Percent Oxygen Sensors

K213933 · Careox, LLC · CCL · Aug 19, 2022 · Anesthesiology

Device Facts

Record IDK213933
Device NamePercent Oxygen Sensors
ApplicantCareox, LLC
Product CodeCCL · Anesthesiology
Decision DateAug 19, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.1720
Device ClassClass 2

Indications for Use

Percent Oxygen Sensors are intended to replace the original oxygen-sensing component of an oxygen analyzer that measures oxygen concentration in breathing gas mixtures.

Device Story

Electrochemical galvanic oxygen sensors; function as replacement components for medical oxygen analyzers. Sensors generate electrical current proportional to oxygen concentration in breathing gas mixtures; output signal used by host equipment to measure and display oxygen levels. Used in hospital, pre-hospital, industrial, and home settings; operated by clinicians or patients depending on host device. Provides real-time oxygen concentration data to healthcare providers or users to monitor inspired oxygen levels; facilitates clinical decision-making regarding respiratory therapy and anesthesia gas management. Benefits include continued accurate monitoring of oxygen delivery.

Clinical Evidence

Bench testing only. Performance evaluated per ISO 80601-2-55, including accuracy, precision, specificity, sensitivity, linearity, range, and response time.

Technological Characteristics

Electrochemical galvanic oxygen sensors. Principle of operation based on Faraday's law of electrolysis; current generation proportional to oxygen volume and lead anode consumption. Variations in physical housing, signal connection, and output to match OEM host equipment. No software or connectivity features.

Indications for Use

Indicated for use as replacement oxygen-sensing components in oxygen analyzers measuring breathing gas concentrations. Patient population follows the equipment to which the sensor is attached; intended for hospital, institutional, pre-hospital, industrial, and home settings.

Regulatory Classification

Identification

An oxygen gas analyzer is a device intended to measure the concentration of oxygen in respiratory gases by techniques such as mass spectrometry, polarography, thermal conductivity, or gas chromatography. This generic type of device also includes paramagnetic analyzers.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that. August 19, 2022 CareOx, LLC % Paul Dryden Consultant ProMedic Consulting, LLC 131 Bav Point Dr NE Saint Petersburg, Florida 33704 Re: K213933 Trade/Device Name: Percent Oxygen Sensors Regulation Number: 21 CFR 868.1720 Regulation Name: Oxygen Gas Analyzer Regulatory Class: Class II Product Code: CCL Dated: July 20, 2022 Received: July 21, 2022 Dear Paul Dryden: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Todd Courtney Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K213933 Device Name Percent Oxygen Sensors Indications for Use (Describe) Percent Oxygen Sensors are intended to replace the original oxygen-sensing component of an oxygen analyzer that measures oxygen concentration in breathing gas mixtures. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instruction of infing data sources, gather and maintain the data needed and complete and review the collection, includion. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." ## PDF Page 14 of 70 Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. {3}------------------------------------------------ | Date: | August 16, 2022 | |----------------------------|---------------------------------------------------------------------| | Sponsor: | CAREOX®, LLC<br>103 Carnegie Center Ste. 300<br>Princeton, NJ 08540 | | | Tel - 609.524.2540 | | Sponsor Contact: | Young Kim – CEO | | Submission Correspondent: | Paul Dryden<br>ProMedic, LLC | | Proprietary or Trade Name: | Percent Oxygen Sensors | | Common/Usual Name: | Oxygen Sensors | | Regulation Number: | 21CFR 868.1720 | | Regulation Code: | Analyzer, Gas, Oxygen, Gaseous-Phase | | Product Code: | CCL | | Regulatory Class: | II | | Predicate Device: | K952736 Analytic Industries Percent Oxygen Sensors | #### Device Description: The subject device is a family of Medical Oxygen Sensors which may be used as industry replacement types with various medical inspired-oxygen measuring devices. The oxygen sensors are all electrochemical galvanic type devices. The difference between models is merely the physical shape of the external housing (the basic oxygen sensor is often time encapsulated in a secondary housing), signal connection, signal output and the response time. The family of oxygen sensors concept extends to OEM manufacturers of anesthesia and respiratory therapy equipment. Again, the difference between models is merely the physical shape of the external housing (the basic oxygen sensor is often time encapsulated in a secondary housing), signal connection, signal output and the response time in order for the OEM to exercise a degree of control over the recurring replacement of oxygen sensors. #### Indications for Use: Percent Oxygen Sensors are intended to replace the original oxygen-sensing components of an oxygen analyzer that measures oxygen concentration in breathing gases. | Patient Population: | The sensor follows the population of the equipment for which it is attached. | |---------------------|------------------------------------------------------------------------------| |---------------------|------------------------------------------------------------------------------| Environment of Use: Hospital/institutional, pre-hospital, industrial and home settings. {4}------------------------------------------------ #### Substantial Equivalence Discussion: The subject devices are manufactured by Analytical Industries Inc. and were cleared under K952736. CareOx, the sponsor of this submission is purchasing final, finished oxygen sensors from Analytical Industries Inc. with the private label of CareOx. Indications - Equivalent to the predicate Technology - The technology is identical Principal of Operation - The principal of operation is identical Operating specifications – Equivalent Environment of Use - Similar Patient Population – Similar Non-clinical Testing We have performed the applicable test from ISO 80601-2-55. - These tests includes: Accuracy Precision Specificity Sensitivity Linearity Range Response time #### Differences: There are no differences between the proposed device and the predicate device. {5}------------------------------------------------ ## Table of Comparison to Predicate | Attribute | Predicate<br>Analytic Industries<br>Percent Oxygen Sensors | Subject Device<br>CareOx | Equivalency to at least one or<br>both devices | |-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------| | K# | K952736 | TBD | Similar | | Product Code<br>Classification Name | Analyzer, Gas, Oxygen, Gaseous-Phase<br>CCL | Analyzer, Gas, Oxygen, Gaseous-Phase<br>CCL | Similar | | Indications for Use | Percent Oxygen Sensors are intended to replace the<br>original oxygen-sensing components of an oxygen<br>analyzer that measures oxygen concentration in<br>breathing gases. | Percent Oxygen Sensors are intended to replace the<br>original oxygen-sensing components of an oxygen<br>analyzer that measures oxygen concentration in<br>breathing gases. | Similar | | Environments of<br>use | Replacement sensor for equipment that provides<br>inspired oxygen concentrations | Replacement sensor for equipment that provides<br>inspired oxygen concentrations | Similar | | Principle of<br>operation | Electrochemical galvanic | Electrochemical galvanic | Similar | | Prescriptive | Yes | Yes | Similar | | Service Life | The life of sensor is based on the Faraday's law of<br>electrolysis, i.e., the current generated by the sensor<br>per unit volume of oxygen and the amount of the<br>lead anode in the sensor determines the life of the<br>sensor. Only 70% of theoretically calculated sensor<br>life is used to claim the useful life of the sensor. For<br>example, if theoretical life is calculated as 90<br>months, the claimed life of the sensor would be 63<br>months. | The life of sensor is based on the Faraday's law of<br>electrolysis, i.e., the current generated by the sensor<br>per unit volume of oxygen and the amount of the<br>lead anode in the sensor determines the life of the<br>sensor. Only 70% of theoretically calculated sensor<br>life is used to claim the useful life of the sensor. For<br>example, if theoretical life is calculated as 90<br>months, the claimed life of the sensor would be 63<br>months. | Similar | ### Substantial Equivalence Conclusion: The sponsor has demonstrated through performance testing, design and non-clinical testing that the proposed device and predicate have been found to substantially equivalent.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...