← Product Code [BZK](/submissions/AN/subpart-b%E2%80%94diagnostic-devices/BZK) · K120087

# EXSPIRON (K120087)

_Respiratory Motion, Inc. · BZK · Sep 26, 2012 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-b%E2%80%94diagnostic-devices/BZK/K120087

## Device Facts

- **Applicant:** Respiratory Motion, Inc.
- **Product Code:** [BZK](/submissions/AN/subpart-b%E2%80%94diagnostic-devices/BZK.md)
- **Decision Date:** Sep 26, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.1850
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology

## Indications for Use

ExSpiron is indicated for use by healthcare professionals in healthcare facilities, such as post-operative care and critical care units, to monitor breathing in adult (at least 21 years of age) patients. ExSpiron is a non-invasive system that graphically displays lung volume against time and reports an approximate value of: Tidal volume, Respiratory rate, and Minute ventilation. ExSpiron measurements are used as an adjunct to other clinical information sources.

## Device Story

ExSpiron is a non-invasive respiratory monitor for use by healthcare professionals in clinical settings. It utilizes a bioimpedance measurement system consisting of six electrodes placed on the patient's torso; a high-frequency current generator delivers current while an adaptive circuit conditions the resulting voltage signal. Firmware performs signal acquisition, relaying data to a Windows 7-based panel PC. The PC processes signals, performs calibration, and runs a GUI to display respiratory traces, tidal volume, minute ventilation, and respiratory rate. The device serves as an adjunct to other clinical information sources to assist in patient monitoring; it does not include alarm functions. Benefits include continuous, non-invasive tracking of respiratory parameters to support clinical decision-making in post-operative and critical care environments.

## Clinical Evidence

Clinical study compared ExSpiron to the Wright spirometer in 20 subjects (age 20-84). Measurements included tidal volume, minute ventilation, and respiratory rate. Results showed high agreement: overall bias for minute ventilation was -2.1%, tidal volume -1.9%, and respiratory rate -0.2%. Precision (overall) was 10.5% for minute ventilation, 10.3% for tidal volume, and 2.0% for respiratory rate. A secondary study compared the Wright spirometer to the Morgan SpiroAir LT to establish a reference baseline. Combined data confirmed ExSpiron accuracy (9.9% for minute ventilation, 9.6% for tidal volume) relative to the diagnostic spirometer.

## Technological Characteristics

Thoracic bioimpedance sensing; six-electrode lead set. Windows 7-based panel PC for signal processing and GUI. Connectivity via digital signal acquisition from adaptive circuit. Electrical/mechanical safety per IEC 60601-1; electromagnetic compatibility per IEC 60601-1-2. Single-patient-use electrode lead set.

## Regulatory Identification

A monitoring spirometer is a device used to measure continuously a patient's tidal volume (volume of gas inhaled by the patient during each respiration cycle) or minute volume (the tidal volume multiplied by the rate of respiration for 1 minute) for the evaluation of the patient's ventilatory status.

## Predicate Devices

- nSpire Wright/Haloscale Respirometer ([K091853](/device/K091853.md))
- Philips MPn0 Intellivue Patient Monitors ([K060541](/device/K060541.md))

## Reference Devices

- Morgan SpiroAir LT diagnostic spirometer

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K120087

Image /page/0/Picture/1 description: The image shows a black and white drawing of a pair of lungs. The lungs are connected to the trachea, which is the tube that carries air to and from the lungs. The lungs are the main organs of the respiratory system, and they are responsible for taking in oxygen and releasing carbon dioxide. The image is a simple illustration of the lungs and trachea.

SEP 2 6 2012

## 510(k) Summary

| 510(k) Owner:                    | Respiratory Motion, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address:                         | 305 Second Avenue, Suite B<br>Waltham, Massachusetts 02451                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Phone:                           | 781-373-1636                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Fax:                             | 781-373-1653                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Contact person:                  | Jenny Freeman, MD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Date 510(k)<br>Summary prepared: | September 20, 2012                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Trade name:                      | ExSpiron™                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Common name:                     | Respiratory Monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Classifications:                 | Monitoring spirometer<br>Product code: BZK<br>Regulation: 21 CFR 868.1850<br><br>Breathing frequency monitor<br>Product code BZQ<br>Regulation: 21 CFR 868.2375                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Predicate devices:               | For product code BZK, nSpire Wright/Haloscale Respirometer, marketed by<br>nSpire Health, Inc. of Longmont, CO 510(k) K091853<br><br>For product code BZQ, Philips MPn0 Intellivue Patient Monitors, marketed by<br>Philips Healthcare 510(k) K060541                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Device Description:              | The ExSpiron™ consists of:<br>Bioimpedance measurement system: A stabilized high frequency<br>current generator is connected to two outer electrodes. The inner four<br>electrodes are connected to an adaptive circuit that conditions the<br>resulting voltage signal and converts it to digital form. Firmware<br>performs signal acquisition and relays data to the panel PC.Panel PC: A Windows 7 PC performs signal processing and calibration,<br>and runs the graphical user interface (GUI). The PC takes user input from<br>a touch screen through a virtual keyboard and mouse. The GUI is used<br>for recording patient data and displaying the respiratory trace as well as<br>scalar values and trends for minute volume, tidal volume, and respiratory<br>rate.Single Patient Use ExSpiron™ Electrode Lead Set: An electrode lead<br>set containing six electrodes to be placed on the torso. It delivers current<br>and records impedance measurements. The electrode lead set is also used<br>to perform subsystem checks prior to patient measurements. |

{1}------------------------------------------------

| Intended use:                                      | ExSpiron™ is indicated for use by healthcare professionals in healthcare facilities,<br>such as post-operative care and critical care units, to monitor breathing in adult (at<br>least 21 years old) patients.<br><br>ExSpiron™ is a non-invasive system that graphically displays lung volume<br>against time and reports an approximate value of:<br>• Tidal volume,<br>• Respiratory rate, and<br>• Minute ventilation<br><br>ExSpiron™ measurements are used as an adjunct to other clinical information<br>sources. |                                                                     |                                                                                                              |                                       |                                                                                                                                                                                                                                                                                                                     |
|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison of<br>technological<br>characteristics: | Characteristic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | ExSpiron™                                                           | nSpire                                                                                                       | Philips<br>Intellivue<br>Monitor      | Comment                                                                                                                                                                                                                                                                                                             |
|                                                    | Intended Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | See above.                                                          | See K091853                                                                                                  | See K060541                           | The clinical application of the<br>ExSpiron™ is equivalent to that of<br>the nSpire.<br><br>The comparison to the intended<br>use of the Philips monitor is<br>restricted to monitoring of<br>respiration rate.<br><br>ExSpiron™ has no alarm function.<br><br>All devices are used by healthcare<br>professionals. |
|                                                    | Technology                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Measurement is<br>by thoracic<br>bioimpedance.                      | Measurement is<br>by in-line<br>turbine flow<br>meter.                                                       | Measurement<br>is by<br>bioimpedance. |                                                                                                                                                                                                                                                                                                                     |
|                                                    | Volume<br>Measurements                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Tidal volume<br>Minute volume<br>Volume vs. time<br>chart           | Tidal volume<br>Minute volume                                                                                | None                                  | The time function for the<br>minute volume measurement of<br>the nSpire is provided by a<br>user's stopwatch.                                                                                                                                                                                                       |
|                                                    | Rate<br>Measurements                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Respiratory rate<br>(breaths/min)                                   | None                                                                                                         | Respiratory<br>rate                   |                                                                                                                                                                                                                                                                                                                     |
|                                                    | Safety                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | IEC 60601-1,<br>including<br>electrical and<br>mechanical<br>safety | Mechanical<br>only. Safety<br>specifications<br>are not given in<br>the user manual<br>or 510(k)<br>summary. | IEC 60601-1,<br>etc.                  |                                                                                                                                                                                                                                                                                                                     |
| Nonclinical<br>performance                         | Performance testing confirmed essential performance by IEC 60601-1 and IEC<br>60601-1-2 tests and confirmed usability by simulated use by physicians and                                                                                                                                                                                                                                                                                                                                                                  |                                                                     |                                                                                                              |                                       |                                                                                                                                                                                                                                                                                                                     |

{2}------------------------------------------------

| Clinical<br>performance<br>testing:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | A clinical study compared simultaneous measurements from the ExSpiron and the<br>predicate, Wright spirometer. (Respiratory rate was calculated using a stop<br>watch.) Twenty subjects representing a broad range of intended patients<br>participated in the study. (Age range: 20-84, BMI range: 19.9-43.2, 11 female, 9<br>male) The study involved two sessions for each subject, an initial session in<br>which electrodes were applied, the ExSpiron was calibrated to the individual, and<br>each subject performed ten 60-second breathing tests. Tidal volume, minute<br>ventilation, and respiratory rate were measured simultaneously by the ExSpiron<br>and the Wright spirometer. Each subject returned twenty-four hours after the first<br>session with the original electrodes still attached. Using the same individual<br>calibration values, a second set of ten 60-second breathing tests was performed. |                   |  |                       |                 |                     |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|--|-----------------------|-----------------|---------------------|
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | The results of ExSpiron-Wright study were:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                   |  |                       |                 |                     |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | ExSpiron-Wright   |  | Minute<br>Ventilation | Tidal<br>Volume | Respiratory<br>Rate |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Bias                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Overall           |  | -2.1%                 | -1.9%           | -0.2%               |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Day 1             |  | -2.3%                 | -1.7%           | -0.6%               |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Day 2             |  | -1.9%                 | -2.0%           | 0.2%                |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Precision                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Overall           |  | 10.5%                 | 10.3%           | 2.0%                |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Day 1             |  | 10.2%                 | 9.7%            | 2.3%                |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Day 2             |  | 10.8%                 | 10.8%           | 1.6%                |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Accuracy                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Overall           |  | 10.7%                 | 10.4%           | 2.0%                |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Day 1             |  | 10.4%                 | 9.8%            | 2.4%                |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Day 2             |  | 10.9%                 | 11.0%           | 1.6%                |
| The results indicate that the individual calibration is stable over a 24-hour period.<br>A similar study was performed comparing the Wright spirometer to the Morgan<br>SpiroAir LT diagnostic spirometer. (The closed cell construction of the Morgan<br>spirometer limits the breathing test time to 30 seconds. A direct comparison of<br>the ExSpiron to the Morgan spirometer is suboptimal.) A different set of twenty<br>subjects participated. (Age range: 20-60, BMI range: 19.0-28.0, 9 female, 11<br>male) Each subject participated in one session performing twenty 30-second<br>breathing tests with volume measurements made simultaneously with the Wright<br>spirometer and the Morgan diagnostic spirometer. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                   |  |                       |                 |                     |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Wright-<br>Morgan |  | Minute<br>Ventilation | Tidal Volume    |                     |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Bias              |  | 0.4%                  | 0.4%            |                     |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Accuracy          |  | 3.9%                  | 3.9%            |                     |
| Continued on next page.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                   |  |                       |                 |                     |

{3}------------------------------------------------

| Clinical<br>performance<br>testing:<br>Continued from<br>previous page | Combining the results of the ExSpiron-Wright study and the Wright-Morgan<br>study provided calculated estimates of ExSpiron bias and accuracy with respect to<br>the Morgan diagnostic spirometer:                                                                                                               |                       |              |
|------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|--------------|
|                                                                        | ExSpiron-<br>Morgan                                                                                                                                                                                                                                                                                              | Minute<br>Ventilation | Tidal Volume |
|                                                                        | Bias                                                                                                                                                                                                                                                                                                             | -1.7%                 | -1.5%        |
|                                                                        | Accuracy                                                                                                                                                                                                                                                                                                         | 9.9%                  | 9.6%         |
| Conclusions<br>regarding safety<br>and effectiveness:                  | Based on the comparisons of intended use, and results of nonclinical and clinical<br>testing, the ExSpironTM is substantially equivalent in intended use, safety, and<br>effectiveness to the nSpire Respirometer and to the breathing frequency monitor<br>function of the Philips Intellivue Patient Monitors. |                       |              |

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is depicted in black, and the text is also in black. The logo is simple and recognizable, and it is often used on official documents and websites related to the U.S. Department of Health & Human Services.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Respiratory Motion, Incorporated C/O Mr. Chas Burr Chas Burr QR Services 11 Mystic Avenue Winchester, Massachusetts 01890

Re: K120087

Trade/Device Name: ExSpiron™ Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: II Product Code: BZQ, BZK Dated: September 20, 2012 Received: September 21, 2012

Dear Mr. Burr:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

## Page 2- Mr. Burr

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm_for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

far.

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and

Radiological Health

Enclosure

{6}------------------------------------------------

## Indications For Use

510(k) Number (if known): K120087 Device Name: ExSpiron Indications for Use:

ExSpiron is indicated for use by healthcare professionals in healthcare facilities, such as post-operative care and critical care units, to monitor breathing in adult (at least 21 years of age) patients.

ExSpiron is a non-invasive system that graphically displays lung volume against time and reports an approximate value of:

- Tidal volume, .
- . Respiratory rate, and
- Minute ventilation. ●

ExSpiron measurements are used as an adjunct to other clinical information sources.

Prescription Use _ × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use_ (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Ar. M21

(Division Sign-Off (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesion Devices Division of Antol, Dental Devices

**510(k) Number:**_

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-b%E2%80%94diagnostic-devices/BZK/K120087](https://fda.innolitics.com/submissions/AN/subpart-b%E2%80%94diagnostic-devices/BZK/K120087)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
