The Vitalograph Model 2120 In2itive eDiary device is a battery-operated spirometer which measures three basic patient respiratory parameters (FVC, MVV and VC). The Vitalograph Model 2120 In2itive eDiary is a handheld spirometer designed for lung function testing in a variety of environments such as hospital wards, health centres and private homes. It is intended for Adults and pediatrics, 5 years and older. The Vitalograph Model 2120 In2itive eDiary can be configured as a stand-alone spirometer or connected to a printer.
Device Story
Handheld, battery-operated diagnostic spirometer; measures FVC, MVV, and VC. Uses Fleisch Pneumotachograph flowhead; transducer converts airflow to electrical signals for processing. Operates in hospital, clinic, or home settings; used by clinicians or patients. Data stored on-device; exported to Spirotrac software or printed. Provides objective respiratory metrics to assist healthcare providers in assessing lung function and managing respiratory conditions.
Clinical Evidence
No clinical or animal testing performed. Bench testing conducted per ATS/ERS standards, ISO 23747 (Peak Flow), and ISO 26782 (Timed Forced Expired Volume). Electrical safety and EMC testing performed per AAMI ANSI ES 60601-1, IEC 60601-1-2, and IEC 60601-1-11.
Technological Characteristics
Fleisch Pneumotachograph flowhead; 4.3" color LCD with capacitive touch; 5V/3.7V Li-ion battery. Connectivity: USB for Spirotrac software. Standards: ISO 10993-1 (biocompatibility), AAMI ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-1-11. Flow integration sampling at 100Hz.
Indications for Use
Indicated for lung function testing (FVC, MVV, VC) in adults and pediatrics 5 years and older in hospital, clinic, or home settings.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 27, 2020
Vitalograph Ireland Ltd. % Paul Dryden Consultant ProMedic. LLC 131 Bay Point Dr. NE St. Petersburg, Florida 33704
Re: K200550
Trade/Device Name: Vitalograph Model 2120 In2itive eDiary Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: Class II Product Code: BZG Dated: June 27, 2020 Received: June 30, 2020
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Michael Ryan Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
## K200550
Device Name
Vitalograph Model 2120 In2itive eDiary
Indications for Use (Describe)
The Vitalograph Model 2120 In2itive eDiary device is a battery-operated spirometer which measures three basic patient respiratory parameters (FVC, MVV and VC). The Vitalograph Model 2120 In2itive eDiary is a handheld spirometer designed for lung function testing in a variety of environments such as hospital wards, health centres and private homes. It is intended for Adults and pediatrics, 5 years and older.
The Vitalograph Model 2120 In2itive eDiary can be configured as a stand-alone spirometer or connected to a printer.
Type of Use (Select one or both, as applicable)
| XX Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number. "
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301)
{3}------------------------------------------------
| Date Prepared: | 27-July-2020 |
|----------------|--------------|
|----------------|--------------|
I. Submitter Vitalograph Ireland Ltd. Gort Road Business Park Ennis Co Clare V95 HFT4 Ireland Tel - +353-65-6864100
| Submitter Contact: | Tony O'Hanlon<br>Regulatory Affairs / Quality Assurance Manager |
|----------------------------|-----------------------------------------------------------------|
| Submission Correspondent: | Paul Dryden ProMedic, LLC |
| II.<br>Device | |
| Proprietary or Trade Name: | Vitalograph Model 2120 In2itive eDiary |
| Common/Usual Name: | Diagnostic spirometer |
| Classification Name: | Diagnostic spirometer (21 CFR 868.1840) |
| Regulatory Class: | II |
| Product Code: | BZG |
#### K100687 -- Vitalograph Model 2120 III. Predicate Device:
#### IV. Device Description:
The Vitalograph Model 2120 In2itive eDiary is a hand-held, battery-operated spirometer which measures the following lung function parameters FVC, MVV and VC in hospital, clinical and home settings.
It can be configured as a standalone spirometer or connected to a printer. Its primary functions and technology are:
- . Spirometry measurements using single breath and multiple-breath testing techniques, the
- display and recording of measured lung volumes and flow rates (including FVC, VC, ● MVV) and other parameters which are subsets of these measured parameters.
- . Record subject data. Storage of data and test results on unit for later printing or export to Spirotrac software
- The Flowhead utilizes a Fleisch Pneumotachograph.
- User Interface navigation via five buttons (Up, Down, Enter/Select, Cancel/Esc and . Power On/Off) or an optional touch screen.
#### V. Indications for Use:
The Vitalograph Model 2120 In2itive eDiary device is a battery-operated spirometer which measures three basic patient respiratory parameters (FVC, MVV and VC). The Vitalograph Model 2120 In2itive eDiary is a handheld spirometer designed for lung function testing in a variety of environments such as hospital wards, health centres and private homes. It is intended for Adults and pediatrics, 5 years and older.
The Vitalograph Model 2120 In2itive eDiary can be configured as a stand-alone spirometer or connected to a printer.
Environments of use: Hospital wards, health centres and private homes
{4}------------------------------------------------
#### VI. Modifications
The subject device includes modifications to the predicate. These modifications include:
## Mechanical:
- • New display and touch panel
- There is a new display which required that the outer housing be revised to o accommodate the larger display and touch panel.
## Electrical:
- Change of user interface to a 4.3" color LCD with capacitive touch panel (was 2.8" . resistive)
- . Inclusion of codec for sound generation, but the unmodified Model 2120 also provided sound
- Update to support Linux Operating System
### Materials:
- Flowhead outer housing material has changed
- o These are patient contacting materials which have been tested via ISO 10993-1
### Software:
- Support for user interface above ●
- Use of an embedded Linux Operating System ●
- Support for Codec as above ●
#### Comparison of Technological Characteristics and Performance with the Predicate VII.
Table 1 is a comparison - Subject Device vs. the Predicate, K100687 including technological characteristics and performance.
{5}------------------------------------------------
| | Predicate | Subject Device | Comparison |
|--------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| | Model 2120 | Vitalograph Model 2120 In2itive eDiary | |
| K# | K100687 | K200550 | - |
| Product Code | BZG | BZG | Same |
| CFR | 868.1840 | 868.1840 | Same |
| Classification | Spirometer, diagnostic | Spirometer, diagnostic | Same |
| Indications for Use | The device is a battery-operated spirometer which<br>measures three basic patient respiratory parameters<br>(FVC, MVV and VC). The Model 2120 is a handheld<br>spirometer designed for lung function testing in a<br>variety of environments such as hospital wards, health<br>centres and private homes.<br>The Model 2120 can be configured as a stand-alone<br>spirometer or connected to a printer. | The Vitalograph Model 2120 In2itive eDiary device<br>is a battery-operated spirometer which measures<br>three basic patient respiratory parameters (FVC,<br>MVV and VC). The Vitalograph Model 2120<br>In2itive eDiary is a handheld spirometer designed for<br>lung function testing in a variety of environments<br>such as hospital wards, health centres and private<br>homes. It is intended for Adults and pediatrics, 5<br>years and older.<br>The Vitalograph Model 2120 In2itive eDiary can be<br>configured as a stand-alone spirometer or connected<br>to a printer. | The indications for use for the<br>Model 2120 In2itive eDiary are<br>identical to the predicate Model<br>2120 with just the device name<br>updated. |
| Fundamental<br>scientific technology | Fleisch Pneumotachograph type Flowhead is<br>connected to a transducer, with related signal-<br>processing circuitry. | Fleisch Pneumotachograph type Flowhead is<br>connected to a transducer, with related signal-<br>processing circuitry. | Same |
| Parameters measured | FVC, MVV and VC | FVC, MVV and VC | Same |
| Hand-held | Yes | Yes | Same |
| Patient Interface | Flowhead cone | Flowhead cone | Same |
| Cleaning | Unit surface cleaning and 70% alcohol | Unit surface cleaning and 70% alcohol | Same |
| Patient use | Single Patient, multi-use | Unit surface cleaning and 70% alcohol | Same |
| Specifications | | | |
| Back pressure | Less than 0.1kPa/L/second @ 14L/s | Less than 0.1kPa/L/second @ 14L/s | Identical |
| Volume detection | Flow integration sampling @ 100Hz | Flow integration sampling @ 100Hz | Identical |
| Maximum displayed<br>volume | 10L | 10L | Identical |
| Volume accuracy<br>+3% or 0.05l | Yes | Yes | Identical |
| | Predicate<br>Model 2120 | Subject Device<br>Vitalograph Model 2120 In2itive eDiary | Comparison |
| Min. Volume | 0.01L | 0.01L | Identical |
| Flow Accuracy when<br>operated in operating<br>temperature range<br>conditions | Flow ±10% or 0.3 L/s<br>Max. flow rate ±16 L/s<br>Min. flow rate ±0.02 L/s | Flow ±10% or 0.3 L/s<br>Max. flow rate ±16 L/s<br>Min. flow rate ±0.02 L/s | Identical |
| Linearity | Better than ±3% | Better than ±3% | Identical |
| Operating<br>temperature range | 10-40°C | 10-40°C | Identical |
| Performance<br>standards | ATS/ERS (2005)<br>ISO 23747<br>ISO 26782 | ATS/ERS (2005)<br>ISO 23747<br>ISO 26782 | Identical |
| Electrical Safety and<br>EMC | ES 60601-1<br>IEC 60601-1-2 | ES 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11 | Similar. Added IEC 60601-1-<br>11 for home use which had not<br>been done for the predicate |
| Storage Temperature | 0-50°C | 0-50°C | Identical |
| Storage Relative<br>Humidity | 10%-95% | 10%-95% | Identical |
| Communications | USB x 1 for connection to Spirotrac | USB x 1 for connection to Spirotrac | Identical |
| Device weight | 0.310Kg | 0.230Kg | Similar |
| Dimensions | 186 x 110 x 48 mm | 160 x 100 x 45mm | Similar |
| Interface | Touchscreen LCD | Touchscreen LCD<br>Buttons, buttons were not used in practice | No.<br>Differences do not raise new<br>concerns |
| Power / Energy<br>Source | 5V/3.7V Li-ion rechargeable battery | 5V/3.7V Li-ion rechargeable battery | Identical |
| Biocompatibility | Surface Contact, Skin / Mucosa<br>Limited Duration | Surface Contact, Skin / Mucosa<br>Limited Duration | Identical patient contact<br>New material tested to ISO<br>10993-1 |
# Table 1: Comparison of the Model 2120 In2itive eDiary vs. the Predicate
{6}------------------------------------------------
{7}------------------------------------------------
#### VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility -
The patient contacting materials were identical to the predicate and review of the geometry and processing did not raise new or different concerns. For those that were different they were considered Surface Contact, Intact skin or Mucosa contacting with limited duration of contact and tested to ISO 10993-5 - Cytotoxicity and ISO 10993-10 - Sensitization and Irritation and found to be noncytotoxic, non-irritating and non-sensitizing.
## Electrical Safetv and EMC
Electrical safety and EMC testing were conducted on the subject device. The system complies with AAMI ANSI ES 60601-1: 2005 + A1: 2012 and IEC 60601-1-2:2014 for EMC and IEC 60601-1-11:2015 for home use.
## Software Verification and Validation Testing
Software verification and validation testing were conducted. The software for this device was considered as a "moderate" level of concern.
## Mechanical, Animal, and Clinical Testing
No mechanical, animal or clinical testing was performed.
## Bench Testing
Performance testing per ATS/ERS for peak flow and timed forced expired volume per ISO 23747:2015- Peak Flow and ISO 26782:2009 – Timed Forced Expired Volume was performed.
#### IX. Conclusions
## Discussion of Differences -
The identified differences do not raise new or different concerns of safety or effectiveness relative to the predicate.
## Substantial Equivalence Conclusion
The performance testing has demonstrated that the subject devise met the applicable standard performance requirements. The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate have been found to substantially equivalent.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.