MINISPIR

K122384 · Mir Medical Intl. Research Srl · BZG · Oct 4, 2012 · Anesthesiology

Device Facts

Record IDK122384
Device NameMINISPIR
ApplicantMir Medical Intl. Research Srl
Product CodeBZG · Anesthesiology
Decision DateOct 4, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 868.1840
Device ClassClass 2
AttributesPediatric

Indications for Use

The Minispir III spirometer and pulse oximeter is intended to be used by either a physician, respiratory therapist or technician. The device is intended to test lung function and can make: - spirometry testing in people of all ages, excluding infants and neonates. . - oximetry testing in people of all ages. . It can be used in any setting.

Device Story

Minispir is a USB-connected spirometer and pulse oximeter; operates via connection to a PC. Device measures respiratory parameters, blood oxygen saturation (SpO2), and heart pulse rate. Used by physicians, respiratory therapists, or technicians in any clinical or non-clinical setting. Input data captured via turbine sensor (spirometry) and oximetry module; processed by internal microprocessor. Output displayed on PC for clinician review to assess lung function and oxygenation status. Benefits include diagnostic support for respiratory conditions. Device updates include new PCB, microprocessor, and oximetry module compared to predicate.

Clinical Evidence

Clinical investigation conducted to validate the new oximetry board. Spirometry performance verified against American Thoracic Society (ATS) standards. Bench testing included electromagnetic compatibility (EMC), electrical safety, mechanical durability, and environmental (temperature/humidity) testing per EN 60601-1:2006 and EN 60601-1-2:2007. Results confirmed compliance with specifications and accuracy requirements.

Technological Characteristics

USB-connected spirometer and pulse oximeter. Features turbine sensor for spirometry and oximetry module for SpO2/pulse rate. Includes new PCB and microprocessor. Operates via PC interface. Complies with EN 60601-1 and EN 60601-1-2 standards. Spirometry algorithms consistent with predicate.

Indications for Use

Indicated for lung function testing in patients of all ages (excluding infants/neonates for spirometry) by physicians, respiratory therapists, or technicians in any setting.

Regulatory Classification

Identification

A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for Medical International Research (MIR). The logo consists of a stylized "M" formed by three overlapping shapes, followed by the letters "MIR" in bold, sans-serif font. Below the logo, the full name "MEDICAL INTERNATIONAL RESEARCH" is printed in a smaller, sans-serif font. # 510 (k) Summary # 1. Applicant Information Date Prepared: March 16, 2012 Submitter: MIR Medical International Research Address: Via del Maggiolino, 125 00155 Roma - Italy Contact Person: Gerda Van Houts 039-06-22754777 ## 2. Device Information Phone Number: Trade Name: Minispir Classification Name: spirometer and oximeter # 3. Identification of legally marketed device to which the submitter claims equivalence: Company Name: MIR Medical International Research Device Name: Minispir 510(k) number: K082766 As Spirometer : Regulation Number 868.1840 As oximeter: Regulation Number 870.2700 # 4. Description of the device: Minispir is a spirometer and pulse oximeter (optional) connected to a Personal Computer using a USB cable. The device measures a range of respiratory parameters, and the saturation of oxygen in the blood and the heart pulse rate. #### 5. Statement of Intended Use: The Minispir III spirometer and pulse oximeter is intended to be used by either a physician, respiratory therapist or technician. The device is intended to test lung function and can make: - spirometry testing in people of all ages, excluding infants and neonates. . - oximetry testing in people of all ages. . It can be used in any setting. 2012 {2}------------------------------------------------ # 6. Summary of the technological characteristics of the new device in comparison to those of the predicate device: The new Minispir uses a new PCB (with a new microprocessor) and a different oximetry module. Also the external aspect and dimensions have slightly changed. # Spirometry function The new Minispir has the - . same intended use - same operating principle . - . same turbine sensor - same spirometry parameters . - same algorithms for spirometry parameters calculation . as the predicate Minispir. # Oximetry function The new Minispir has the - . same intended use - same operating principle . - . same sensors - same oximetry parameters (SpO2, pulse rate) . - same algorithms for oximetry parameters calculation . as the predicate Minispir. Minispir uses a different oximetry module. ### 7. Brief discussion of the clinical and non clinical tests relied on for a determination of SE. Testing was done to ensure that the Minispir would perform safely and accurately within the environments for which it is to be marketed. Safety and environmental testing was conducted in accordance with EN 60601-1:2006 and EN 60601-1-2:2007. Electromagnetic compatibility (EMC), electrical, mechanical durability, safety (operator and patient), and temperature/humidity testing have been completed. The results demonstrates that the Minispir is in compliance with the guidelines and standards referenced and that it performs within its specifications. Spirometry testing was performed according with American Thoracic Society (ATS) Standards. The results obtained were within the range of accuracy required by ATS. A clinical investigation has been carried out to validate the new oximetry board. The results demonstrated required accuracy. #### 8. Conclusions Based on these results, it is our determination that the device is safe, effective, and performs as well as the legally marketed device. This summary on 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with outstretched wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the bird figure. The logo is black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 OCT 4 2012 M.I.R. Medical International Research Ms. Gerda Van Houts Export Area Manager Via Del Maggiolino 125 Rome, Italy 00155 Re: K122384 Trade/Device Name: Minispir- Spirolab III Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: II Product Code: BZQ, DQA Dated: September 11, 2012 Received: September 14, 2012 Dear Ms. Van Houts: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Nh for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): Device Name: Minispir -Spirolab III Indications for Use: The Minispir and Spirolab III spirometers and pulse oximeters are intended to be used by either a physician, respiratory therapist or technician. The devices are intended to test lung function and can make: - spirometry testing in people of all ages, excluding infants and neonates. . - oximetry testing in people of all ages. ● They can be used in any setting. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) L. Shulto (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devic Page 1 of 1 510(k) Number:
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...