The intended use of the microstream nasal cannula filterline sample line is to conduct a sample of the patient's breathing from the patient's nose to the gas measurement device for measuring the percentage of co2 In the patient's exhalation.
Device Story
Device is a gas sampling nasal cannula used to transport exhaled breath from a non-intubated patient to a capnograph for CO2 analysis. It consists of a nasal cannula, plastic tubing, and an integrated in-line hydrophobic filter. Operation involves a vacuum pump within the capnograph (approx. 30mbar) drawing breath samples through the line. The hydrophobic filter prevents moisture from entering the capnograph. Used in clinical settings by healthcare providers to monitor patient CO2 levels, aiding in respiratory assessment and clinical decision-making.
Indicated for CO2 measurement in non-intubated patients requiring respiratory monitoring.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
{0}------------------------------------------------
K980325
APR 1 6 1998
בסיד
- ഗ് Summary of safety and effectiveness information in this submission
- 5. 1. Product name
PROPRIETARY: Microstream nasal cannula Filterline COMMON: Gas sampling line for capnograph
- 5. 2. Establishment registration number Establishment registration number: 8044004
- 5. 3. Establishment Address:
Oridion Medical Products Ltd. Har Hotzvim Science Based Industrial Park POB 45025 91450 Jerusalem, Israel
- 5. 4. Device Listing Fda Form 2892:
A 733250
5.5. Product classification
The microstream nasal cannula filterline sample line has not been classified. It was included as an accessory in FDA submission K964239. We believe that it falls under 21CFR868 anesthesiology devices.
- 5. 6. Intended use:
The intended use of the microstream nasal cannula filterline sample line is to conduct a sample of the patient's breathing from the patient's nose to the gas measurement device for measuring the percentage of co2 In the patient's exhalation.
## 5.7. Device description
The common product name for this device is a gas sampling nasal cannula . The gas sampling nasal cannula is used with a capnograph (carbon dioxide analyzer 21CFR 868.1400). There is a nasal cannula at one end of the device for connecting to the patient's nose and a female Luer lock on the Other end for connecting to the capnograph. The design and construction of the nasal cannula is almost identical to the nasal oxygen cannula 21 CFR 868.5340. The main difference is that instead of flowing oxygen through the cannula to the patient we use a vacuum to draw a sample of the breathing from the patient.
The two connectors are joined by a plastic tube and an in line hydrophobic filter.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Oridion. The logo consists of a stylized globe-like shape made up of horizontal lines, positioned above the word "Oridion" in a serif typeface. The globe is not a perfect sphere, and the lines give it a textured appearance.
One end of the tube is connected to the source of the patient's breathing (exhalation) and the other end of the tube is connected to a capnograph. The capnograph has a pump that creates a vacuum of approximately 30mbar which draws a sample of the patient's Breathing (exhalation) through the sampling tube into the capnograph for analysis of the CO2 content of the patient's exhalation.
The anesthesiologist sometimes places a hydrophobic filter between the sample line and the capnograph to keep moisture from entering the capnograph. The microstream nasal cannula filterline has integrated an in line hydrophobic filter between the patient and the capnograph to reduce the amount of patient generated moisture that can enter the capnograph.
## 5. 8. Substantial equivalence
The microstream nasal cannula filterline is identical to the same device described as an accessory to the NPB-75/microcap capnograph/pulse oximeter in approved submittal K964239.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of three human profiles facing to the right, stacked on top of each other. The profiles are rendered in a simple, flowing line style.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 1 6 1998
Mr. Sanford Brown Oridion Medical Ltd. P.O. Box 45025 Jerusalem 91450 Israel
Re: K980325 Microstream Nasal Cannula Filterline Regulatory Class: II (two) Product Code: 73 CCK Dated: January 26, 1998 Received: January 28, 1998
Dear Mr. Brown:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
## Page 2 - Mr. Sanford Brown
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".
Sincerely yours,
sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Oridion. The logo consists of a stylized globe made up of horizontal lines, with the word "Oridion" written in a serif font below it. The globe is positioned above the word, creating a visual hierarchy.
בסייד JANUARY 26, 1998
3. INDICATIONS FOR USE
510(K) NUMBER (IF KNOWN): ____________________________________________________________________________________________________________________________________________________
DEVICE NAME: MICROSTREAM FILTERLINE NC
INDICATIONS FOR USE:
THE MICROSTREAM FILTERLINE NC DEVICE IS USED WHENEVER THE PHYSICIAN NEEDS TO MEASURE THE CO₂ IN A PATIENT'S BREATHING IN A NON INTUBATED PATIENT.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)
OR
PRESCRIPTION USE X (PER 21 CFR 801.109)
over-the-counter use
(OPTIONAL FORMAT 1-2-96)
. …
Mm. P
(Division Sign-Off) (Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number .
ORIDION MEDICAL LTD. • POB 45025 • JERUSALEM 91450 ISRAEL • VOICE: +972 2 589-911 • FAX:+972 2 582-8868