MICRO SPIROMETER

K963035 · Micro Direct, Inc. · BZG · Nov 1, 1996 · Anesthesiology

Device Facts

Record IDK963035
Device NameMICRO SPIROMETER
ApplicantMicro Direct, Inc.
Product CodeBZG · Anesthesiology
Decision DateNov 1, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.1840
Device ClassClass 2

Indications for Use

The intended device takes spirometer readings of a patient during exhalation. The output readings are FEV1 and FVC.

Device Story

Hand-held portable electronic spirometer; measures FEV1 and FVC during patient exhalation. Uses flow-sensing turbine transducer; LCD display for output. Operated by clinicians in home or hospital settings. Device provides lung function measurements to assist clinical evaluation.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Hand-held electronic spirometer; flow-sensing turbine transducer; LCD display. Flow range 0-12 lps BTPS; volume range 0-9.99 liters. Accuracy +/- 3%. Utilizes paper mouthpiece; transducer requires cleaning after use.

Indications for Use

Indicated for patients requiring lung function evaluations; intended for use by clinicians in environments such as home or hospital.

Regulatory Classification

Identification

A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.

Predicate Devices

Submission Summary (Full Text)

{0} K963035 Micro Direct NOV - 1 1996 # Non-Confidential Summary of Safety and Effectiveness August 1, 1996 page 1 of 2 Micro Direct, Inc. 840 Pownal Rd. Auburn, ME 04210 Tel - (207) 786-7808 Fax - (207) 786-7280 Official contact: David R. Staszak, President Proprietary or Trade Name: Micro Spirometer Common/Usual Name: Diagnostic Spirometer Classification Name: 73 BZG - Diagnostic Spirometer Intended device: Diagnostic spirometer Predicate devices: Spirometrics Model 2600 Micro Spirometer - K901057 Device description: A hand-held portable electronic spirometer measuring FEV1 and FVC. ## Indicated use: The intended device takes spirometer readings of a patient during exhalation. The output readings are FEV1 and FVC. ## Targeted population: Patients requiring lung function evaluations. ## Environment of use: Places where a qualified clinician desires to take these lung measurements - FEV1 and FVC. ## Comparison to predicate devices: | Attribute | Intended device | Spirometrics Model 2600 | | --- | --- | --- | | Use | | | | Intended as a spirometer | Yes | Yes | | Intended to measure lung function FEV1 and FVC | Yes | Yes | page 3 of 29 Micro Direct, Inc. • 840 Pownal Road • P. O. Box 239 • Auburn, ME 04212-0239 {1} Non-Confidential Summary of Safety and Effectiveness August 1, 1996 page 2 of 2 Comparison to predicate devices: (continued) | Attribute | Intended device | Spirometrics Model 2600 | | --- | --- | --- | | Use (continued) | | | | Indicated for measuring patients lung function | Yes | Yes | | Indicated for use by a clinician in environments such as the home or hospital | Yes | Yes | | Manufactured by Micro Medical | Yes | Yes | | Exact same unit as supplied to and marketed by Spirometrics and approved under K901057 | Yes | -- | | Design | | | | Transducer - flow sensing turbine | Yes | Yes | | No output | Yes | Yes | | Flow range Ips BTPS 0-12 | Yes | Yes | | Volume range liters 0-9.99 | Yes | Yes | | Measures - FVC - FEV1 | Yes | Yes | | Accuracy + / - 3 | Yes | Yes | | Display LCD | Yes | Yes | | Transducer cleaned after use | Yes | Yes | | Utilizes paper mouthpiece | Yes | Yes | | Materials | | | | Same materials as K901057 | Yes | Yes | | Performance Testing | | | | None applicable | Yes | Yes | ## Differences There is no difference between the intended device and the listed predicate device. page 4 of 29
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...