Peak Flow Meter

K241338 · Chongqing Moffy Innovation Technology Co., Ltd. · BZH · Dec 31, 2024 · Anesthesiology

Device Facts

Record IDK241338
Device NamePeak Flow Meter
ApplicantChongqing Moffy Innovation Technology Co., Ltd.
Product CodeBZH · Anesthesiology
Decision DateDec 31, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.1860
Device ClassClass 2
AttributesPediatric

Indications for Use

SMPF-1A, SMPF-1B, SMPF-2A, SMPF-2B, SMPF-3A, SMPF-3B: This device is intended to monitor a patient's Peak Expiratory Flow (PEF) at home. The device is designed for adults and children over 5 years of age with caregiver supervision. The device is intended for monitoring respiratory onditions such as asthma. The device is for Over-The-Counter Use. SMPF-1C, SMPF-1D, SMPF-2C, SMPF-2D, SMPF-3C, SMPF-3D: This device is intended to monitor a patient's Peak Expiratory Flow (PEF) and Forced Expiratory Volume (FEVI) in one second at home. The device is designed for adults and children over 5 years of age with caregiver supervision. The device is intended for monitoring respiratory conditions such as asthma. The device is for Over-The-Counter Use.

Device Story

Peak Flow Meter (models SMPF-1A through 3D) is a handheld device for home monitoring of respiratory conditions like asthma. Patients perform forced expiratory maneuvers into the device; it measures Peak Expiratory Flow (PEF) and, for certain models, Forced Expiratory Volume in one second (FEV1). Used by adults and children over 5 years (with caregiver supervision) to track respiratory health. Output allows patients and healthcare providers to monitor disease status and adjust management plans. Device provides objective data to assist in clinical decision-making regarding asthma control.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Handheld peak flow meter for respiratory monitoring. Models vary by measurement capability (PEF only vs. PEF and FEV1). Designed for home use. No specific materials, energy sources, or software algorithm details provided in the source text.

Indications for Use

Indicated for adults and children >5 years (with caregiver supervision) for home monitoring of respiratory conditions like asthma, specifically measuring Peak Expiratory Flow (PEF) and/or Forced Expiratory Volume in one second (FEV1).

Regulatory Classification

Identification

A peak-flow meter for spirometry is a device used to measure a patient's maximum ventilatory flow rate.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 31, 2024 Chongqing Moffy Innovation Technology Co.,Ltd. % Eva Li Consultant Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave. Shanghai, 200122 China Re: K241338 Trade/Device Name: Peak Flow Meter Regulation Number: 21 CFR 868.1860 Regulation Name: Peak-Flow Meter For Spirometry Regulatory Class: Class II Product Code: BZH Dated: November 26, 2024 Received: November 26, 2024 Dear Eva Li: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory {2}------------------------------------------------ assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Rachana Visaria -S Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K241338 Device Name Peak Flow Meter Indications for Use (Describe) SMPF-1A, SMPF-1B, SMPF-2A, SMPF-2B, SMPF-3A, SMPF-3B: This device is intended to monitor a patient's Peak Expiratory Flow (PEF) at home. The device is designed for adults and children over 5 years of age with caregiver supervision. The device is intended for monitoring respiratory onditions such as asthma. The device is for Over-The-Counter Use. SMPF-1C, SMPF-1D, SMPF-2C, SMPF-2D, SMPF-3C, SMPF-3D: This device is intended to monitor a patient's Peak Expiratory Flow (PEF) and Forced Expiratory Volume (FEVI) in one second at home. The device is designed for adults and children over 5 years of age with caregiver supervision. The device is intended for monitoring respiratory conditions such as asthma. The device is for Over-The-Counter Use. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|-----------------------------| | Prescription Use (Part 21 CFR 801 Subpart D) | | | Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <span> </span> </div> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...