The Capnomask is intended to deliver supplemental oxygen to patients and monitor breathing by sampling exhaled carbon dioxide. Environment of use - hospital and sub-acute settings. Patient population - Non-intubated patients who are breathing spontaneously.
Device Story
Capnomask is a single-use, disposable oxygen mask with integrated gas sampling capabilities. Device features a face mask, elastic head strap, air entrainment holes, and a split connector with ports for standard oxygen tubing and a luer fitting for a gas sampling line. The sampling port extends into the mask to collect exhaled gases. Used in hospital and sub-acute settings by clinicians to deliver supplemental oxygen while simultaneously monitoring breathing via an external gas monitor. The device provides a qualitative assessment of breathing presence and supplemental oxygen delivery. It does not provide an airtight seal; air leakage occurs around the mask and through entrainment ports. Benefits include simultaneous oxygen therapy and respiratory monitoring for non-intubated patients.
Clinical Evidence
Bench testing only. Performance testing evaluated CO2 sampling and waveform performance at various breathing rates, tidal volumes, and oxygen flow rates (1, 5, and 10 lpm) with CO2 concentrations of 1% and 5%. Additional testing included real-time aging, pressure drop, resistance, environmental/mechanical testing, and ISO 10993 biocompatibility (cytotoxicity, sensitization, intracutaneous irritation, total VOC).
Technological Characteristics
Single-use face mask with elastic head strap. Materials meet ISO 10993 standards for surface and externally communicating mucosal contact. Features air entrainment ports and a split connector for oxygen tubing and luer-lock gas sampling. Dimensions: Large (121x85x72mm, 160cc) and Small (94x74x59mm, 72cc). Passive device; no energy source or software.
Indications for Use
Indicated for non-intubated, spontaneously breathing patients in hospital or sub-acute settings requiring supplemental oxygen and CO2 monitoring.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
{0}------------------------------------------------
NOV 2 5 2013
510(k) Summary Page 1 of 4 20-Nov-2013
#### Date Prepared:
Mediplus Ltd. Unit 7. The Gateway Centre Coronation Road, Cressex Business Park High Wycombe, Bucks, HP12 3SU UK
Tel - +44 (0) 1494 551299 Fax - +44 (0) 1494 536333
| Official Contact: | Tim Ward |
|----------------------------|------------------------------------------------------------|
| | Operations Manager |
| Proprietary or Trade Name: | Capnomask |
| Common/Usual Name: | Oxygen mask with gas sampling |
| Classification Name: | analyzer, gas, carbon-dioxide, gaseous phase (accessories) |
| | CCK - CFR 868.1400 |
| | Class II |
| Predicate Devices: | K971229 - Capnoxygen LLC, Capnoxygen mask |
#### Device Description:
The proposed Capnomask design is like a standard oxygeu mask with a port for connecting a gas sampling line which is then to be comected to a gas monitor to measure exhaled carbon dioxide.
The mask has a connector which incorporates ports for connecting standard oxygen tubing and a luer fitting to counect a standard gas sampling line.
In addition, it has air entrainment holes to supply room air if needed.
It is available in sizes ranging from large to snuall, similar to the predicate device.
#### Indications for Use:
The Capnomask is intended to deliver supplemental oxygen to patients and monitor breathing by sampling exhaled carbon dioxide.
Environment of use - hospital and sub-acute settings.
Patient population - Non-intubated patients who are breathing spontaneously.
{1}------------------------------------------------
**510(k) Summary**
510(k) Summary
20-Nov-2013
| 20-Nov-2013 | | |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Attribute | Predicate<br>Capnoxygen LLC<br>Capnoxygen Mask K971229 | Proposed<br>Capnomask |
| Indications for Use | The Capnoxygen mask is a medium<br>concentration single -use mask intended to<br>be used for the delivery of supplemental<br>oxygen and monitoring breathing by<br>sampling exhaled carbon dioxide. Standard<br>connectors for the oxygen tubing and a<br>standard female luer connector for the gas<br>sample line are provided. | The Capnomask is intended to deliver<br>supplemental oxygen to patients and monitor<br>breathing by sampling exhaled carbon<br>dioxide. |
| Environments of use | Hospitals and sub-acute setting | Hospitals and sub-acute setting |
| Prescriptive | Yes | Yes |
| Patient population | Non-intubated patients who are breathing<br>spontaneously | Non-intubated patients who are breathing<br>spontaneously |
| Single patient use,<br>disposable | Yes | Yes |
| Basic components | Face mask<br>Elastic head strap<br>Connector for oxygen tubing and gas<br>sampling line<br>Entrainment ports | Face mask<br>Elastic head strap<br>Connector for oxygen tubing and gas<br>sampling line<br>Entrainment ports |
| Sizes | Large and Small | Large and Small |
| Overall dimensions | Large -<br>Length - 140 mm<br>Width - 89 mm (at widest point)<br>Height - 67 mm<br>Small -<br>Length - 95 mm<br>Width - 76 mm (at widest point)<br>Height - 59 mm | Large -<br>Length - 121 mm<br>Width - 85 mm (at widest point)<br>Height - 72 mm<br>Small -<br>Length - 94 mm<br>Width - 74 mm (at widest point)<br>Height - 59 mm |
| Internal Volume | Large - 155 cc<br>Small - 65 cc | Large - 160 cc<br>Small - 72 cc |
| Entrainment holes | Yes | Yes |
| Typical Flow Rate<br>Range | < 10 lpm however this is set by the clinician | < 10 lpm however this is set by the clinician |
| Method of oxygen<br>delivery and Gas<br>sampling | Split connector with barbed oxygen tubing<br>fitting on top of Luer fitting for attaching a<br>gas sampling line.<br>The gas sampling port extends into the mask | Split connector with barbed oxygen tubing<br>fitting on top of Luer fitting for attaching a<br>gas sampling line.<br>The gas sampling port extends into the mask |
| Method of holding<br>mask on patient | Elastic band | Elastic band |
| Performance<br>Testing | Comparative End-tidal CO₂ results and<br>waveform at various settings of oxygen flow<br>rate, tidal volume and breathes per minute | Comparative End-tidal CO₂ results and<br>waveform at various settings of oxygen flow<br>rate, tidal volume and breathes per minute<br>Environmental and Mechanical testing<br>Real-time aging<br>Internal Volune |
{2}------------------------------------------------
#### 510(k) Summary Page 3 of 4 20-Nov-2013
### Substantial Equivalence Discussion
The above table compares the key features of the vroposed Capnomask with the identified predicate and how the proposed device can be found to be substantially equivalent.
In summary one can conclude that substantial equivalence is met based upon the following:
### Indications for Use -
The indications for use are identical for the proposed device when commared to the predicate -K971229 - Capnoxygen LLC Capnoxygen mask.
Discussion - Each device is indicated for use delivering supplemental oxygen and sampling expired gases.
## Technology and construction -
The design, fabrication, shape, size, etc. are equivalent to the predicate - K971229 - Capnoxygen LLC Capnoxygen mask. This has been confirmed by dimensional verification, internal volume, entrainment port size, and configuration of the sampling connector.
Discussion - The design incorporates a standard face mask shape, connector for oxygen tubing and luer fitting for connecting a gas sampling line. The minor diffreences in dimenrsions and internal volume are not clinical significant as this mask does not provide an air tight seal and air will leak around the surface as well as through the entrainment holes.
#### Environment of Use -
The environments of use are identical to predicate - K971229 - Capnoxygen LLC Capnoxygen mask
Discussion - The environments of use are identical to the predicate K971229 - Capnoxygen LLC Capnoxygen mask
## Patient Population -
The patient population is Non-intubated patients who are breathing spontaneously which is identical to the predicate - K971229 - Capnoxygen LLC Capnoxygen mask Discussion - The patient populations are identical to the predicate K971229 - Capnoxygen LLC Capnoxygen mask.
## Non-Clinical Testing Summary -
We have performed a number of performance tests which include:
- Courative CO-sampling and waveform performance at various breathing rates, tidal . volumes and oxygen flow rates. The testing was performed at oxygen flow rates of 1 Ipm, 5 lpm, and 10 lpm and CO2% of 1% and 5%.
- Real-time aging was performed and the device evaluated for performance to ● specifications which included pressure drop and resistance.
- Environmental and mechanical testing �
- . Total VOC
All testing demonstrated that the proposed device is substantially equivalent to the predicate device. It was noted that with higher oxygen flow rates and breath rates that the measured EtCO2 would be diluted which is as expected. As the proposed devices are a tool to detect the qualitative presence of breathing and delivery supplemental oxygen.
{3}------------------------------------------------
510(k) Summary Page 4 of 4 20-Nov-2013
## Materials -
We have performed ISO 10993 testing on the component materials of the Capnomask which has parts which are considered as Surface (direct) contact and Externally Communicating (Indirect) mucosal contact with the patient which means the following tests are required if a material certification cannot be provided.
- Cytotoxicity .
- Sensitization .
- Intracutaneous Irritation .
- Total VOC �
# Substantial Equivalence Conclusion -
The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate have been found to substantially equivalent.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure embracing or supporting another figure, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 25, 2013
Mediplus, Limited C/O Mr. Paul Dryden President, Regulatory Consultant ProMedic, Incorporated 24301 Woodsage Drive Bonita Springs, FL 34134
Re: K131339
Trade/Device Name: Capnomask Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon dioxide gas analyzer Regulatory Class: II Product Code: CCK Dated: October 21, 2013 Received: October 24, 2013
## Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
## Page 2 - Mr. Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default,htm.
Sincerely vours.
Image /page/5/Picture/8 description: The image shows the text "ErinFDith" in a bold, sans-serif font. The letters are closely spaced together, and the word appears to be a combination of two names, "Erin" and "Edith". The letters "FDA" are in the middle of the word and are stylized with a decorative border.
Erin Keith M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use Statement
Page 1 of 1
510(k) Number: K131339 .
Indications for Use:
The Capnomask is intended to deliver supplemental oxygen to patients and monitor breathing by sampling exhaled carbon dioxide.
Environment of use - hospital and sub-acute settings.
Patient population - Non-intubated patients who are breathing spontaneously.
Prescription Use XX (Part 21 CFR 801 Subpart D) or
Over-the-counter use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Navar 2013 05.00,
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.