Nasal CO2 Sample Line, Oral Nasal CO2 Sampling Cannula, O2 Oral/Nasal Co2 Cannula, Sample Lines
Applicant
Salter Labs
Product Code
CCK · Anesthesiology
Decision Date
Mar 10, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1400
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Nasal CO2 Sample Line is intended to be used to sample exhaled gas. The O2 Delivery / CO2 Sampling Nasal Cannula is intended to be used where low flow oxygen is administered and exhaled gas is monitored. The Oral / Nasal CO2 Sampling Cannula is intended to be used where exhaled gas is monitored. The O2 Delivery with Oral / Nasal CO2 Sampling Cannula is intended to be used where low flow oxygen is administered and exhaled gas is monitored. The Divided O2 Delivery / CO2 Sampling Nasal Cannula is intended to be used where low flow oxygen is administered and exhaled gas is monitored. The Sample Lines are intended to be used where exhaled gas is monitored.
Device Story
Non-sterile, disposable, single-patient-use cannulas and sample lines; used for supplemental oxygen delivery and/or sampling of exhaled gases (CO2) for monitoring. Devices feature nasal inserts, some with divided channels for simultaneous O2 delivery and CO2 sampling; includes Salter Eyes® (aperture to minimize occlusions) and Oral-Trac® (oral sampling feature). Operated by clinicians in hospital, sub-acute, or pre-hospital settings. Output is gas sample delivered to external CO2 analyzer; enables real-time monitoring of patient respiratory status; assists clinical decision-making regarding oxygenation and ventilation.
Clinical Evidence
No clinical data. Bench testing only. Biocompatibility testing (irritation, sensitization, cytotoxicity, particulate, volatile organic compounds) passed. Performance testing verified back pressure (<2 psi) and bond strength (>2x minimum requirement) met specifications after simulated 3-year aging.
Technological Characteristics
Materials: PVC with DINCH, DOTP, and DINCH plasticizers. Form factor: Flexible extruded plastic tubing and nasal cannulas. Connectivity: Passive interface to external gas analyzers. Sterilization: Non-sterile. Design: Includes Salter Eyes® aperture and Oral-Trac® sampling features.
Indications for Use
Indicated for patients requiring supplemental oxygen and/or expired gas monitoring, ranging from pediatric to adult populations, in hospital, sub-acute, and pre-hospital settings.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure embracing a bird, which is a symbol of care and protection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 10, 2016
Salter Labs Margaret Caler Director, Regulatory Affairs 2365 Camino Vida Roble Carlsbad, California 92011
Re: K151421
Trade/Device Name: Nasal CO2 Sample Line; 02 Delivery / CO2 Sampling Nasal Cannula; Oral / Nasal CO2 Sampling Cannula; O2 Delivery with Oral / Nasal CO2 Sampling Cannula; Divided O2 Delivery / CO2 Sampling Nasal Cannula and Sample Lines Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: Class II Product Code: CCK Dated: March 1. 2016 Received: March 3, 2016
Dear Margaret Caler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{2}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
# 510(k) Number (if known) K151421
## Device Name
Nasal CO2 Sample Line; O2 Delivery / CO2 Sampling Nasal Cannula; Oral / Nasal CO2 Sampling Cannula; O2 Delivery with Oral / Nasal CO2 Sampling Cannula; Divided O2 Delivery / CO2 Sampling Nasal Cannula and Sample Lines
## Indications for Use (Describe)
The Nasal CO2 Sample Line is intended to be used to sample exhaled gas.
The O2 Delivery / CO2 Sampling Nasal Cannula is intended to be used where low flow oxygen is administered and exhaled gas is monitored.
The Oral / Nasal CO2 Sampling Cannula is intended to be used where exhaled gas is monitored.
The O2 Delivery with Oral / Nasal CO2 Sampling Cannula is intended to be used where low flow oxygen is administered and exhaled gas is monitored.
The Divided O2 Delivery / CO2 Sampling Nasal Cannula is intended to be used where low flow oxygen is administered and exhaled gas is monitored.
The Sample Lines are intended to be used where exhaled gas is monitored.
Environment of use - hospital, sub-acute, and pre-hospital settings.
Patient population - Patients requiring supplemental oxygen and / or expired gas monitoring, adult to pediatrics.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size:10pt"> <span style="font-family:Wingdings">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|--------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:10pt"> <span style="font-family:Wingdings">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
In accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title 21 of the Code of Federal Regulations Part 807 (21 CFR §807), and in particular §807.92, the following summary of safety and effectiveness information is provided:
# 510 (k) Summary
### A. Submitter:
Salter Labs 2365 Camino Vida Robles Carlsbad, CA 92011 Telephone: 760-795-7100 Fax: 760-683-6797
## B. Contact Person:
Mara Caler Director, Regulatory Affairs
## C. Date Prepared:
09 March 2016
#### Device Names: D.
Tra
| Trade Name: | Nasal CO2 Sample Line; O2 Delivery / CO2 Sampling<br>Nasal Cannula; Oral / Nasal CO2 Sampling Cannula; O2<br>Delivery with Oral / Nasal CO2 Sampling Cannula;<br>Divided O2 Delivery / CO2 Sampling Nasal Cannula and<br>Sample Lines |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Name: | Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase |
| Product Code: | CCK |
| Regulation Number: | 868.1400 |
| Classification: | II |
| Classification Panel: | Anesthesiology (73) |
| Trade Name | Common Name |
|----------------------------------------------------|----------------------------------------------|
| Nasal CO2 Sample Line | Nasal CO2 Sample Line |
| O2 Delivery / CO2 Sampling Nasal Cannula | Oxygen delivery / CO2 Sampling Nasal Cannula |
| Oral / Nasal CO2 Sampling Cannula | Oral / Nasal CO2 Sampling Cannula |
| O2 Delivery with Oral / Nasal CO2 Sampling Cannula | Oxygen Delivery-Oral /Nasal CO2 Sampling |
| Divided O2 Delivery / CO2 Sampling Nasal Cannula | Infusion / Aspiration Cannula |
| Sample Lines | Sample Lines |
{5}------------------------------------------------
# Predicate Devices:
This submission demonstrates substantial equivalence to the predicate devices: K863703, Nasal CO2 Sample Line K863883, O2 Delivery / CO2 Sampling Nasal Cannula K864199, Oral / Nasal CO2 Sampling Cannula K864902, O2 Delivery with Oral / Nasal CO2 Sampling Cannula K892406, Divided O2 Delivery / CO2 Sampling Nasal Cannula K894350, Sample Lines The proposed devices that correspond to these predicates have identical names and are listed in the table below.
### Device Descriptions E.
| Device Name | Device Description |
|-----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Nasal CO2 Sample<br>Line | The device is a non-sterile, disposable, single patient use<br>cannula that allows sampling of patients exhaled gases. |
| O2 Delivery / CO2<br>Sampling Nasal<br>Cannula | The device is a non-sterile, disposable, single patient use cannula<br>that provides supplemental oxygen while sampling patients<br>exhaled gases. |
| Oral / Nasal CO2<br>Sampling Cannula | The device is a non-sterile, disposable, single patient use<br>cannula that allows sampling of patients exhaled gases from<br>both the nares and orally. |
| O2 Delivery with<br>Oral / Nasal CO2<br>Sampling Cannula | The device is a non-sterile, disposable, single patient use<br>cannula that provides supplemental oxygen while sampling<br>patients exhaled gases, using two nasal inserts. One insert<br>samples the CO2 while the other delivers the oxygen. The<br>Cannula includes Salter Eyes®, an aperature in the nasal prong<br>intended to minimize occlusions and Oral-Trac, a method of<br>sampling orally exhaled gases. The ETCO2 Cannula with<br>oxygen delivery can also be purchased with the Oral-Trac®<br>feature to allow oral (mouth) ETCO2 sampling. |
| Divided O2<br>Delivery / CO2<br>Sampling Nasal<br>Cannula | The Salter Labs ETCO2 Cannula provides O2 delivery in one<br>nostril and samples exhaled CO2 from the other nostril using<br>a divided nasal insert. |
| Sample Lines | The Salter Labs ETCO2 Tubing consists of flexible extruded<br>plastic tubes with connectors on each end, with differing<br>configurations. |
{6}------------------------------------------------
#### F. Indications for Use
The Intended/Indications for Use Statement is described below:
# Intended / Indications for Use
The Nasal CO2 Sample Line is intended to be used to sample exhaled gas.
The On Delivery / CO2 Sampling Nasal Cannula is intended to be used where low flow oxygen is administered and exhaled gas is monitored.
The Oral / Nasal CO2 Sampling Cannula is intended to be used where exhaled gas is monitored.
The On Delivery with Oral / Nasal CO2 Sampling Cannula is intended to be used where low flow oxygen is administered and exhaled gas is monitored.
The Divided O2 Delivery / CO2 Sampling Nasal Cannula is intended to be used where low flow oxygen is administered and exhaled gas is monitored.
The Sample Lines are intended to be used where exhaled gas is monitored.
Environment of use - hospital, sub-acute, and pre-hospital settings
Patient population - Patients requiring supplemental oxygen and / or expired gas monitoring, adult to pediatrics.
#### Comparison of Technological Characteristics with the Predicate Device G.
The proposed Modified ETCO2 Devices are substantially equivalent to the predicate devices listed above in that the indications for use, the intended use, device dimensions, device specifications and fundamental scientific technology remain unchanged.
The differences between the Modified ETCO2 Devices and predicate devices are:
| Features | Predicate ETCO2 Devices | Modified ETCO2 Devices | Performance Testing |
|----------------------|-------------------------|------------------------|-----------------------------------------------------------|
| Material Formulation | PVC, DIDP | PVC, DINCH | Biocompatibility |
| Material Formulation | PVC, DEHP | PVC, DOTP & DINCH | Biocompatibility |
| Shelf life | None | 3 years | Performance<br>testing after<br>simulated 3 year<br>aging |
### H. Performance Data
The proposed device was tested to verify that the new material did not affect the bond and performance characteristics of flow rate, back pressure. The test results demonstrate that the device is substantially equivalent to the predicate devices
{7}------------------------------------------------
# Performance Testing Summary
| Criteria | Predicate<br>specifications | Modified devices | Comments |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| Back pressure<br>(flow rates): | Shall not have a back<br>pressure that exceeds<br>3 psi at a maximum<br>flow rate in ambient<br>of 5°C, 20°C, and<br>40°C. | Maximum back<br>pressure was found<br>to be less than 2 psi. | The modified<br>devices passed all<br>specifications |
| Bond Strength<br>(tensile<br>strength): | The bonded<br>components of the set<br>will have a bond<br>strength that is ≥ 2<br>lbs. when pulled at a<br>rate of 5 inches per<br>minute. | The bond strength<br>test achieved over 2<br>times the minimum<br>allowable value. | The modified<br>devices passed<br>all specifications |
| Device<br>dimensions: | Unchanged. | Unchanged. | The modified<br>devices use the<br>same molds and are<br>unchanged from the<br>predicate devices |
The modified ETCO2 Devices meet established Salter Labs performance specifications.
## Clinical / Non-Clinical I.
No clinical testing was required for this submission.
The following biocompatibility testing was performed. The materials passed all parameters:
- Irritation .
- Sensitization ●
- Cytotoxicity .
- Particulate ●
- Volatile organic compounds ●
### J. Conclusions
ETCO2 device data and test results demonstrate that the device is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.