K130036 · Capnia, Inc. · CCJ · Jan 14, 2014 · Anesthesiology
Device Facts
Record ID
K130036
Device Name
COSENSE CO MONITOR
Applicant
Capnia, Inc.
Product Code
CCJ · Anesthesiology
Decision Date
Jan 14, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1430
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K130036 · Jan 14, 2014
COSENSE CO MONITOR
Capnia, Inc.
Published clinical literature
The sponsor conducted an analysis of published clinical data concerning the use of predicate and reference devices in routine clinical practice to demonstrate that the subject device's performance specifications are within the range of devices currently used for detecting endogenous and exogenous CO sources.
Literature review; Clinical performance; Substantial equivalence
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Analysis of published clinical data; Literature review/analysis of clinical use
Patients undergoing CO monitoring for poisoning, smoke inhalation, or hemolysis detection
Not applicable for this study
Functional and performance specifications (accuracy, measurement, resolution)
Indications for Use
The CoSense ETCO Monitor is indicated for the monitoring of carbon monoxide from endogenous sources (including hemolysis) and exogenous sources (including CO poisoning and smoke inhalation) in exhaled breath. The end tidal carbon monoxide level can be used for the monitoring of carbon monoxide in medical conditions in which the rate of hemolysis may be relevant. It is also for use in smoking cessation programs and can be used for the screening of CO poisoning and smoke inhalation.
Device Story
CoSense ETCO Monitor is a portable, battery-operated device for measuring end-tidal carbon monoxide (ETCO) concentration. It utilizes an infrared capnometer to identify the end-tidal portion of a patient's breath and an electrochemical sensor to quantify CO concentration. The device is operated by healthcare professionals in clinical settings. Input consists of a single breath sample collected via a disposable nasal cannula. The device processes this input to provide a quantitative CO measurement, which is displayed on an LCD screen. The output assists clinicians in assessing CO levels related to hemolysis, smoke inhalation, or CO poisoning, thereby informing clinical management and diagnostic decisions. The device features a software-controlled menu for patient identification and monitoring parameters.
Clinical Evidence
No new clinical trials were conducted. Evidence consists of an analysis of published clinical data regarding the use of FDA-cleared CO monitoring devices for screening CO poisoning, smoke inhalation, and hemolysis detection. Results demonstrate that the CoSense device's functional and performance specifications are consistent with those of devices used in the cited studies, specifically the CO-STAT End Tidal Breath Analyzer, and fall within established clinical ranges for accuracy and resolution.
Technological Characteristics
Portable monitor; electrochemical CO sensor; infrared capnometer for breath detection. Powered by rechargeable Li-Ion battery. Patient interface: non-DEHP PVC nasal cannula. Dimensions: 246mm x 197mm x 68mm; Weight: 3.3lbs. Connectivity: standalone with LCD interface. Software: microprocessor-based with menu-driven control. Performance: 1.0-25.0 ppm range, 0.1 ppm resolution, <5 min measurement time.
Indications for Use
Indicated for monitoring endogenous (e.g., hemolysis) and exogenous (e.g., CO poisoning, smoke inhalation) carbon monoxide in exhaled breath. Used for smoking cessation programs and screening for CO poisoning/smoke inhalation. Applicable to patients where hemolysis rate monitoring is clinically relevant.
Regulatory Classification
Identification
A carbon monoxide gas analyzer is a device intended to measure the concentration of carbon monoxide in a gas mixture to aid in determining the patient's ventilatory status. The device may use techniques such as infrared absorption or gas chromatography.
{0}------------------------------------------------
14130036
JAN 1 4 2014
# 5.0 510(k) Summary
| Submitter: | Capnia, Inc.<br>2445 Faber Place, Suite 250<br>Palo Alto CA 94303<br>(650) 213-8444 phone<br>(650) 213-8383 fax |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Julie Blacklock<br>Director Quality and Regulatory Affairs<br>(650) 213-8444 ext 41<br>julie@capnia.com |
| Date Prepared: | 01/13/2014 |
| Trade Name: | CoSense™ ETCO Monitor |
| Common Name: | Carbon Monoxide Monitor |
| Classification Name: | Carbon Monoxide Gas Analyzer<br>(21 CFR 868.1430, Product Code CCJ) |
| Predicate Devices: | CoSense™ CO Monitor (K121768) and<br>CO-STATTM End Tidal Breath Analyzer (K974805)<br>(reference predicate) |
| Device Description: | The CoSense ETCO Monitor is a battery-operated carbon<br>monoxide (CO) monitor. It uses an infrared capnometer to<br>detect the end-tidal portion of the breath and an<br>electrochemical carbon monoxide sensor to measure the<br>end-tidal breath CO concentration. The device consists of<br>a portable unit with software controlled menu (date, time,<br>patient identification, measurement time of monitoring),<br>single-use nasal cannula, replaceable CO Sensor, and a |
Capnia, Inc.
battery charger/power supply.
{1}------------------------------------------------
### Indications for Use:
Technological Characteristics: The CoSense ETCO Monitor is indicated for the monitoring of carbon monoxide from endogenous sources (including hemolysis) and exogenous sources (including CO poisoning and smoke inhalation) in exhaled breath. The end tidal carbon monoxide level can be used for the monitoring of carbon monoxide in medical conditions in which the rate of hemolysis may be relevant. It is also for use in smoking cessation programs and can be used for the screening of CO poisoning and smoke inhalation.
The CoSense ETCO Monitor uses the identical performance specifications (accuracy, range, and resolution), software algorithms, sensors, and accessories as our predicate device, the CoSense CO Monitor. The CoSense ETCO Monitor also has similar performance specifications (accuracy, range, and resolution) as the reference predicate, CO-STAT End Tidal Breath Analyzer.
### Comparison to Predicate Device (K121768) and Reference Predicate (K974805):
| | Capnia CoSense<br>ETCO Monitor<br>(Subject Device) | Capnia CoSense CO<br>Monitor<br>(Predicate Device) | Natus CO-STAT, End<br>Tidal Breath Analyzer<br>(Reference Predicate<br>Device) |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>Number | K130036 | K121768 | K974805 |
| Manufacturer | Capnia, Inc. | Capnia, Inc. | Natus Medical, Inc. |
| Classification | Class II | Class II | Class II |
| Product Code | CCJ | CCJ | CCJ |
| Regulation | 21 CFR 868.1430 | 21 CFR 868.1430 | 21 CFR 868.1430 |
| Indications for<br>Use | The CoSense ETCO<br>Monitor is indicated<br>for the monitoring<br>of carbon monoxide<br>from endogenous<br>sources (including<br>hemolysis) and<br>exogenous sources | The CoSense CO<br>Monitor is indicated<br>for the monitoring of<br>Carbon Monoxide<br>from endogenous and<br>exogenous sources in<br>exhaled breath. It is<br>for use in smoking | The Natus Breath Analyzer<br>is intended for non-<br>invasive, quantitative<br>measurement of respiratory<br>rate, end tidal carbon<br>dioxide concentration, and<br>end tidal carbon monoxide<br>(corrected for background |
| Capnia CoSense<br>ETCO Monitor<br>(Subject Device) | Capnia CoSense CO<br>Monitor<br>(Predicate Device) | Natus CO-STAT End<br>Tidal Breath Analyzer<br>(Reference Predicate<br>Device) | |
| (including CO<br>poisoning and<br>smoke inhalation)<br>in exhaled breath.<br>The end tidal<br>carbon monoxide<br>level can be used<br>for the monitoring<br>of carbon monoxide<br>in medical<br>conditions in which<br>the rate of<br>hemolysis may be<br>relevant. It is also<br>for use in smoking<br>cessation programs<br>and can be used for<br>the screening of CO<br>poisoning and<br>smoke inhalation. | cessation programs<br>and can be used for<br>the screening of CO<br>poisoning and smoke<br>inhalation. It is for use<br>by health<br>professionals. | carbon monoxide)<br>concentration in the breath.<br>The analyzer is intended for<br>use with neonates, children,<br>and adults breathing<br>spontaneously.<br>The analyzer measures the<br>carbon monoxide<br>concentration in end tidal<br>breath, as an indicator of<br>the blood level of COHb.<br>The level of COHb (and<br>consequently the<br>concentration of carbon<br>monoxide in the end tidal<br>breath) can be affected by<br>endogenous sources (for<br>example the rate of<br>hemolysis), exogenous<br>sources (for example,<br>combustion engine<br>exhaust), or in some cases<br>both. The COHb level,<br>elevated or normal, can be<br>used in the diagnosis of<br>medical conditions in<br>which the rate of hemolysis<br>may be relevant, and in the<br>monitoring of patient<br>populations affected by the<br>rate of hemolysis. The<br>analyzer is also indicated<br>for use in respiratory status<br>evaluation, whenever<br>measurement of respiratory<br>rate and end tidal carbon<br>dioxide concentration are<br>desired.<br>The analyzer is intended for<br>use under the direction of a<br>physician in hospitals and a<br>variety of health care<br>settings. | |
CONFIDENTIAL K 130036 Pre-Market Notification for the CoSense ETCO Monitor
{2}------------------------------------------------
Capnia, Inc.
CONFIDENTIAL K 130036 Pre-Market Notification for the CoSense ETCO Monitor
.
{3}------------------------------------------------
| | Capnia CoSense<br>ETCO Monitor<br>(Subject Device) | Capnia CoSense CO<br>Monitor<br>(Predicate Device) | Natus CO-STAT End<br>Tidal Breath Analyzer<br>(Reference Predicate<br>Device) |
|-------------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| Patient<br>Interface | Nasal cannula | Nasal cannula | Nasal cannula |
| Dimensions<br>(LxWxH) | 246mm x 197mm x<br>68mm | 244mm x 183mm x<br>58mm | 222mm x 292mm x 152mm |
| Weight | 3.3lbs | 2.3lbs | 12lbs |
| Materials | | | |
| CO Sensor<br>Cell Type | Electrochemical | Electrochemical | Electrochemical |
| Cannula | Non-DEHP PVC | Non-DEHP PVC | Not Available |
| Battery | Li-Ion | Li-Ion | AC Power connected to UPS |
| Performance<br>Specifications | | | |
| Accuracy | +/- 10% or +/-0.5ppm<br>whichever is greater | +/- 10% or +/-0.5ppm<br>whichever is greater | +/- 10% or +/-0.3ppm<br>(whichever is greater) at<br>8-60 bpm |
| CO<br>Measurement<br>Range | 1.0 - 25.0ppm | 1.0 - 25.0ppm | 0-25 ppm |
| Resolution | 0.1 ppm | 0.1 ppm | 0.1 ppm |
| Measurement<br>Time | Less than 5 minutes | Less than 5 minutes | 90% of final reading in 30<br>seconds; total time<br>unknown |
| Sample<br>Collection | Collection of a normal<br>breath using a<br>disposable nasal<br>cannula | Collection of a normal<br>breath using a<br>disposable nasal<br>cannula | Collection of a normal<br>breath using a disposable<br>nasal cannula |
| Modes | Expired | Expired | Expired |
| Device Shelf<br>Life | 1 year before servicing | 1 year before servicing | 1 year before servicing |
| CO Sensor<br>Shelf Life | 6 months | 6 months | 30 days before calibration |
| Cannula Shelf<br>Life | 8 months | 8 months | Not Available |
| Screen | LCD | LCD | Not Available |
| Software/<br>Hardware | Analog and digital<br>electronics with<br>microprocessor | Analog and digital<br>electronics with<br>microprocessor | Not Available |
Capnia, Inc.
CONFIDENTIAL K 130036 Pre-Market Notification for the CoSense ETCO Monitor
{4}------------------------------------------------
| | Capnia CoSense<br>ETCO Monitor<br>(Subject Device) | Capnia CoSense CO<br>Monitor<br>(Predicate Device) | Natus CO-STAT End<br>Tidal Breath Analyzer<br>(Reference Predicate<br>Device) |
|--------------|----------------------------------------------------|----------------------------------------------------|-------------------------------------------------------------------------------|
| Power Source | Rechargeable Battery | Rechargeable Battery | AC |
## Non-clinical Performance The design and performance specifications are identical Data: to our predicate device. No additional non-clinical performance data is provided. Clinical Performance An analysis of published clinical data was conducted. Data: This analysis was performed on the uses of other FDA-cleared CO monitoring devices for measurement of CO in screening of CO poisoning, smoke inhalation and detection of hemolysis. Results provide objective evidence that the functional and performance specifications of CoSense device are similar to the devices in the published studies, specifically the CO-STAT End Tidal Breath Analyzer, and are within the range of accuracy, measurement, and resolution of the devices currently used clinically for detection of endogenous and exogenous sources of elevated CO. Conclusion: Capnia considers the revised CoSense ETCO Monitor device to be equivalent to our predicate device and the reference predicate listed above. This conclusion is based upon the devices' similarities in indications for use and identical principles of operation, technology, and performance. The accuracy, measurement range, and resolution specifications are unchanged from our predicate CoSense device. These performance specifications are similar to the CO-STAT End Tidal Breath Analyzer (reference predicate). The proposed CoSense device indication is also similar to the indication for CO-STAT End Tidal Breath Analyzer, as both indications use hemolysis as examples of endogenous CO. Published clinical data and previous design verification data demonstrate that the device is equivalent to our predicate device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three wing-like strokes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol. The logo is presented in black and white.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 14, 2014
Capnia, Incorporated Ms. Julie Blacklock Director, Quality and Regulatory Affairs 2445 Faber Place, Suite 250 Palo Alto, CA 94303
Re: K130036
Trade/Device Name: CoSense™ ETCO Monitor Regulation Number: 21 CFR 868.1430 Regulation Name: Carbon Monoxide Gas Analyzer Regulatory Class: II Product Code: CCJ Dated: December 13, 2013 Received: December 16, 2013
Dear Ms. Blacklock:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Ms. Blacklock
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Image /page/6/Picture/6 description: The image contains a signature and some text. The text reads "Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID FOR". The signature is illegible and is located to the left of the text. The text is slightly blurred.
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
K130036
Device Name
CoSense ETCO Monitor
#### Indications for Use (Describe)
The CoSense ETCO Monitor is indicated for the monitoring of carbon monoxide from endogenous sources (including hemolysis) and exogenous sources (including CO poisoning and smoke inhalation) in exhaled breath. The end tide level can be used for the monitoring of carbon monoxide in medical conditions in which the rate of hemolysis may be relevant. It is also for use in smoking cessation programs and can be used for the screening of CO poisoning and smoke inhalation.
Type of Use (Select one or both, as applicable)
[ Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement on last page.
# PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.
THE ONLY FOR FOR THE SERVED A FOR FOR FOR FOR FOR FOR FOR FOR 이 가지 이용 대한민국 대학 Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/7/Picture/13 description: The image shows the name "Nayan J. Patel" in bold, black font. Below the name is the date and time "2014.01.13 18:05:50 -05'00'" also in bold, black font. There is a logo to the right of the name and date.
**4-2**
FORM FDA 3881 (1/14)
{8}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.