The CO2 Sidestream Module is designed to provide carbon dioxide monitoring system during anesthesia / recovery, in the intensive care unit (ICU), and in Respiratory care. The CO2 Sidestream Module is used for adult patients.
Device Story
CO2 Sidestream Module (Capno-S, Capno-S+) is a non-dispersive infrared (NDIR) gas analyzer; measures CO2 concentration in breathing gases via infrared energy absorption. Device uses multi-channel infrared detector (Capno-S) or single-beam dual-wavelength optics (Capno-S+); includes autozero system with three-way valve and atmospheric pressure compensation. Input: gas sample from patient breathing circuit. Output: EtCO2 (mmHg, %, kPa) and respiration rate (breaths/min) transmitted to host monitoring system. Device is a component intended for integration into host monitors; operated by clinical personnel in hospital settings. Provides real-time CO2 data to support clinical decision-making regarding patient ventilation and respiratory status.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by bench testing, including performance verification (EtCO2 accuracy, respiration rate accuracy), electromagnetic compatibility (IEC 60601-1-2), electrical safety (IEC 60601-1), and comprehensive biocompatibility testing (ISO 10993 and ISO 18562 series).
Technological Characteristics
NDIR gas analysis; multi-channel (Capno-S) or single-beam dual-wavelength (Capno-S+) optics; no moving parts. Materials: ABS plastic (upper/lower case). Connectivity: Module-to-host interface. Power: DC. Standards: ISO 80601-2-55, IEC 60601-1, IEC 60601-1-2, ISO 10993-1, ISO 18562. Non-sterile.
Indications for Use
Indicated for adult patients requiring non-invasive continuous monitoring of EtCO2, InsCO2, and respiration rate during anesthesia, recovery, ICU, or respiratory care.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Beijing Kingst Commercial & Trade Co.,Ltd. % Charlie Mack Principal Engineer Irc 2950 E Lindrick Drive Chandler, Arizona 85249
Re: K192488
Trade/Device Name: CO2 Sidestream Module, Capno-S, Capno-S+ Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: Class II Product Code: CCK Dated: September 3, 2019 Received: September 11, 2019
#### Dear Charlie Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K192488
Device Name
CO2 Sidestream Module, Capno-S, Capno-S+
Indications for Use (Describe)
The CO2 Sidestream Module is designed to provide carbon dioxide monitoring system during anesthesia / recovery, in the intensive care unit (ICU), and in Respiratory care. The CO2 Sidestream Module is used for adult patients.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows a logo with the word "lingst" in a stylized font. The word is in blue and is slanted upwards from left to right. Below the word "lingst" are three Chinese characters, also in blue. The logo appears to be for a company or organization.
## Beijing Kingst Commercial & Trade Co., Ltd.
510(k) Summary (21 CFR §807.92)
## Submitter Information:
| Submitter Name: | Beijing Kingst Commercial & Trade Co., Ltd. |
|----------------------|--------------------------------------------------------|
| Address: | 6th Floor of No.2-7, Beibinhe Road, Xibianmen, Xicheng |
| | Dist, Beijing, P.R.C. 100053 |
| Telephone: | +86-10-68158013 |
| Fax: | +86-10-68156857 |
| Contact Person: | Mr. Judong Gao |
| | General Manager |
| Email: | charliemack@irc-us.com |
| | |
| Date of Preparation: | May 21, 2020 |
{4}------------------------------------------------
## Subject Devices:
Trade/proprietary name: CO2 Sidestream Module Model Capno-S, Capno-S+ Common Name: CO2 Sidestream Module Regulation Number: 21CFR868.1400 Regulatory Class: Class II Classification Code: CCK
## Predicate Device:
| Trade Name: | Capnograph and Oximeter |
|--------------------|------------------------------------|
| 510(k) Reference: | K170820 |
| Common Name: | CO2 Sidestream Module and Oximeter |
| Regulation Number: | 21CFR868.1400 |
| Regulatory Class: | Class II |
| Manufacturer: | CMI Health Inc. |
| Trade Name: | LoFlo C5 CO2 sensor |
| 510(k) Reference: | K053174 |
| 510(k) Reference: | R053174 |
| Common Name: | CO2 Sidestream Module |
| Regulation Number: | 21CFR868.1400 |
| Regulatory Class: | Class II |
| Manufacturer: | Respironics Novametrix, LLC |
## Purpose of Submission
This is a new traditional 510(K) submission of CO2 Sidestream Module.
## Device Description
The CO2 Sidestream module is a non-dispersive infrared gas analyzer with an autozero adjustment system and gain control.
The operation of the CO2 Sidestream module is based on CO2 energy absorption rates. CO2 molecules absorb infrared light energy of specific wavelengths, with the amount of energy absorbed being directly related to the CO2 concentration. When an IR light beam is passed through a gas sample containing CO2, the electronic signal from the infrared sensor (which measures the remaining light energy), can be obtained. This signal is then compared to the energy of the IR source, and calibrated to reflect CO2
{5}------------------------------------------------
concentration in the sample accurately. Calibration is performed using the infrared sensor's response to a known concentration of CO2 stored in the Module's memory. As the sample gas passes through a three-way valve, with the change of temperature and time, the valve leading to the pure air will close for 3-4 seconds to adjust the zero point.
The circuit module retains the atmospheric absolute pressure sensors and control of the pressure sensor. Modules can measure atmospheric pressure, and atmospheric can compensate for the calculation for the concentrations of carbon dioxide, which improves the design accuracy.
The Module then determines CO2 concentration in the breathing gases by measuring the amount of light absorbed by these gases. EtCO2 displays a numerical value in millimeters of mercury (mmHg), percent (%), or kilopascals (kPa). Respiration rate is calculated by measuring the time interval between detected breaths.
## Indication for use
The CO2 Sidestream Module is designed to provide carbon dioxide monitoring to a host monitoring system during anesthesia/recovery, in the intensive care unit (ICU), and in Respiratory care.
The CO2 Sidestream Module is used for adult patients.
## Comparison with the predicate device:
Beijing Kingst Commercial & Trade Co., Ltd. believes that the CO2 Sidestream Module is substantially equivalent to the CMI Health Inc. Capnograph and Oximeter (K170820) and the Respironics Novametrix LoFlo C5 CO2 sensor (K053174). The submitted Beijing Kingst Commercial & Trade Co., Ltd. CO2 Sidestream Module, Models Capno-S, and Capno-S+ were components within the predicate CMI Health's Capnograph and Oximeter for that device's FDA clearance. The subject device is also compared to the Respironics Novametrix LoFlo C5 sensor. Beijing Kingst Commercial & Trade Co., Ltd. is submitting the subject device for individual clearance, so that the manufacturer may market the Module separately. The submitted device will be installed in a host monitor, where the host monitor manufacturer will determine fit, form, and function.
{6}------------------------------------------------
The differences noted between the submitted CO2 Sidestream Module and the predicate Capnograph and Oximeter and Respironics Novametrix LoFlo C5 are based on the use. The Capnograph and Oximeter is a complete device that measures CO2 gas and also measures oxygen saturation. The Beijing Kingst Commercial & Trade Co., Ltd. CO2 Sidestream Module only measures CO2 gas. The display, any alarms, power supply, and physical configuration differences are due to the use in the end product. The subject device is manufactured to be a component of a complete device and not intended to be a stand-alone device and relies on a host device to display information or graphics. Please refer to the following pages for specific difference details.
{7}------------------------------------------------
| Element of<br>comparison | Capno-S | Capno-S+ | Conclusion |
|-----------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Appearance | Image: Capno-S device | Image: Capno-S+ device | different |
| Indications for<br>Use | The CO2Sidestream Module is designed for<br>monitoring vital physiological signs of the patient.<br>It's intended to be used for non-invasive<br>continuous monitoring of EtCO2, InsCO2, and<br>Respiration Rate. The Module is indicated for<br>inpatients from newborn (neonate) to adult in a<br>hospital environment. It is intended to be used only<br>under the regular supervision of clinical personnel. | The CO2Sidestream Module is designed for<br>monitoring vital physiological signs of the patient.<br>It's intended to be used for non-invasive continuous<br>monitoring of EtCO2, InsCO2, and Respiration Rate.<br>The Module is indicated for inpatients from newborn<br>(neonate) to adult in a hospital environment. It is<br>intended to be used only under the regular<br>supervision of clinical personnel. | same |
| Intended<br>patient type | intubated patients and non-intubated patients | intubated patients and non-intubated patients | same |
| Operation<br>principles | Non-dispersive infrared gas analysis, multi-channel<br>infrared detector, no moving parts | Non-dispersive infrared (NDIR) single beam optics,<br>dual-wavelength, no moving parts | same |
| Units | mmHg, kPa or Vol% | mmHg, kPa or Vol% | same |
| EtCO2<br>Measurement<br>Range | 0-150 mmHg<br>0-19.7%<br>0-20 kPa | 0-150 mmHg<br>0-19.7%<br>0-20 kPa | same |
| EtCO2<br>Accuracy<br>(at 760<br>mmHg,<br>ambient<br>temperature<br>of 25°C) | 0~40 mmHg ±2 mmHg<br>41~70 mmHg ±5% of reading<br>71~100 mmHg ±8% of reading<br>101~150mmHg±10% of reading | 0~40 mmHg ±2 mmHg<br>41~70 mmHg ±5% of reading<br>71~100 mmHg ±8% of reading<br>101~150mmHg±10% of reading | same |
| Respiration<br>Rate<br>measure<br>range | 3 - 150 breaths/minute (RPM) | 3 - 150 breaths/minute (RPM) | same |
| Respiration<br>Rate<br>accuracy | ±1% of reading or ±1 breaths/min whichever is<br>greater | ±1% of reading or ±1 breaths/min whichever is<br>greater | same |
## Comparison Table of Capno-S and Capno-S+
{8}------------------------------------------------
# Comparison to Predicate Devices
| Element of comparison | Subject Device | Predicate Device | Reference Device | Discussion |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company | Beijing Kingst Commercial & Trade | Respironics Novametrix, LLC | CMI Health Inc. | N/A |
| | Co., Ltd. | | | |
| FDA510(K) Number | N/A | K053174 | K170820 | N/A |
| Device Name | CO2 Sidestream Module | LoFlo C5 CO2 sensor | Capnograph and Oximeter | N/A |
| Model Number | Capno-S, Capno-S+ | LoFlo C5 | Capno-H | N/A |
| CO2 Module | CapnoCore | LoFlo C5 CO2 sensor | CapnoCore | N/A |
| Indications for Use | The CO2 Sidestream Module is<br>designed to provide carbon dioxide<br>monitoring to a host monitoring<br>system during anesthesia/recovery<br>in the intensive care unit (ICU) and in<br>Respiratory care.<br>The CO2 Sidestream Module is used<br>for adult patients. | The intended use of the LoFlo C5 CO2<br>sensor is to provide carbon dioxide<br>monitoring to a host monitoring system<br>during anesthesia/recovery in the<br>intensive care unit (ICU) and<br>Emergency Medicine/Transport or<br>Respiratory care. | The Capnograph and Oximeter are<br>designed for monitoring the vital<br>physiological signs of the patient. It is<br>used for non-invasive continuous<br>monitoring of oxygen saturation (SpO2),<br>Pulse Rate, EtCO2, InsCO2, and<br>Respiration Rate.<br>The Capnograph and Oximeter are<br>adaptable to adult usage in a hospital<br>environment. It is intended to be used<br>only under regular supervision of<br>clinical personnel | Identical to the predicate device.<br><br>The indications for the use of the subject<br>device are the same as the predicate<br>device, LoFlo C5 CO2 sensor, and add<br>the limit to the adult patient population<br>only.<br><br>The subject CO2 Sidestream Module is<br>the same as the CO2 sidestream module<br>used in the Reference Device, and they<br>are the same manufacturer -Beijing<br>Kingst.<br><br>The Reference device has an additional<br>SpO2 module for additional function, but<br>the subject device has no SpO2 module<br>and function. |
| Power Supply | DC | DC | Battery or AC | Identical<br>All Comply with IEC requirement |
| The type of protection<br>against electric shock | Class II | Class II | Class I and internally powered per IEC<br>60601-1. | Identical<br>All Comply with IEC requirement |
| The degree of protection<br>against electric shock | Type BF | Type BF | Type BF | Identical |
| Display | No display | No display | LED and LCD display | Identical<br>Module's measurement data can be<br>displayed on the primary device. |
| Element of comparison | Subject Device | Predicate Device | Reference Device | Discussion |
| Intended patient<br>population | Adult | From newborn (neonate) to adult | The whole device claim for Adult<br>SpO2 Measurement available for Adult<br>CO2 Measurement available for Adult<br>and Neonatal | The target population for CO2<br>Measurement is smaller than the<br>predicate device. |
| CO2 measurement<br>method | Infrared absorption method | Infrared absorption method | Infrared absorption method | Infrared absorption method |
| CO2 measure mode | Sidestream | Sidestream | Sidestream | Sidestream |
| Measuring parameters | EtCO2 and Respiration Rate | EtCO2 and Respiration Rate | EtCO2 and Respiration Rate | EtCO2 and Respiration Rate |
| Units | mmHg, kPa or Vol% | mmHg, kPa or Vol% | mmHg, kPa or Vol% | mmHg, kPa or Vol% |
| EtCO2 measure range | 0-150 mmHg<br>0-19.7%<br>0-20 kPa | 0-150 mmHg<br>0-19.7%<br>0-20 kPa | 0-150 mmHg<br>0-19.7%<br>0-20 kPa | 0-150 mmHg<br>0-19.7%<br>0-20 kPa |
| CO2 Accuracy | 0~40 mmHg ±2 mmHg<br>41~70 mmHg ±5% of reading<br>71~100 mmHg ±8% of reading<br>101~150mmHg±10% of reading | 0~40 mmHg ±2 mmHg<br>41~70 mmHg ±5% of reading<br>71~100 mmHg ±8% of reading<br>101~150mmHg±10% of reading | 0~40 mmHg ±2 mmHg<br>41~70 mmHg ±5% of reading<br>71~100 mmHg ±8% of reading<br>101~150mmHg±10% of reading | 0~40 mmHg ±2 mmHg<br>41~70 mmHg ±5% of reading<br>71~100 mmHg ±8% of reading<br>101~150mmHg±10% of reading |
| Respiration Rate<br>measure range | 3 - 150 breaths/minute | 2 - 150 breaths per minute | 3 - 150 breaths/minute | 2 RPM of the respiration rate will be<br>abnormal. So, it doesn't matter whether<br>the minimum range is 2 or 3.<br>It is complying with ISO80601-2-55 |
| Respiration Rate<br>accuracy | ±1% of reading or ±1 breaths/min<br>whichever is greater | ±1 breath per minute | ±1% of reading or ±1 breaths/min<br>whichever is greater | ±1% of reading or ±1 breaths/min<br>whichever is greater |
| Flow Rate | 50±10 mL/min | 50±10 mL/min | 50–250 mL/min | 50–250 mL/min |
| NO CO2 DETECTED<br>Alarm Delay | N/A | N/A | 15~39s off | Compare with predicate device: Identical<br>Compare with the reference device: The<br>Module itself has no alarm function. |
| Alarm of EtCO2 | N/A | N/A | High and lower alarms.<br>The limits are adjustable. | Compare with predicate device: Identical<br>Compare with the reference device: The<br>Module itself has no alarm function. |
| Alarm of RR | N/A | N/A | High and lower alarms.<br>The limits are adjustable. | Compare with predicate device: Identical<br>Compare with the reference device: The<br>Module itself has no alarm function. |
| Operation principles | Non-dispersive infrared gas analysis,<br>multi-channel infrared detector, no<br>moving parts. | Non-dispersive infrared (NDIR) single<br>beam optics, dual-wavelength, no<br>moving parts | Non-dispersive infrared gas analysis,<br>multi-channel infrared detector, no<br>moving parts. | Non-dispersive infrared gas analysis,<br>multi-channel infrared detector, no<br>moving parts. |
| Materials | the upper case: ABS<br>the lower case: ABS | the upper case: ABS<br>the lower case: ABS | the upper case: ABS<br>battery cover: ABS<br>the lower case: ABS<br>SpO2 Module: K063641 | Identical.<br>All comply with ISO10993-1<br>Biocompatibility requirement. |
{9}------------------------------------------------
{10}------------------------------------------------
## Safety and Performance Data:
To establish substantial equivalence to the identified predicate devices, tests were completed as defined below to the subject devices, CO2 Sidestream Module. The results of the testing demonstrate that the device complies with the requirements of the applicable standard, and the device is substantially equivalent to the predicate device.
## Non-Clinical Study:
## Safety and EMC
- ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and -A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005. MODI
- IEC 60601-1-2: 2014 Medical devices part 1-2: General requirements for basic safety and essential performance – Collateral standards: electromagnetic compatibility - Test and requirements
## Performance Data:
- ISO 80601-2-55 Second edition 2018-02 Medical electrical equipment Part 2-55: -Particular requirements for the basic safety and essential performance of respiratory gas monitors
| Test | Test Method | Test Result | Test Report No. |
|-------------------------------------|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| EtCO2 | See test report | Range: 0-150mmHg<br>Accuracy:<br>0 - 40 mmHg: $\pm$ 2 mm Hg;<br>41- 70 mmHg: $\pm$ 5% of reading; 71 -<br>100 mmHg: $\pm$ 8% of reading; 101<br>-150 mmHg: $\pm$ 10% of reading | JJX-FDA-A/0-06-04 |
| Inspired Dioxide Carbon<br>(InsCO2) | See test report | | JJX-FDA-A/0-06-06 |
| Respiration Rate | See test report | Range: 3bpm~150bpm<br>Accuracy: $\pm$ 1% of reading or $\pm$ 1<br>breaths/min whichever is greater | JJX-FDA-A/0-06-07 |
{11}------------------------------------------------
## Biocompatibility
- ISO 10993-5: 2009 Biological evaluation of medical devices Part 5 Tests for In Vitro Cytotoxicity
- ISO 10993-10: 2010 Biological evaluation of medical Devices Part 10: Tests for - Irritation and Delayed-Type Hypersensitivity
- ISO 10993-11: 2017 Biological evaluation of medical Devices – Part 11: Tests for systemic toxicity
- ISO 18562-1 First edition 2017-03 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 1: Evaluation and testing within a risk management process
- ISO 18562-2 First edition 2017-03 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 2: Tests for emissions of particulate matter
- ISO 18562-3 First edition 2017-03 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 3: Tests for emissions of volatile organic compounds
- ISO 18562-4 First edition 2017-03 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 4: Tests for leachables in condensate
| Device | Category and Contact | Duration | Test Conducted |
|---------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Capno-S, Capno-S+ | Surface Device →Skin | A — Limited≤( 24<br>h) | In Vitro Cytotoxicity (ISO10993-5)<br>Irritation (ISO10993-10)<br>Delayed-Type Hypersensitivity (ISO10993-10) |
| Accessories<br>(nasal tubes for<br>adult and infant,<br>filters, L- and<br>T-Connectors,<br>sampling line,<br>etc.) | external<br>communicating<br>devices →Tissue<br>contact<br>(contacting<br>the<br>patient's<br>inhaled<br>and exhaled gases) | A — Limited≤( 24<br>h) | In Vitro Cytotoxicity (ISO10993-5)<br>Irritation (ISO10993-10)<br>Delayed-Type Hypersensitivity (ISO10993-10)<br>Acute systemic toxicity(ISO10993-11) Material<br>mediated pyrogenicity ISO10993-11) |
{12}------------------------------------------------
| Test | Test Method | Result | Test Report No. | Page 10 of |
|-------------------------------------------------------------------|---------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|------------|
| In Vitro<br>Cytotoxicity Test | ISO10993-5 MTT method<br>extracts MEM with 10%<br>FBS Extract | Viability%: 87.9%<br>No toxicity. | SDWH-M201602887-1 | |
| Skin Sensitization<br>Test | ISO10993-10<br>Guinea Pig<br>Maximization Test<br>0.9% Sodium Chloride<br>Injection Extract | Skin sensitization rate:<br>0% No evidence of<br>sensitization. | SDWH-M201602887-2 | skin |
| Skin Sensitization<br>Test | ISO10993-10<br>Guinea Pig<br>Maximization Test<br>Sesame Oil Extract | Skin sensitization rate:<br>0% No evidence<br>of sensitization. | SDWH-M201602887-3 | skin |
| Skin Irritation Test | ISO10993-10<br>0.9% Sodium Chloride<br>Injection Extract | Primary irritation index:<br>0 Negligible. | SDWH-M201602887-4 | |
| Skin Irritation Test | ISO10993-10<br>Sesame Oil Extract | Primary irritation index:<br>0 Negligible. | SDWH-M201602887-5 | |
| Test for emissions<br>of particulate | ISO18562-2 | Particulate Matter:<br>$3.2 \times 10^{-3}$ µg/m3(Particle size<br>≤2.5 μm)<br>$3.2 \times 10^{-3}$ µg/m3(Particle size<br>≤10 μm)<br>Qualified | SDWH-M201901183-2(E) | |
| Tests for emissions<br>of volatile<br>organic<br>compounds (VOCs) | ISO18562-3 | Each VOC < 360 µg/d<br>Qualified | SDWH-M201901183-1(E) | |
| Tests for<br>leachable<br>substances<br>in condensate | ISO18562-4 | Organic<br>impurities(acetophenone):<br><0.03 µg/d<br>Metal ions (Ba): 4.83 µg/d<br>Metal ions(others): <0.8 µg/<br>d Qualified | SDWH-M201901183-3(E) | |
| Acute Systemic<br>Toxicity | ISO 10993-11<br>Intravenous 0.9%<br>Sodium Chloride Extract | All animals appeared clinical<br>normal throughout the study.<br>Qualified | SDWH-M201901183-4(E) | |
| Acute Systemic<br>Toxicity | ISO 10993-11<br>Intraperitoneal Sesame<br>Oil Extract | All animals appeared clinical<br>normal throughout the study.<br>Qualified | SDWH-M201901183-5(E) | |
| Pyrogen Test | ISO 10993-11<br>0.9% Sodium Chloride<br>Injection Extract Rabbit | No rabbit shows an individual<br>rise in temperature of 0.5°C or<br>more. | SDWH-M201901183-6(E) | |
Clinical Study:
- No clinical studies were performed.
{13}------------------------------------------------
## Sterility Information:
- This device is not delivered sterile and is not sterile during operation. -
## Package and Shelf Life:
- The CO2 Sidestream module (Capno-S, Capno-S+) is not subject to shelf life, as the device does not contain any sterile or degradable elements.
## Conclusion:
The differences between subject device and predicate devices do not raise issues of safety and effectiveness based on the indication for use, technological characteristics, and performance testing. The subject device complies with the same applicable standards as the predicate device.
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, the subject device, CO2 Sidestream Module, models Capno-S and Capno-S+ is safe and effective and substantially equivalent to predicate devices as described herein.
END
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.