The Vitalograph Model 6600 COMPACT is intended for use by or on the order of a physician in a hospital or clinic setting. The product is designed for use on adults and paediatrics, 5 years and older. The device is intended to be used as a spirometer which measures patient respiratory parameters including FVC, FEV1, FEV6, PEF, MVV and VC or connected to compatible Vitalograph or third party devices to acquire, view, store and print the device output.
Device Story
Mains-powered desktop spirometer; utilizes Fleisch pneumotachograph for single/multiple-breath testing. Inputs: patient respiratory flow/volume; data from compatible third-party devices (Nonin iPod, A&D blood pressure/scales, Corscience BT12). Operation: touch screen interface; runs Spirotrac software (K141546). Outputs: lung volumes/flow rates; display/print/storage of respiratory and connected device data. Used in hospitals/clinics by physicians/staff. Benefits: aids measurement of lung disease effects on pulmonary function; enables integrated patient data management. Healthcare providers use output for clinical assessment of respiratory function.
Clinical Evidence
Bench testing only. Device validated against international performance standards: ISO 26782 (spirometers), ATS/ERS 2005 (lung function testing), and ISO 23747 (peak expiratory flow). Mechanical testing included drop, random vibration, sinusoidal vibration, bump tests, and environmental (storage/operating temperature) testing. All tests demonstrated satisfactory results meeting performance requirements.
Indicated for adults and pediatric patients 5 years and older for the assessment of respiratory function via measurement of dynamic lung volumes (FVC, FEV1, FEV6, PEF, MVV, VC) in hospital or clinic settings.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around an emblem. The emblem consists of a stylized human figure with three faces, representing the department's focus on health and well-being. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 30, 2015
Vitalograph Ireland Ltd. Tom J. Healy Regulatory Affairs/OA Manager Gort Road Business Park Ennis, Co Clare Ireland
Re: K142642
Trade/Device Name: Vitalograph Model 6600 Compact Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: II Product Code: BZG Dated: May 25, 2015 Received: May 27, 2015
Dear Mr. Healy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Tom J. Healy
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K142642
#### Device Name
Vitalograph Model 6600 Compact
#### Indications for Use (Describe)
The Vitalograph Model 6600 COMPACT is intended for use by or on the order of a physician in a hospital or clinic setting. The product is designed for use on adults and paediatrics, 5 years and older. The device is intended to be used as a spirometer which measures patient respiratory parameters including FVC, FEV1, FEV6, PEF, MVV and VC or connected to compatible Vitalograph or third party devices to acquire, view, store and print the device output.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510K Summary
### as required by 21 CFR 807.92
- 1. Company Information: Name: Vitalograph (Ireland) Ltd Address: Gort Road Business Park, Ennis, Co Clare, Ireland.
Tel: +353656864100 Fax: +353656829289.
- 2. Contact Person / Official Correspondent: Mr. Tom J Healy Regulatory Affairs / Quality Assurance Manager
- 3. Date prepared: 29th June 2015.
- 4. Device Trade Name: Vitalograph Model 6600 Compact
- 5. Common / Usual name: Diagnostic Spirometer, Vitalograph Compact Expert
- 6. Classification number:
Class 2 Diagnostic Spirometer as classified per 21 CFR 868.1840. Product Code BZG.
## 7. Predicate Device:
| Manufacturer | : Vitalograph |
|--------------|---------------------------------------|
| Device Name | : Model 7000 Spirotrac |
| 510(k) No | : K141546, Class 2, Product Code BZG. |
#### 8. Description of Device:
Vitalograph Model 6600 Compact, running Vitalograph Model 7000 Spirotrac software, ref 510(k) K141546, shall provide a mains-powered desktop spirometer for creating, adding and recalling subjects and performing Spirometry testing on those subjects to aid in the measuring the effect of lung disease on pulmonary function.
Model 6600 Compact will also, via the Spirotrac software, connect to compatible third party devices to read and display the output from these devices to allow the
{6}------------------------------------------------
information to be retained with the subject. All connected compatible third party devices are those cleared via Model 7000 510(k) K141546.
Compatible third party devices are:
- Nonin iPod
- . A&D Blood Pressure
- Corscience BT12 ●
- A&D weighing scales ●
The intended use of the Compact is in the simple assessment of respiratory function through the measurement of dynamic lung volumes i.e. spirometry.
Its primary functions are:
- 1. Interaction will be via the touch Screen interface.
- 2. Running the Spirotrac software, as cleared in K141546, The model 6600 Compact performs spirometric measurements using the established fleisch Pneumotachograph, using single breath and multiple-breath testing techniques, to display and record lung volumes and flow rates (including FVC, FEV1, FEV6, PEF, MVV and VC) and their sub-divisions to aid in the measuring the effect of lung disease on pulmonary function
- 3. Record subject demographic data.
- 4. Produce printed reports to external printers.
- 9. Indications for Use:
The Vitalograph Model 6600 COMPACT is intended for use by or on the order of a physician in a hospital or clinic setting. The product is designed for use on adults and paediatrics. 5 years and older. The device is intended to be used as a spirometer which measures patient respiratory parameters including FVC, FEV1, FEV6, PEF, MVV and VC or connected to compatible Vitalograph or third party devices to acquire, view, store and print the device output.
- 10. Technological Characteristics
The differences between the Model 6600 Compact and the predicate devices are the software which is being used as the primary interface with the user. This software was previously cleared, ref K141546. The profile and weight of the Model 6600 Compact also differs where the Model 6600 is larger.
These differences do not pose any risks to the intended use of the device. The Model 7000 Spirotrac software (K141546) to be used independently of a PC within a controlled environment with all of the functions of a PC still available.
For performance, the same flow measurement and operating principles are used on the predicate devices as are on the Vitalograph Model 6600 Compact device. The flow circuit will use the same exact circuit and transducer as used in the Vitalograph Model 2120 {ref K100687 }. While, a new industrial design has been created, this has the
{7}------------------------------------------------
same touchscreen input mechanism, interface and functionality as was cleared for the Model 7000 Spirotrac, {ref K141546}.
The indications for use for the Vitalograph Pneumotrac now include pediatric population in line with the updated FDA guidance. The Model 6600 has the same indications for use, including pediatric population, as the Model 7000 as cleared under K141546. No new testing was required for this revised indication for use. No new risks have been introduced as a result of pediatric population inclusion. The device complies with the existing international performance standards to cater for all population groups.
Materials used continue to be those materials that have been previously used and do not introduce any new risks in relation to safety and effectiveness.
In relation to Patient interface accessories the Model 6600 will use the established Model 2820 and Model 2020 mouthpieces which have their own previous clearances and active device listings. See comparison table below.
The characteristics of the Model 6600 Compact are similar to those of the predicate devices listed in comparison table below. The similarities are
- Identical user interface, and software, as cleared via model 7000 Spirotrac, ref . 510(k) K141546.
- Same indications for use (Clarification of pediatric population and parameters ● measured)
- Some operating principle and flow measurement principles.
- Same parameters calculation.
- Same method of use ●
- Same performance when bench tested against performance standards. ●
- Same patient interface accessories ●
Risks have been evaluated and the performance has been validated. This validation is on file for all devices.
| | Compact Model<br>6600 | K141546<br>Spirotrac Model<br>7000 {predicate} | K100687 Model<br>2120 {predicate} |
|-------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|-----------------------------------------------------------------------|------------------------------------------------------------------|
| Spirometry -<br>acquire, view,<br>store and print<br>measures and<br>waveforms of<br>pulmonary<br>function | Yes | Yes | Yes |
| ECG waveforms -<br>view, store, print | Yes | Yes | No |
| ECG waveforms -<br>acquire | Yes, From<br>compatible device. | Yes, From<br>compatible device. | No |
| ECG waveform-<br>view, store, print | Yes | Yes | No |
| ECG<br>Interpretation via<br>algorithms | Yes | Yes | No |
| Ambulatory Blood<br>Pressure -<br>retrieve, view,<br>store and print<br>patient<br>ambulatory blood<br>pressure history | Yes, from compatible<br>device. | Yes, from<br>compatible device. | No |
| Spot Oximetry<br>download, view. | Yes, From<br>compatible device. | Yes, From<br>compatible device. | No |
| Weight<br>(measurement) | Manual entry or<br>download via<br>connection to<br>compatible device | Manual entry or<br>download via<br>connection to<br>compatible device | No |
| Microsoft<br>windows<br>Operating<br>Systems<br>Supported: | Yes | Yes | No |
| Database: | MS SQL Server | MS SQL Server | MS SQL when<br>downloaded to<br>Spirotrac {K100687<br>& K141546} |
| Where used | Hospital, Health<br>centre, primary<br>care practices and clinics | Hospital, Health<br>centre, primary care<br>practices and clinics | Hospital wards,<br>health centres and<br>homes |
| Networked<br>operation | Yes | Yes | No |
| Subject<br>Management:<br>Demographic<br>Entry,<br>Maintenance and<br>Deletion | Yes | Yes | Yes |
| Report Printing | Yes | Yes | Yes |
| Spirometry<br>testing | Yes | Yes | Yes |
| Trending Graphs<br>for Spirometry<br>Results | Yes | Yes | Yes, when<br>downloaded to<br>Spirotrac {K100687<br>& K141546} |
| Spirometry<br>Predicted Value<br>Equations | Yes | Yes | Yes |
{8}------------------------------------------------
{9}------------------------------------------------
| Population Group Management | | | |
|-----------------------------------------------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| Data Import/Export | Yes | Yes | Yes |
| Subject and Spirometry Data Export | Yes | Yes | Yes |
| Manual data entry of results | Yes | Yes | Yes |
| Data export via Email | Yes | Yes | Yes, when downloaded to Spirotrac {K100687 & K141546} |
| Database Management | Yes | Yes | Yes, when downloaded to Spirotrac {K100687 & K141546} |
| Colour Display | Yes | Yes | No |
| Target Population | Adult & Paediatric | Adult & Paediatric | Adult, Paediatric |
| Communication | Bluetooth, USB, | Bluetooth, USB, | USB, Micro SD card |
| Storage | Dependent on storage media | Dependent on storage media | Non-volatile data storage |
| Sterile | No | No | No |
| Regulatory (USA): | FDA - 510(k) Class 2 | FDA - 510(k) Class 2. K141546 | FDA - 510(k) Class 2 K100687 |
| Performance standards | ATS/ERS 2005<br>ISO 26782,<br>ISO 23747, | ATS/ERS 2005<br>ISO 26782,<br>ISO 23747, | ATS/ERS 2005<br>ISO 26782,<br>ISO 23747, |
| Safety Standards IEC / EN 60601 {EN 60601-1-1 and EN 60601-1-2} | Yes | Yes, for connected compatible devices | Yes |
| Performance Standards {incl bench tests}: | ATS ERS 2005, ISO 23747:2009 for PEF {formerly EN13826:2003}. EN ISO 26782:2009 | ATS ERS 2005, ISO 23747:2009 for PEF {formerly EN13826:2003}. EN ISO 26782:2009 | ATS ERS 2005, ISO 23747:2009 for PEF {formerly EN13826:2003}. EN ISO 26782:2009 |
| | Drop test.<br>Vibration testing<br>Bump Tests. | | Drop test. |
{10}------------------------------------------------
| | Operating<br>temperature | | Operating<br>temperature |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device materials | ABS plastic Body,<br>Silicone Rubber,<br>Stainless Steel,<br>Aluminium, TPX<br>plastic | N/A. Software | ABS plastic Body,<br>Silicone Rubber,<br>Stainless Steel,<br>Aluminium, TPX<br>plastic |
| Device weight | 2.5KG | N/A. Software | 0.230Kg |
| Dimensions | 375*235*110mm | N/A. Software | 160x100x45mm |
| Patient interface<br>accessories | Model 2820 BVF<br>Mouthpiece,<br>Singe; patient use.<br>510(k) K942779.<br>Product Code BZG | Dependent on<br>connecting<br>compatible third<br>party device. | Model 2820 BVF<br>Mouthpiece,<br>Singe; patient use.<br>510(k) K942779.<br>Product Code BZG<br><br>Model 2020<br>SafeTway<br>Mouthpiece.<br>Singe; patient use<br>{Class 1, 510(k)<br>exempt, with an<br>active device listing.<br>Device listing:<br>D141382. Product<br>Code BYP,<br><br>Vitalograph Nose<br>Clip, Singe; patient<br>use {Class 1, 510(k)<br>exempt, with an<br>active device listing.<br>Device listing<br>D130170. Product<br>Code BXJ |
| Product Code,<br>Class, CFR | BZG, Class 2,<br>868.1840 | BZG, Class 2,<br>868.1840 | BZG, Class 2,<br>868.1840 |
| Indications for<br>Use | The Vitalograph<br>Model 6600<br>COMPACT is<br>intended for use by<br>or on the order of a<br>physician in a<br>hospital or clinic<br>setting. The product<br>is designed for use<br>on adults and<br>paediatrics, 5 years<br>and older. The<br>device is intended to<br>be used as a | The Vitalograph<br>Model 7000<br>Spirotrac is<br>intended for use<br>by or on the order<br>of a physician in a<br>hospital or clinic<br>setting. The<br>product is<br>designed for use<br>on both adult and<br>pediatric patients.<br>The device is a PC<br>based software | The device is a<br>battery operated<br>spirometer which<br>measures three<br>basic patient<br>respiratory<br>parameters {FVC,<br>MVV and VC}. The<br>model 2120 is a<br>hand held<br>spirometer<br>designed for lung<br>function testing in a<br>variety of |
{11}------------------------------------------------
| | spirometer which<br>measures patient<br>respiratory<br>parameters including<br>FVC, FEV1, FEV6,<br>PEF, MVV and VC or<br>connected to<br>compatible<br>Vitalograph or third<br>party devices to<br>acquire, view, store<br>and print the device<br>output. | application which<br>is intended to be<br>used as a<br>spirometer or<br>connect to<br>compatible<br>Vitalograph or<br>third party devices<br>to acquire, view,<br>store and print the<br>device output. | environments such<br>as hospital wards,<br>health centres and<br>private homes. The<br>model 2120 can be<br>configured as a<br>stand-alone<br>spirometer or<br>connected to a<br>printer. |
|--|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
The Vitalograph Model 6600 underwent validation testing to ensure performance according to its specifications against current standards. These tests included performance testing against international standards such as
- . ISO 26782 { Anaesthetic and respiratory equipment -- Spirometers intended for the measurement of time forced expired volumes in humans},
- ATS/ERS 2005 { ATS/ERS Task Force: Standardisation of Lung Function ● Testing} and
- . ISO 23747 { Anaesthetic and respiratory equipment -- Peak expiratory flow meters for the assessment of pulmonary function in spontaneously breathing humans}.
Mechanical shock testing was also performed to evaluate the effects on the device during transport.
These tests included:
- Drop test of the packaged device from a specified height onto all corners and edges.
- Random Vibration, sinusoidal Vibration, and Bump Tests. ●
- Storage conditions testing. ●
- Operating temperature limits testing. ●
All tests and validations demonstrated satisfactory results.
As with the predicate device the Model 6600 Compact successfully passed the performance requirements of these tests. Evidence of successful completion of tests and validations has been provided with this submission.
## 11. Conclusion:
The characteristics of the Model 6600 Compact are similar to those of the predicate devices listed.
Based on the above, including the successful completion of all device testing Vitalograph conclude that this device is as safe and as effective as the predicate devices.
No new issues of safety or effectiveness have been introduced as a result.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.