← Product Code [BZG](/submissions/AN/subpart-b%E2%80%94diagnostic-devices/BZG) · K103575

# SPIRO PD (K103575)

_Pmd Healthcare · BZG · Nov 9, 2011 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-b%E2%80%94diagnostic-devices/BZG/K103575

## Device Facts

- **Applicant:** Pmd Healthcare
- **Product Code:** [BZG](/submissions/AN/subpart-b%E2%80%94diagnostic-devices/BZG.md)
- **Decision Date:** Nov 9, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.1840
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Pediatric

## Indications for Use

The SPIRO PD spirometer is intended to be used by a patient under the instruction of a physician or respiratory therapist to test lung function in child, adolescent and adult. It is also intended to be used as a single-patient device only and can be used in the home, factory, hospital or physician's office. The SPIRO PD spirometer is indicated for the following age groups: 2-12 years Child 13-21 years Adolescent 22 and over- Adult

## Device Story

SPIRO PD is a handheld, battery-powered diagnostic spirometer for measuring patient breath flow and volume. It uses a turbine transducer to convert mechanical rotation—caused by exhaled air—into electrical signals via infrared light interruption. The device is intended for single-patient use in homes, clinics, or hospitals. Patients operate the device under physician or respiratory therapist instruction. The device processes flow data to calculate pulmonary parameters (FVC, FEV1, FEV1/FVC, PEF, FEF 25-75%) and estimates lung age. Results, including flow/volume curves and severity indices, are displayed on a color touchscreen LCD. Data can be downloaded to a computer for review. By providing objective lung function metrics and symptom scores, the device aids healthcare providers in monitoring respiratory health and making clinical decisions regarding conditions like asthma or COPD.

## Clinical Evidence

No clinical data was collected or assessed. Substantial equivalence was established through bench testing, including electrical safety (IEC 60601-1, -1-2, -1-4, -1-6), biocompatibility (ISO 10993), and a human factors study (n=45) demonstrating 92.5% ease-of-use and 97% label comprehension across all age groups.

## Technological Characteristics

Handheld, battery-powered (3.7V Li-ion) diagnostic spirometer. Flow measurement via digital turbine transducer with infrared interruption. Features 320x240 color touchscreen LCD. Dimensions: 200x96x40mm; weight: 256g. Adheres to ATS guidelines. Predictive models: NHANES III. Biocompatibility per ISO 10993. Electrical safety per IEC 60601-1 series. Connectivity: data download to PC.

## Regulatory Identification

A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.

## Predicate Devices

- Spirotel ([K031643](/device/K031643.md))
- Vitalograph Model 4000 Series ([K073155](/device/K073155.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K103575

NOV - 9 2011

## 510(k) Summary

Sponsor:

PMD Healthcare

6620 Grant Way Company Address: Allentown, PA 18106

484-664-7600 Telephone: Fax: 484-664-7500

Contact Person:

Wayne Meng

Summary Preparation Date: November 7, 2011

Device Name: Trade Name: Common/Usual Name: Classification Name: Regulation Number: Product Code: Device Class: Review Panel:

SPIRO PD Spirometer Spirometer, Diagnostic 868.1840 BZG Class 2 Anesthesiology

### Predicate Device

| Company Name    | Product Name      | 510(k) Number |
|-----------------|-------------------|---------------|
| SDI Diagnostics | Spirotel          | K031643       |
| Vitalograph     | Model 4000 Series | K073155       |

#### Device Description

The SPIRO PD spirometer is an instrument that acquires mechanical signals and processes the information provided by the signal related to the pulmonary function. For processing purposes, mechanical signals must be changed to electrical signals. The devices responsible for this change are called transducers. The SPIRO PD has a Turbinetype transducer. The turbine transducer performs transduction in two stages: The volume to be measured crosses the turbine vane and records its rotation that is proportional to that Volume, this rotation is detected by the interrupting of a beam of infrared light, the sensor of which converts the light received into a digital-type electrical signal.

{1}------------------------------------------------

The SPIRO PD performs the following tests:

- . FVC,
- FEV1 .
- FEV1/FVC ●
- PEF
- . FEF 25-75%
- Estimated Lung Age and Flow/Volume Curve .

The Spiro PD performs the following calculations:

- % Predicted of FVC ●
- Severity of FVC .
- . % Predicted of FEV1
- Severity of FEV1 .
- % Predicted of FEV1/FVC .
- Severity of FEV1/FVC .
- % Predicted of FEF 25-75% .
- . Severity of FEF 25-75%
- % Predicted of PEF, Severity of PEF .

The physical configuration of the Spiro PD:

- Display Color, Touch screen LCD 320 x 240 resolution .
- . Keyboard - Touch screen
- Power supply 3.7V Lithium-ion battery 650 mAh .
- Dimensions 200 x 96 x 40mm .
- Weight 256g with battery .

Device Indications for Use

The SPIRO PD spirometer is intended to be used by a patient under the instruction of a physician or respiratory therapist to test lung function in child, adolescent and adult. It is also intended to be used as a single-patient device only and can be used in the home, factory, hospital or physician's office.

The Spiro PD spirometer is indicated for the following age groups:

- 2-12 years Child ●
- 13-21 years Adolescent �
- 22 and over- Adult .

{2}------------------------------------------------

## Predicate Product Comparison

| Predicate Product Comparison Chart                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                          |
|------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Parameter                                                              | SPIRO PD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | SDI Diagnostics'<br>Spirotel                                                                                                                                                                                                                                                                                                                                                                | Vitalograph                                                                                                                                                              |
| 510(k) Number                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | K031643                                                                                                                                                                                                                                                                                                                                                                                     | K073155                                                                                                                                                                  |
| Indications for Use<br>(IFU)                                           | Indications For Use:<br>The SPIRO PD<br>spirometer is<br>intended to be used<br>by a patient under<br>the instruction of a<br>physician or<br>respiratory<br>therapist to test<br>lung function in<br>people of all ages. It<br>is also intended to<br>be used as a single-<br>patient device only<br>and can be used in<br>the home, factory,<br>hospital or<br>physician's office.<br>The Spiro PD<br>spirometer is<br>indicated for the<br>following age<br>groups:<br>• 2-12 years -<br>Child<br>• 13-21 years<br>- Adolescent<br>• 22 and over-<br>Adult | Indications For Use:<br>The Spirotel<br>spirometer is<br>intended to be used<br>by a patient under<br>the instruction of a<br>physician or<br>respiratory therapist<br>to test lung function<br>in people of all ages.<br>It is also intended to<br>be used as a single-<br>patient device only<br>and can be used in<br>any setting - home,<br>factory, hospital or<br>physician's office. | The Vitalograph<br>Model 4000 is a<br>battery powered,<br>handheld electronic<br>spirometer used to<br>measure Peak Flow<br>(PEF) and Forced<br>expired volume<br>(FEV). |
| Device Description                                                     | The SPIRO PD<br>spirometer is a<br>hand-held portable<br>diagnostic<br>spirometer for the<br>measurement of<br>patient breath flow                                                                                                                                                                                                                                                                                                                                                                                                                            | The Spirotel<br>Spirometer is a<br>hand-held portable<br>diagnostic<br>spirometer for the<br>measurement of<br>patient breath flow                                                                                                                                                                                                                                                          | The Vitalograph<br>Model 4000 series<br>{ama-1 and copd-6}<br>are battery<br>powered handheld<br>electronic<br>spirometers used to                                       |
|                                                                        | and volume. The                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | and volume. The                                                                                                                                                                                                                                                                                                                                                                             | measure expired                                                                                                                                                          |
|                                                                        | device uses a                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | device uses a                                                                                                                                                                                                                                                                                                                                                                               | Peak Flow and                                                                                                                                                            |
|                                                                        | turbine transducer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | turbine transducer                                                                                                                                                                                                                                                                                                                                                                          | Forced Expired                                                                                                                                                           |
|                                                                        | that measures flow                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | that measures                                                                                                                                                                                                                                                                                                                                                                               | Volume after one                                                                                                                                                         |
|                                                                        | via infrared                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | flow via infrared                                                                                                                                                                                                                                                                                                                                                                           | second. The results                                                                                                                                                      |
|                                                                        | interruption.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | interruption.                                                                                                                                                                                                                                                                                                                                                                               | can aid in the                                                                                                                                                           |
|                                                                        | Algorithms are used                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Algorithms are used                                                                                                                                                                                                                                                                                                                                                                         | diagnosis of Asthma                                                                                                                                                      |
|                                                                        | to determine values                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | to determine                                                                                                                                                                                                                                                                                                                                                                                | and COPD in                                                                                                                                                              |
|                                                                        | based on this flow                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | values based on this                                                                                                                                                                                                                                                                                                                                                                        | patients.                                                                                                                                                                |
|                                                                        | measurement.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | flow measurement.                                                                                                                                                                                                                                                                                                                                                                           | All variants {asma-1                                                                                                                                                     |
|                                                                        | Tabular and                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Tabular and                                                                                                                                                                                                                                                                                                                                                                                 | and copd-6} within                                                                                                                                                       |
|                                                                        | graphical data are                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | graphical data are                                                                                                                                                                                                                                                                                                                                                                          | the range use the                                                                                                                                                        |
|                                                                        | displayed on the                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | displayed on the                                                                                                                                                                                                                                                                                                                                                                            | very same                                                                                                                                                                |
|                                                                        | spirometer LCD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | spirometer LCD                                                                                                                                                                                                                                                                                                                                                                              | operating principle,                                                                                                                                                     |
|                                                                        | display.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | display                                                                                                                                                                                                                                                                                                                                                                                     | LCD, Buttons, and                                                                                                                                                        |
|                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                             | Mouldings. Items                                                                                                                                                         |
|                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                             | that may vary                                                                                                                                                            |
|                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                             | within the range are                                                                                                                                                     |
|                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                             | list of parameters<br>that the different                                                                                                                                 |
|                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                             | variants display. I.e.,                                                                                                                                                  |
|                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                             | Asma-1 displays                                                                                                                                                          |
|                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                             | FEV1 and PEF whilst                                                                                                                                                      |
|                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                             | the copd-6 displays                                                                                                                                                      |
|                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                             | FEV1 and FEV6 only.                                                                                                                                                      |
|                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                             | A uni-directional                                                                                                                                                        |
|                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                             | rotating vane with                                                                                                                                                       |
|                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                             | flow sensor to                                                                                                                                                           |
|                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                             | measure lung                                                                                                                                                             |
|                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                             | function is used.                                                                                                                                                        |
|                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                             | The measurements                                                                                                                                                         |
|                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                             | are taken via                                                                                                                                                            |
|                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                             | expiration into the                                                                                                                                                      |
|                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                             | unit flowhead,                                                                                                                                                           |
|                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                             | which is in-turn                                                                                                                                                         |
|                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                             | displayed onto and                                                                                                                                                       |
|                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                             | LCD.                                                                                                                                                                     |
| Physical<br>Configuration                                              | Display:<br>Color,<br>Touchscreen LCD<br>320 x 240<br>resolution<br>Keyboard:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Display:<br>STN LCD, 2 lines x<br>16 alphanumeric<br>characters<br><br>Keyboard:                                                                                                                                                                                                                                                                                                            | Display:<br>Color LCD<br><br>Keyboard;                                                                                                                                   |
|                                                                        | Touchscreen                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 5 keys membrane                                                                                                                                                                                                                                                                                                                                                                             | User Buttons                                                                                                                                                             |
|                                                                        | Power supply:<br>3.7V Lithium-ion<br>battery 650 mAh<br>Dimensions:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Power supply:<br>3V Lithium battery<br>CR123A<br>Dimensions:<br>70 x 80 x 30mm                                                                                                                                                                                                                                                                                                              | Power Supply:<br>2 x 1.5v AAA<br>batteries<br>Dimensions                                                                                                                 |
|                                                                        | 200 x 96 x 40mm<br>Weight:<br>256g with battery                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Weight:<br>100g with battery                                                                                                                                                                                                                                                                                                                                                                | 113 x 63 x 48mm<br>Weight:<br>83g with battery                                                                                                                           |
| Power Source                                                           | 3.7 V, 650 mAh<br>Lithium-ion Battery                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 3V Lithium Battery                                                                                                                                                                                                                                                                                                                                                                          | 2x1.5v AAA Battery                                                                                                                                                       |
| ATS Spirometry<br>Performance<br>Recommendations                       | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                      |
| Cross<br>Contamination<br>Control                                      | Yes, single patient<br>device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                      |
| Flow Detection<br>Principle                                            | Digital Turbine                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Digital Turbine                                                                                                                                                                                                                                                                                                                                                                             | Rotor Stator Design                                                                                                                                                      |
| Flowmeter<br>Calibration Method                                        | Factory Calibration                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Factory Calibration                                                                                                                                                                                                                                                                                                                                                                         | Factory Calibration                                                                                                                                                      |
| Display and Printer<br>Used                                            | The SPIRO PD uses<br>a color touchscreen<br>LCD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | STN LCD, 2 Lines x<br>16 alphanumeric<br>characters                                                                                                                                                                                                                                                                                                                                         | The Vitalograph<br>Model 4000 uses a<br>color LCD Display                                                                                                                |
| Graphic Output                                                         | Graphic output can<br>be shown on the<br>LCD screen and<br>downloaded to a<br>computer                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Graphic output can<br>be shown on the<br>LCD screen and<br>downloaded to a<br>computer                                                                                                                                                                                                                                                                                                      | Graphic Output<br>shown on LCD<br>screen and<br>downloaded to a<br>computer                                                                                              |
| Tests Performed                                                        | FVC, FEV1,<br>FEV1/FVC, PEF, FEF<br>25-75%, Estimated<br>Lung Age and<br>Flow/Volume Curve                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | FVC, FEV1, FEV1%,<br>PEF, FEF25-75%,<br>FET, and<br>Flow/Volume Curve                                                                                                                                                                                                                                                                                                                       | FEV1, FEV6,<br>FEV1/FEV6 and<br>estimated Lung Age                                                                                                                       |
| Indices Calculated                                                     | % Predicted of FVC,<br>Severity of FVC, %<br>Predicted of FEV1,<br>Severity of FEV1, %<br>Predicted of<br>FEV1/FVC, Severity<br>of FEV1/FVC, %<br>Predicted of FEF 25-                                                                                                                                                                                                                                                                                                                                                                                        | % Predicted of FVC,<br>Severity of FVC, %<br>Predicted of FEV1,<br>Severity of FEV1, %<br>Predicted of<br>FEV1/FVC, Severity<br>of FEV1/FVC, %<br>Predicted of FEF 25-                                                                                                                                                                                                                      | FEV1 and FEV1 %<br>Predicted<br>FEV6 and FEV6 %<br>Predicted<br>FEV1/FEV6 and<br>FEV1/FEV6<br>Predicted<br>FEV1/FEV6 ration                                              |
|                                                                        | 75%, Severity of FEF<br>25-75%, %<br>Predicted of PEF,<br>Severity of PEF                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 75%, Severity of FEF<br>25-75%, % Predicted<br>of PEF, Severity of<br>PEF                                                                                                                                                                                                                                                                                                                   | 75%, Severity of FEF<br>25-75%, % Predicted<br>of PEF, Severity of<br>PEF                                                                                                |
| Predictive Models<br>Used                                              | NHANES III                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | NHANES III, Crapo,<br>Morris                                                                                                                                                                                                                                                                                                                                                                | NHANES III                                                                                                                                                               |
| General Electrical<br>Safety Tests IEC<br>60601-1 and IEC<br>60601-1-2 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                      |
| Provides Test date<br>and time, plus<br>symptom scores                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                      |
| Diary storage                                                          | 180 days                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 100 Days                                                                                                                                                                                                                                                                                                                                                                                    | None                                                                                                                                                                     |
| Replies to simple<br>questions and test<br>quality control             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                                                                                                                                         | No                                                                                                                                                                       |
| Maximum Volume                                                         | 0 - 10 L                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 10 L                                                                                                                                                                                                                                                                                                                                                                                        | 0.99 L BTPS                                                                                                                                                              |
| Flow Range                                                             | 0 -16 L/s                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | +16 L/s                                                                                                                                                                                                                                                                                                                                                                                     | 0.15 kPa/L/s at 14<br>L/s                                                                                                                                                |
| Flow Accuracy                                                          | $\u00b15%$ or 200 mL/s                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | $\u00b15%$ or 200 mL/s                                                                                                                                                                                                                                                                                                                                                                      | $\u00b13%$                                                                                                                                                               |
| Volume Accuracy                                                        | $\u00b13%$ or 50 mL,<br>whichever is<br>greater                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | $\u00b13%$ or 50 mL,<br>whichever is greater                                                                                                                                                                                                                                                                                                                                                | $\u00b13%$                                                                                                                                                               |
| Lung Age                                                               | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | NO                                                                                                                                                                                                                                                                                                                                                                                          | Yes…

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-b%E2%80%94diagnostic-devices/BZG/K103575](https://fda.innolitics.com/submissions/AN/subpart-b%E2%80%94diagnostic-devices/BZG/K103575)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
