The DeRoyal Esophageal Stethoscope with Temperature Sensor is to be used for routine monitoring of core body temperature, as well as heart and respiratory sounds, in an anesthetized patient by inserting the stethoscope tube into the esophagus.
Device Story
Device is a sterile, single-use esophageal stethoscope with integrated temperature sensor. Input: acoustic signals from heart/respiratory sounds and thermal energy from esophageal environment. Operation: stethoscope tube with distal cuff transmits acoustic sounds to proximal ear piece; internal thermistor chip modifies electrical current based on temperature. Output: acoustic sounds for clinician auscultation and electrical signal for external temperature monitor. Used in hospital settings by clinicians during anesthesia. Benefits: enables simultaneous core temperature monitoring and auscultation in anesthetized patients.
Clinical Evidence
Bench testing only. Evaluated per ISO 10993-1 (biocompatibility), ISO 80601-2-56 (accuracy/time response), IEC 60601-1 (electrical safety), and IEC 60601-1-2 (EMC). Acoustic responsiveness and leakage current tests confirmed the modified encapsulation method maintains device performance and safety. Accuracy is ± 0.2°C.
Indicated for routine monitoring of core body temperature, heart sounds, and respiratory sounds in anesthetized patients via esophageal insertion.
Regulatory Classification
Identification
An esophageal stethoscope with electrical conductors is a device that is inserted into the esophagus to listen to a patient's heart and breath sounds and to monitor electrophysiological signals. The device may also incorporate a thermistor for temperature measurement.
Predicate Devices
DeBusk Temperature Systems Esophageal Stethoscope (K925789)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
DeRoyal Industries, Inc. Sarah Bennett Senior Regulatory Affairs Specialist 200 DeBusk Lane Powell, Tennessee 37849
Re: K193027
Trade/Device Name: Esophageal Stethoscope with Temperature Sensors Regulation Number: 21 CFR 868.1920 Regulation Name: Esophageal Stethoscope with Electrical Conductors Regulatory Class: Class II Product Code: BZT Dated: February 25, 2020 Received: February 27, 2020
Dear Sarah Bennett:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or post-marketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193027
#### Device Name
Esophageal Stethoscope with Temperature Sensor
Indications for Use (Describe)
The DeRoyal Esophageal Stethoscope with Temperature Sensor is to be used for routine monitoring of core body temperature, as well as heart and respiratory sounds in an anesthetized patient by inserting the stethoscope tube into the esophagus.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
DeRoyal Industries, Inc. Traditional 510(k) Submission – Esophageal Stethoscope with Temperature Sensor October 25, 2019
## 510(k) Summary
| Date prepared: | October 25, 2019 |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Owner: | DeRoyal Industries, Inc.<br>200 DeBusk Lane<br>Powell, TN 37849<br>Owner/Operator #1044833 |
| 510(k) Contact: | Sarah Bennett<br>Senior Regulatory Affairs Specialist<br>P: 865-362-6112<br>F: 865-362-3741<br>sabennett@deroyal.com |
| Contract Manufacturer: | DeRoyal Cientifica de Latinoamerica<br>S.R.L.<br>Global Park<br>Box 180-3006, 602 Parkway<br>La Aurora, Heredia Heredia, Costa<br>Rica 146 |
| Trade Name: | Esophageal Stethoscope with<br>Temperature Sensor |
| Common Name: | Esophageal Stethoscope with<br>Temperature Sensor |
| Classification Name: | Stethoscope, Esophageal, with<br>Electrical Conductors |
| Device Class: | Class II |
| Device Product Code: | BZT |
| Regulation Number: | 21 CFR 868.1920 |
| Classification Panel: | Anesthesiology |
| Predicate Devices: | DeBusk Temperature Systems<br>Esophageal Stethoscope [K925789] |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for DeRoyal. The logo consists of a blue swoosh-like graphic on the left, followed by the word "DeRoyal" in blue, with a registered trademark symbol next to it. The logo is simple and modern, with a clean design.
### Device Description
The DeRoyal Esophageal Stethoscope with Temperature Sensor is inserted into the esophagus of anesthetized patients to monitor core body temperature as well as heart and respiratory sounds. The probe consists of a small diameter plastic tube upon which a blue cuff is placed. When inserted in the esophagus, the cuff facilitates transmission of sound to an ear piece attached at the proximal end of the tube. Additionally, a wire set is placed inside the plastic tube. This wire set contains a thermistor chip at the distal end that passively modifies the electrical current traveling through the probe. A blue connector at the proximal end of the wire set interfaces with a cable that is connected to an independent temperature monitor that is used to display the temperature readings.
The probe is sterile and individually packaged. It is available in 9, 12, 18, and 24 French sizes and may come with 400 or 700 series sensors. The stethoscope has a male luer fitting for attachment to standard ear pieces.
## Intended Use
The DeRoyal Esophageal Stethoscope with Temperature Sensor is to be used for routine monitoring of core body temperature, as well as heart and respiratory sounds, in an anesthetized patient by inserting the stethoscope tube into the esophagus.
| Characteristic | DeRoyal Esophageal<br>Stethoscope with<br>Temperature Sensor | DeBusk<br>Temperature<br>Systems Esophageal<br>Stethoscope<br>(K925789) |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The DeRoyal<br>Esophageal<br>Stethoscope with<br>Temperature Sensor<br>is to be used for<br>routine monitoring of<br>core body<br>temperature, as well<br>as heart and<br>respiratory sounds, in<br>an anesthetized<br>patient by inserting<br>the stethoscope tube<br>into the esophagus. | The EXAC-TEMP<br>Esophageal<br>Stethoscope with<br>temperature sensor is<br>to be used for routine<br>monitoring of<br>temperature, as well<br>as for heart and<br>respiratory sounds, in<br>an anesthetized<br>patient by inserting<br>the stethoscope tube<br>into the esophagus. |
| Prescription<br>Only | Yes | Yes |
#### Summary of Technological Characteristics
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logo for DeRoyal. The logo consists of a blue swoosh-like graphic on the left, followed by the word "DeRoyal" in blue, with a registered trademark symbol next to it. The logo is simple and modern, with a clean design.
| Mode of<br>Operation | Direct according to<br>ISO 80601-2-56 | Direct according to<br>ISO 80601-2-56 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Measuring Site | Esophagus | Esophagus |
| Reference Body<br>Site | Core body<br>temperature | Core body<br>temperature |
| Use Environment | Hospital | Hospital |
| Design | A wire set inserted<br>into a plastic tube<br>with a blue cuff at the<br>distal end. The wire<br>set has a thermistor<br>chip at the distal end<br>and a blue connector<br>at the proximal end. | A wire set inserted<br>into a plastic tube<br>with a blue cuff at the<br>distal end. The wire<br>set has a thermistor<br>chip at the distal end<br>and a blue connector<br>at the proximal end. |
| Materials | Tube – PVC<br>Cuff – PVC<br>Wire – Copper with<br>PVC insulation<br>Thermistor - Ceramic<br>Ear Piece - K-Resin<br>Connector – PVC-<br>molded brass<br>Strain Relief – PVC<br>Cap - UV-Curing<br>Adhesive | Tube – PVC<br>Cuff – PVC<br>Wire - Copper with<br>PVC insulation<br>Thermistor - Ceramic<br>Ear Piece - K-Resin<br>Connector - PVC-<br>molded brass<br>Strain Relief - PVC<br>Cap - PVC and<br>epoxy glue |
| Accuracy | $\pm$ 0.2°C | $\pm$ 0.2°C |
| Sterilization | Sterilized with<br>ethylene oxide | Sterilized with<br>ethylene oxide |
| Biocompatibility | ISO 10993-1<br>compliant | ISO 10993-1<br>compliant |
## Summary of Performance Tests
The proposed device has been tested and/or evaluated according to the following standards: ISO 10993-1, ISO 80601-2-56, IEC 60601-1, and IEC 60601-1-2.
The following biocompatibility tests were performed on the predicate device: cytotoxicity, skin sensitization, irritation, acute systemic toxicity, and material-mediated pyrogenicity. The proposed change was evaluated in a biological risk assessment, and it was determined that, because the change is to a non-patient contacting material that does not directly or indirectly contact tissue, it was unnecessary to perform new testing on a device manufactured with the proposed encapsulation method.
Testing according to IEC 60601-1 and IEC 60601-1-2 was performed to ensure the change in the encapsulation did not affect the electrical safety of the device. Accuracy and time response
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for DeRoyal. The logo consists of a blue swoosh-like graphic above the text "DeRoyal" in blue. The "R" in DeRoyal has a registered trademark symbol next to it.
testing according to ISO 80601-2-56 were performed to ensure the proposed device performed its essential performance safely and effectively. A leakague current test also was performed after submerging the device in solution to ensure the encapsulation method is effective. Additionally, an acoustic responsiveness test was done to demonstrate the proposed change did not affect the stethoscope functions of the device.
All testing was performed on a final, finished device manufactured with the proposed modification. The test results met the requirements of the aforementioned standards and demonstrate that the proposed modification does not impact the safety or efficacy of the device.
#### Conclusion
The results of performance testing demonstrate the Esophageal Stethoscope with Temperature Sensor do not raise different questions of safety or effectiveness. With the exception of the proposed change to the encapsulation method, the proposed device is identical to the predicate, which has been on the market since 1993. Therefore, the proposed device is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.