases that are related to user need, requirement, and improvement.
- Static code analysis
#### Cybersecurity Documentation
Cybersecurity documentation as described in the FDA guidance for Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submission was provided. Detailed information on cybersecurity of the device was reviewed and found to be acceptable.
#### Usability/Human Factors:
Human factors validation testing for the EndoTool Glucose Management System was previously conducted and submitted to FDA in earlier clearances, including K132547, K200443, K201619, and K241088. These prior studies evaluated the ability of intended users, including healthcare professionals such as physicians, nurses, and clinical staff, to safely and effectively perform critical tasks associated with insulin dosing management using the EndoTool system in hospital care environments. For the subject device, an evaluation of the need for additional Human Factors/Usability (HF/U) validation testing was conducted in accordance with FDA guidance Applying Human Factors and Usability Engineering to Medical Devices (2016) and Content of Human Factors Information in Medical Device Marketing Submissions (2023). A comparative use-related risk analysis was performed to evaluate all UI changes introduced in EndoTool Cloud 1.0. The analysis confirmed the modifications do not affect critical tasks or use-related risk and therefore do not require additional human factors validation testing.
### B. Clinical Studies:
N/A. The subject device is an updated cloud-hosted implementation of the most recently cleared predicate device, EndoTool IV (K241088). The intended users, intended use, clinical
K260910 - Page 4 of 5
{4}
workflow, and core insulin dosing algorithm remain unchanged from the predicate device. Therefore, additional clinical testing was not necessary to demonstrate substantial equivalence.
## C. Other Supportive Device Performance Characteristics Data
None
## VII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
## VIII Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K260910 - Page 5 of 5