TRUPEAK PEAK FLOW METER

K963095 · Monaghan Medical Corp. · BZH · Nov 4, 1996 · Anesthesiology

Device Facts

Record IDK963095
Device NameTRUPEAK PEAK FLOW METER
ApplicantMonaghan Medical Corp.
Product CodeBZH · Anesthesiology
Decision DateNov 4, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.1860
Device ClassClass 2

Indications for Use

The TruPeak Peak Flow Meter is a simple mechanical device that responds to and indicates the Peak Expiratory Flow Rate (PEFR) generated by the user during a forced exhalation maneuver.

Device Story

Mechanical peak flow meter; measures Peak Expiratory Flow Rate (PEFR). User places inlet in mouth; performs forced exhalation from full lungs. Internal piston extends spring against pressure spike; indicator moves to maximum displacement position. Indicator remains at peak position for reading on logarithmic scale (liters per minute). Provides higher resolution at lower PEFR readings. Used by patients for self-monitoring or clinicians for assessment. Output informs respiratory status and clinical decision-making.

Clinical Evidence

No clinical data; bench testing only to verify conformance to November 1994 ATS Standards.

Technological Characteristics

Mechanical device; internal piston and spring mechanism. Logarithmic scale for PEFR quantification. Dimensions modified to meet November 1994 ATS Standards.

Indications for Use

Indicated for users requiring measurement of Peak Expiratory Flow Rate (PEFR) via forced exhalation maneuver.

Regulatory Classification

Identification

A peak-flow meter for spirometry is a device used to measure a patient's maximum ventilatory flow rate.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K963095 NOV 4 1996 # SECTION 2 - SUMMARY AND CERTIFICATION ## 510(k) SUMMARY Monaghan Medical Corporation 5 Latour Ave., Suite 1600 P.O. Box 2805 Plattsburgh, NY 12901-0299 Phone: 518-561-7330 Fax: 518-561-5660 Contact: James A. Cochie QA/RA Manager **Date Prepared:** July 23, 1996 **Device Proprietary Name:** TruPeak Peak Flow Meter **Device Classification name:** Peak Flow Meter for Spirometry, 73 BZH **Legally Marketed Device Identity:** TruPeak Peak Flow Meter, 510(k) #955234 **Device Description:** The TruPeak Peak Flow Meter is a simple mechanical device that responds to and indicates the Peak Expiratory Flow Rate (PEFR) generated by the user during a forced exhalation maneuver. The inlet of the device is placed into the user's mouth after which the maximum rate at exhalation is attempted from nearly full lungs. As this forced exhalation commences an internal piston extends a spring as reaction to an internal pressure spike. An indicator is pushed by the moving piston and remains at the maximum displacement position of the piston after the maneuver is terminated. This position corresponds to the PEFR and is quantified in liters per minute by reading a scale next to the indicator. The piston displacement function is logarithmic thereby providing greater resolution of scale at lower PEFR readings. The TruPeak Peak Flow Meter is identical in form and function to the TruPeak Peak Flow Meter as described in its 510(k) #955234 except for changes in component dimensions necessary for the performance of the device to conform to the November 1994 ATS Standards update.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...