AIM is a bite block intended for use in patients 18 years and older who require supplemental oxygen and CO2 monitoring during procedures where the patient is expected to be minimally or moderately sedated. AIM is not indicated for use during procedures that are expected to require deep sedation.
Device Story
AIM is a single-use, non-sterile bite block with integrated oxygen delivery and expired gas sampling tubing. It is used in clinical settings (hospitals, clinics, physician offices, MRI, transport) for patients undergoing procedures requiring supplemental oxygen and capnography. The device is placed in the patient's mouth (secured with tape) to deliver oxygen to the oropharynx and sample exhaled CO2. It connects to standard external oxygen supplies and capnography monitors. The healthcare provider uses the capnography output to monitor the patient's respiratory status during sedation. By providing a stable airway access point and reliable gas monitoring, the device assists in maintaining patient oxygenation and ventilation safety during procedures.
Clinical Evidence
Bench testing only. No clinical data. Performance evaluated via simulated respiratory conditions (varying respiratory rates, tidal volumes, and oxygen flow rates) comparing FiO2, EtCO2, and CO2 waveforms against the predicate. Mechanical testing included bite force, dislodgement force, and axial separation force. Biocompatibility testing performed per ISO 10993 and ISO 18562 standards.
Technological Characteristics
Single-use, non-sterile bite block. Materials evaluated for biocompatibility per ISO 10993-5, -10, -23 and ISO 18562-2, -3. Mechanical design includes integrated O2 delivery and CO2 sampling lines. Connects to standard capnography monitors and oxygen sources. Complies with ISO 80369-2, ISO 594-2, and ISO 11607-1. No software or electronic components; passive gas sampling and delivery.
Indications for Use
Indicated for patients >= 18 years old requiring supplemental oxygen and CO2 monitoring during minimally or moderately sedated procedures. Not indicated for deep sedation procedures.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 16, 2024
Inventeur, LLC % Paul Dryden Consultant ProMedic Consulting LLC 131 Bay Point Dr NE Saint Petersburg, Florida 33704
Re: K240937
Trade/Device Name: AIM Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: Class II Product Code: CCK, MNK, CAT Dated: November 19, 2024 Received: November 19, 2024
### Dear Paul Dryden:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bradley Q. Quinn -S
Bradley Quinn Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
| Submission Number (if known) | K240937 | | |
|--------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Device Name | AIM | | |
| Indications for Use (Describe) | AIM is a bite block intended for use in patients 18 years and older who require supplemental oxygen and CO2 monitoring during procedures where the patient is expected to be minimally or moderately sedated. AIM is not indicated for use during procedures that are expected to require deep sedation. | | |
| Type of Use (Select one or both, as applicable) | <table><tr><td><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</td><td><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr></table> | <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) Summary Page 1 of 4 12-Dec-2024
| 510k Sponsor: | Inventeur, LLC 16192<br>Coastal Hwy<br>Lewes, DE 19958 |
|--------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact: | Tariq Chaudhry<br>Email: tacmd1@gmail.com<br>Phone: 813.843.4408 |
| Submission Correspondent: | Paul Dryden, ProMedic, LLC<br>Paul.Dryden@promedic.cc |
| Proprietary or Trade Name:<br>Common/Usual Name:<br>Classification Name: | AIM<br>Bite block with gas sampling and oxygen delivery<br>21 CFR §862.1345 - CCK - Analyzer, Gas, Carbon-<br>Dioxide, Gaseous- Phase<br>21 CFR §876.1500 - MNK – Endoscopic Bite Block<br>21 CFR §868.5340 - CAT - Cannula, Nasal, Oxygen |
| Predicate Device:<br>Common/Usual Name:<br>Classification Name: | Dualguard™ (K140473)<br>Endoscopic bite block and nasal oxygen cannula with<br>CO2 monitoring accessory<br>21 CFR §862.1345 - CCK - Analyzer, Gas, Carbon-<br>Dioxide, Gaseous- Phase |
### 1. Device Description
AIM is a single-use, non-sterile bite block with integrated oxygen (O2) delivery and expired gas sampling tubing for patients undergoing procedures where supplemental oxygen and expired gas sampling is required expired. When paired with an oxygen supply and a capnography monitor, AIM can be left in place after the procedure to deliver oxygen and monitor CO2 levels.
AIM consists of a bite block, an attached oxygen delivery line and an attached CO2 sampling line. It delivers oxygen and samples exhaled CO2 in the oropharynx.
{5}------------------------------------------------
## 510(k) Summary Page 2 of 4
12-Dec-2024
| | Subject Device: AIM | Predicate Device: DualGuard™<br>K140473 | Comparison |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | AIM is a bite block intended for use in<br>patients 18 years and older who require<br>supplemental oxygen and CO2 monitoring<br>during procedures where the patient is<br>expected to be minimally or moderately<br>sedated. AIM is not indicated for use during<br>procedures that are expected to require deep<br>sedation. | DualGuard™ is intended for use in Adult<br>patients who require supplemental oxygen<br>delivery and CO2 monitoring during<br>endoscopy procedures and recovery.<br>The complete device with bite block is to<br>be used during endoscopy procedures. The<br>DualGuard™ Bite Block is to be removed<br>after endoscopy, leaving the O2<br>delivering/CO2 sampling cannula in place<br>for the patient recovery period. | AIM is not indicated for use during<br>procedures that are expected to require deep<br>sedation.<br>AIM specifies patients greater than 18 years<br>old whereas the DualGuard specifies adult<br>patients.<br>These differences do not result in a new<br>intended use. |
| Environments of Use | Locations where procedures are performed<br>in which the patient requires supplemental<br>oxygen and monitoring of exhaled gases.<br>Hospitals, Sub-acute facilities, Clinics,<br>Physician offices, MRI, Patient Transport<br>(ambulances, etc.) | Locations where endoscopy procedures are<br>performed in which the patient requires<br>supplemental oxygen and monitoring of<br>exhaled gases. | Similar |
| Duration of Use | Single patient, disposable | Single patient, disposable | Same |
| Mode of Operation | AIM is a biteblock that includes<br>methods to deliver oxygen and<br>sample expired gases in the oral<br>cavity. | DualGuard delivers oxygen through a<br>nasal/oral cannula and samples CO2<br>nasally and orally. | Similar: Both devices are placed in the<br>mouth and serve as bite blocks that deliver<br>oxygen and monitor CO2 in the oral cavity.<br>AIM does not include an integrated nasal<br>cannula. |
| 12-Dec-2024 | | | |
| | Subject Device: AIM | Predicate Device: DualGuard™ K140473 | Comparison |
| Biocompatibility | Surface contact, skin and mucosal.<br>Externally communicating, tissue.<br>Limited use (<24hr) | Surface contact, skin and mucosal.<br>Externally communicating, tissue.<br>Limited use (<24hr) | Similar patient contact and duration of use |
| % Fraction of Inspired O2<br>(FiO2) | Comparative testing of simulated oxygen delivery was completed at several respiratory<br>rates, tidal volumes, and oxygen flow rates. | | AIM maintained FiO2 equivalent to the<br>predicate device in all simulated conditions. |
| % End Tidal CO2<br>(EtCO2) | Comparative testing was completed under two simulated conditions at multiple simulated<br>EtCO2 values. | | AIM indicated EtCO2 is equivalent to the<br>predicate device in all simulated conditions. |
| CO2 Waveforms | Comparative testing was completed under several simulated respiratory conditions,<br>oxygen flow rates, and simulated EtCO2 values. | | AIM captured CO2 waveforms as well as<br>or better than the predicated device in all<br>simulated conditions |
| Target population | >/= 18 y.o. | Adult | Similar |
| Connectable to<br>capnograph? | Yes | Yes | Same |
| Expired gas sampling<br>location | Oral - multiple ports | Nasal and Oral | Similar |
| Design | Bite block with attached oxygen delivery<br>tubing which delivers oxygen to the back of<br>the mouth. Multiple sampling ports which are<br>connected to a gas sampling line which is<br>attached a capnograph | Oval shape inserted into the mouth with a<br>central opening for insertion of instruments.<br>Includes sampling nasal cannula and oxygen<br>delivery tubing. It has an elastic head band | Both design include means for delivering<br>oxygen and sampling CO2 and method of<br>securement. |
| Location of use | Inserted on either the left or right side of the<br>mouth | Inserted in the center of the mouth | The differences do not raise a different<br>question of safety and effectiveness. |
| Mechanism for securement | AIM inserted into mouth and secured with<br>tape | Secured with an elastic band around the<br>head | Each has a means of securing from potential<br>dislodgement |
{6}------------------------------------------------
## 510(k) Summary Page 3 of 4
12-Dec-2024
{7}------------------------------------------------
### 510(k) Summary Page 4 of 4 12-Dec-2024
### 3. Substantial Equivalence
AIM is considered substantially equivalent to the predicate DualGuard device based on the following discussion.
AIM has a similar intended use as the predicate device. AIM specifies patients greater than 18 years old whereas the DualGuard specifies adult patients. This difference does not affect the risk profile of the device.
DualGuard device supplies oxygen and samples CO2 through the nose and mouth, while AIM delivers oxygen and samples CO2 for capnography in the oropharynx. During testing, AIM delivered oxygen and sampled CO2 for capnography as effectively as the DualGuard device over a broad range of oxygen flow rates, simulated end-tidal CO2 levels, tidal volumes, and respiratory rates. Both devices are used with standard oxygen sources and capnography machines.
Neither device is life-supporting or life-sustaining.
AIM and DualGuard have similar intended use and technological characteristics. The differences between the two devices do not raise additional questions of safety and effectiveness.
### 4. Summary of Testing
AIM has been evaluated according to the standards listed in the table below. EtCO2 and FiO2 tests were performed at a range of respiratory rates, tidal volumes, and oxygen flow settings. All AIM samples passed the non-clinical tests, and all samples passed the performance tests at least as well as the predicate device.
| Test | Standard / Pre-determined<br>Acceptance Criteria |
|---------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| Biocompatibility | ISO 10993-5:2009<br>ISO 10993-23:2021<br>ISO 10993-10:2021<br>ISO 18562-2:2017<br>ISO 18562-3:2017 |
| Performance Tests (FiO2, EtCO2)<br>including CO2 waveforms | Comparison to predicate |
| Mechanical Tests (Bite Force Finite<br>Element Analysis, Dislodgement Force,<br>Axial Separation Force) | Minimal deformation<br>Comparison to predicate<br>ISO 80369-2 |
| Accelerated aging | Comparison of pre and post-aging<br>ISO 80369-2<br>ISO 594-2<br>ISO 11607-1 |
| Ship Testing | ISTA 3A |
### 5. Conclusion
The testing and comparison to the predicate support a decision of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.