ExSpiron is indicated for use by healthcare professionals in healthcare facilities, such as post-operative care and critical care units, to monitor breathing in adult (at least 21 years of age) patients. ExSpiron is a non-invasive system that graphically displays lung volume against time and reports an approximate value of: Tidal volume, Respiratory rate, and Minute ventilation. ExSpiron measurements are used as an adjunct to other clinical information sources.
Device Story
ExSpiron is a non-invasive respiratory monitor for use by healthcare professionals in clinical settings. It utilizes a bioimpedance measurement system consisting of six electrodes placed on the patient's torso; a high-frequency current generator delivers current while an adaptive circuit conditions the resulting voltage signal. Firmware performs signal acquisition, relaying data to a Windows 7-based panel PC. The PC processes signals, performs calibration, and runs a GUI to display respiratory traces, tidal volume, minute ventilation, and respiratory rate. The device serves as an adjunct to other clinical information sources to assist in patient monitoring; it does not include alarm functions. Benefits include continuous, non-invasive tracking of respiratory parameters to support clinical decision-making in post-operative and critical care environments.
Clinical Evidence
Clinical study compared ExSpiron to the Wright spirometer in 20 subjects (age 20-84). Measurements included tidal volume, minute ventilation, and respiratory rate. Results showed high agreement: overall bias for minute ventilation was -2.1%, tidal volume -1.9%, and respiratory rate -0.2%. Precision (overall) was 10.5% for minute ventilation, 10.3% for tidal volume, and 2.0% for respiratory rate. A secondary study compared the Wright spirometer to the Morgan SpiroAir LT to establish a reference baseline. Combined data confirmed ExSpiron accuracy (9.9% for minute ventilation, 9.6% for tidal volume) relative to the diagnostic spirometer.
Technological Characteristics
Thoracic bioimpedance sensing; six-electrode lead set. Windows 7-based panel PC for signal processing and GUI. Connectivity via digital signal acquisition from adaptive circuit. Electrical/mechanical safety per IEC 60601-1; electromagnetic compatibility per IEC 60601-1-2. Single-patient-use electrode lead set.
Indications for Use
Indicated for monitoring breathing in adult patients (≥21 years) in healthcare facilities (e.g., post-operative, critical care).
Regulatory Classification
Identification
A monitoring spirometer is a device used to measure continuously a patient's tidal volume (volume of gas inhaled by the patient during each respiration cycle) or minute volume (the tidal volume multiplied by the rate of respiration for 1 minute) for the evaluation of the patient's ventilatory status.
{0}------------------------------------------------
K120087
Image /page/0/Picture/1 description: The image shows a black and white drawing of a pair of lungs. The lungs are connected to the trachea, which is the tube that carries air to and from the lungs. The lungs are the main organs of the respiratory system, and they are responsible for taking in oxygen and releasing carbon dioxide. The image is a simple illustration of the lungs and trachea.
SEP 2 6 2012
## 510(k) Summary
| 510(k) Owner: | Respiratory Motion, Inc. |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 305 Second Avenue, Suite B<br>Waltham, Massachusetts 02451 |
| Phone: | 781-373-1636 |
| Fax: | 781-373-1653 |
| Contact person: | Jenny Freeman, MD |
| Date 510(k)<br>Summary prepared: | September 20, 2012 |
| Trade name: | ExSpiron™ |
| Common name: | Respiratory Monitor |
| Classifications: | Monitoring spirometer<br>Product code: BZK<br>Regulation: 21 CFR 868.1850<br><br>Breathing frequency monitor<br>Product code BZQ<br>Regulation: 21 CFR 868.2375 |
| Predicate devices: | For product code BZK, nSpire Wright/Haloscale Respirometer, marketed by<br>nSpire Health, Inc. of Longmont, CO 510(k) K091853<br><br>For product code BZQ, Philips MPn0 Intellivue Patient Monitors, marketed by<br>Philips Healthcare 510(k) K060541 |
| Device Description: | The ExSpiron™ consists of:<br>Bioimpedance measurement system: A stabilized high frequency<br>current generator is connected to two outer electrodes. The inner four<br>electrodes are connected to an adaptive circuit that conditions the<br>resulting voltage signal and converts it to digital form. Firmware<br>performs signal acquisition and relays data to the panel PC.Panel PC: A Windows 7 PC performs signal processing and calibration,<br>and runs the graphical user interface (GUI). The PC takes user input from<br>a touch screen through a virtual keyboard and mouse. The GUI is used<br>for recording patient data and displaying the respiratory trace as well as<br>scalar values and trends for minute volume, tidal volume, and respiratory<br>rate.Single Patient Use ExSpiron™ Electrode Lead Set: An electrode lead<br>set containing six electrodes to be placed on the torso. It delivers current<br>and records impedance measurements. The electrode lead set is also used<br>to perform subsystem checks prior to patient measurements. |
{1}------------------------------------------------
| Intended use: | ExSpiron™ is indicated for use by healthcare professionals in healthcare facilities,<br>such as post-operative care and critical care units, to monitor breathing in adult (at<br>least 21 years old) patients.<br><br>ExSpiron™ is a non-invasive system that graphically displays lung volume<br>against time and reports an approximate value of:<br>• Tidal volume,<br>• Respiratory rate, and<br>• Minute ventilation<br><br>ExSpiron™ measurements are used as an adjunct to other clinical information<br>sources. | | | | |
|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison of<br>technological<br>characteristics: | Characteristic | ExSpiron™ | nSpire | Philips<br>Intellivue<br>Monitor | Comment |
| | Intended Use | See above. | See K091853 | See K060541 | The clinical application of the<br>ExSpiron™ is equivalent to that of<br>the nSpire.<br><br>The comparison to the intended<br>use of the Philips monitor is<br>restricted to monitoring of<br>respiration rate.<br><br>ExSpiron™ has no alarm function.<br><br>All devices are used by healthcare<br>professionals. |
| | Technology | Measurement is<br>by thoracic<br>bioimpedance. | Measurement is<br>by in-line<br>turbine flow<br>meter. | Measurement<br>is by<br>bioimpedance. | |
| | Volume<br>Measurements | Tidal volume<br>Minute volume<br>Volume vs. time<br>chart | Tidal volume<br>Minute volume | None | The time function for the<br>minute volume measurement of<br>the nSpire is provided by a<br>user's stopwatch. |
| | Rate<br>Measurements | Respiratory rate<br>(breaths/min) | None | Respiratory<br>rate | |
| | Safety | IEC 60601-1,<br>including<br>electrical and<br>mechanical<br>safety | Mechanical<br>only. Safety<br>specifications<br>are not given in<br>the user manual<br>or 510(k)<br>summary. | IEC 60601-1,<br>etc. | |
| Nonclinical<br>performance | Performance testing confirmed essential performance by IEC 60601-1 and IEC<br>60601-1-2 tests and confirmed usability by simulated use by physicians and | | | | |
{2}------------------------------------------------
| Clinical<br>performance<br>testing: | A clinical study compared simultaneous measurements from the ExSpiron and the<br>predicate, Wright spirometer. (Respiratory rate was calculated using a stop<br>watch.) Twenty subjects representing a broad range of intended patients<br>participated in the study. (Age range: 20-84, BMI range: 19.9-43.2, 11 female, 9<br>male) The study involved two sessions for each subject, an initial session in<br>which electrodes were applied, the ExSpiron was calibrated to the individual, and<br>each subject performed ten 60-second breathing tests. Tidal volume, minute<br>ventilation, and respiratory rate were measured simultaneously by the ExSpiron<br>and the Wright spirometer. Each subject returned twenty-four hours after the first<br>session with the original electrodes still attached. Using the same individual<br>calibration values, a second set of ten 60-second breathing tests was performed. | | | | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|--|-----------------------|-----------------|---------------------|
| | The results of ExSpiron-Wright study were: | | | | | |
| | | ExSpiron-Wright | | Minute<br>Ventilation | Tidal<br>Volume | Respiratory<br>Rate |
| | Bias | Overall | | -2.1% | -1.9% | -0.2% |
| | | Day 1 | | -2.3% | -1.7% | -0.6% |
| | | Day 2 | | -1.9% | -2.0% | 0.2% |
| | Precision | Overall | | 10.5% | 10.3% | 2.0% |
| | | Day 1 | | 10.2% | 9.7% | 2.3% |
| | | Day 2 | | 10.8% | 10.8% | 1.6% |
| | Accuracy | Overall | | 10.7% | 10.4% | 2.0% |
| | | Day 1 | | 10.4% | 9.8% | 2.4% |
| | | Day 2 | | 10.9% | 11.0% | 1.6% |
| The results indicate that the individual calibration is stable over a 24-hour period.<br>A similar study was performed comparing the Wright spirometer to the Morgan<br>SpiroAir LT diagnostic spirometer. (The closed cell construction of the Morgan<br>spirometer limits the breathing test time to 30 seconds. A direct comparison of<br>the ExSpiron to the Morgan spirometer is suboptimal.) A different set of twenty<br>subjects participated. (Age range: 20-60, BMI range: 19.0-28.0, 9 female, 11<br>male) Each subject participated in one session performing twenty 30-second<br>breathing tests with volume measurements made simultaneously with the Wright<br>spirometer and the Morgan diagnostic spirometer. | | | | | | |
| | | Wright-<br>Morgan | | Minute<br>Ventilation | Tidal Volume | |
| | | Bias | | 0.4% | 0.4% | |
| | | Accuracy | | 3.9% | 3.9% | |
| Continued on next page. | | | | | | |
{3}------------------------------------------------
| Clinical<br>performance<br>testing:<br>Continued from<br>previous page | Combining the results of the ExSpiron-Wright study and the Wright-Morgan<br>study provided calculated estimates of ExSpiron bias and accuracy with respect to<br>the Morgan diagnostic spirometer: | | |
|------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|--------------|
| | ExSpiron-<br>Morgan | Minute<br>Ventilation | Tidal Volume |
| | Bias | -1.7% | -1.5% |
| | Accuracy | 9.9% | 9.6% |
| Conclusions<br>regarding safety<br>and effectiveness: | Based on the comparisons of intended use, and results of nonclinical and clinical<br>testing, the ExSpironTM is substantially equivalent in intended use, safety, and<br>effectiveness to the nSpire Respirometer and to the breathing frequency monitor<br>function of the Philips Intellivue Patient Monitors. | | |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is depicted in black, and the text is also in black. The logo is simple and recognizable, and it is often used on official documents and websites related to the U.S. Department of Health & Human Services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Respiratory Motion, Incorporated C/O Mr. Chas Burr Chas Burr QR Services 11 Mystic Avenue Winchester, Massachusetts 01890
Re: K120087
Trade/Device Name: ExSpiron™ Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: II Product Code: BZQ, BZK Dated: September 20, 2012 Received: September 21, 2012
Dear Mr. Burr:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
## Page 2- Mr. Burr
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm_for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
far.
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{6}------------------------------------------------
## Indications For Use
510(k) Number (if known): K120087 Device Name: ExSpiron Indications for Use:
ExSpiron is indicated for use by healthcare professionals in healthcare facilities, such as post-operative care and critical care units, to monitor breathing in adult (at least 21 years of age) patients.
ExSpiron is a non-invasive system that graphically displays lung volume against time and reports an approximate value of:
- Tidal volume, .
- . Respiratory rate, and
- Minute ventilation. ●
ExSpiron measurements are used as an adjunct to other clinical information sources.
Prescription Use _ × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use_ (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Ar. M21
(Division Sign-Off (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesion Devices Division of Antol, Dental Devices
**510(k) Number:**_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.