PowerCube+ Series (PowerCube Body+); PowerCube+ Series (PowerCube Diffusion+); PowerCube+ Series (PowerCube Body+ / Diffusion+)
Applicant
Ganshorn Medizin Electronic GmbH
Product Code
JEH · Anesthesiology
Decision Date
Dec 9, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1760
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The PowerCube+ Series is indicated for use in the measurement and data collection of lung function parameters. The system performs cooperation-dependent pulmonary function tests which include Spirometry/Flow-Volume measurement, Body Plethysmography measurement, Lung Diffusion measurement, Occlusive Resistance measurement and Respiratory Muscle Strength measurement. The device provides information that aids in a diagnosis by a clinician. The PowerCube+ Series is indicated for use by a clinician in a professional healthcare setting on adult and pediatric patients who are 5 years and older that can cooperate in the testing.
Device Story
PowerCube+ Series performs cooperation-dependent pulmonary function tests; inputs include patient-generated airflow, chamber pressure, and gas concentrations (He/CO). Spirometry uses ultrasound transit-time transducers; Body Plethysmography uses air-tight chamber pressure and mouth pressure (Boyle's Law); Lung Diffusion uses gas analyzers to measure concentration differences after 10-second breath-hold. Device is mains-powered; used in professional healthcare settings by clinicians. Output consists of physiological parameters (e.g., FVC, TGV, DLCO, sRaw, respiratory muscle strength) displayed via PC software. Clinicians use data to aid in diagnosis of pulmonary conditions. Benefits include non-invasive assessment of lung function and respiratory mechanics.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including comparative testing against the predicate device using a calibrated Hans-Rudolph flow/volume simulator. Compliance with ISO 23747, ISO 26782, and ATS/ERS standards for all measurement modalities (Spirometry, Body Plethysmography, Lung Diffusion, Occlusive Resistance, Respiratory Muscle Strength) was verified.
Technological Characteristics
Mains-powered (110-240 VAC) pulmonary function system. Spirometry: ultrasound transit-time flow measurement. Body Plethysmography: air-tight chamber with pressure sensors. Lung Diffusion: NDIR CO analyzer and ultrasound He analyzer. Occlusive Resistance/Respiratory Muscle Strength: shutter mechanism and mouth pressure sensors. Software-based data processing on PC. Biocompatibility per ISO 10993 and ISO 18562. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2.
Indications for Use
Indicated for adult and pediatric patients 5 years and older capable of cooperation in testing, for measurement and data collection of lung function parameters including spirometry, body plethysmography, lung diffusion, occlusive resistance, and respiratory muscle strength.
Regulatory Classification
Identification
A volume plethysmograph is an airtight box, in which a patient sits, that is used to determine the patient's lung volume changes.
Predicate Devices
MASTERSCREEN PFT, MASTERSCREEN PFT CT, MASTERSCREEN PFT BODY (K072061)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 9, 2024
GANSHORN Medizin Electronic GmbH % Jim Chickering Regulatory Consultant 15 Belleau Woods Georgetown, New Hampshire 01833
Re: K240706
Trade/Device Name: PowerCube+ Series (PowerCube Body+); PowerCube+ Series (PowerCube Diffusion+); PowerCube+ Series (PowerCube Body+ / Diffusion+) Regulation Number: 21 CFR 868.1760 Regulation Name: Volume Plethysmograph Regulatory Class: Class II Product Code: JEH, BZG Dated: November 4, 2024 Received: November 4, 2024
Dear Jim Chickering:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Rachana Visaria -S
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K240706
Device Name
PowerCube+ Series (PowerCube Body+) PowerCube+ Series (PowerCube Diffusion+) PowerCube+ Series (PowerCube Body+ / Diffusion+)
Indications for Use (Describe)
The PowerCube+ Series is indicated for use in the measurement and data collection of lung function parameters. The system performs cooperation-dependent pulmonary function tests which include Spirometry/Flow-Volume measurement, Body Plethysmography measurement, Lung Diffusion measurement, Occlusive Resistance measurement and Respiratory Muscle Strength measurement. The device provides information that aids in a diaician.
The PowerCube+ Series is indicated for use by a clinician in a professional healthcare setting on adult and pediatric patients who are 5 years and older that can cooperate in the testing.
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## K240706 510(k) Summary
### Contact Details
Date Prepared:
Applicant Name: Applicant Address: Applicant Contact Telephone: Applicant Contact: Applicant Contact Email:
Correspondent Name: Correspondent Address: Correspondent Contact Telephone: Correspondent Contact Email:
December 9, 2024
GANSHORN Medizin Electronic GmbH Industriestr 6-8, Niederlauer D-97618 Germany +49 (9771) 6222 Dr. Stefan Ponto Stefan.Ponto@ganshorn.de
Jim Chickering 15 Belleau Woods, Georgetown MA USA (978) 808-8196 jimchickering@gmail.com
#### Device Name
Device Trade Name: PowerCube+ Series (PowerCube Body+) PowerCube+ Series (PowerCube Diffusion+) PowerCube+ Series (PowerCube Body+ / Diffusion+) Common Name: Volume Plethysmograph Volume Plethysmograph Classification Name: Regulation Number: 868.1760, 868.1840 JEH, BZG Product Codes: Device Class: 2
Legally Marketed Predicate Device
| Predicate #: | K072061 |
|-----------------------------|----------------------------------------|
| Trade Name: | MASTERSCREEN PFT, MASTERSCREEN PFT CT, |
| | MASTERSCREEN PFT BODY |
| Common/Classification Name: | Volume plethysmograph |
| Regulation Number: | 868.1760 |
| Product Codes: | JEH |
| Device Class: | 2 |
{5}------------------------------------------------
### Legally Marketed Reference Device
| Reference #: | K160116 |
|-----------------------------|------------------------|
| Trade Name: | SpiroScout |
| Common/Classification Name: | Diagnostic spirometer. |
| Regulation Number: | 868.1840 |
| Product Codes: | BZG |
| Device Class: | 2 |
## Device Description Summary
The GANSHORN PowerCube+ Series is a device that performs cooperation-dependent pulmonary function tests, including Spirometry, Body Plethysmography, Lung Diffusion measurement, Occlusive Resistance measurement and Respiratory Muscle Strength measurement. The device provides information that aids in a diagnosis by a clinician.
Spirometry is a set of non-invasive pulmonary function tests where the flow of inhaled/exhaled air is measured to determine physiological parameters such as Peak Expiratory Flow and Forced Vital Capacity. The patient's breathing flow is measured with ultrasound technology inside a breathing insert. Two ultrasound transducers measure the difference in ultrasound wave transit time to calculate breathing flow direction, speed, and volume.
Body Plethysmography provides for the measurement of physiological parameters such as Functional Residual Capacity and Specific Airway Resistance. The patient is seated in an air-tight chamber which has a fixed shape and volume. Pressure sensors measure the chamber pressure and the pressure close to the mouth, which is used as a proxy for alveolar pressure when measured under a zero-flow condition. Boyle's Law is used to infer the volume in the lungs from changes in chamber pressure.
Lung Diffusion testing is a non-invasive process for measuring diffusion capacity and lung volume. The patient inhales a test gas with known concentrations of helium and carbon monoxide. The patient's breath is held for 10 seconds during which time the helium dilutes into the lungs and the carbon monoxide diffuses through the alveoli into the blood. After 10 seconds of breath-hold time, the patient exhales and the difference between inhaled and exhaled gas concentrations is measured with a gas analyzer. The differences in gas concentration are used to determine physiological parameters such as DLCO (diffusing capacity of the lungs for carbon monoxide) and Alveolar Volume.
Occlusive Resistance measurement is an established method for measuring airway resistance during tidal breathing, using a shutter and mouth pressure sensor.
Respiratory Muscle Strength measurement is an established method for measuring the maximal strength of respiratory muscles, using a shutter and mouth pressure sensor.
The PowerCube+ Series has the following product model configurations:
{6}------------------------------------------------
- . PowerCube Body+ includes Spirometry and Body Plethysmography measurement
- . PowerCube Diffusion+ includes Spirometry and Lung Diffusion measurement
- . PowerCube Body+ / Diffusion+ includes Spirometry, Body Plethysmography, and Lung Diffusion measurement
All product model configurations support Occlusive Resistance measurement and Respiratory Muscle Strength measurement.
The PowerCube+ Series is mains-powered and not intended for mobile/transportable use.
#### Intended Use/Indications for Use
The PowerCube+ Series is indicated for use in the measurement and data collection of lung function parameters. The system performs cooperation-dependent pulmonary function tests which include Spirometry/Flow-Volume measurement, Body Plethysmography measurement, Lung Diffusion measurement, Occlusive Resistance measurement and Respiratory Muscle Strength measurement. The device provides information that aids in a diagnosis by a clinician.
The PowerCube+ Series is indicated for use by a clinician in a professional healthcare setting on adult and pediatric patients who are 5 years and older that can cooperate in the testing.
{7}------------------------------------------------
## Technological Comparison
| Characteristic | Subject Device<br>Ganshorn PowerCube+ Series | Predicate Device<br>K072061 - Viasys MasterScreen | Discussion of Differences |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The PowerCube+ Series is indicated for<br>use in the measurement and data<br>collection of lung function parameters.<br>The system performs cooperation-<br>dependent pulmonary function tests<br>which include Spirometry/Flow- Volume<br>measurement, Body Plethysmography<br>measurement, and Lung Diffusion<br>measurement. The device provides<br>information that aids in a diagnosis by a<br>clinician.<br>The PowerCube+ Series is indicated for<br>use by a clinician in a professional<br>healthcare setting on adult and pediatric<br>patients who are 5 years and older that<br>can cooperate in the testing. | The Masterscreen PFT Body is intended<br>to be used for measurement and data<br>collection of lung function parameters.<br>The system performs cooperation-<br>dependent pulmonary function tests<br>which include Spirometry/Flow-<br>Volume/Resistance measurements, lung<br>diffusion measurements and body<br>plethysmography measurement. The<br>device provides data / information and<br>supports help for a diagnosis.<br>MasterScreen PFT CT (Clinical Trial<br>version) includes Spirometry/Flow-<br>Volume/Resistance measurements and<br>lung diffusion measurements with<br>individual access rights defined for<br>different user roles (e.g. Investigator,<br>doctor, study nurse, trainer and service<br>personnel).<br>MasterScreen PFT includes<br>Spirometry/Flow-Volume/Resistance and<br>lung diffusion measurements.<br>Measurements will be performed under<br>the direction of a physician in the clinic,<br>doctor's office or hospital. It can be<br>utilized for patients from 4 years on and<br>older as long as they can cooperate in the<br>performance.<br>The MS-PFT Body is powered from<br>100-240V / 50-60Hz wall outlets. No<br>energy is transferred to the patient. | Substantially equivalent.<br>Both devices offer Spirometry,<br>Body Plethysmography and<br>Lung Diffusion measurement<br>programs.<br>Both devices provide<br>information that aids in a<br>diagnosis by a clinician.<br>The predicate device offers a<br>special clinical trial version that<br>is outside of the Intended Use<br>scope. |
| Environment of use | Professional healthcare setting | Same | N/A |
{8}------------------------------------------------
| K240706 | |
|---------|--|
|---------|--|
| Characteristic | Subject Device<br>Ganshorn PowerCube+ Series | Predicate Device<br>K072061 - Viasys MasterScreen | Discussion of Differences |
|-----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prescription Use only | Yes | Same | N/A |
| Patient population | 5 years and older | 4 years and older | Substantially equivalent |
| Measurement programs | Spirometry (Slow Vital Capacity,<br>Forced Vital Capacity,<br>Maximum Voluntary Ventilation) Body Plethysmography Lung Diffusion (Single Breath<br>Diffusion method) Occlusive Resistance<br>measurement Respiratory Muscle Strength<br>measurement | Same | The predicate device offers<br>the same lung function<br>measurement programs as<br>the subject device. |
| Spirometry technology<br>(flow measurement) | Ultrasound<br>(transit-time between 2 ultrasound<br>sensors) | Pneumotachograph<br>(pressure drop across a resistive<br>element) | Different technologies, but<br>for same Intended Use.<br>Subject device uses the<br>same technology as the<br>SpiroScout reference<br>device (K160116). |
| Flow Parameters | Range: ±18 l/s Accuracy: ±2% or 50 ml/s,<br>whichever is greater Resolution: 10 ml/s | Similar | Conforms to ISO<br>23747:2015 |
| Spirometry technology<br>(volume measurement) | Integration of flow measurement<br>data | Same | N/A |
| Characteristic | Subject Device<br>Ganshorn PowerCube+ Series | Predicate Device<br>K072061 - Viasys MasterScreen | Discussion of Differences |
| Volume Parameters | Range: 0-20 litersAccuracy: ±2% or 50 ml, whichever is greaterResolution: 1 ml | Similar | Conforms to ISO<br>26782:2009 |
| Body Plethysmography<br>technology | An air-tight chamber of fixed shape<br>and volume.<br>A pressure sensor that measures<br>the pressure inside the chamber<br>relative to ambient pressure.<br><br>A pressure sensor that measures<br>mouth pressure, which is used as a<br>proxy for alveolar pressure when<br>measured under a zero-flow<br>condition.<br><br>The use of Boyle's Law to infer the<br>volume in the lungs from changes in<br>chamber pressure. | Similar | The subject device<br>incorporates an airtight<br>chamber, mouth pressure<br>sensor, and application of<br>Boyle's Law, which are<br>standard and well-<br>established principles for<br>body plethysmography. |
| Characteristic | Subject Device<br>Ganshorn PowerCube+ Series | Predicate Device<br>K072061 - Viasys MasterScreen | Discussion of Differences |
| Lung Diffusion (Single<br>Breath Diffusion method)<br>technology | A test gas that includes known<br>quantities of helium and carbon<br>monoxide (provided by the<br>customer, not part of the subject<br>device).<br>A gas analyzer that measures<br>helium concentration during<br>inhalation and exhalation.<br>A gas analyzer that measures<br>carbon monoxide concentration<br>during inhalation and exhalation.<br>A comparison of the inhaled and<br>exhaled gas concentrations to infer<br>lung diffusing capacity. | Similar | The subject device uses a<br>helium and carbon<br>monoxide test gas,<br>alongside gas analyzers, to<br>measure lung diffusion<br>capacity based on<br>established ATS/ERS<br>standards. |
| Gas mixture: | helium: 18% carbon monoxide: 0.25% oxygen: 17-21% nitrogen: balance | Similar | Helium and carbon<br>monoxide concentrations<br>are consistent with<br>ATS/ERS guidelines |
| Carbon monoxide gas<br>analyzer | Type: non-dispersive infrared absorption Range: 0-3000 ppm CO Accuracy: ±2.5% FSO | Similar | Sensor measurement range<br>in the subject device is<br>sufficient for Intended Use |
| Helium gas analyzer | Type: ultrasound Range: 0-20 Vol% He Accuracy: ±2.5% FSO | Similar | Sensor measurement range<br>in the subject device is<br>sufficient for Intended Use |
| Characteristic | Subject Device<br>Ganshorn PowerCube+ Series | Predicate Device<br>K072061 - Viasys MasterScreen | Discussion of Differences |
| Occlusive Resistance<br>Measurement technology | A pressure sensor that measures<br>mouth pressure, which is used as a<br>proxy for alveolar pressure when<br>measured under a zero-flow<br>condition.<br>A shutter mechanism that<br>momentarily blocks the airflow at<br>the mouth by transient occlusion. | Similar | The subject device's use of<br>pressure sensors and a<br>shutter mechanism for<br>transient occlusion aligns<br>with well-established<br>principles in pulmonary<br>diagnostics and does not<br>introduce new questions of<br>safety or effectiveness. |
| Respiratory Muscle<br>Strength Measurement<br>technology | A pressure sensor that measures<br>mouth pressure, which is used as a<br>proxy for alveolar pressure when<br>measured under a zero-flow<br>condition.<br>A shutter mechanism that<br>momentarily blocks the airflow at<br>the mouth by transient occlusion. | Similar | The subject device's use of<br>pressure sensors and a<br>shutter mechanism for<br>transient occlusion aligns<br>with well-established<br>principles in pulmonary<br>diagnostics and does not<br>introduce new questions of<br>safety or effectiveness. |
| Characteristic | Subject Device<br>Ganshorn PowerCube+ Series | Predicate Device<br>K072061 - Viasys MasterScreen | Discussion of Differences |
| Spirometry Slow Vital<br>Capacity reported<br>measurements | • VT<br>• ERV<br>• IRV<br>• VC max<br>• VC ex<br>• VC in<br>• VC<br>• IC<br>• BF | Similar | Mathematically derived<br>based on physiological<br>parameters. Although data<br>for the predicate device<br>(K072061 – Viasys<br>MasterScreen) is not<br>publicly available, the<br>derivation of these<br>measurements adheres to<br>standard spirometric<br>calculation principles,<br>ensuring equivalent clinical<br>relevance and accuracy. |
| Spirometry Forced Vital<br>Capacity reported<br>measurements | • FVC<br>• FIVC<br>• FEV1<br>• FEV6<br>• FEV1 / FVC<br>• FEV1 / VCmax<br>• FEV1 / FEV6<br>• FEF 25 (MEF 75)<br>• FEF 50 (MEF 50)<br>• FEF 75 (MEF 25)<br>• FEF 25-75<br>• PEF<br>• PIF<br>• VC max | Similar | Bench testing in<br>compliance with ATS/ERS<br>standards and ISO 26782<br>and ISO 23747. Testing<br>validated accurate<br>measurements of key<br>parameters, including<br>FEV1, FEV6, FVC, and<br>PEF. Other reported<br>parameters are<br>mathematically derived<br>from these core<br>measurements, |
| Characteristic | Subject Device<br>Ganshorn PowerCube+ Series | Predicate Device<br>K072061 - Viasys MasterScreen | Discussion of Differences |
| Spirometry Maximum<br>Voluntary Ventilation<br>reported measurements | • MVV | Similar | Mathematically derived<br>based on validated<br>spirometric measurements,<br>such as FEV1 and tidal<br>volume. |
| Body Plethysmography<br>reported measurements | • TGV<br>• FRCpleth<br>• ERV<br>• RV<br>• TLC<br>• VC<br>• IC<br>• sRaw eff<br>• sRaw tot<br>• sRaw mid<br>• sRaw peak<br>• sRaw 0.5<br>• Raw eff<br>• Raw tot<br>• Raw 0.5<br>• RV%TLC | Similar | SE demonstrated for the<br>Body Plethysmography<br>component by conducting<br>bench testing comparing<br>the PowerCube+ Series to<br>the FDA-cleared<br>MasterScreen Body device,<br>demonstrating that<br>measurements for Thoracic<br>Gas Volume (TGV) and<br>Specific Airway Resistance<br>(sRaw) fall within a 5%<br>deviation threshold,<br>consistent with ATS/ERS<br>guidelines and regulatory<br>standards. Additionally, the<br>device's piezo-resistive<br>sensors and chamber<br>design were validated to<br>provide accurate and<br>reliable measurements<br>under clinical conditions |
| Characteristic | Subject Device<br>Ganshorn PowerCube+ Series | Predicate Device<br>K072061 - Viasys MasterScreen | Discussion of Differences |
| Lung Diffusion (Single<br>Breath Diffusion method)<br>reported measurements | DLCO DLCOc VA KCO TLC FRC RV RV%TLC | Similar | Bench testing using<br>simulated inhalation and<br>exhalation with calibrated<br>gas mixtures, demonstrated<br>compliance with ATS/ERS<br>2017 standards for single-<br>breath carbon monoxide<br>uptake. The testing<br>confirmed that all<br>measurements, including<br>Diffusing Capacity of the<br>Lungs for Carbon Monoxide<br>(DLCO) and Alveolar<br>Volume (VA), fall within<br>acceptable tolerances,<br>validating the accuracy and<br>reliability of the device for<br>clinical use. |
| Characteristic | Subject Device<br>Ganshorn PowerCube+ Series | Predicate Device<br>K072061 - Viasys MasterScreen | Discussion of Differences |
| Occlusive Resistance<br>Measurement<br>reported measurements | ROcc GOcc | Similar | Bench testing that<br>demonstrated compliance<br>with a strict acceptance<br>criterion of ≤2% deviation.<br>Comparative testing with<br>the FDA-cleared<br>MasterScreen Body device<br>confirmed that<br>measurements of Rocc and<br>derived parameters such as<br>Occlusive Conductance<br>(Gocc) were consistent and<br>accurate, aligning with<br>ATS/ERS standards. |
| Respiratory Muscle<br>Strength Measurement<br>reported measurements | Plmax PEmax P0.1 P0.1max P0.1 / Plmax P0.1 / MV | Similar | Bench testing that adhered<br>to a strict acceptance<br>criterion of ≤2% deviation.<br>Comparative testing with<br>the FDA-cleared<br>MasterScreen Body device<br>confirmed the accuracy and<br>reliability of these<br>measurements,<br>demonstrating compliance<br>with ATS/ERS standards<br>and validating their clinical<br>utility. |
| Characteristic | Subject Device<br>Ganshorn PowerCube+ Series | Predicate Device<br>K072061 - Viasys MasterScreen | Discussion of Differences |
| Components | Spirometry flow meter Body Plethysmography chamber Breathing gas analyzers Patient-applied parts, used for the acquisition of breathing gases Software module, running on a PC | Similar | The components of the subject device are consistent with those typically found in volume plethysmograph devices. These include a spirometry flow meter, body plethysmography chamber, breathing gas analyzers, patient-applied parts, and a software module running on a PC. |
| Patient-applied parts | Spirometry breathing tube Spirometry bacterial filter Nose clip Gas analysis intermediate adapter (draws an air sample for Lung Diffusion measurement) | Similar | Subject device applied parts are similar to other cleared devices of this type. |
| Device energy type | 110-240 VAC 50/60 Hz | Same | N/A |
| Characteristic | Subject Device<br>Ganshorn PowerCube+ Series | Predicate Device<br>K072061 - Viasys MasterScreen | Discussion of Differences |
| Body Plethysmography<br>chamber characteristics | Security glass, aluminum frame Door lock: magnetic Standard-sized, XL-sized | Similar | Chamber is fixed in shape<br>and volume, which is<br>required for the<br>measurement process.<br><br>A specific volume (standard<br>vs XL) is not required for<br>the measurement process,<br>as long as the volume is<br>known and fixed in nature. |
| Software –<br>user interface | LFX software module,<br>running on a PC | Similar | Each device has its own<br>software program,<br>supporting the same<br>Intended Use |
| Software –<br>signal processing | LFX software module,<br>running on a PC | Similar | Each device has its own<br>software program,<br>supporting the same<br>Intended Use |
| EMR connectivity support | Yes | Similar | The Ganshorn<br>PowerCube+ Series<br>includes EMR connectivity<br>support, enabling seamless<br>integration with electronic<br>medical record systems to<br>streamline data<br>management and improve<br>clinical workflows. |
| Characteristic | Subject Device<br>Ganshorn PowerCube+ Series | Predicate Device<br>K072061 - Viasys MasterScreen | Discussion of Differences |
| Environmental Use<br>Conditions (Operating) | Temperature: +15 - +35°C Relative Humidity: 10-95% (non-condensing) Atmospheric Pressure: 700-1060 hPa | Similar | The Ganshorn<br>PowerCube+ Series, the<br>specified environmental use<br>conditions for both<br>operating and transport<br>include a wide range of<br>temperature, humidity, and<br>atmospheric pressure<br>values, ensuring reliable<br>performance in various<br>environment. |
| Environmental Use<br>Conditions (Transport) | Temperature: -20 - +50°C Relative Humidity: 10-95% (non-condensing) Atmospheric Pressure: 600-1060 hPa | Similar | The Ganshorn<br>PowerCube+ Series, the<br>specified environmental use<br>conditions for both<br>operating and transport<br>include a wide range of<br>temperature, humidity, and<br>atmospheric pressure<br>values, ensuring reliable<br>performance in various<br>environments. |
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
{13}------------------------------------------------
{14}------------------------------------------------
{15}------------------------------------------------
{16}------------------------------------------------
{17}------------------------------------------------
{18}------------------------------------------------
{19}------------------------------------------------
## Non-Clinical and Clinical Tests Summary
| Category | Testing Summary |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cleaning Validation | The subject device and the cleaning / disinfection instructions<br>for the subject device have been tested in accordance with<br>FDA Guidance and the following standards:<br>ISO 17664-1:2021 Processing of health care products —<br>Information to be provided by the medical device<br>manufacturer for the processing of medical devices - Part<br>1: Critical and semi-critical medical devices |
| Shelf-Life Testing | The subject device has a disposable bacterial filter patient-<br>applied part that has been tested for shelf-life stability. Test<br>results indicate that the subject device complies with its<br>predetermined specification. |
| Biocompatibility Testing | The subject device has direct surface contact with patient skin<br>for a limited (< 24 hours) duration and indirect contact with<br>patient gas pathway.<br>The patient-contacting materials in the subject device have<br>been tested for biocompatibility in accordance with the<br>following standards: ISO 10993-1:2020 Biological evaluation of medical devices<br>- Part 1: Evaluation and testing within a risk management<br>system ISO 18562-1 First edition - Biocompatibility evaluation of<br>breathing gas pathways in healthcare applications - Part 1:<br>Evaluation and testing within a risk management process ISO 18562-2 First edition - Biocompatibility evaluation of<br>breathing gas pathways in healthcare applications - Part 2:<br>Tests for emissions of particulate matter ISO 18562-3 First edition - Biocompatibility evaluation of<br>breathing gas pathways in healthcare applications - Part 3:<br>Tests for emissions of volatile organic compounds ISO 18562-4 First edition - Biocompatibility evaluation of<br>breathing gas pathways in healthcare applications - Part 4:<br>Tests for leachables in condensate…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.