K100928 · Viasys Respiratory Care, Inc. · BZG · Mar 7, 2011 · Anesthesiology
Device Facts
Record ID
K100928
Device Name
MICRO DIARY SPIROMETER
Applicant
Viasys Respiratory Care, Inc.
Product Code
BZG · Anesthesiology
Decision Date
Mar 7, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1840
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Micro Diary Spirometer is used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs. Specifically, the Micro Diary Spirometer measures the following lung function parameters: FEV1, FVC, FEV6 and PEF.
Device Story
Micro Diary Spirometer is a compact, battery-operated, handheld recording spirometer. Device utilizes a digital turbine volume transducer; exhaled breath rotates a vane; rotation frequency/count proportional to flow rate and volume. Measures FEV1, FVC, FEV6, and PEF. Used in hospitals, clinics, or patient homes under medical practitioner supervision. Operated by patients or clinicians. Device records test data for later review and supports transfer to compatible computers. Healthcare providers use output to monitor lung function and manage respiratory conditions like asthma or COPD. Benefits include portable, longitudinal tracking of pulmonary function.
Clinical Evidence
No clinical data. Bench testing only; device verified for accurate measurement of FEV1, FVC, FEV6, and PEF. Electrical safety and electromagnetic compatibility testing passed.
Technological Characteristics
Handheld diagnostic spirometer; turbine-based digital volume transducer; battery-powered; 5-button keypad; B&W LCD display. Data recording and transfer capability. Technological design, materials, and energy source identical to predicate.
Indications for Use
Indicated for patients 4 years and older requiring lung function measurements, including those with asthma or chronic obstructive pulmonary disorder, provided they can follow practitioner instructions.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
{0}------------------------------------------------
## 510(k) Summary
Image /page/0/Picture/1 description: The image shows the logo for CareFusion. The logo consists of a circular icon on the left and the word "CareFusion" on the right. The icon contains a shield shape inside the circle.
K100925
CareFusion 22745 Savi Ranch Pa Yorba Linda, CA 92 714-283-2228 te 714-283-8424 fax www.carefusion.com
# 510(k) Summary
MAR - 7 2011
Device Name: Device Trade Name: Common/Classification Name: Diagnostic Spirometer Micro Diary Spirometer Diagnostic Spirometer
Contact Person and Submitter: Merritt Girgis CareFusion 22745 Savi Ranch Parkway Yorba Linda, CA 92887 Phone: (714) 919-3286 Fax: (714)-283-8420 Email: merritt.girgis@carefusion.com
#### Substantial Equivalence
The Micro Diary Spirometer is substantially equivalent to the previous version of the Micro Diary Spirometer (K965042).
#### Description of the device
The Micro Diary Spirometer is a compact, hand held, and battery operated recording spirometer. The device features a five button keypad and an LCD display. The device utilizes a digital volume transducer. As a patient's exhaled breath is passed through the transducer, the vane rotates. The number of rotations is proportional to the volume of air passed through the transducer, and the frequency of rotation is proportional to the flow rate.
#### Indications for Use Statement
The Micro Diary Spirometer is used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs. Specifically, the Micro Diary Spirometer measures the following lung function parameters: FEV1, FVC, FEV6 and PEF.
#### Target Population
The device can be used on patients who require lung function measurements. These patients are usually suffering from diseases such as asthma and chronic obstructive pulmonary disorder. It can be utilized for patients from 4 years and older, providing that they are able to follow the medical practitioner's instructions.
#### Environment of Use
The environment of use is the hospital or a doctor's or medical practitioner's office or clinic or the patient's home.
{1}------------------------------------------------
### Technological Characteristics
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
The updated version of the Micro Diary Spirometer has the same technological characteristics (i.e., design, material, chemical composition, energy source) as the predicate device, which is previous version of the Micro Diary Spirometer (K965042).
### Comparison of Technological Characteristics
As a result of tooling and component obsolescence, it was necessary to produce a direct replacement for the previous version of the Micro Diary Spirometer using current tooling and available copnoponents. Both devices perform standard spirometric measurements, and both are capable of recording the test results for review at a later time. The MicroDiary (new) and the older MicroDiary perform exactly the same functions and tests.
| | Micro Diary (new) | Micro Diary (previous) |
|-------------------------------|--------------------------------------------------------------------|------------------------------------------------------------------|
| Parent Company | CareFusion | CareFusion(Formerly<br>Cardinal Health, Viasys,<br>Micro Medical |
| Device Classification<br>name | Diagnostic Spirometer | Diagnostic Spirometer |
| 510(k) Number | New | K965042 |
| Indications for Use | Measures lung function<br>as follows: FEV1, FVC,<br>FEV6, and PEF. | Measures lung function<br>as follows: FEV1, FVC,<br>FEV6, PEF |
| Display Screen | LCD Display, B&W | LCD Display, B&W |
| User interface | 5 Keys – on, up, down,<br>enter, and delete | 5 Keys – on, up, down,<br>enter, and delete |
| Measurement Transducer | Turbine | Turbine |
| Power supply | Battery | Battery |
### Nonclinical Testing
The Micro Diary Spirometer was tested to assure that it accurately measures lung function across the following standard spirometric measures: FEV1, FVC, FEV6, and PEF. All tests were passed successfully. Electrical Safety and Electromagnetic Compatibility testing was also performed and passed.
#### Clinical Testing
No clinical testing was submitted for device.
#### Conclusion
The testing has demonstrated that the device is as safe, as effective, and performs as well as or better than the legally marketed Micro Diary Spirometer (K965042).
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Merritt Girgis Manager, Regulatory Affairs CareFusion 22745 Savi Ranch Parkway Yorba Linda, California 92887
MAR - 7 2011
- Re: K100928
Trade/Device Name: Micro Diary Spirometer Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: II Product Code: BZG Dated: March 2, 2011 Received: March 3, 2011
Dear Mr. Girgis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registerion and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Anthony D. Avena
Anthony Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## EXHIBIT A Statement of Indications for Use
510(k) Number (if known): Not known.
Device Name: Micro Diary Spirometer
Indications for Use: The Micro Diary Spirometer measures lung function using the following standard spirometric measures: FEV1, FVC, FEV6, and PEF. The device also records the test data for later review and has the ability to transfer these test records to a compatible computer.
Target Population: Patients requiring lung function measurements.
Environment of Use: Under the supervision of a medical practitioner.
AND/OR Over-The-Counter Use Prescription Use × (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
L. Schultheis
(Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital
Division of Anest Franks Devices Division Controi, Dental Devices
510(k) Number: K/00928
{5}------------------------------------------------
### Statement of Indications for Use
510(k) Number: K100928.
Device Name: Micro Diary Spirometer
Indications for Use: The Micro Diary Spirometer is used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs. Specifically, the Micro Diary Spirometer measures the following lung function parameters: FEV1, FVC, FEV6 and PEF.
Target Population: The device can be used on patients who require lung function measurements. These patients are usually suffering from diseases such as asthma and chronic obstructive pulmonary disorder. It can be utilized for patients from 4 years and older, providing that they are able to follow the medical practitioner's instructions.
Environment of Use: The environment of use is the hospital or a doctor's or medical practitioner's office or clinic or the patient's home.
X Prescription Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
L. Schutter
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control and Dental Devices
§10(k) Number:
K 100 928
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.