The UltraMaxO2 Oxygen Analyzer is a tool used to measure oxygen purity, flow and pressure at the outlet of an oxygen concentrator. The UltraMaxO2 Oxygen Analyzer is intended to be used in an environment where oxygen concentrators are being serviced or repaired. This includes Hospitals, Nursing Homes, Extended Care Facilities, Patient Homes, and Respiratory Device Service and Repair Centers.
Device Story
UltraMaxO2 is a handheld, battery-powered ultrasonic oxygen analyzer used by technicians to verify oxygen concentrator performance. Device inputs include gas samples from concentrator outlets; it measures transit time of ultrasonic pulses through a sample chamber to determine gas concentration based on molecular mass. Flow is calculated via differential transit time (against vs. with flow); pressure is measured via a separate sensor when the output port is blocked. Output is displayed on an LCD screen. Used in clinical and home settings for maintenance/repair; not for patient monitoring. Provides objective data on purity, flow, and pressure to ensure concentrator functionality, aiding in service decisions.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by non-clinical functional and performance testing comparing the device to legally marketed predicates.
Technological Characteristics
Ultrasonic oxygen sensing; ABS plastic enclosure; PVC tubing; nickel-plated brass fittings; pressure, temperature, and humidity sensors; LCD display. Powered by 2x AA alkaline batteries. Dimensions: 3.16" x 5.10" x 1.04".
Indications for Use
Indicated for use by service technicians or healthcare professionals in hospitals, nursing homes, extended care facilities, patient homes, and repair centers to verify performance (purity, flow, pressure) of oxygen concentrators during service or repair. Not for patient use or continuous monitoring of oxygen delivery.
Regulatory Classification
Identification
An oxygen gas analyzer is a device intended to measure the concentration of oxygen in respiratory gases by techniques such as mass spectrometry, polarography, thermal conductivity, or gas chromatography. This generic type of device also includes paramagnetic analyzers.
{0}------------------------------------------------
DEC 1 4 2011 510(k) Summary of Safety and Effectiveness This summary of safety and effectiveness is provided as part of this Pre-Market Notification in compliance with 21 CFR Part 807, Subpart E, section 807.92, as a means of providing sufficient detail to provide understanding of the basis for a determination of substantial equivalence. } } Submitter's name, address, telephone number, a contact person and the date the summary was prepared. Maxtec, LLC Submitter's name/address: 6526 South Cottonwood Street Salt Lake City, Utah 84107 (801) 266-5300 Phone: (801) 270-5590 Fax: Tammy Lavery, RAC Contact Name: Director of Regulatory and Quality Contact Title: Contact Address: Maxtec, LLC Salt Lake City, UT 84107 (801) 327-9870 Phone: (801) 270-5590 Fax: Date Summary prepared: 08/19/2011 2) Subject device information: Device Name: Ultrasonic Oxygen Gas Analyzer UltraMaxO2 Oxygen Analyzer Trade Name(s): Oxygen Gas Analyzer Common/Usual Name: Analyzer, Gas, Oxygen, Gaseous-phase Classification Names: Classification: II CCL Product Code: CFR Reference: 21 CFR 868.1720
3) The following predicate devices were used to establish substantial equivalence for the UltraMaxO2 Oxygen Analyzer:
Anesthesiology
- 510(k) # K072469: "DigiFLO Concentrator Analyzer" .
- 510(k) # K984295: "Check O2 Plus Oxygen Analyzer" .
- 510(k) # K983500: "Pro2 Check Oxygen Indicator" .
4) Description of the subject device:
Classification Panel:
ুন প্রায়
The UltraMaxO2 device is used for checking oxygen concentrator performance with the measure of oxygen purity, and flow at the outlet of an oxygen concentrator. Both the subject and predicate devices are capable of measuring oxygen concentration with an accuracy of at least +/- 3%. In addition, the subject and predicate devices are capable of measuring flow from 0 to 10 LPM with an accuracy of +/-
1 of 3
6526 South Cottonwood Street · Salt Lake City, Utah 84107 · www.maxtecinc.com · 866.4.maxtec
{1}------------------------------------------------
0.2 LPM. The subject and predicate devices are battery powered and use ultrasonic based oxygen sensing. These devices are not intended to be used by patients who are prescribed oxygen, nor are they intended to continuously monitor or confirm oxygen delivery to a patient. As per predicates listed above, the indication noted has been cleared by the FDA.
The UltraMaxQ2 Oxygen Analyzer functions by passing an ultrasonic pulse through the gas sample and measuring the amount of time required for the pulse to transit the sample chamber. The transit time is converted into a gas concentration via calibration data stored in the device. This can be done because the transit time varies according to the molecular mass of the gas in the chamber. Flow is determined using the difference between the ultrasonic pulse traveling against the flow and the ultrasonic pulse traveling with the flow. Pressure is measured using a separate pressure sensor that measures the pressure build-up in the device when the output port is blocked. The pressure is displayed in either kPa or PSI determined by a user operable switch in the battery compartment.
The materials of the UltraMaxO2 include ABS plastic for the enclosures and the ultrasonic oxygen sensor tube body, adhesive backed Polycarbonate labels, PVC tubing and nickel plated brass fittings for the flow path, electronic circuitry including transducers/receivers, pressure, temperature and humidity sensors and an LCD screen. Further detail concerning components is included as appropriate within the 510k submission.
5) Statement of indication for use:
The UltraMaxO2 Oxygen Analyzer is a tool used to measure oxygen purity, flow and pressure of an oxygen concentrator. The UltraMaxO2 Oxygen Analyzer is intended to be used in an environment where oxygen concentrators are being serviced or repaired. This includes Hospitals, Nursing Homes, Extended Care Facilities, Patient Homes, and Respiratory Device Service and Repair Centers.
6) Technological characteristics:
The technological characteristics of the design, materials, chemical composition, and energy source, etc. of the UltraMaxO2 Oxygen Analyzer are, for the most part, the same as the predicate devices identified in 3) above. Maxtec has not introduced any new technological characteristics for the UltraMaxO2 device.
- Non-clinical functional and performance tests for substantial equivalence testing and the results a. are noted in the attached table.
- No clinical studies were performed No clinical studies were performed for the UltraMaxO2 as the b. device represents a well known technology for a recognized indication as evidenced in Section 5.0 by comparison to the predicate devices currently cleared for sale in the US market.
- 7) The conclusion drawn from the non-clinical and clinical tests demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed predicate devices identified in 3) above.
{2}------------------------------------------------
| | UltraMax O2 Oxygen Analyzer<br>- Subject Device | DigiFlow<br>Concentrator<br>Analyzer-<br>K072469 | Check O2<br>Plus Oxygen<br>Analyzer -<br>K984295 | Pro2 Check<br>Oxygen<br>Indicator-<br>K983500 |
|--------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|--------------------------------------------------|---------------------------------------------------------------|
| Indications for Use | The UltraMaxO₂ Oxygen Analyzer is a tool<br>used to measure oxygen purity, flow and<br>pressure of an oxygen concentrator. The<br>UltraMaxO₂ Oxygen Analyzer is intended<br>to be used in an environment where oxygen<br>concentrators are being serviced or repaired.<br>This includes hospitals, nursing homes,<br>extended care facilities, patient homes, and<br>respiratory device service and repair<br>centers. | Same | Same | Same |
| Sensor | Ultrasonic | Same | Same | Same |
| Low Battery Alarm<br>or Indictor | Low Battery Indicator | Same | Same | Same |
| Measurement<br>Capability - O₂<br>Measurement<br>Range | 20.9 – 96% | 20.8 – 95.7% | 73 – 95.6% | 20.9 – 100% |
| Accuracy of<br>Oxygen<br>Concentration | $\pm$ 1.5% Full Scale<br>(at constant temp. & optimal flow) | $\pm$ 1.8% | $\pm$ 2%<br>Full Scale | $\pm$ 2%<br>Full Scale (at<br>constant temp.<br>and pressure) |
| Measurement<br>Capability - Flow<br>Range | 0 – 10 LPM | O₂ 0 – 20 LPM<br>Air 0 – 10 LPM | 0 – 6 LPM | 0 – 10 LPM |
| Accuracy of Flow<br>Measurement | $\pm$ 0.2 LPM | $\pm$ 0.2 LPM<br>Both O₂ and Air | $\pm$ 0.3 LPM<br>(5% of full<br>scale) | $\pm$ 0.3 LPM<br>(3% of full<br>scale) |
| Measurement<br>Capability -<br>Pressure Range | 0.5 – 50 PSI<br>3.4 – 344 kPa | 0 – 35 PSI<br>(as noted in 510(k)<br>and 0-30 noted in<br>attached labeling)<br>0 – 207 kPa | 0 – 10 PSI<br>0 – 68.95 kPa | 0 – 10 PSI |
| Accuracy of<br>Pressure | $\pm$ 0.5% | $\pm$ 0.5% | $\pm$ 2%<br>Full Scale | $\pm$ 1%<br>Full Scale |
| Power Source &<br>Requirements | 2 Batteries:<br>AA (Alkaline) – 2 x 1.5 V | 1 Battery:<br>Alkaline 9 V | 1 Battery:<br>Alkaline 9 V | 1 Battery:<br>Alkaline 9 V |
| Display and Low<br>Battery Indicator | LCD | Same | Same | Same |
| Dimensions | 3.16" x 5.10" x 1.04"<br>(80.3mm<br>x 129.5mm x 26.4mm) | 9" x 1.5" x 1" | 3.3" x 7.5" x<br>1.25" | 3.60" x 5.75" x<br>1.29" |
| Response Time | Less than or equal to 17 sec. | 0.1 sec | 1 min and 45<br>sec $\pm$ 5 sec | 10 sec. |
| Weight | 0.4 lbs (181 g) | 179 g | 10 oz. (295 g) | 9 oz. (255.15 g) |
| Operating<br>Temperature | 15 – 40 C°<br>(59 – 104 F°) | 10 – 40 C°<br>(From Original<br>Submission and<br>10 – 45 C° in<br>attached labeling.) | 15 – 35 C°<br>(59 – 95 F°) | 0 – 41 C°<br>(32 – 105 F°) |
1
Page 3of 3
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Tammy Lavery Director of Regulatory and Quality Maxtec. LLC 6526 South Cottonwood Street Salt Lake City, Utah 84107
DEC 1 4 2011
Re: K112402
Trade/Device Name: UltraMaxO2 Oxygen Analyzer Regulation Number: 21 CFR 868.1720 Regulation Name: Oxygen Gas Analyzer Regulatory Class: II Product Code: CCL Dated: December 9, 2011 Received: December 12, 2011
Dear Ms. Lavery:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Lavery
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Antron D.m
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## 1.3 Indications for Use Statement
510(k) Number (if known): K112402
Device Name: UltraMaxO2 Oxygen Analyzer
Indications for Use:
The Indications for use for the Maxtec Handheld Oxygen Analyzer and Accessories are as follows:
The UltraMaxO2 Oxygen Analyzer is a tool used to measure oxygen purity, flow and pressure at the outlet of an oxygen concentrator. The UltraMaxO2 Oxygen Analyzer is intended to be used in an environment where oxygen concentrators are being serviced or repaired. This includes Hospitals, Nursing Homes, Extended Care Facilities, Patient Homes, and Respiratory Device Service and Repair Centers.
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)
X Prescription Use _____________________________________________________________________________________________________________________________________________________________
AND/OR
Over-The-Counter Use
L. Schultheis
Division Sign-Offi Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.