Breezing Med

K200076 · Tf Health Co. · BZL · Sep 25, 2020 · Anesthesiology

Device Facts

Record IDK200076
Device NameBreezing Med
ApplicantTf Health Co.
Product CodeBZL · Anesthesiology
Decision DateSep 25, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.1730
Device ClassClass 2

Indications for Use

Breezing Med™ Metabolism Analyzer measures metabolic functions including resting energy expenditure and respiratory quotient during resting conditions via direct measurement of oxygen uptake and carbon dioxide production to provide nutritional assessment, to optimize nutritional supplements, and to quantify substrate utilization. The nutritional assessment enables the healthcare professional to further assess energy expenditure and caloric intake for weight management, and to aid in the diagnosis of diseases related to abnormal metabolic parameters. Breezing Med is intended for use with adults breathing normally on their own in a sitting position in a healthcare environment. Since the device is designed for the patient to breathe ambient air, it is not intended for patients where supplementary oxygen is being provided. Breezing Med is not intended to use as sole means for any diagnosis, and ultimately will provide data for evaluation by the professional.

Device Story

Breezing Med is a wearable, mask-based indirect calorimeter for professional healthcare use. It measures resting metabolic rate by capturing exhaled air to determine oxygen uptake (VO2) and carbon dioxide production (VCO2). The device consists of a body, mask, headgear, and a single-use sensor cartridge. It connects to a mobile application for data processing and display. The system calculates Resting Energy Expenditure (REE) and Respiratory Quotient (RQ) from measured gas volumes. Healthcare professionals use these outputs to assess energy expenditure, guide nutritional planning, and assist in diagnosing metabolic disorders. The device provides objective data to support clinical decision-making regarding weight management and metabolic health.

Clinical Evidence

Bench testing only. Verification testing compared Breezing Med performance against the predicate Express Series. Results showed O2 and CO2 volume measurement accuracy within +/-0.8% and flow rate accuracy within +/-0.3 Lpm (<10 Lpm) or +/-3% (>=10 Lpm). Biocompatibility (ISO 10993), electrical safety (IEC 60601-1), and EMC (IEC 60601-1-2) testing were performed. A simulated clinical use study confirmed the device functions according to its intended use.

Technological Characteristics

Wearable indirect calorimeter; mask-based; fixed-orifice differential pressure pneumotach flowmeter; colorimetric O2 and CO2 sensors; single-use sensor cartridge; rechargeable Lithium-Ion battery; mobile application interface; non-sterile; IEC 60601-1 Class II/Type BF; EN60601-1-2 EMC compliant; biocompatible materials per ISO 10993-1.

Indications for Use

Indicated for adults breathing normally in a sitting position in a healthcare environment for nutritional assessment, optimization of nutritional supplements, quantification of substrate utilization, and aiding in the diagnosis of diseases related to abnormal metabolic parameters. Contraindicated for patients receiving supplementary oxygen.

Regulatory Classification

Identification

An oxygen uptake computer is a device intended to compute the amount of oxygen consumed by a patient and may include components for determining expired gas volume and composition.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 25, 2020 TF Health Co. % Roberta (Bobbi) Druyor-Sanchez Regulatory Consultant NDA Partners 1761 W. University Dr., Suite 140 Tempe, Arizona 85281 Re: K200076 Trade/Device Name: Breezing Med Regulation Number: 21 CFR 868.1730 Regulation Name: Oxygen Uptake Computer Regulatory Class: Class II Product Code: BZL Dated: August 28, 2020 Received: September 3, 2020 Dear Roberta (Bobbi) Druyor-Sanchez: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration #### Indications for Use 510(k) Number (if known) ### K200076 Device Name Breezing MedTM Metabolism Analyzer #### Indications for Use (Describe) Breezing Med™ Metabolism Analyzer measures metabolic functions including resting energy expenditure and respiratory quotient during resting conditions via direct measurement of oxygen uptake and carbon dioxide production to provide nutritional assessment, to optimize nutritional supplements, and to quantify substrate utilization. The nutritional assessment enables the healthcare professional to further assess energy expenditure and caloric intake for weight management, and to aid in the diagnosis of diseases related to abnormal metabolic parameters. Breezing Med is intended for use with adults breathing normally on their own in a sitting position in a healthcare environment. Since the device is designed for the patient to breathe ambient air, it is not intended for patients where supplementary oxygen is being provided. Breezing Med is not intended to use as sole means for any diagnosis, and ultimately will provide data for evaluation by the professional. | Type of Use (Select one or both, as applicable) | <span> <span style="padding-right: 20px;"> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> | |-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (7/17) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. {3}------------------------------------------------ # TF Healt Corporation Traditional Premarket Notification Breezing Med™ Metabolism Analyzer ## 510(K) SUMMARY (21 CFR 807.92(C)) 1. Applicant TF Health Co. 1761 W University Dr. Suite 140 Tempe, AZ 85281 Primary Contact Erik Herrmann VP Operations, TF Health 480-629-5360 Email: erik.herrmann@breezing.co ### 2. Device - Trade Name: Breezing MedTM Metabolism Analyzer ● - Common Name: Indirect calorimeter ● - Classification Name: Computer, Oxygen Update (21 CFR 868.1730, product code ● BZL) - 3. Predicate Device: Express Series (K070858) ## 4. Device Description Breezing Med™ Metabolism Analyzer is a resting metabolic rate measurement system with an application installed on mobile devices that provides measurements of exhalation rate, oxygen uptake and carbon dioxide production. It is used to determine resting metabolic rates, teach energy balance, and pinpoint accurate caloric intake for weight management. - 5. Intended Use Breezing Med™ Metabolism Analyzer is intended for professional use, to determine resting metabolic rates, teach energy balance, and pinpoint accurate caloric intake for weight management. ## 6. Indications for Use Breezing Med™ Metabolism Analyzer measures metabolic functions including resting energy expenditure and respiratory quotient during resting conditions via direct measurement of oxygen uptake and carbon dioxide production to provide nutritional assessment, to optimize nutritional supplements, and to quantify substrate utilization. The nutritional assessment enables the healthcare professional to further assess energy expenditure and caloric intake for weight {4}------------------------------------------------ Traditional Premarket Notification Breezing Med™ Metabolism Analyzer management, and to aid in the diagnosis of diseases related to abnormal metabolic parameters. Breezing Med 100 is intended for use with adults breathing normally on their own in a sitting position in a healthcare environment. Since the device is designed for the patient to breathe ambient air. it is not intended for patients where supplementary oxygen is being provided. Breezing Med™ is not intended to use as sole means for any diagnosis, and ultimately will provide data for evaluation by the professional. - 7. Technological Characteristics Breezing Med™ is a stand-alone and fully integrated mask-like wearable metabolic analyzer based on the principle of indirect calorimetry, which is a gold standard for energy expenditure measurement. Breezing Med 104 is designed for metabolic rate and respiratory quotient measurement and the main outputs include Resting Energy Expenditure (REE), Respiratory Quotient (RQ), Volume of Oxygen consumption (VO2), and Volume of Carbon Dioxide production (VCO2), where REE and RQ are calculated from measured parameters VO2 and VC02. Breezing Med M measures the expiratory volume similar to the predicate device Medical Graphics Express® Series. Breezing Med "" consists of five components: 1) a device body; 2) a mask; 3) headgear; 4) a sensor cartridge; and 5) an application. ## 8. Performance Data A verification test of the Breezing Med™ to performance specifications derived from the predicate Medical Graphics Express series was conducted. The result of the test demonstrates substantial equivalence of the proposed to the predicate. The summary of the Breezing Med™ test results are as follows: - Both devices measured O2 volumes within the specified accuracy of +/-0.8%. ● - Both devices measured CO2 volumes within the specified accuracy of +/-0.8%. . - Both devices measured Flow Rate within +/- 0.3 Lpm < 10 Lpm and +/-3% >= 10 Lpm. We have also performed testing to ISO10993, IEC 60601-1 General requirements for basic safety and essential performance and IEC 60601-1-2 Electromagnetic disturbances. ## 9. Predicate Device Comparison The Breezing Med Metabolism Analyzer is substantially equivalent to the Express Series (K070858). Both Breezing Med Me Metabolism Analyzer and its predicate device have a similar intended use. A summary comparison of technological characteristics is provided in Table 5.1. ## Table 5.1 - Breezing Med 111 Metabolism Analyzer and Predicate Comparative Analysis {5}------------------------------------------------ # TF Health Corporation Traditional Premarket Notification Breezing Med™ Metabolism Analyzer | Feature | Subject Device | Predicate Device | |---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | Breezing MedTM | Express® Series (indirect<br>calorimeter) | | Device Manufacturer | TF Health Co. | MGC Diagnostics Corporation | | 510(k) Reference | This submission | K070858 | | FDA Product Code | BZL | BZL | | FDA Classification | Computer, Oxygen Uptake | Computer, Oxygen Uptake | | FDA Regulation<br>Number | 868.1730 | 868.1730 | | Device Description | Breezing MedTM is a resting<br>metabolic rate measurement<br>system with an application<br>installed on mobile devices<br>that provides measurements<br>of exhalation rate, oxygen<br>uptake and carbon dioxide<br>production. | The Express® Series is a<br>cardiopulmonary exercise or<br>resting metabolic rate<br>measurement system with an<br>integrated touch screen computer<br>that provides breath by breath<br>measurements of flow, oxygen<br>uptake and carbon dioxide<br>production. | | Indications for use | Breezing MedTM Metabolism<br>Analyzer measures metabolic<br>functions including resting<br>energy expenditure and<br>respiratory quotient during<br>resting conditions via direct<br>measurement of oxygen<br>uptake and carbon dioxide<br>production to provide<br>nutritional assessment, to<br>optimize nutritional<br>supplements, and to quantify<br>substrate utilization. The<br>nutritional assessment<br>enables the healthcare<br>professional to further assess<br>energy expenditure and<br>caloric intake for weight<br>management, and to aid in<br>the diagnosis of diseases<br>related to abnormal<br>metabolic parameters.<br>Breezing MedTM is intended<br>for use with adults breathing | The Express® Series uses the<br>direct measurement of oxygen<br>uptake to objectively and<br>noninvasively assess cardiac and<br>pulmonary function during<br>exercise. The system can be used<br>to screen for early signs of<br>cardiac and pulmonary<br>dysfunction; differentiate heart<br>and lung disease; assess dyspnea<br>complaints; classify patients<br>according to disease as a guide<br>for patient management; establish<br>an optimal exercise prescription<br>and training program; evaluate<br>the efficiency of prescribed<br>therapy. The Express® Series<br>uses the direct measurement of<br>oxygen uptake and carbon<br>dioxide production to provide<br>nutritional assessment, to<br>optimize nutritional supplements<br>and quantify substrate utilization.<br>The Express® Series can also | | Feature | Subject Device | Predicate Device | | | sitting position in a<br>healthcare environment.<br>Since the device is designed<br>for the patient to breathe<br>ambient air, it is not intended<br>for patients where<br>supplementary oxygen is<br>being provided. Breezing<br>MedTM is not intended to use<br>as sole means for any<br>diagnosis, and ultimately will<br>provide data for evaluation<br>by the professional. | of cardiac output using real time<br>oxygen uptake measurements. | | Target population | Adults | Unspecified | | Use Environment | Professional healthcare<br>facilities | Professional healthcare facilities | | Device<br>Measurements | Metabolic functions during<br>resting conditions | Non-invasive assessment of the<br>cardiopulmonary response to<br>exercise or measurement of<br>energy expenditure using indirect<br>calorimetry. | | Measurement<br>Mode(s) | Mixing chamber | Breath by breath | | Output Parameters | Main outputs include resting<br>energy expenditure (REE),<br>respiratory quotient (RQ),<br>oxygen consumption (VO2),<br>carbon dioxide production<br>(VCO2), and respiratory<br>parameters. | Main outputs include resting<br>energy expenditure (REE),<br>respiratory quotient (RQ), oxygen<br>consumption (VO2), carbon<br>dioxide production (VCO2), and<br>respiratory parameters. | | Major Separate<br>System Components | Device body, disposable<br>mask, headgear, sensor<br>cartridge, app in a mobile<br>device | Express® Series indirect<br>calorimeter, face tent/face mask,<br>printer, mobile cart | | Flowmeter<br>Technology | Fixed-orifice differential<br>pressure pneumotach | Bi-directional Pitot Tube Flow<br>Sensor | | O2 Sensor<br>Technology | Colorimetry | Galvanic | | CO2 Sensor<br>Technology | Colorimetry | Non-dispersive infrared (NDIR) | | User Interface | App on Mobile device | Touchscreen display | | Software | Application | Intuitive icon-based software | | Sterility | Non-sterile | Non-sterile | | Feature | Subject Device | Predicate Device | | Biocompatibility | All user contact components<br>are biocompatible and used<br>in already legally marketed<br>devices with the same<br>intended use | All patient contact components<br>are biocompatible and used in<br>already legally marketed devices<br>with the same intended use | | Mechanical Safety | No mechanical safety risk<br>identified. | No mechanical safety risk<br>identified. | | Chemical Safety | No chemical safety risks<br>identified. | No chemical safety risks<br>identified. | | Anatomical Sites | Patient's mouth and nose,<br>through face mask | Patient's mouth and nose, through<br>face tent/face mask | | Human factors | Requires patient to keep<br>mask sealed. | Requires patient to keep face<br>tent/face mask sealed. | | Energy used | External or internal power<br>supply | External power supply | | Energy delivered | No energy delivered to the<br>patient. | No energy delivered to the<br>patient. | | Battery | Lithium-Ion, rechargeable | NA | | Compatibility with<br>environment | No known issues. | No known issues. | | Compatibility with<br>other devices | No known devices for<br>compatibility issues. | No known devices for<br>compatibility issues. | | Standards Electrical<br>safety | IEC 60601-1: class II /<br>Internal Electric Power<br>Source - type BF | Unspecified | | Standards: EMC<br>Testing | EN60601-1-2 | Unspecified | | Thermal Safety | No known thermal safety<br>issues | No known thermal safety issues | | Radiation Safety | No known radiation safety<br>issues. | No known radiation safety issues. | | Sensor cartridge | One-time use sensor<br>cartridge for simultaneously<br>O2 and CO2 sensing | O2 and CO2 sensors are for long<br>time use (e.g. 1 year), and change<br>during maintenance periods | | Dimensions | 80 x 110 x 160 mm | 240 x 190 x 267 mm* | | Weight | 170 gm | 9.2 lbs. * | | Measurement Range | Flow: 0-150 l/min<br>O2: 10-21%<br>CO2: 0-10% | Flow: 0-40 l/min*<br>O2: 5-85%*<br>CO2: 0-10% * | | Feature | Subject Device | Predicate Device | | Measurement<br>Accuracy<br>Error/Resolution | Flow Accuracy: <3% or 10<br>ml/min…
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...