← Product Code [BZG](/submissions/AN/subpart-b%E2%80%94diagnostic-devices/BZG) · K201562

# Spirotrac (K201562)

_Vitalograph Ireland, Ltd. · BZG · Oct 14, 2020 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-b%E2%80%94diagnostic-devices/BZG/K201562

## Device Facts

- **Applicant:** Vitalograph Ireland, Ltd.
- **Product Code:** [BZG](/submissions/AN/subpart-b%E2%80%94diagnostic-devices/BZG.md)
- **Decision Date:** Oct 14, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.1840
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Software as a Medical Device, Pediatric

## Indications for Use

The Vitalograph Spirotrac Model 7000 is a PC-based software application intended to be used as a spirometer or connect to compatible Vitalograph or third party devices to acquire, view, store and print the device output. The product is designed for use on adults and pediatrics, 5 years and older, in a variety of professional healthcare environments, e.g. primary care, hospitals and occupational health centers under the supervision of a healthcare provider.

## Device Story

PC-based software application; functions as spirometer or interface for compatible Vitalograph/third-party diagnostic devices (pulse oximetry, blood pressure, weight, ECG). Inputs: patient demographic data; digital signals from connected devices. Processing: data acquisition, storage, display, and printing of pulmonary function tests (VC, FIVC, FVC, PEF, FEFx, MVV) and connected device outputs. Used in primary care, hospitals, and occupational health centers by healthcare providers. Output: visual display of waveforms/measurements, printed reports, and database records. Facilitates clinical decision-making through longitudinal tracking and trending of pulmonary and diagnostic data. Benefits: centralized patient data management and standardized pulmonary function assessment.

## Clinical Evidence

Bench testing only. Verification and validation performed per FDA software guidance. Compliance with ATS/ERS (2005) waveform simulator testing, ISO 23747:2015 (peak expiratory flow meters), and ISO 26782:2009 (spirometers) confirmed data accuracy and performance.

## Technological Characteristics

Software-only device; no physical hardware, electrical, or biocompatibility components. Operates on Windows 7, 8, and 10. Connectivity via USB and Bluetooth. Database management via MS SQL Server. Supports networked operation and standard PC peripherals (keyboard, mouse, touchscreen).

## Regulatory Identification

A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.

## Predicate Devices

- Vitalograph Model 7000 ([K141546](/device/K141546.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 14, 2020

Vitalograph Ireland Ltd. % Paul Dryden Consultant ProMedic. LLC 131 Bay Point Dr. NE St. Petersburg, Florida 33704

Re: K201562

Trade/Device Name: Vitalograph Spirotrac Model 7000 Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: Class II Product Code: BZG Dated: September 10, 2020 Received: September 11, 2020

Dear Paul Dryden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Malvina B. Eydelman, M.D. Director OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

#### K201562

#### Device Name

## Vitalograph Spirotrac Model 7000

Indications for Use (Describe)

The Vitalograph Spirotrac Model 7000 is a PC-based software application intended to be used as a spirometer or connect to compatible Vitalograph or third party devices to acquire, view, store and print the device output. The product is designed for use on adults and pediatrics, 5 years and older, in a variety of professional healthcare environments, e.g. primary care, hospitals and occupational health centers under the supervision of a healthcare provider.

Type of Use (Select one or both, as applicable)

XX Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.'

FORM FDA 3881 (7/17)

Page 1 of 1

PSC Publishing Services (301)

{3}------------------------------------------------

#### K201562

### 510(k) Summary Page 1 of 5

| Date Prepared: | 13-Oct-2020 |
|----------------|-------------|
|----------------|-------------|

I Submitter

| Vitalograph Ireland Ltd.<br>Gort Road Business Park<br>Ennis Co Clare V95 HFT4 Ireland<br>Tel - +353-65-6864100 |
|-----------------------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------------|

| Submitter Contact:        | Tony O'Hanlon<br>Regulatory Affairs / Quality Assurance Manager |
|---------------------------|-----------------------------------------------------------------|
| Submission Correspondent: | Paul Dryden<br>ProMedic, LLC                                    |

II Device

| Proprietary or Trade Name: | Vitalograph Spirotrac Model 7000        |
|----------------------------|-----------------------------------------|
| Common/Usual Name:         | Diagnostic spirometer                   |
| Classification Name:       | Diagnostic spirometer (21 CFR 868.1840) |
| Regulatory Class:          | II                                      |
| Product Code:              | BZG                                     |

III Predicate Device:

|  | Vitalograph Model 7000, K141546 |
|--|---------------------------------|
|--|---------------------------------|

#### IV Device Description:

The Vitalograph Spirotrac Model 7000 provides a secure PC based medical device software application for creating, adding and recalling subjects and performing Spirometry testing on those subjects. Spirotrac will also link to compatible third party devices to read and display the output from these devices to allow the information to be retained with the subject.

Spirotrac integrates and reads / displays information from compatible Pulse Oximetry devices, Blood Pressure and Weight measurements devices, and ECG test devices.

Its primary functions are:

Spirometry measurements using single breath and multiple-breath testing techniques, the display and recording of measured lung volumes and flow rates (including VC, FIVC, FVC) and its subdivisions. The unit also allows for the measurements of Inspiratory and Expiratory Flow rates (PEF, FEFx, etc.), indirect measures (e.g. MVV) and Pre-post testing (e.g. Challenge, work shift).

Record subject demographic data as input.

Interact with existing Vitalograph and compatible third party devices via standard PC communication methods for download of data for storage within the Spirotrac database.

Navigation is allowed via the use of a standard PC keyboard and mouse or touchscreen.

{4}------------------------------------------------

#### 510(k) Summary Page 2 of 5

#### V Indications for Use:

The Vitalograph Spirotrac Model 7000 is a PC-based software application intended to be used as a spirometer or connect to compatible Vitalograph or third party devices to acquire, view, store and print the device output. The product is designed for use on adults and pediatrics, 5 years and older, in a variety of professional healthcare environments, e.g. primary care, hospitals and occupational health centers under the supervision of a healthcare provider.

Environments of use: Professional healthcare environments, e.g. primary care, hospitals and occupational health centers.

#### VI Modifications

A new code base, rewrite of the application, use of interfacing with latest Windows technology, User Interface etc. There were no changes to the algorithm.

#### Mechanical:

None, this is a software only device

#### Electrical:

None, this is a software only device

#### Materials:

None, this is a software only device

#### Software:

A new code base, rewrite of the application, use of interfacing with latest Windows technology, User Interface etc.

#### VII Comparison of Technological Characteristics and Performance with the Predicate

Table 1 is a comparison - Subject Device vs. the Predicate, K141546 including technological characteristics and performance.

{5}------------------------------------------------

# 510(k) Summary

### Table 1: Comparison of the Model 7000 Spirotrac vs. the Predicate

|                                                                                                                | Predicate                                                                                                                                                                                                                                                                                                                                                                                                                          | Subject                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Comparison                                                                                                                                                                                                                                                 |
|----------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                                                                                | Model 7000 Spirotrac                                                                                                                                                                                                                                                                                                                                                                                                               | Vitalograph Spirotrac Model 7000                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                            |
| K#                                                                                                             | K141546                                                                                                                                                                                                                                                                                                                                                                                                                            | K201562                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                            |
| Product Code                                                                                                   | BZG                                                                                                                                                                                                                                                                                                                                                                                                                                | BZG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                       |
| CFR                                                                                                            | 868.1840                                                                                                                                                                                                                                                                                                                                                                                                                           | 868.1840                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                       |
| Classification                                                                                                 | Spirometer, diagnostic                                                                                                                                                                                                                                                                                                                                                                                                             | Spirometer, diagnostic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                       |
| Indications for Use                                                                                            | The Vitalograph Model 7000 Spirotrac is intended<br>for use by or on the order of a physician in a<br>hospital or clinic setting. The product is designed<br>for use on both adult and pediatric patients. The<br>device is a PC based software application which is<br>intended to be used as a spirometer or connect to<br>compatible Vitalograph or third party devices to<br>acquire, view, store and print the device output. | The Vitalograph Spirotrac Model 7000 is a<br>PC-based software application intended to be<br>used as a spirometer or connect to compatible<br>Vitalograph or third party devices to acquire,<br>view, store and print the device output. The<br>product is designed for use on adults and<br>pediatrics, 5 years and older, in a variety of<br>professional healthcare environments, e.g.<br>primary care, hospitals and occupational<br>health centers under the supervision of a<br>healthcare provider. | The indications for use<br>have been updated from<br>the predicate K141546<br>which was not specific<br>regarding the type of<br>testing performed,<br>patient population or<br>environments of use.<br>Otherwise the indications<br>for use are the same. |
| Fundamental scientific technology                                                                              | Digital data communication                                                                                                                                                                                                                                                                                                                                                                                                         | Digital data communication                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                       |
| Spirometry - acquire, view,<br>store and print measures and<br>waveforms of pulmonary<br>function              | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                       |
| ECG waveforms -<br>view, store, print                                                                          | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                       |
| ECG waveforms - acquire                                                                                        | Yes, From compatible device.                                                                                                                                                                                                                                                                                                                                                                                                       | Yes, From compatible device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                                                                                                                                       |
| ECG waveform- view, store, print                                                                               | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                       |
| ECG Interpretation via<br>algorithms                                                                           | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                       |
| Ambulatory Blood Pressure -<br>retrieve, view, store and print<br>patient ambulatory blood<br>pressure history | Yes, from compatible device.                                                                                                                                                                                                                                                                                                                                                                                                       | Yes, from compatible device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                                                                                                                                       |
|                                                                                                                | Predicate                                                                                                                                                                                                                                                                                                                                                                                                                          | Subject                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Comparison                                                                                                                                                                                                                                                 |
|                                                                                                                | Model 7000 Spirotrac                                                                                                                                                                                                                                                                                                                                                                                                               | Vitalograph Spirotrac<br>Model 7000                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                            |
| K#                                                                                                             | K141546                                                                                                                                                                                                                                                                                                                                                                                                                            | K201562                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                            |
| Spot Oximetry download, view.<br>Weight                                                                        | Yes, From compatible device.<br>Manual entry or download via connection to<br>compatible device                                                                                                                                                                                                                                                                                                                                    | Yes, From compatible device.<br>Manual entry or download via connection<br>to compatible device                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                       |
| Microsoft windows Operating<br>Systems Supported:                                                              | Yes, Windows 7 and 8                                                                                                                                                                                                                                                                                                                                                                                                               | Yes, Windows 7, 8 and 10                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Support for newer<br>operating system                                                                                                                                                                                                                      |
| Database:                                                                                                      | MS SQL Server                                                                                                                                                                                                                                                                                                                                                                                                                      | MS SQL Server                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                                                                                                       |
| Where used                                                                                                     | Hospital, Health center, primary care practices<br>and clinics                                                                                                                                                                                                                                                                                                                                                                     | Hospital, Health center, primary care<br>practices and clinics                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                       |
| Networked operation                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                       |
| Subject Management:<br>Demographic Entry, Maintenance<br>and Deletion                                          | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                       |
| Report Printing                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                       |
| Spirometry testing                                                                                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                       |
| Trending Graphs for Spirometry<br>Results                                                                      | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                       |
| Spirometry Predicted Value<br>Equations                                                                        | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                       |
| Population Group Management                                                                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                       |
| Data Import/Export                                                                                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                       |
| Subject and Spirometry Data<br>Export                                                                          | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                       |
| Manual data entry of results                                                                                   | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                       |
| Data export via Email                                                                                          | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                       |
| Database Management                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                       |
| Color Display                                                                                                  | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                       |
| Population groups                                                                                              | Adult, Pediatric                                                                                                                                                                                                                                                                                                                                                                                                                   | Adult, Pediatric (5 years and above)                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                                                       |
| Communication                                                                                                  | Bluetooth, USB,                                                                                                                                                                                                                                                                                                                                                                                                                    | Bluetooth, USB,                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                       |
| Storage                                                                                                        | Dependent on storage media                                                                                                                                                                                                                                                                                                                                                                                                         | Dependent on storage media                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                       |
| Biocompatibility                                                                                               | No patient contact                                                                                                                                                                                                                                                                                                                                                                                                                 | No patient contact                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                                                                                                                                                                                                                                                       |

{6}------------------------------------------------

## K201562

# 510(k) Summary

{7}------------------------------------------------

#### VIII Performance Data

#### Non-clinical Testing

Performance testing demonstrated that the subject device met its acceptance criteria. Testing included:

#### Software

Verification and Validation was performed following Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 2005 and General Principles of Software Validation Guidance. January 2020.

#### Bench

- ATS / ERS (2005) waveform simulator testing
- ISO 23747:2015- Anaesthetic and respiratory equipment -- Peak expiratory flow meters ● for the assessment of pulmonary function in spontaneously breathing humans
- ISO 26782:2009 - Anaesthetic and respiratory equipment - Spirometers intended for the measurement of time forced expired volumes in humans.

Note that this testing was performed to demonstrate that the data displayed along with any limits set by user or by default, are aligned with these performance standards.

There is no hardware so electrical safety, electromagnetic compatibility, biocompatibility, and cleaning testing were not required.

#### Substantial Equivalence Conclusion

The performance testing has demonstrated that the subject devise met the applicable standard performance requirements. The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device are substantially equivalent.

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-b%E2%80%94diagnostic-devices/BZG/K201562](https://fda.innolitics.com/submissions/AN/subpart-b%E2%80%94diagnostic-devices/BZG/K201562)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
